Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Graph calibration

The calibration graph for the probe using a strength machine, has been shown in Fig. 7 It can be observed that the dependence of indications of the device of Wirotest type on the loading is linear within the proportionality limit scope. After unloading the indications do not return to zero, but show own stress caused in effect of plastic deformation of the tested sample... [Pg.387]

PVSA-SG film was used for determination of Fe(Phen) + and Zn + as ternary complex Zn +-Phen-bengal rose by spectrophotometric method. The calibration graph was linear in the concentration 5T0 -5T0 mol/lfor Fe(II) and FlO - 5T0 mol/1 for Zn(II). The film can be regenerated and reused. LG-PDMDA-SG film was shown to be perspective modificator of the PG electrode surface and used for voltammetric detection of Mo(VI) at ppb level. [Pg.306]

It was shown that Zn + adsorbed onto SG-PVSA composite film as Zn(Phen) complex. It can be detected spectrophotometrically after treatment with anionic dye Bengal Rose (BR). Ternary complex Zn + - Phen-BR formed on the surface under optimal conditions. SG-PVSA film was used for determination of Zn + by spectrophotometric method. The calibration graph was linear in the concentration range 2,5T0 - STO mol/l. [Pg.317]

Following this procedure urea can be determined with a linear calibration graph from 0.143 p.g-ml To 1.43 p.g-ml and a detection limit of 0.04 p.g-ml based on 3o criterion. Results show precision, as well as a satisfactory analytical recovery. The selectivity of the kinetic method itself is improved due to the great specificity that urease has for urea. There were no significant interferences in urea determination among the various substances tested. Method was applied for the determination of urea in semm. [Pg.371]

Internal standard of Rh with 2 ppb concentration was added to all the solutions. Three certified Multi-element solutions (CLMS-1,-2,-4, SPEX, USA) were employed for constmcting calibration graphs. [Pg.454]

FIGURE 22.12 Theoretical HdC/SEC calibration graphs for different particle diameters. Pore diameter = 10 nm particle diameters dotted line, S /xm dashed line, 3 /xm solid line, I /xm. (Reprinted from J. Chromatogr., 550, 728, Copyright 1991, with permission from Elsevier Science.)... [Pg.608]

Once a linear relationship has been shown to have a high probability by the value of the correlation coefficient (r), then the best straight line through the data points has to be estimated. This can often be done by visual inspection of the calibration graph but in many cases it is far better practice to evaluate the best straight line by linear regression (the method of least squares). [Pg.145]

Calculate the amount of boron present by reference to a calibration graph of absorbance against boron concentration (mg L 1). Multiply the result obtained by the appropriate volume correction factor arising from neutralisation of the sample. [Pg.177]

Calibration. Take 5, 10, 25, 50, 75 and lOOmL of the standard boric acid solution (2.5 x 10 4M) and make each up to lOOmL with distilled water this yields a boron concentration range up to 2.70mgL 1. Continue with each solution as described under procedure (b), i.e. one-hour reaction time, except that the initial neutralisation of the boron solution to pH 5.5 is not necessary. Construct a calibration graph of absorbance at 516 nm against boron concentration, mg L 1. For maximum accuracy, the calibration should be carried out immediately prior to the analysis of samples. [Pg.177]

Determine aluminium (present as its acetylacetonate) in the sample solution by injecting 0.30 L into the column. Record the peak area obtained and read off the aluminium concentration from the calibration graph (see Note). [Pg.249]

The following procedure has been recommended by the Analytical Methods Committee of the Society for Analytical Chemistry for the determination of small amounts of arsenic in organic matter.20 Organic matter is destroyed by wet oxidation, and the arsenic, after extraction with diethylammonium diethyldithiocarbamate in chloroform, is converted into the arsenomolybdate complex the latter is reduced by means of hydrazinium sulphate to a molybdenum blue complex and determined spectrophotometrically at 840 nm and referred to a calibration graph in the usual manner. [Pg.683]

Determine the phosphate content of an unknown solution, containing say ca 0.005 mg P205 per mL use the calibration graph. [Pg.730]

Procedure. Measure the fluorescence intensities of each of the series of standard solutions at 345 nm, with excitation at 285 nm. Construct a calibration graph for each of the four series (a), (b), (c), and (d) above. [Pg.740]

