Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Optical parameter instrumentation

The separation of absorption and scattering often requires specific types of instrumentation, based on the theoretical approach that is used. In this section, four main designs will be discussed integrating sphere-based reflectance and transmittance, spatially resolved spectroscopy, time-resolved spectroscopy, and frequency domain photon migration. [Pg.27]

Spatially resolved spectroscopy consists in measuring the light reflected at different distances from the incident light. Typical designs use a set of aligned optical fibers, with one fiber being the source and others the detection systems. Setups based on hyperspectral imaging have [Pg.27]

While these instrument designs may appear most suitable for laboratory analysis, some manufacturers have proposed apparatus suitable for real-time online process monitoring. It is the case for spatially resolved spectroscopy. A very good review on the topic on pharmaceutical applications to the separation of absorption and scattering was written by Shi and Anderson [34]. A comprehensive review of those techniques and applications in tissue optics can be found in a review by Wilson [35]. [Pg.28]


Microscopy. A broad definition of microscopy is the observation and measurement of optical parameters with any instrument that uses energy sources such as photons, electrons or X-rays to... [Pg.143]

A Malvern Mastersizer (Malvern Instruments Ltd, Malvern, UK) with optical parameters defined by the manufacturer s presentation code 0505 was used to determine the droplet size distribution. The measurement was made in triplicate at room temperature. Water was used to disperse the emulsion droplets. [Pg.273]

The lifetime detection techniques are self-referenced in a sense that fluorescence decay is one of the characteristics of the emitter and of its environment and does not depend upon its concentration. Moreover, the results are not sensitive to optical parameters of the instrument, so that the attenuation of the signal in the optical path does not distort it. The light scattering produces also much lesser problems, since the scattered light decays on a very fast time scale and does not interfere with fluorescence decay observed at longer times. [Pg.12]

For accurate measurements it is important to enter the correct optical constants of the dispersed and continuous phase into the theory used to calculate the droplet size distribution. In most instruments, it is necessary to enter the refractive index and absorptivity of the component phases at the appropriate wavelength of the laser. Significant errors in the measured droplet size distribution will occur if incorrect optical parameters are used, particularly if the... [Pg.586]

Microscopy. A broad definition of microscopy is the observation and measurement of optical parameters with any instrument that uses energy sources such as photons, electrons or X-rays to provide an enlarged image of an object. Energetic material parameters that have been observed and measured include quantity, size, shape and color (Expls Refs 12-15, 19, 20, 25, 25a, 27, 28,... [Pg.144]

The correction for the refractive indices of the sample and reference solutions, (s) and ra(r), allows for the variation in the angles of the luminescence emerging from the solution to air. The correction may be substantial, for example D2(benzene)/ D2(water) 1.27, and may depend somewhat on optical parameters of the instrument. Moreover, internal reflections within the cell also depend on the refractive index. It is therefore preferable to dissolve both sample and reference in the same solvent to avoid errors from these sources. [Pg.119]

In conformity with all ellipsometers, a DOAP detects the change in polarization of an initially polarized laser beam (of well-defined and known polarization) after reflection from the surface of interest, by dividing the total signal into four polarization components. After an elaborate calibration routine [99], the optical constants of Ae sample were obtained from the Fresnel equations, provided that die angle of incidence and the refractive index of the ambient medium, e.g. air, were known. The whole instrument was originally designed to measure emittance and the optic parameters result as a by-product fi om this process, see e.g., [100] obtained with a / -DOAP. [Pg.321]

The thermal system includes the Warm Optics Module and the Cold Optics Module, which given the optical set-up and the optical parameters of the different optical elements calculates the transmission of the sky map through the instmment. At this point the physical properties of the instrument are defined and the Double Fourier Modulation can be performed at the Double Fourier Module. Here is where the interferograms are computed analytically for different baseline positions. If pointing errors are selected, the Pointing Errors Module generates them and they are fed to the Double Fourier Moduie. [Pg.75]

It would be of obvious interest to have a theoretically underpinned function that describes the observed frequency distribution shown in Fig. 1.9. A number of such distributions (symmetrical or skewed) are described in the statistical literature in full mathematical detail apart from the normal- and the f-distributions, none is used in analytical chemistry except under very special circumstances, e.g. the Poisson and the binomial distributions. Instrumental methods of analysis that have Powjon-distributed noise are optical and mass spectroscopy, for instance. For an introduction to parameter estimation under conditions of linked mean and variance, see Ref. 41. [Pg.29]

