Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Mass selection method

Using appropriate analytical methods, the type and concentration of the different functionality fractions must be determined and, within each functionality, the molar mass distribution has to be obtained. To do this, two different methods must be combined, each of which is preferably selective towards one type of heterogeneity. For example, a chromatographic method separating solely with respect to functionality could be combined with a molar-mass-selective method. Another approach would be the separation of the sample into different molar mass fractions which are then analyzed with respect to functionality. [Pg.7]

Fleming CM, et al. Windowed mass selection method A new data processing algorithm for liquid chromatography—mass spectrometry data. J Chromatogr A 1999 849 71-85. [Pg.717]

The running methods for liquid chromatography, as well as all current detection modes, are described in more detail in the following section and are summarized in Table 2, with the exception of mass selective methods, which are discussed later in this chapter. [Pg.254]

In TOF-SARS [9], a low-keV, monoenergetic, mass-selected, pulsed noble gas ion beam is focused onto a sample surface. The velocity distributions of scattered and recoiled particles are measured by standard TOF methods. A chaimel electron multiplier is used to detect fast (>800 eV) neutrals and ions. This type of detector has a small acceptance solid angle. A fixed angle is used between the pulsed ion beam and detector directions with respect to the sample as shown in figure Bl.23.4. The sample has to be rotated to measure ion scattering... [Pg.1805]

It is crucial in quantitative GC to obtain a good separation of the components of interest. Although this is not critical when a mass spectrometer is used as the detector (because ions for identification can be mass selected), it is nevertheless good practice. If the GC effluent is split between the mass spectrometer and FID detector, either detector can be used for quantitation. Because the response for any individual compound will differ, it is necessary to obtain relative response factors for those compounds for which quantitation is needed. Care should be taken to prevent contamination of the sample with the reference standards. This is a major source of error in trace quantitative analysis. To prevent such contamination, a method blank should be run, following all steps in the method of preparation of a sample except the addition of the sample. To ensure that there is no contamination or carryover in the GC column or the ion source, the method blank should be run prior to each sample. [Pg.215]

Several methods are available for the analysis of trichloroethylene in biological media. The method of choice depends on the nature of the sample matrix cost of analysis required precision, accuracy, and detection limit and turnaround time of the method. The main analytical method used to analyze for the presence of trichloroethylene and its metabolites, trichloroethanol and TCA, in biological samples is separation by gas chromatography (GC) combined with detection by mass spectrometry (MS) or electron capture detection (ECD). Trichloroethylene and/or its metabolites have been detected in exhaled air, blood, urine, breast milk, and tissues. Details on sample preparation, analytical method, and sensitivity and accuracy of selected methods are provided in Table 6-1. [Pg.229]

Conditions apparatus, Hewlett-Packard HP5890 equipped with an HP5972 mass-selective ion detector (quadruple) column, PTE-5 (30 m x 0.25-mm i.d.) with 0.25- am film thickness column temperature, 50 °C (1 min), increased at 20 °C min to 150 °C(5 min) and then at 4 °Cmin to 280 °C (30 min) inlet and detector (GC/MS transfer line) temperature, 250 and 280 °C, respectively gas flow rate, He carrier gas ImLmin" injection method, splitless mode solvent delay, 3 min electron ionization voltage, 70eV scan rate, 1.5 scanss scanned-mass range, m/z 50-550. The retention times of benfluralin, pendimethalin and trifluralin are 15.2, 25.1 and... [Pg.393]

There are methods available to quantify the total mass of americium in environmental samples. Mass spectrometric methods provide total mass measurements of americium isotopes (Dacheux and Aupiais 1997, 1998 Halverson 1984 Harvey et al. 1993) however, these detection methods have not gained the same popularity as is found for the radiochemical detection methods. This may relate to the higher purchase price of a MS system, the increased knowledge required to operate the equipment, and the selection by EPA of a-spectrometry for use in its standard analytical methods. Fluorimetric methods, which are commonly used to determine the total mass of uranium and curium in environmental samples, have limited utility to quantify americium, due to the low quantum yield of fluorescence for americium (Thouvenout et al. 1993). [Pg.213]

Undoubtedly, the technique most suited to tackle polyatomic multichannel reactions is the crossed molecular beam (CMB) scattering technique with mass spectrometric detection and time-of-flight (TOF) analysis. This technique, based on universal electron-impact (El) ionization coupled with a quadrupole mass filter for mass selection, has been central in the investigation of the dynamics of bimolecular reactions during the past 35 years.1,9-11 El ionization affords, in principle, a universal detection method for all possible reaction products of even a complex reaction exhibiting multiple reaction pathways. Although the technique is not usually able to provide state-resolved information, especially on a polyatomic... [Pg.331]

Greaves et al. [74] used a selected ion-monitoring assay method for the determination of primaquine in plasma and urine using gas chromatography-mass spectrometric method and a deuterated internal standard. After freeze-drying and extraction with trichloroethylene, the sample plus internal standard was reacted with Tri Sil TBT (a 3 3 2 by volume mixture of trimethylsilylimidazole, A/O-bis-(trimethylsilylacetamide and trimethylchlorosilane) and an aliquot injected to the gas chromatograph-mass spectrometer. The gas chromatographic effluent was monitored at m/z 403, and m/z 406, the molecular ions of the bis-tetramethylsilane ethers of primaquine and 6-trideuteromethoxy primaquine. [Pg.187]