Calibration graph (d) may be used to correct for the small fluorescence intensity due to morphine in NaOH this is not negligible when the morphine concentration is high and the codeine concentration is low. [Pg.740]

A) The use of a calibration graph. This overcomes any problems created due to non-linear absorbance/concentration features and means that any unknown concentration run under the same conditions as the series of standards can be determined directly from the graph. The procedure requires that all standards and samples are measured in the same fixed-path-length cell, although the dimensions of the cell and the molar absorption coefficient for the chosen absorption band are not needed as these are constant throughout all the measurements. [Pg.753]

Take some crude cresol mixture (1 g) and dissolve it in cyclohexane (20 mL). Obtain the infrared spectrum for the mixture if necessary, dilute the solution further with cyclohexane to obtain absorbances which will lie on the calibration graphs. From the selected absorption peaks calculate the absorbances for the three individual isomers and use the calibration graphs to calculate the percentage composition of the cresol mixture. [Pg.757]

Take 10 mL of commercial propan-2-ol and dilute to 100 mL with carbon tetrachloride in a graduated flask. Record the infrared spectrum and calculate the absorbance for the peak at 1718 cm-1. Obtain a value for the acetone concentration from the calibration graph. The true value for the acetone in the propan-2-ol will be 10 times the figure obtained from the graph (this allows for the dilution) and the percentage v/v value can be converted to a molar concentration (mol L-1) by dividing the percentage v/v by 7.326 e.g. 1.25 per cent v/v = 1.25/7.326 = 0.171 molL-1. [Pg.757]

Procedure (ii). Make certain that the instrument is fitted with the correct burner for an acetylene-nitrous oxide flame, then set the instrument up with the calcium hollow cathode lamp, select the resonance line of wavelength 422.7 nm, and adjust the gas controls as specified in the instrument manual to give a fuel-rich flame. Take measurements with the blank, and the standard solutions, and with the test solution, all of which contain the ionisation buffer the need, mentioned under procedure (i), for adequate treatment with de-ionised water after each measurement applies with equal force in this case. Plot the calibration graph and ascertain the concentration of the unknown solution. [Pg.807]

It is important for obtaining precise results that the signals from the samples to be determined should lie on the linear part of the calibration graph as elsewhere within the dynamic range a small change in detector response corresponds to a relatively large range of concentrations. [Pg.42]

In many cases when methods involve internal or external standards, the solutions used to construct the calibration graph are made up in pure solvents and the signal intensities obtained will not reflect any interaction of the analyte and internal standard with the matrix found in unknown samples or the effect that the matrix may have on the performance of the mass spectrometer. One way of overcoming this is to make up the calibration standards in solutions thought to reflect the matrix in which the samples are found. The major limitation of this is that the composition of the matrix may well vary widely and there can be no guarantee that the matrix effects found in the sample to be determined are identical to those in the calibration standards. [Pg.270]

In the author s experience, such confirmation is not appropriate when the calibration range is greater than one order of magniffide or calibration points are not chosen carefully. The reason is that lower concentration levels of a calibration graph influence the correlation coefficient to a much smaller extent than higher concentrations. The hypothetical example of calibration results presented in Table 3 demonstrates this very simply. If the amount injected is correlated with the observed peak area in the second column in Table 3, the calibration graph in Figure 2 is obtained. [Pg.103]

Consequently, the proof of calibration should never be limited to the presentation of a calibration graph and confirmed by the calculation of the correlation coefficient. When raw calibration data are not presented in such a situation, most often a validation study cannot be evaluated. Once again it should be noted that nonlinearity is not a problem. It is not necessary to work within the linear range only. Any other calibration function can be accepted if it is a continuous function. [Pg.104]

The highly dispersible calix[4]arene neutral carriers can also improve the durability for neutral-carrier-type ion sensors. Time-course changes in both sensitivity (slope for Na+ calibration graph) and selectivity (selectivity coefficient for Na" " with respect to K+) were followed in the Na -ISFETs based on ion-sensing membranes of silicone rubber-(l), plasticized PVC-(l), and plasticized PVC-(2) (Fig. 2). Deterioration proceeded quite quickly in the Na -ISFETs of plasticized PVC-(2) both the Na+ sensitivity and selectivity... [Pg.589]

