Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Fluorescence Spectrometers

In contrast to absorption spectroscopy the use of a reference is rather difficult. It requires a sample for which absorption and fluorescence spectra both exhibit the same spectral distribution as the probe. This is quite impossible. For this reason in some instrumentation one tries to eliminate fluctuations in the intensity of the light source. The excitation beam is split and a small amount of the intensity of excitation is focused on a photodiode. Nevertheless neither the changes in the optics of the instrumentation nor the variance of the transmission curve with wavelength can be overcome. However there is a way to calibrate the spectral distribution of a fluorimeter in- [Pg.277]

Fortunately in reaction kinetics the absolute value of fluorescence intensity does not have to be known. Since fluorescence intensity is proportional to the concentration of the fluorophore in dilute solutions, changes in time dependence of fluorescence can be correlated with a change in concentration. If even one of the components fluoresces, photokinetic evaluation becomes very simple. [Pg.278]

In principle the wavelength of irradiation and excitation are the same. The excitaticm light source is used to cause the photoreaction. By these means monochromatic irradiation is used automatically. A typical cell compartment is given in Fig. 4.22. The sample is stirred and temperature controlled. Three positions are mounted tilted at 45 relative to the optical axis of the excitation light beam. One position is for the cell with the solvent, one for the fluorescence standard, and the third for the sample itself. These three positions [Pg.279]

Combined absorption, fluorescence measurement, and irradiation compartment. /, [Pg.279]

Is controlled by a split-beam and a photodiode. The 45 tilted bar contains three positions (1) for a fluorescence standard (S), (2) for the sample (R), (3) for a reference blank (v). The sample can be stirred. The cell holder is temperature controlled. The irradiation source is either a [Pg.279]


A widely used procedure for determining trace amounts of selenium involves separating selenium from solution by reduction to elemental selenium using tellurium (as a carrier) and hypophosphorous acid as reductant. The precipitated selenium, together with the carrier, are collected by filtration and the filtered soflds examined directly in the wavelength-dispersive x-ray fluorescence spectrometer (70). Numerous spectrophotometric and other methods have been pubHshed for the deterruination of trace amounts of selenium (71—88). [Pg.335]

As atomic fluorescence spectrometer a mercury analyzer Mercur , (Analytik-Jena, Germany) was used. In the amalgamation mode an increase of sensitivity by a factor of approximately 7-8 is obtained compared with direct introduction, resulting in a detection limit of 0,09 ng/1. This detection limit has been improved further by pre-concentration of larger volumes of samples and optimization of instrumental parameters. Detection limit 0,02 ng/1 was achieved, RSD = 1-6 %. [Pg.171]

Atomic Fluorescence System - Millennium Excalibur PSA 10.055 -was used in our work. This system consists of the autosampler, the integrated continuous flow vapour generator and the atomic fluorescence spectrometer with the boosted dischar ge hollow cathode lamp and a control computer. [Pg.208]

The mercury vapour was detected by atomic fluorescence spectrometer Millennium Merlin PSA 10.025. [Pg.211]

Procedures for calibrating both monochromators in a fluorescence spectrometer using narrow line sources have been discussed (IS,18) care must be taken with placement of the calibration source. [Pg.100]

Room-temperature fluorescence (RTF) has been used to determine the emission characteristics of a wide variety of materials relative to the wavelengths of selected Fraunhofer lines in support of the Fraunhofer luminescence detector remote-sensing instrument. RTF techniques are now used in the compilation of excitation-emission-matrix (EEM) fluorescence "signatures" of materials. The spectral data are collected with a Perkin-Elraer MPF-44B Fluorescence Spectrometer interfaced to an Apple 11+ personal computer. EEM fluorescence data can be displayed as 3-D perspective plots, contour plots, or "color-contour" images. The integrated intensity for selected Fraunhofer lines can also be directly extracted from the EEM data rather than being collected with a separate procedure. Fluorescence, chemical, and mineralogical data will be statistically analyzed to determine the probable physical and/or chemical causes of the fluorescence. [Pg.228]

The Fe-B nanocomposite was synthesized by the so-called pillaring technique using layered bentonite clay as the starting material. The detailed procedures were described in our previous study [4]. X-ray diffraction (XRD) analysis revealed that the Fe-B nanocomposite mainly consists of Fc203 (hematite) and Si02 (quartz). The bulk Fe concentration of the Fe-B nanocomposite measured by a JOEL X-ray Reflective Fluorescence spectrometer (Model JSX 3201Z) is 31.8%. The Fe surface atomic concentration of Fe-B nanocomposite determined by an X-ray photoelectron spectrometer (Model PHI5600) is 12.25 (at%). The BET specific surface area is 280 m /g. The particle size determined by a transmission electron microscope (JOEL 2010) is from 20 to 200 nm. [Pg.389]

