Inductively coupled plasma power

If a sample solution is introduced into the center of the plasma, the constituent molecules are bombarded by the energetic atoms, ions, electrons, and even photons from the plasma itself. Under these vigorous conditions, sample molecules are both ionized and fragmented repeatedly until only their constituent elemental atoms or ions survive. The ions are drawn off into a mass analyzer for measurement of abundances and mJz values. Plasma torches provide a powerful method for introducing and ionizing a wide range of sample types into a mass spectrometer (inductively coupled plasma mass spectrometry, ICP/MS). [Pg.87]

Laser based mass spectrometric methods, such as laser ionization (LIMS) and laser ablation in combination with inductively coupled plasma mass spectrometry (LA-ICP-MS) are powerful analytical techniques for survey analysis of solid substances. To realize the analytical performances methods for the direct trace analysis of synthetic and natural crystals modification of a traditional analytical technique was necessary and suitable standard reference materials (SRM) were required. Recent developments allowed extending the range of analytical applications of LIMS and LA-ICP-MS will be presented and discussed. For example ... [Pg.425]

The inductively coupled plasma source (Fig. 20.11) comprises three concentric silica quartz tubes, each of which is open at the top. The argon stream that carries the sample, in the form of an aerosol, passes through the central tube. The excitation is provided by two or three turns of a metal induction tube through which flows a radio-frequency current (frequency 27 MHz). The second gas flow of argon of rate between 10 and 15 L min-1 maintains the plasma. It is this gas stream that is excited by the radio-frequency power. The plasma gas flows in a helical pattern which provides stability and helps to isolate thermally the outside quartz tube. [Pg.774]

Inductively coupled plasma Plasmas generated by application of radiofrequency power to a nonresonant inductive coil and maintained by an inductive electromagnetic field. Low-pressure ICP is a high-density plasma source. [Pg.10]

At present, inductively coupled plasma mass spectrometry provides a unique, powerful alternative for the determination of rare earths in natural samples [638,639]. Nevertheless, its application to the determination of rare earths at ultratrace concentration level in seawater is limited, because highly saline samples can cause both spectral interferences and matrix effects [640]. Therefore, a separation of the matrix components and preconcentration of the analytes are prerequisites. To achieve this goal, many preconcentration techniques have been used, including coprecipitation with... [Pg.214]

Wendt and Fassel [2] reported early experiments with a tear-drop shaped inductively coupled plasma but later described the medium power l-3kW 18mm annular plasma now favoured in modern analytical instruments [3]. [Pg.39]

Chemically pure reagents were used. Cadmium was added as its sulfate salt in concentrations of about 50 ppm. Lanthanides were added as nitrates. For the experiments with other metal ions so-called "black acid from a Nissan-H process was used. In this acid a large number of metal ions were present. To achieve calcium sulfate precipitation two solutions, one consisting of calcium phosphate in phosphoric acid and the other of a phosphoric acid/sulfuric acid mixture, were fed simultaneously in the 1 liter MSMPR crystallizer. The power input by the turbine stirrer was 1 kW/m. The solid content was about 10%. Each experiment was conducted for at least 8 residence times to obtain a steady state. During the experiments lic iid and solid samples were taken for analysis by ICP (Inductively Coupled Plasma spectrometry, based on atomic emission) and/or INAA (Instrumental Neutron Activation Analysis). The solid samples were washed with saturated gypsum solution (3x) and with acetone (3x), and subsequently dried at 30 C. The details of the continuous crystallization experiments are given in ref. [5]. [Pg.384]

The distributions of trace elements between minerals and within a suite of related rocks provide powerful tools for constraining the origin and history of rocks and meteorites. Trace-element abundances for rocks typically are part of the data set collected when determining bulk compositions. Trace element compositions of minerals require more powerful techniques such as the ion microprobe or the laser-ablation inductively coupled plasma mass spectrometer (ICPMS). [Pg.22]

The temperature of the inductively coupled plasma varies with the distance from the load coil and according to the setting of the ICP rf power and nebulizer gas flow rate. A typical profile of the plasma gas temperature along the torch axis as a function of distance from the load coil is shown in Figure 2.4. With increasing distance from the load coil and with a reduction of ICP rf power the gas plasma temperature decreases. [Pg.30]

The most frequently applied analytical methods used for characterizing bulk and layered systems (wafers and layers for microelectronics see the example in the schematic on the right-hand side) are summarized in Figure 9.4. Besides mass spectrometric techniques there are a multitude of alternative powerful analytical techniques for characterizing such multi-layered systems. The analytical methods used for determining trace and ultratrace elements in, for example, high purity materials for microelectronic applications include AAS (atomic absorption spectrometry), XRF (X-ray fluorescence analysis), ICP-OES (optical emission spectroscopy with inductively coupled plasma), NAA (neutron activation analysis) and others. For the characterization of layered systems or for the determination of surface contamination, XPS (X-ray photon electron spectroscopy), SEM-EDX (secondary electron microscopy combined with energy disperse X-ray analysis) and... [Pg.259]

A serious problem in LA-ICP-MS described in the literature on many occasions is the time-dependent elemental fraction (so-called ablation fractionation ) occurring during laser ablation and the transport process of ablated material, or during atomization and ionization processes in the inductively coupled plasma.20-22 Numerous papers focus on the study of fraction effects in LA-ICP-MS as a function of experimental parameters applied during laser ablation (such as laser energy, laser power density, laser pulse duration, wave length of laser beam, ablation spot size,... [Pg.390]

