Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Response ratios

The detector response ratio is elevated with respect to the homopolymer line but very near parallel to it less than = 0.74,or number average styrene sequence lengths less... [Pg.165]

Table 13 Percentage of response ratios above 2 (response in matrix extract divided by the response in solvent data from about 40 GC instruments)... Table 13 Percentage of response ratios above 2 (response in matrix extract divided by the response in solvent data from about 40 GC instruments)...
Instrumental response ratios (tebuconazole/i NsJtebuconazole) versus concentration of tebuconazole present should be proven in solvent and each matrix analyzed up to the highest undiluted final sample extract concentration expected. Once proven linear, final sample extract residues found to lie above the range of linearity are to be diluted prior to addition of IS solution and re-analyzed. [Pg.1239]

Since IS solution is added to the sample at a point just after extraction, no addition of IS to the final sample extract is required. The concentration (mg kg ) of tebuconazole is calculated by applying the peak-area ratio (tebuconazole/l Nsj tebuconazole) with a calibration curve analyzed throughout the analytical sequence. All sample extract response ratios must fall within the limits of the calibration curve injected over the course of sample analyses. Samples showing a response ratio above the limit of the coanalyzed calibration curve must be re-analyzed with a new curve including standards having higher concentrations of tebuconazole that bracket that of the sample. [Pg.1239]

R = (response ratio sample)/(av. standard response ratio) x std cone, x dil. factor where... [Pg.1240]

Detectors are usually conpued in terns of their operational characteristics defined by the nininvin detectable quantity of standards, the selectivity response ratio between standards of different conpositlon or structure, and the range of the linear portion of the detector-response calibration curve. These terns are wid. y used to neasure the perfomance of different chronatographic detectors and were fomally defined in section 1.8.1. [Pg.135]

A compound or element added to all calibration standards and samples in a constant known amount. Sometimes a major constituent of the samples to be analysed can be used for this purpose. Instead of preparing a conventional calibration curve of instrument response as a function of analyte mass, volume or concentration, a response ratio is computed for each calibration standard and sample, i.e. the instrument response for the analyte is divided by the corresponding response for the fixed amount of added internal standard. Ideally, the latter will be the same for each pair of measurements but variations in experimental conditions may alter the responses of both analyte and internal standard. However, their ratio should be unaffected and should therefore be a more reliable function of... [Pg.620]

A comparative study of the res-ponce of neomycin B and neomycin C in the turbidi-metric assay using K. pneumo n-tae. (ATCC 10031) has been reported239. The response-ratio neomycin B neomycin C was 100 38.9. [Pg.468]

The explanation of this result is illustrated in Fig. 1.37, which shows selected reaction-monitoring traces of the sample at 10 ng mL f It becomes obvious that the response ratio between the analyte and the IS has dramatically changed. On one side there is enhancement of the analyte s response and on the other side suppression of the internal standard (IS) signal. These effects are mainly caused... [Pg.43]

Fig. 3.10 Examples of isosteric binding competition. (A) ALIS-MS results for the titration of 5 pM Zap-70 by staurosporine in the presence of a 5 m concentration of its structural congener K252a and (B) titration of 5 pM DHFR with the known DHFR inhibitor trimethoprim in the presence of ligand NCD-157 at 5 pm concentration. Linear MS response ratios in these experiments are consistent with direct binding competition. (C) Compound structures. Fig. 3.10 Examples of isosteric binding competition. (A) ALIS-MS results for the titration of 5 pM Zap-70 by staurosporine in the presence of a 5 m concentration of its structural congener K252a and (B) titration of 5 pM DHFR with the known DHFR inhibitor trimethoprim in the presence of ligand NCD-157 at 5 pm concentration. Linear MS response ratios in these experiments are consistent with direct binding competition. (C) Compound structures.
Fig. 3.11 Examples of allosteric binding competition. Titration of 5 pM Akt-1 plus 5 pM staurosporine by (A) NGD-28835 and (B) Merck-1 does not yield linear response ratios for the two competing ligands. Asymptotically bound response ratios indicate allosteric binding between these two... Fig. 3.11 Examples of allosteric binding competition. Titration of 5 pM Akt-1 plus 5 pM staurosporine by (A) NGD-28835 and (B) Merck-1 does not yield linear response ratios for the two competing ligands. Asymptotically bound response ratios indicate allosteric binding between these two...
It is also noteworthy that the ACE50 technique for affinity ranking also allows mixture components to be classified as either allosteric or directly competitive with another ligand of interest. In the M2 example, reanalyzing the sigmoidal AGE50 curves in Pig. 3.13 as the ratio plots instead shows that the response ratios... [Pg.141]

It was then possible to screen some possible candidates and come up with the probable presence of a chlorinated paraffin. This material was found to elute at the same retention volume as di-2-ethylhexyl-phtha-late and showed no response at 254 nm (Figure 6). For quantitation UV/RI response ratio provided all the data required. However, in order to confirm the presence of this additional component, the sample was hydrolyzed and the acid converted to its dimethyl ester. The products were then examined by GPC. The UV response for the peak of interest was completely... [Pg.249]