Figure 5.7 Layout of a typical fluorescence spectrometer. After Rendell [132]. Reprinted from D. Rendell, Fluorescence and Phosphorescence Spectroscopy. Copyright 1987 John Wiley Sons, Limited. Reproduced with permission... Figure 5.7 Layout of a typical fluorescence spectrometer. After Rendell [132]. Reprinted from D. Rendell, Fluorescence and Phosphorescence Spectroscopy. Copyright 1987 John Wiley Sons, Limited. Reproduced with permission...
A schematic diagram of a typical fluorescence spectrometer is shown in Figure 2.13. [Pg.321]

Due to the separation between excitation and emission in a fluorescence spectrometer, concentrations can be detected down to picomolar, with a wide linear range over up to five orders of magnitude. As a consequence of,e.g.,vibra-tional relaxation, the amount of energy which is released as fluorescence (quantum yield) is strongly dependent on solvent and temperature. [Pg.80]

T Damerau and M Hennecke, Determination of orientational order parameters of uniaxial films with a commercial 90°C-angle fluorescence spectrometer, J. Chem. Phys., 103 6232-6240, 1995. [Pg.476]

Desnica, V. and Schreiner, M. (2006). A LabVIEW-controlled portable X-ray fluorescence spectrometer for the analysis of art objects. X-Ray Spectrometry 35 280-286. [Pg.71]

An X-ray fluorescence spectrometer needs to resolve the different peaks, identify them and measure their area to quantify the data. There are two forms of X-ray spectrometers (Fig. 5.5), which differ in the way in which they characterize the secondary radiation - wavelength dispersive (WD), which measures the wavelength, and energy dispersive (ED), which measures the energy of the fluorescent X-ray (an illustration of the particle-wave duality nature of electromagnetic radiation, described in Section 12.2). [Pg.102]

Ti content in the polymer films was measured with a Princeton Gamma Tech System 4 x-ray Fluorescence Spectrometer. The conditions employed were Cr target, 50 keV source operating at 3 mA, 0.75 mm aperture, 4.8 mm beam stop, helium atmosphere and 100 sec. counting time. A calibration curve was constructed by plotting the fluorescence counts versus the amount of Ti in HB-HPR 206 films determined by Rutherford Backscattering Spectroscopic (RBS) analysis. [Pg.194]

Figure 1. Flow Cell for Monitoring solvent Permeation and PMMA Film Dissolution Simultaneously. The cell is placed in the sample chamber of a fluorescence spectrometer. (Reproduced with permission from Ref. ll. Copyright 1988 Wiley Sons.)... Figure 1. Flow Cell for Monitoring solvent Permeation and PMMA Film Dissolution Simultaneously. The cell is placed in the sample chamber of a fluorescence spectrometer. (Reproduced with permission from Ref. ll. Copyright 1988 Wiley Sons.)...
The LS-3B is a fluorescence spectrometer with separate scanning monochromators for excitation and emission, and digital displays of both monochromator wavelengths and signal intensity. The LS-5B is a ratioing luminescence spectrometer with the capability of measuring fluorescence, phosphorescence and bio- and chemiluminescence. Delay time (t) and gate width (t) are variable via the keypad in lOps intervals. It corrects excitation and emission spectra. [Pg.29]

Fluorescence spectrometers are equivalent in their performance to singlebeam UV-visible spectrometers in that the spectra they produce are affected by solvent background and the optical characteristics of the instrument. These effects can be overcome by using software built into the Perkin-Elmer LS-5B instrument or by using application software for use with the Perkin-Elmer models 3700 and 7700 computers. [Pg.30]

A wide variety of X-ray fluorescence spectrometers may be used, depending on the nature and complexity of the sample, and on the number of samples to be analysed. To prove this and to indicate the substantial influence which the sample has on the choice of measuring instrument, let us consider some of the main characteristics of some X-ray fluorescence instruments used today [38]. These are shown in Table 14.11. [Pg.451]

Recent developments in laser technology and fast detection methods now allow the kinetic behaviour of the excited state species arising from absorption of radiation by polymers to be studied on time-scales down to the picosecond region ( ). An example of a time-resolved fluorescence spectrometer which can be used to study such ultrafast phenomena is illustrated in Figure 5 Q). [Pg.31]