In isotope dilution inductively coupled plasma-mass spectrometry (ID-ICP-MS) the spike, the unspiked and a spiked sample are measured by ICP-MS in order to determine the isotope ratio. Using this technique, more precise and accurate results can be obtained than by using a calibration graph or by standard addition. This is due to elimination of various systematic errors. Isotopes behave identically in most chemical and physical processes. Signal suppression and enhancement due to the matrix in ICP-MS affects both isotopes equally. The same holds for most long-term instrumental fluctuations and drift. Accuracy and precision obtained with ID-ICP-QMS are better than with other ICP-QMS calibration... [Pg.660]

For their current density technique at the dme, Kies and Van Dam82c reported a standard deviation for a single solution (8 10 5 AfTlN03), based on 11 recordings, of 0.5% and rectilinear calibration graphs in the range 20-1000 nM. [Pg.191]

Wienke D, Lucasius C, Ehrlich M, Kateman G (1993) Multicriteria target vector optimization of analytical procedures using a genetic algorithm. Part II. Polyoptimization of the photometric calibration graph of dry glucose sensors for quantitative clinical analysis. Anal Chim Acta 271 253... [Pg.148]

Erk [20] described a spectrophotometric method for the simultaneous determination of metronidazole and miconazole nitrate in ovules. Five capsules were melted together in a steam bath, the product was cooled and weighed, and the equivalent of one capsule was dissolved to 100 mL in methanol this solution was then diluted 500-fold with methanol. In the first method, the two drugs were determined from their measure d%/dk values at 328.6 and 230.8 nm, respectively, in the first derivative spectrum. The calibration graphs were linear for 6.2—17.5 pg/mL of metronidazole and 0.7—13.5 pg/mL of miconazole nitrate. In the second (absorbance ratio) method, the absorbance was measured at 310.4 nm for metronidazole, at 272 nm for miconazole nitrate and at 280.6 nm (isoabsorptive point). The calibration graphs were linear over the same ranges as in the first method. [Pg.39]


See other pages where Graph calibration is mentioned: [Pg.87]    [Pg.128]    [Pg.215]    [Pg.240]    [Pg.240]    [Pg.570]    [Pg.740]    [Pg.752]    [Pg.753]    [Pg.754]    [Pg.756]    [Pg.43]    [Pg.44]    [Pg.275]    [Pg.286]    [Pg.359]    [Pg.24]    [Pg.605]    [Pg.605]    [Pg.332]    [Pg.661]    [Pg.337]    [Pg.39]    [Pg.44]   
See also in sourсe #XX -- [ Pg.86 , Pg.105 , Pg.111 ]

See also in sourсe #XX -- [ Pg.30 , Pg.54 ]

See also in sourсe #XX -- [ Pg.527 ]

See also in sourсe #XX -- [ Pg.75 ]

See also in sourсe #XX -- [ Pg.171 , Pg.177 , Pg.180 ]

See also in sourсe #XX -- [ Pg.292 , Pg.295 ]

See also in sourсe #XX -- [ Pg.219 ]

See also in sourсe #XX -- [ Pg.527 ]

See also in sourсe #XX -- [ Pg.45 , Pg.47 , Pg.48 , Pg.49 , Pg.50 , Pg.73 , Pg.75 , Pg.79 , Pg.80 , Pg.81 , Pg.82 , Pg.106 , Pg.107 , Pg.110 , Pg.111 , Pg.141 , Pg.142 , Pg.242 ]

See also in sourсe #XX -- [ Pg.261 ]

See also in sourсe #XX -- [ Pg.163 ]




SEARCH



Amount intervals calibration graphs

Analytical Calibration graph

Calibration Graph Method

Calibration graph errors

Calibration graphs in instrumental analysis

Calibration graphs standard addition method used

Calibration graphs, nonlinear

Estimated amount interval calibration graphs

Linear regression analysis, calibration graphs

Linearity, calibration graphs

Phenol calibration graph

Quantitative Analysis Using Calibration Graph Method

Quantitative analysis calibration graph method

Regression analysis, calibration graphs

Standard addition method calibration graphs using

Temperature calibration graph

© 2024 chempedia.info