Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Quantitation absolute calibration

The analysis of the collected average shift samples was made with a method validated by us, based on NIOSH method 1600/1994 (National Institute for Occupational Safety and Health). After eluation of the activated charcoal with toluene the test was made with a gas chromatograph with mass selective detector Perkin Elmer and capillary column DB-5, 30 m long and thickness of the coating 0.25pm. Helium was used as carrier gas. Apparatus conditions injector temperature 250°C, detector temperature 250°C, carrier gas pressure - 7 psig. The quantitative assessment of the samples was performed after absolute calibration with standard solutions of carbon disulfide in toluene. The limit of detection of the method is 0.01 mg/m3 at 25dm3 air sample. [Pg.403]

In contrast to NIR spectroscopy, the absolute values of the y-axis of a Raman measurement are difficult to quantify. Possible specific methods are described in USP chapter <1120> [21]. However, it is most common for quantitative Raman measurements to be done using the ratio of two peaks or other approaches which eliminate the need for absolute calibration of the y-axis. The USP chapter on Raman specifies that the photometric consistency or reproducibility specific integrated Raman band intensity should be on the order of 10%. [Pg.236]

Quantitative analysis of a series of copolymers of 2-((phenylamino)carbonyl)oxyethyl methacrylate and 1-methyl-2-phenoxyethyl methacrylate was carried out by FTIR spectroscopy. Emphasis was placed on aspects optimising the accuracy of the technique. The irreproducibility resulting from the use of cast films was shown to be overcome when using solutions. The criteria for solvent selection were established. Highly satisfactory calibration was afforded by blends of the two homopolymers in solution, thus providing an absolute calibration and obviating the use of model compounds or copolymers of known composition. 23 refs. [Pg.120]

Here G(f), F(f) and W(f) are Fourier transforms of g(t), f(t) and w(t), respectively. w(t) and W(f) are named the transfer function and the function of frequency response. A calibration of AE sensor is equivalent to determination of function W(f). On the other hand, it implies that fiequency contents of AE waves are usually smeared by function W(f) of AE sensor. Thus, the absolute calibration means quantitative estimation of function w(t) or W0. [Pg.22]

Susceptibility to radiation damage must be considered seriously if reference samples are to be calibrated for use in place of absolute systems. For the measurement of absolute C He, H) cross sections, films of polystyrene (CH) (which is relatively radiation hard) have been used successfiiUy, the RBS determination of carbon providing implied quantitation for the hydrogen present in the film. For a durable laboratory reference sample, however, there is much to recommend a known ion-implanted dose of H deep within Si or SiC, where the loss of hydrogen under room temperature irradiation will be neghgible. [Pg.498]

In a different example, traceability in the amount-of-substance analysis of natural potassium, thorium, and uranium by the method of passive gamma-ray spectrometry was demonstrated by Nir-El (1997). For an absolute quantitative determination, accurate values of two parameters were required (i) the emission probability of a gamma-ray in the decay of the respective indicator radionuclides, and (2) the detection efficiency of that gamma-ray. This work employed a number of CRMs in the critical calibration of the detection efficiency of the gamma-ray spectrometer and the establishment of precise emission probabilities. The latter results compared well with literature values and provided smaller uncertainties for several gamma-rays that were critical for the traceabUity claim. The amount-of-substance analytical results of the long lived naturally occurring radionucHdes K, Th, and... [Pg.251]

As XRF is not an absolute but a comparative method, sensitivity factors are needed, which differ for each spectrometer geometry. For quantification, matrix-matched standards or matrix-correction calculations are necessary. Quantitative XRF makes ample use of calibration standards (now available with the calibrating power of some 200 international reference materials). Table 8.41 shows the quantitative procedures commonly employed in XRF analysis. Quantitation is more difficult for the determination of a single element in an unknown than in a known matrix, and is most complex for all elements in an unknown matrix. In the latter case, full qualitative analysis is required before any attempt is made to quantitate the matrix elements. [Pg.632]

PIXE is a primary analytical technique, like NAA, and permits absolute determinations of concentrations. The basis for quantitative PIXE is, as in all X-ray methods, that there exists a relationship between the net peak area of an X-ray line in the spectrum and the amount of element in the sample. One of two methods can be applied to calibration ... [Pg.640]

Quantitative and standardless (absolute method no calibration curves)... [Pg.664]

