Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Interferences internal standard

A compound for quantitative reference not contained in the sample with chemical characteristics similar to those of the analyte. Labeled analogues of the target analytes are typically applied, for example, in dioxin analysis. The ISTD provides an analytical response that is distinct from the analyte and not subject to interference. Internal standards are used to adjust for variations in the analytical response due to... [Pg.798]

The method is based on the international standard ISO 4053/IV. A small amount of the radioactive tracer is injected instantaneously into the flare gas flow through e.g. a valve, representing the only physical interference with the process. Radiation detectors are mounted outside the pipe and the variation of tracer concentration with time is recorded as the tracer moves with the gas stream and passes by the detectors. A control, supply and data registration unit including PC is used for on site data treatment... [Pg.1054]

When possible, quantitative analyses are best conducted using external standards. Emission intensity, however, is affected significantly by many parameters, including the temperature of the excitation source and the efficiency of atomization. An increase in temperature of 10 K, for example, results in a 4% change in the fraction of Na atoms present in the 3p excited state. The method of internal standards can be used when variations in source parameters are difficult to control. In this case an internal standard is selected that has an emission line close to that of the analyte to compensate for changes in the temperature of the excitation source. In addition, the internal standard should be subject to the same chemical interferences to compensate for changes in atomization efficiency. To accurately compensate for these errors, the analyte and internal standard emission lines must be monitored simultaneously. The method of standard additions also can be used. [Pg.438]

Errors due to nonspectral interferences can be reduced via matrix matching, the method of standard additions (and its multivariant extensions), and the use of internal standards. ... [Pg.642]

At this point, the solution containing the component to be measured (Ax) also contains any other compounds from the original matrix that are soluble in the solvent used in the analysis. For the analysis to be accurate, other components in the matrix cannot interfere by eluting at the same retention time as the components to be measured. For accurate MS analyses, the matrix component must not interfere with production of the ions being measured for either the internal standard or the component to be measured. In some cases, to eliminate interferences, it may be necessary to increase the resolution of the mass spectrometer by narrowing the mass window being monitored. Alternatively, MS/MS can be used to avoid chemical interference (see Chapter 1). [Pg.26]

The use of an internal standard probably gives the most accurate quantitative results. However, the procedure depends upon finding an appropriate substance that will elute in a position on the chromatogram where it will not interfere or merge with any of the natural components of the mixture. If the sample contains numerous components, this may be difficult. Having identified a reference standard, the response factors for each component of interest in the mixture to be analyzed must be determined. A synthetic mixture is made up containing known concentrations of each of the components of interest and the standard. If there are (n) components, and the (r) component is present at concentration (Cr) and the standard at a concentration (Cst). [Pg.268]

Figure 4.22. Correlation of assay values for components A and B, for three dosage levels of A, with 10 samples per group. The comer symbols indicate the 10% specification limits for each component. For manual injection (left panel) only relative standard deviations of 1-2% are found, but no correlation. Automatic injection (right panel) has a lower intrinsic relative standard deviation, but the data are smeared out along the proportionality line because no internal standard was used to correct for variability of the injected volume. The proportionality line does not go through the comers of the specification box because component B is either somewhat overdosed (2.4%). analytical bias, or because an interference results in too high area readings for B. The... Figure 4.22. Correlation of assay values for components A and B, for three dosage levels of A, with 10 samples per group. The comer symbols indicate the 10% specification limits for each component. For manual injection (left panel) only relative standard deviations of 1-2% are found, but no correlation. Automatic injection (right panel) has a lower intrinsic relative standard deviation, but the data are smeared out along the proportionality line because no internal standard was used to correct for variability of the injected volume. The proportionality line does not go through the comers of the specification box because component B is either somewhat overdosed (2.4%). analytical bias, or because an interference results in too high area readings for B. The...
An internal standard is desirable in any quantitative trace environmental analysis. The ideal internal standard should behave in a manner identical to that of the analyte in all the procedures followed for isolation, purification, and determination without producing interference. This is a difficult requirement to meet for nitrosamines, especially for NDMA. [Pg.334]