Quinones and Hydroqulnones. In the analysis of quin-ones and hydroqulnones, the use of two different dual detector systems was required. The retention data for hydroqulnones shows the normal behavior of hydroxyl groups associating with the solvent, THF. Thus octyl quinone and hydroquinone elute almost together. Similarly dioctylquinone and octyl hydroquinone elute together (Figure 7). The UV/RI response ratio for benzoquinone is 3.75. Hydroquinone and dioctylquinone show similar disparities in the UV/RI responses. This information provides a very good method for detecting impurities in dioctyl hydroquinone. [Pg.252]

To understand the characteristics of the MSG response over the known mineralization zone, Lines 20, 28, 32, 40 and 48 were sampled in 2007. MSG signatures vary greatly with time, climate, devices and sampling team. Response ratios are calculated by dividing each sample value by the predetermined background value for that element under the same sampling conditions to reduce these errors. [Pg.44]

The uncertainties of K and the response ratio depend upon the ability to measure detector response. Peak height ratios can be used with excellent results (generally better than areas) when the peaks are symmetrical and sharp. Peak hight ratios are also more useful than peak areas in overlapping peaks. For automated systems, peak areas are preferred since the ratios are readily measured, and the data are calculated with electronic integrators and computers. [Pg.72]

Table B1.1.5 demonstrates the relative degree of protein-to-protein variation that can be expected with the different protein assay methods. This differential may be a consideration in selecting a protein assay method, especially if the relative color response ratio of the protein in the samples is unknown. As expected, the protein assay methods that share the same basic chemistry show similar protein-to-protein variation. Table B1.1.5 demonstrates the relative degree of protein-to-protein variation that can be expected with the different protein assay methods. This differential may be a consideration in selecting a protein assay method, especially if the relative color response ratio of the protein in the samples is unknown. As expected, the protein assay methods that share the same basic chemistry show similar protein-to-protein variation.
Plus Protein Assay Reagent was used, the total protein (IgG) concentration in the sample would be underestimated by -40%. (From Table B1.1. 5, the response ratio for IgG is -0.58 for IgG compared to 1.00 for BSA.) If the BCA Protein Assay Reagent was used, the total protein (IgG) concentration in the sample would be overestimated by -15%. (From Table B1.1.5, the response ratio for IgG is -1.15 for IgG compared to 1.00 for BSA.) On the other hand, if BGG had been used for both standard curves, the total protein estimates for the sample would have been in much greater agreement between the two methods. [Pg.99]

While Table B1.1.5 is useful because it provides an estimate of the protein-to-protein variation in color response that can be expected with each method, it does not tell the whole story. Because the comparisons were done at a single protein concentration, it is not apparent that the color response ratio also varies with changes in protein concentration. [Pg.99]

Subject to silylation (steps 15 to 17) and analyze (steps 18 to 20). Analyze data 23. Divide the area ratio of cholesterol in the samples by that of the internal standard to obtain a standard response ratio. 24. Plot the average response ratio against the ratio of cholesterol to internal standard in the standards. [Pg.456]

This same qualitative difference between adrenaline and noradrenaline obtains for the splenic artery/splenic vein equipressor dosage-response ratios as well, and both observations quite possibly may find their explanation in the recent work of Euler (64-70). He has found that the pressor substance isolated from the heart, blood, liver, and spleen has predominantly the characteristics of noradrenaline. Thus, he has considered Sympathin E to be identical with Z-noradrenaline. [Pg.41]

The composition of the hydrolysate will affect growth rate, specific substrate uptake rates, and yield coefficients in a not easily predictable way (14). It is thus not possible to use a predetermined feed profile, but instead a closed-loop procedure should be used. One approach is to maximize the productivity at every point in time, and this can relatively easily be implemented in a closed-loop control. This approach was used by Taherzadeh et al. (14), who applied step changes in the feed rate and gradually increased the feed rate as long as there was a relative response in CER corresponding to at least include 50% of the relative increase in feed rate. The major drawbacks of this control were that the increase in feed rate was fixed and that the CER-response ratio was somewhat sensitive to measurement errors. [Pg.606]

For the internal standard method, a substance is added at the earliest possible point in the analytical scheme. This compensates for sample losses during extraction, cleanup, and final chromatographic analysis. There are two variations in the use of the internal standard technique. One involves the determination of response factors which are the ratios of the analyte peak response to the internal standard peak response. The second is referred to as response ratios which are calculated by dividing the weight of the analyte by the corresponding peak response. [Pg.10]


See other pages where Response ratios is mentioned: [Pg.128]    [Pg.517]    [Pg.1239]    [Pg.1286]    [Pg.148]    [Pg.669]    [Pg.878]    [Pg.248]    [Pg.35]    [Pg.37]    [Pg.227]    [Pg.228]    [Pg.621]    [Pg.218]    [Pg.137]    [Pg.138]    [Pg.45]    [Pg.271]    [Pg.86]    [Pg.228]    [Pg.1284]    [Pg.631]    [Pg.37]    [Pg.38]    [Pg.403]   
See also in sourсe #XX -- [ Pg.47 ]




SEARCH



Frequency response experimental amplitude ratio

Saturates aromatics response ratio

© 2024 chempedia.info