In this method, inorganic lead in seawater samples are converted to tetra-ethylead using sodium tetraethylboron (NaB(C2H5)4) which is then trapped in a graphite furnace at 400 °C. Quantitation is achieved by using a simple calibration graph prepared from aqueous standards. An absolute detection limit of (3relative standard deviation. [Pg.189]

Internal Standard for KBr-Disc Technique In quantitative analysis it is essential to examine absolutely uniform discs of identical weights. To achieve this, known weights of both KBr and analyte are required in the preparation of the KBr-disc and finally from the absorption data a calibration-curve may be obtained. In this process, it is a must to weigh the discs and also to measure their thickness at different points... [Pg.329]

Quantitative methods using flame emission photometry cannot be absolute because an unknown, although relatively constant, proportion of the sample will reach the flame of which only a further small proportion of atoms will actually be excited and subsequently emit radiation. Hence it is essential to construct calibration curves for any analysis. The radiation emitted by the flame when pure solvent is sprayed is used to zero the instrument and the maximum reading set when the standard with the highest concentration is sprayed. [Pg.78]

The use of a continuous GPC viscosity detector in conjunction with a DRI detector permits the quantitative determination of absolute molecular weight distribution in polymers. Furthermore, from this combination one can obtain Mark-Houwink parameters and the bulk intrinsic viscosity of a given polymer with a GPC calibration curve based only on polystyrene standards. Coupling these two detectors with ultraviolet and infrared detectors then will permit the concurrent determination of polymer composition as a function of molecular weight and... [Pg.294]

This short overview illustrates the large complexity of the SEC processes and explains the absence of a quantitative theory, which would a priori express dependence between pore size distribution of the column packing—determined for example by mercury porosimetry—and distribution constant K in Equation 16.4. Therefore SEC is not an absolute method. The SEC columns must be either calibrated or the molar mass of polymer species in the column effluent continuously monitored (Section 16.9.1). [Pg.471]

In conventional electrochemistry in solution, quantitation of analytes can be obtained by using several techniques. Thus, exhaustive electrolysis provides an absolute quantitation of an electroactive component in the sample. Voltammetric measurements (linear potential scan, cyclic, pulse, and square-wave techniques) can be used for determination of analytes in solution via calibration because peak currents (and peak areas) are usually proportional to the analyte concentration under fixed electrochemical conditions. [Pg.95]

In summary, the overall rate of reductive dehalogenation of a given compound in a given system may be determined by various rather complex steps, and may, therefore, be influenced by several compound properties. Furthermore, even within a series of structurally related compounds, the relative importance of the various steps may differ, thus rendering any quantitative structure reactivity relationships (QSARs) rather difficult. This also means that calibration of a given system with a small set of model compounds for estimating absolute reaction rates will be even more difficult as compared to the situation with NAC reduction (see above). [Pg.595]

The accurate determination of incident light intensity is of pivotal importance in any quantitative photochemical experiment. While various physical devices are available for making absolute intensity measurements,168 these devices can be difficult to calibrate and usually are rather expensive. A much simpler approach involves the use of a chemical actinometer. This type of system is based upon a photochemical reaction for which product quantum yields are reasonably insensitive to variations in reactant concentration, temperature, light intensity and excitation wavelength. Once the quantum yield is calibrated by an absolute method, a chemical actinometer becomes a rapid, inexpensive and highly accurate secondary standard for light intensity measurements. [Pg.409]

Other factors, such as mechanical clamping, damping in the electrical circuit, and temperature also affect the absolute accuracy. For this reason it is necessary to use calibration curves for quantitative work. In spite of these limitations, the quartz microbalance is an extremely sensitive and versatile sensor. [Pg.70]


See other pages where Quantitation absolute calibration is mentioned: [Pg.16]    [Pg.314]    [Pg.445]    [Pg.354]    [Pg.342]    [Pg.449]    [Pg.311]    [Pg.102]    [Pg.360]    [Pg.833]    [Pg.145]    [Pg.290]    [Pg.589]    [Pg.62]    [Pg.162]    [Pg.405]    [Pg.40]    [Pg.108]    [Pg.503]    [Pg.126]    [Pg.48]    [Pg.232]    [Pg.3]    [Pg.18]    [Pg.186]    [Pg.118]    [Pg.4]    [Pg.166]    [Pg.390]    [Pg.258]    [Pg.102]   
See also in sourсe #XX -- [ Pg.44 ]




SEARCH



Absolute calibration

Absolute quantitation

© 2024 chempedia.info