Atomic emission spectroscopy can be employed, generally with an inductively coupled plasma for thermal excitation. The sample is introduced into the plasma as a mist of ultrafine droplets, and the monochromator and detector are set to measure the intensity of an atomic emission line characteristic of the element. This technique is powerful, general, sensitive, linear, and able to measure over 70 elements, and, as a result, is widely used. Response is typically linear over four orders of magnitude in concentration with relative standard deviations of 1 to 3%. In low-salt aqueous solutions, detection limits range from 10 to 1000 nanomolar without preconcentration. Significant problems with saline samples remain, but use of Babington nebulizers alleviates these problems somewhat. [Pg.60]

A very intense inductively coupled plasma (ICP) could be created, using a 13.56-MHz RF-source and coupling the power through a quartz tube into the plasma. Working pressure was 2 x 10-2 mbar and the substrate could be heated up to 800 °C [222]. The precursor was trimethylborazine ((HBN-CH3)3), which was transported in a nitrogen/argon carrier gas. Similar to other deposition processes, prior to the c-BN nucleation an oriented h-BN layer was formed [223]. [Pg.32]

Thompson and Zao [170] have described a solvent extraction-inductively coupled plasma atomic emission spectrometric method for the determination of down to 0.02 - 0.03 xg/g of molybdenum in soils. The soil sample is pressure-leached with 6 M hydrochloric acid and at 120 °C for 15 minutes. The digest is then extracted with heptan-2-one to separate molybdenum from potentially interfering elements such as iron, aluminium, calcium and magnesium. This organic extract is then directly sprayed into an inductively coupled plasma atomic emission spectrometer operated at 1.65 to 1.7 kW power. [Pg.49]

In this method the soil sample is dried overnight at 85 °C and ground into an homogeneous mixture. A 1 g soil sample is placed into a beaker and 10 ml of concentrated nitric acid added. The solution is heated to dryness and 5 ml of concentrated nitric acid is added. The uranium is redissolved in 5 ml of 8 N nitric acid and diluted to 25 ml with distilled water. The inductively coupled plasma mass spectrometry system used was an ELAN Model 250. The ion source consists of a modified plasma Thermal Model 2500 control box. The forward power was set at 1200 W with the plasma flow, auxiliary flow and nebuliser pressure set at 131/min, 1.0l/min and 0.27 MPa, respectively. The focusing lenses B, El, P and S2 are set at +5.3 V, -12.5 V, -18.0 V and -7.6 V, respectively. The m/z238 ion was monitored for two sec-... [Pg.58]

In principle, all powerful element-specific methods that are able to monitor continuously the effluents of separation processes commonly in the range of a few mimin-1 and in element concentrations of some Klpg liter-1. A well-suited method is based on modern element-specific quadrupole mass spectrometry (MS) with an inductively coupled plasma (ICP) interface to the separation unit [e.g., liquid chromatography (LC) or field-flow fractionation (FFF)].Tlie ICP-MS detection can also be used for continuously characterizing the effluent of any kind of packed column (Metreveli and Frimmel, 2007). By this, the transport and elution properties of... [Pg.375]

Inductively coupled plasma-mass spectrometry (ICP-MS) is a powerful technique that uses an inductively coupled plasma as an ion source and a mass spectrometer as an ion analyzer. It can measure the presence of more than 75 elements in a single scan, and can achieve detection limits down to parts per trillion (ppt) levels for many elements—levels that are two or three orders of magnitude lower than those obtained by ICP-AES (Keeler 1991). It is more expensive than ICP-AES and requires more highly skilled technical operation. Aluminum levels in urine and saliva were detected down to 0.02 g/mL and in blood serum to 0.001 g/mL using ICP-MS (Ward 1989). Speciation studies have employed ICP-MS as a detector for aluminum in tissue fractions separated by size-exclusion chromatography (SEC) with detection limits of 0.04 g/g in femur, kidney and brain (Owen et al. 1994). [Pg.263]

Atomic absorption remained the technique of choice until relatively recently. However, with the introduction of plasma sources, atomic emission, in the form of inductively coupled plasma spectroscopy, has made a comeback. This development is now receiving historical attention, and was the subject of a symposium held in 1999. Papers discussed atomic emission analysis prior to 1950,206 the fact that emission techniques developed continuously, even in the period when absorption methods were dominant,207 and the development of the plasma sources on which the new techniques depend.208 Also discussed was the powerful hyphenated technique of ICP-MS,209 and the history of one of the leading manufacturers of atomic emission instruments.210... [Pg.165]


See other pages where Inductively coupled plasma power is mentioned: [Pg.382]    [Pg.2]    [Pg.665]    [Pg.614]    [Pg.649]    [Pg.29]    [Pg.31]    [Pg.186]    [Pg.357]    [Pg.17]    [Pg.160]    [Pg.327]    [Pg.20]    [Pg.83]    [Pg.153]    [Pg.6]    [Pg.30]    [Pg.31]    [Pg.34]    [Pg.35]    [Pg.84]    [Pg.122]    [Pg.131]    [Pg.158]    [Pg.241]    [Pg.340]    [Pg.355]    [Pg.443]    [Pg.358]    [Pg.451]    [Pg.28]    [Pg.430]   