Alternatively, LC is used for the separation and quantification of PAHs using both UV and fluorescence detection. The analytes are identified based on their relative retention times and UV and/or fluorescence emission spectra. For UV detection an efficient cleanup is a prerequisite since this detection method is not very selective (almost universal for PAHs), and hence it also responds to many coeluting compounds. Due to the high specificity of fluorescence detection for most PAHs, this LC detection method is less susceptible to potential interferences. As in the case of GC the apphcation of internal standard(s) is mandatory since solvents have to be evaporated during the cleanup, which may result in partial losses of some of the more volatile analytes. [Pg.100]

The method using GC/MS with selected ion monitoring (SIM) in the electron ionization (El) mode can determine concentrations of alachlor, acetochlor, and metolachlor and other major corn herbicides in raw and finished surface water and groundwater samples. This GC/MS method eliminates interferences and provides similar sensitivity and superior specificity compared with conventional methods such as GC/ECD or GC/NPD, eliminating the need for a confirmatory method by collection of data on numerous ions simultaneously. If there are interferences with the quantitation ion, a confirmation ion is substituted for quantitation purposes. Deuterated analogs of each analyte may be used as internal standards, which compensate for matrix effects and allow for the correction of losses that occur during the analytical procedure. A known amount of the deuterium-labeled compound, which is an ideal internal standard because its chemical and physical properties are essentially identical with those of the unlabeled compound, is carried through the analytical procedure. SPE is required to concentrate the water samples before analysis to determine concentrations reliably at or below 0.05 qg (ppb) and to recover/extract the various analytes from the water samples into a suitable solvent for GC analysis. [Pg.349]

Unlike GC/MS methods, internai standards are not appropriate for immunoassays. Internal standards that would react with the antibody but would not interfere with the assay are nonexistent. In the place of internal standards, external QC must be maintained. [Pg.647]

Under some conditions, it is difficult to incorporate an internal standard into a method. If the chromatogram is very complex, an internal standard may interfere with quantitation of a peak of interest. The development of highly precise sample transfer techniques, including modem autoinjectors, reduces the dependence of the experimentalist on the use of an internal standard to correct for effects of dilution and transfer losses. In many cases, external standardization can be used effectively. The weight percent purity is determined by comparing the area of each peak in a chromatogram with those generated by separately injected pure standards of known concentration. [Pg.186]

In a bioanalytical method, analyses of blank samples (plasma, urine, or other matrix) should be obtained from at least six sources. Each blank sample should be tested for the possible interference of endogenous substances, metabolites, or degradation products. The response of the peaks interfering at the retention time of the analyte should be less than 20% of the response of a lower quantitation limit standard, and should be less than 5% of the response of the internal standard that was used [18, 19]. For dissolution studies, the dissolution media or excipients should not give a peak or spot that has an identical Rt or Rf value with the analyte [20]. [Pg.248]

Sichere et al. [25] determined bromine concentrations in the 0.06-120mg/1 range in brines, directly by X-ray fluorescence using selenium as an internal standard to eliminate interference effects. Lower concentrations of bromine must be concentrated on filter paper containing an ion exchange resin. The same concentrations of chlorine can be determined with the addition of barium to reduce the interferences from carbonates and sulfates. Relative standard deviation was better than 1%. The interference of some other ions (e.g., calcium, potassium, magnesium, sodium, and iron) was examined. [Pg.65]

Sample preconcentration was performed by means of an automated on-line SPE sample processor Prospekt-2 (Spark Holland, Emmen, The Netherlands). Oasis HLB cartridges (Waters, Barcelona, Spain) were used to preconcentrate cannabi-noids present in the water samples whereas isolation of the rest of the compounds was done in PLRPs cartridges (Spark Holland). Before extraction, influent samples were diluted with HPLC water (1 9, v/v) to reduce matrix interferences and to fit some analyte concentrations, e.g., cocaine (CO) and benzoylecgonine (BE), within the linear calibration range. A sample volume of 5 mL was spiked with the internal standard mixture (at 20 ng/L) in order to correct for potential losses during the analytical procedure, as well as for matrix effects. Elution of the analytes to the LC system was done with the chromatographic mobile phase. [Pg.193]

In addition to online filters and precolumns, a simple protein precipitation step often precedes online SPE LC/MS/MS to prolong cartridge life. Protein precipitation can also reduce analytical interference and shorten chromatographic separation time. Since an internal standard (IS) solution is often added to plasma samples and centrifugation is used to remove possible particles before loading into the autosampler, protein precipitation does not add labor to the process. [Pg.283]