High performance liquid chromatography-mass spectrometric methods Nitin et al. [75] developed and validated a sensitive and selective liquid chromatography-tandem mass spectrometric method (LC MS MS) for the simultaneous estimation of bulaquine and its metabolites primaquine in monkey plasma. The mobile phase consisted of acetonitrile ammonium acetate buffer (20 mM, pH 6) (50 50, v/v) at a flow rate of 1 mL/min. The chromatographic separations were achieved on two Spheri cyano columns (5 pm, 30 cm x 4.6 mm), connected in... [Pg.187]

Conductometric transducers, as the oldest electrochemical devices, seem not to enjoy wide applications due to their poor selectivity. For example, Yagiuda et al. proposed a conductometric immunosensor for the determination of methamphetamine (MA) in urine [89], The decrease in the conductivity between a pair of platinum electrodes might result from the direct attachment of MA onto the anti-MA antibodies immobilized on the electrode surface. The system was claimed to be a useful detection technique of MA in comparison with a gas chromatography-mass spectrometry method. [Pg.267]

A number of other laser spectroscopic techniques are of interest but space does not permit their discussion. A few specialized methods of detecting laser absorption worthy of mention include multiphoton ionization/mass spectrometry (28), which is extremely sensitive as well as mass selective for gas-phase systems optically detected magnetic resonance (29) laser intracavity absorption, which can be extremely sensitive and is applicable to gases or solutions (30) thermal blooming, which is also applicable to very weak absorbances in gases or liquids (31) and... [Pg.468]

Schmidt, L.J. and R.J. Hesselberg. 1992. A mass spectroscopic method for analysis of AHH-inducing and other polychlorinated biphenyl congeners and selected pesticides in fish. Arch. Environ. Contam. Toxicol. 23 37-44. [Pg.1336]

HPLC is a universal separation technique that is capable of separating both volatiles and non-volatiles without the need for derivatization. We are developing methods that employ both on-line photodiode array (PDA) detection and mass selective detection, HPLC/PDA/MS. This approach also utilizes an ion-trap mass... [Pg.41]

Tandem mass spectrometric methods have demonstrated superb specificity because of their ability to isolate analytes selectively in the presence of endogenous interferences. Attempts to further increase sample throughput led to the idea of using LC/MS/MS without the LC. Traditional chromatographic separations were replaced with flow injection analysis (FLA) or nanoelectrospray infusion techniques. The MS-based columnless methods attracted a lot of attention because of their inherent fast cycle times and no need for LC method development. [Pg.76]

One method of acquiring a mass spectrum is the mass selective instability scan. As the RF voltage increases, the ions with lowest m/z become unstable and are ejected through small holes in the end cap to hit a detector. As the RF voltage is further increased, heavier ions become successively unstable and are ejected, thus yielding a mass spectrum. [Pg.53]

To examine variation in the quality of methyl ketones expressed by females and newly-emerged males, we determined the number of unique methyl ketones expressed by individual snakes and compared the relative concentrations of individual methyl ketones comprising the overall pheromone profiles for the two groups. The methyl ketones present in the pheromone extracts were identified by gas chromatography / mass spectrometry (Hewlett Packard 5890 Series II gas chromatograph coupled with a Hewlett Packard 5971 Series mass selective detector— see LeMaster and Mason 2003 for full description of the GC/MS platform and methods). [Pg.225]

The dissociation of gas-phase ions can be a guide to the structure of the ions. There are two ways to dissociate gas-phase ions, either by collision-induced dissociation (CID) sometimes termed collision-activation dissociation (CAD) or by photodissociation. In each method, a mass selected ion is dissociated and the fragment ion (often called a daughter ion) is measured the neutral fragment cannot be experimentally observed. [Pg.358]

The deposition of mass and charge selected ions onto surfaces is underway but is in its infancy. How do the ions survive the collision with a surface This question has a myriad of answers depending on many variables and will have a future in investigative studies. A soft landing is now a possibility (280) and allows the potential spectroscopic investigation of trapped ions. So far no transition metal ions have been examined using this method but it is only a matter of time. Soft landings via inert gas matrices also have potential in the surface deposition of mass selected clusters. [Pg.419]

The isotope dilution gas chromatography-mass spectrometry method described by Lopez-Avila et al. [16] and discussed in section 5.3.1.3 has been applied to the determination of Atrazine in soil. In this method known amounts of labelled Atrazine were specked into soil samples before extraction with acetone-hexane. The ratio of the naturally abundant compound and the stable-labelled isotope was determined by high-resolution gas chromatography-mass spectrometry with the mass spectrometer in the selected ion monitoring mode. Detection limits of 0.1-l.Oppb were achieved. Accuracy was >86% and precision better than 8%. [Pg.238]


See other pages where Mass selection method is mentioned: [Pg.199]    [Pg.134]    [Pg.134]    [Pg.61]    [Pg.370]    [Pg.199]    [Pg.134]    [Pg.134]    [Pg.61]    [Pg.370]    [Pg.872]    [Pg.423]    [Pg.622]    [Pg.71]    [Pg.171]    [Pg.119]    [Pg.4]    [Pg.317]    [Pg.397]    [Pg.409]    [Pg.412]    [Pg.466]    [Pg.510]    [Pg.112]    [Pg.164]    [Pg.343]    [Pg.187]    [Pg.173]    [Pg.277]    [Pg.9]    [Pg.658]   


SEARCH



Method selection

Method selectivity

SELECT method

Selective methods

© 2024 chempedia.info