Instrumental measurement of whiteness has been the subject of much research. The parameters needed for unambiguous characterisation in the assessment of whiteness and tint of fluorescent substrates have been reviewed [21]. The importance of seeking good correlation between different instruments is stressed [20]. Various trials have demonstrated that it is possible to adjust modern instruments used to measure the optical characteristics of FBA-treated samples of paper so that the results agree with a standard deviation of the order of one CIE whiteness unit [22]. [Pg.302]

Fig. 5 shows the instrumental arrangement of the commercially most successful optical chemical sensor between 1984 and 2000. It is used in about 70% of all critical care operations in the US to monitor pH, pC02 and p02 in the cardiopulmonary bypass operations35. It contains 3 fluorescent spots, each sensitive for one parameter, in contact with blood. Fluorescence intensity is measured at two wavelengths and the signals are then submitted to internal referencing and data processing. [Pg.36]

Figure 6.1. Mass spectra of synthetic peptide, FLFQPQRF-NH2. Both spectra were recorded on an ESI ion trap mass spectrometer at different instrument settings (modifications of ion optics and ion-trap parameters). Figure 6.1. Mass spectra of synthetic peptide, FLFQPQRF-NH2. Both spectra were recorded on an ESI ion trap mass spectrometer at different instrument settings (modifications of ion optics and ion-trap parameters).
These instruments feature keyboard entry of instrument parameters which combined with digital displays, simplifying instrument operation. A high-output pulsed xenon lamp, having low power consumption and minimal ozone production, is incorporated within the optical module. [Pg.29]

This is a very high-performance instrument in which instrument control resides in a multiprocessor system manager leaving only the analytically important parameters to be defined by the operator. It utilizes a completely new concept of ion optics for double focusing and this gives the instrument unmatched performance. [Pg.73]

Problems with the mechanics of a procedure can involve an improperly diluted sample (perhaps manifested by an absorbance reading that is greater than specified or expected), an obstruction in the sample or reference beam, an improperly aligned source or mirror, the incomplete programming of a scan, or improper or inappropriate software entry. In these cases, the operator will need to carefully examine his or her technique or procedure, or instrumental parameters, such as the optical path, perhaps with the help of the instrument troubleshooting guide, to solve the problem. [Pg.215]

The proportionality factor k depends on several parameters, in particular on the optical configuration for observation (i.e. the solid angle through which the instrument collects fluorescence, which is in fact emitted in all directions) and on the bandwidth of the monochromators (i.e. the entrance and exit widths see Chapter 5). [Pg.50]

Some of the typical parameters or properties utilized for NIR detection are potentiometry,(5) absorbance,(52 54) refractometry/18,19) or fluorescence spectros-copy.(55) Of these, has proven to be the most valuable detection method in fiber optic applications/2,56) In standard spectroscopic techniques, the detection limits of a method are greatly determined by the instrument and by the chemical method used for the analysis. However, in OFCD research the detection limits are governed by a series of other variables including the dye, the matrix, and the instrument. By optimizing these variables, low detection limits can be obtained with this technique. [Pg.191]

Instrumentation and methods currently available provide limited means for realtime measurements of the continuous wave (CW) and transient characteristics of luminescent substances. The measurement, in real time, of the spatial distribution of a parameter of interests, for instance, in cells and in tissue cannot be attained with present technology. Optical fibers are used to monitor the response of a sensor in limited regions in space. [Pg.255]


See other pages where Optical parameter instrumentation is mentioned: [Pg.27]    [Pg.27]    [Pg.6]    [Pg.427]    [Pg.88]    [Pg.79]    [Pg.9]    [Pg.414]    [Pg.75]    [Pg.100]    [Pg.307]    [Pg.48]    [Pg.4]    [Pg.24]    [Pg.353]    [Pg.244]    [Pg.233]    [Pg.343]    [Pg.81]    [Pg.492]    [Pg.513]    [Pg.95]    [Pg.101]    [Pg.51]    [Pg.425]    [Pg.112]    [Pg.64]    [Pg.524]    [Pg.447]    [Pg.157]    [Pg.195]    [Pg.309]   


SEARCH



Instrument optics

Instrument parameters

Optical Instruments

Optical parameter

© 2024 chempedia.info