Quantification is usually achieved by a standard addition method, use of labeled internal standards, and/or external calibration curves. In order to allow for matrix interferences the most reliable method for a correct quantitation of the analytes is the isotope dilution method, which takes into account intrinsic matrix responses, using a deuterated internal standard or carbon-13-labeled internal standard with the same chemistry as the pesticide being analyzed (i.e., d-5 atrazine for atrazine analysis). Quality analytical parameters are usually achieved by participation in interlaboratory exercises and/or the analysis of certified reference materials [21]. [Pg.63]

During a single run, which may take all day if a large number of samples are to be analyzed, the instrument may drift from its optimum settings. To detect this drift in solution-based techniques, and also to compensate for some matrix effects, a known amount of an element may be added to each sample before analysis. This internal standard (also called a spike) is added to all the samples and blanks, with the exception of the instrument blank (which is defined as zero concentration for all elements see below). It is important that the element (or isotope) chosen as the spike is not an element which is to be determined in the samples, and preferably which does not occur naturally in the samples. It must not be an element which will cause, or suffer from, interference with the other elements to be determined. In solution ICP-MS,... [Pg.308]

The sample is extracted with a mixture of hexane, acetone and water. After separation, the hexane phase is reduced in volume and divided into two aliquots, one of which is first shaken with 7% fuming sulphuric acid to remove lipids, and then with cyanide to eliminate interference by elemental sulphur. The other aliquot is evaporated to dryness and heated with ethanolic potassium hydroxide. The two aliquots are injected into a gas chromatograph fitted with a glass capillary column and an electron capture detector. Hexabromobenzene is used as an internal standard. Polychlorinated biphenyls are determined quantitatively by comparing the peaks of the sample with those of Clophen A... [Pg.220]

The first two points are best dealt with as part of the process for developing vahdated analytical methods. Vafidation should include testing the robustness of a method in repeated use over a period of time determining the precision and accuracy and study of potential interferences. As an example, it would be expected that in the capillary GC—TEA method for organic explosives, a peak should be at least three times the basefine noise to be counted as a real signal, and that the relative retention time should be within 1.0% of the standard for volatile compounds and within 0.5% for the rest. The relative retention time is simply the ratio of the analyte s retention time compared with that of an internal standard. Use of relative retention times significantly improves the repeatabdity of GC analysis... [Pg.237]

Gas Chromatography Analysis of Water for Pesticides. All analyses for pesticides in water were done by gas chromatography. Solvents used for extraction were checked by gas chromatography for purity and interferences and all glassware used in the extraction was cleaned in a chromic acid/sulfuric acid mixture. Standards consisted of mixtures of various pesticides (actual commercial formulations) suspended or dissolved in water. These aqueous standards were extracted in the same manner as unknown solutions. The standard concentrations encompassed the concentration of unknowns to be determined. A standard curve normally consisted of a set of four pesticide concentrations. Blanks were run and an internal standard (eicosane) was used. The internal standard concentration was kept constant for all analyses. The conditions for GC analysis were guided by the pesticides expected in the water. For the more complex mixtures, such as those employed in the synthetic waste and those encountered in the field, a 6 ft., 3 percent SE-30 on GAS CHROM Q column sufficed. A typical chromatogram of a complex pesticide mixture is shown in Figure 2. ( )... [Pg.128]


See other pages where Interferences internal standard is mentioned: [Pg.47]    [Pg.312]    [Pg.370]    [Pg.47]    [Pg.312]    [Pg.370]    [Pg.426]    [Pg.642]    [Pg.427]    [Pg.604]    [Pg.866]    [Pg.207]    [Pg.285]    [Pg.54]    [Pg.148]    [Pg.609]    [Pg.651]    [Pg.49]    [Pg.212]    [Pg.263]    [Pg.355]    [Pg.87]    [Pg.30]    [Pg.245]    [Pg.228]    [Pg.541]    [Pg.276]    [Pg.188]    [Pg.16]    [Pg.75]   
See also in sourсe #XX -- [ Pg.170 , Pg.171 ]




SEARCH



Interferences internal standard method

Internal standards

International Standardization

International Standards

Standardization international standards

© 2024 chempedia.info