Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Preparation measurements

M. Wen2el and P. E. Schul2e, Tritium Markierung, Preparation, Measurement and Uses ofWil ach Uahelled Compounds, Berlin, Walter de Gru2ter, 1962. [Pg.17]

Fig. 12a—c NOE experiments carried out at 200 MHz on compound 3. a Normal spectrum, with expansion of methine doublet b selective NOE spectrum, total time required 18 min c NOE difference spectrum, total time required (preparation, measurement) 42 min... [Pg.18]

Chromatographic system. The gas chromatograph is equipped with a flamioniza-tion detector and a 2 mm x 1.8 m glass column packed with 10% phase G34 on 80- to 100-mesh support SI A. The column temperature is maintained at about 150 °C, and the injection port and the detector block temperatures are maintained at about 250 °C. Dry helium is used as the carrier gas at a flow rate of about 40 mL/min. Chromatograph the Standard preparation, measure the peak responses, and calculate the ratio, Rs, as directed for procedure the relative retention times are about 0.5 for valproic acid and 1.0 for biphenyl the resolution, R, between valproic acid and biphenyl is not less than 3.0 the relative standard deviation for replicate injections is not more than 2.0%. [Pg.227]

Normal fire protection practices and standards recommend fire pumps be tested annually to determine performance levels. Common practice in the petroleum industry is to trend the result of flow performance to prepare predictive maintenance and replacement forecasts. Such forecast can predict poor pump performance and prepare measures to implement corrective actions before this occurs. [Pg.250]

As indicated in Section 4.1 (and as should be apparent from the discussion thus far in this chapter), titrimetric analysis methods heavily utilize solution chemistry, and therefore volumes of solutions are prepared, measured, transferred, and analyzed with some degree of frequency in this type of analysis. It should not be surprising that analytical laboratory workers need to be well versed in the selection and proper use of the glassware and devices used for precise volume measurement. [Pg.79]

The author is especially grateful to Robert Evans for making numerous preparations, measurements and related studies. [Pg.217]

Many methods have been used to produce and characterize protein foams. The foaming characteristics of proteins are markedly influenced by conditions of preparation, measurement, and so forth (13-18). Because of the variety of methods employed, it is difficult to compare data from different sources. [Pg.154]

ZPC data were selected according to the criteria used to study simple oxides and hydroxides previously (72). The data are summarized in Table III. Insofar as possible, details of sample preparation, measurement technique, and environmental conditions are given. It is assumed that phase identification was correct. [Pg.148]

Fig. 15.1. LCM concentrations during Filmix suspension preparation measured using optical particle counters. (Taken from ref. 717. See text for further discussion.)... Fig. 15.1. LCM concentrations during Filmix suspension preparation measured using optical particle counters. (Taken from ref. 717. See text for further discussion.)...
Pietra et al. (1990) studied the effects of some antidepressants by the in vitro inhibition of carbachol-induced contractions of rat detrusor strip preparations. The detrusor muscle tissue (bladder dome) was cut in a semicircular direction and further dissected into strip preparations measuring approximately 2 x 20 mm. [Pg.137]

The same technique is applied for the determination of the pressure in the combustion chamber. The pressure cell must be attached to the previously prepared measuring points on the combustion chamber. [Pg.400]

Similarly, inject 1.0 xL of Sample Preparation. Measure its response ratio, and determine from the standard curve the micrograms of DCP in the sample taken. [Pg.14]

Procedure Use a gas chromatograph equipped with a flame-ionization detector, a cold-on-column injector, a suitable deactivated precolumn, and a 10-m x 0.32-mm (id) capillary column coated with an apolar stationary phase 0.12-p.L film thickness. Program the column to heat to 60°, hold for 1 min, heat to 300° at 20°/min, and hold for 3 min. Set the flame-ionization detector to 320°. Chromatograph five injections of the Standard Preparation. Measure the responses. The relative standard deviation for each peak should be below 2%. The peaks for brassicasterol and campesterol should be baseline resolved (Rs > 1.0) and show no tailing. Measure the response of the Internal Standard Solution and all the individual sterols eluting in the relative retention window of 0.98 to 1.13. [Pg.493]

Sample Preparation Measure 50 mg of sample into a 50-mL volumetric flask. Dissolve the sample, with stirring, in approximately 25 mL of Sodium Phosphate Buffer Solution. Dilute to volume with Sodium Phosphate Buffer Solution, and mix the solution thoroughly. Quantitatively transfer 3 mL of the solution to a 100-mL volumetric flask, and dilute to volume with Sodium Phosphate Buffer Solution. [Pg.916]

M. Wenzel andP. E. Schulze, Tritium Markierung Preparation Measurement and Uses of Wilzbach Labelled Compounds. De Gruyter, Berlin, 1962. [Pg.141]

For analytical sample preparation, measurement of the final pH of the sample solution (excipients, API(s), and sample solvent mixed together) will be helpful in the development of any analytical procedure. If an API is known to be stable in acidic pH (pH 1-2), then an analytical chemist will try to utilize a certain sample solvent that has a pH in the required range. However, when a dosage form is dissolved in a sample solvent, the excipients present in the formulation (and even the API) will change the pH of the solution. The final pH of the solution must be measured in order to determine the optimal pH of sample solution to achieve longest solution stability. This is particularly important for a long sequence of injections on autosamplers for analysis, so solutions do not need to be made daily. [Pg.683]

W. J. Whitehouse, J. L. Putman, Radioactive Isotopes - An Introduction to their Preparation, Measurement and Use, Clarendon Press, Oxford, 1953 H. A. C. McKay, Principles of Radiochemistry, Butterworths, London, 1971 International Atomic Energy Agency, Radioisotope Production and Quality Control, IAEA Technical Reports Series No. 128, Vienna, 1971... [Pg.261]

UV photometer UV photometer ca. 6700 Glass bottles, filters Preparation, measurement and sampling 20 min. [Pg.390]

Sample preparation Measure out S3Tup equivalent to about 5 mg hydrocodone bitartrate, add 5 mL water, add 1 mL 1.8 M sulfuric acid, wash twice with 40 mL portions of chloroform. Make the aqueous layer alkaline with 5 mL 1 M NaOH, extract twice with 25... [Pg.699]


See other pages where Preparation measurements is mentioned: [Pg.292]    [Pg.65]    [Pg.423]    [Pg.561]    [Pg.68]    [Pg.1153]    [Pg.71]    [Pg.49]    [Pg.156]    [Pg.17]    [Pg.420]    [Pg.224]    [Pg.68]    [Pg.2182]    [Pg.453]    [Pg.671]    [Pg.22]    [Pg.655]    [Pg.19]    [Pg.21]    [Pg.142]    [Pg.37]    [Pg.43]    [Pg.397]    [Pg.377]    [Pg.189]    [Pg.435]    [Pg.144]   
See also in sourсe #XX -- [ Pg.156 ]




SEARCH



Cell preparation measurements

Compressive measurement sample preparation

Conditional measurement / preparation

Diffuse-reflection measurements sample-preparation methods

Electrophoresis and Sample Preparation for SRXRF Measurement

Fluorescence measurements preparation

Measurement error, preparative chromatography

Measurement sample preparation

Measuring preparations

Measuring preparations

Particle size, measurement sample preparation

Sample Preparation and Measurement

Sample preparation, generally measurement process

Sample-preparation methods measurements

Single Crystal Preparation for In Situ STM Measurements

Some Remarks on Sample Preparations for Voltammetric Measurements

Tissue preparation measurement

© 2024 chempedia.info