Figure 5. Schematic diagram of a time-resolved fluorescence spectrometer using a picosecond laser as an excitation source. Inset diagram intensity/time/ wavelength surface for poly (acenaphthalene) in benzene at 20°C. Excitation wavelength 295 nm. (Reproduced with permission from Ref. 21. Copyright 1987 Chemistry in Australia.)... Figure 5. Schematic diagram of a time-resolved fluorescence spectrometer using a picosecond laser as an excitation source. Inset diagram intensity/time/ wavelength surface for poly (acenaphthalene) in benzene at 20°C. Excitation wavelength 295 nm. (Reproduced with permission from Ref. 21. Copyright 1987 Chemistry in Australia.)...
X-ray fluorescence is a type of atomic spectroscopy since the energy transitions occur in atoms. However, it is distinguished from other atomic techniques in that it is nondestructive. Samples are not dissolved. They are analyzed as solids or liquids. If the sample is a solid material in the first place, it only needs to be polished well, or pressed into a pellet with a smooth surface. If it is a liquid or a solution, it is often cast on the surface of a solid substrate. If it is a gas, it is drawn through a filter that captures the solid particulates and the filter is then tested. In any case, the solid or liquid material is positioned in the fluorescence spectrometer in such a way that the x-rays impinge on a sample surface and the emissions are measured. The fluorescence occurs on the surface, and emissions originating from this surface are measured. [Pg.280]

Figure 12.1 shows the classic L-format of the most commonly used fluorescence spectrometer configuration which is topologically the same for the measurement of both steady-state spectra and lifetimes. The source and detector options of relevance to IR fluorescence measurements are discussed in Sections 12.3 and 12.4, respectively. The other optical components comprised of the lenses for focusing and collection and monochromators for wavelength selection contain few peculiarities in the near-IR as... [Pg.378]

S. Arnold and L. M. Folan, Fluorescence spectrometer for a single electrodynamically levitated microparticle, Rev. Sci. Instrum 57, 2250-2253 (1986). [Pg.384]

Introduction into a DC plasma requires rather more care and attention owing to its inherent design features. As the hydride is being introduced into the plasma, it is necessary to provide a controlled sheath of argon to contain the hydride and direct it into the plasma. This chimney effect significantly improves the sensitivity for hydride-forming elements. This interface has also formed the basis of an introduction system for mercury vapour into an atomic-fluorescence spectrometer as described by Godden and Stockwell [12]. [Pg.146]

For low-level mercury measurements, the sensitivity of the fluorescence spectrometer offers the most attractive route for analysis it is also possible to analyse air samples. In the... [Pg.148]

Application of field-portable x-ray fluorescence spectrometers in mineral exploration, with examples from the Abitibi Greenstone Belt... [Pg.83]

Fig. 5. Effect of WH on [Ca +], in Ang II-, CaCU or Bay K8644-stimulated VSMC. Ang II (10 UM)-induced [Ca " ]/ elevation (a) in Fura-2/AM-loaded VSMC was measured in the absence (Control) or presence of WH (300 juM). One pM PD 123177 as an AT2-receptor antagonist was added 10 min before Ang II addition. CaCf (2.5 mM) was added to VSMC treated with 2-APB (200 /xM IP3-R blocker)-containing Ca free PSS buffer (b). Effect of WH on [Ca ]/ in Bay K8644-stimulated VSMC was measured using a con-focal microscope and a fluorescent spectrometer (c). Bay K8644 (50 /xM)-induced [Ca ]/ elevation in VSMC was measured in the absence (Control) or presence of WH (300 /xM). Fig. 5. Effect of WH on [Ca +], in Ang II-, CaCU or Bay K8644-stimulated VSMC. Ang II (10 UM)-induced [Ca " ]/ elevation (a) in Fura-2/AM-loaded VSMC was measured in the absence (Control) or presence of WH (300 juM). One pM PD 123177 as an AT2-receptor antagonist was added 10 min before Ang II addition. CaCf (2.5 mM) was added to VSMC treated with 2-APB (200 /xM IP3-R blocker)-containing Ca free PSS buffer (b). Effect of WH on [Ca ]/ in Bay K8644-stimulated VSMC was measured using a con-focal microscope and a fluorescent spectrometer (c). Bay K8644 (50 /xM)-induced [Ca ]/ elevation in VSMC was measured in the absence (Control) or presence of WH (300 /xM).
Then, k equals the observed reaction rate divided by the initial reactant concentration i.e., k = Vinitiai/[Ainitiai])-This method is most useful when one has an assay method that is sufficiently sensitive to ensure that only a small fraction, say 3-5%, of the reactant is depleted during the rate measurements. Typically, this is satisfactorily achieved with a UV-visible spectrophotometer, a fluorescence spectrometer, or a radioactively labeled reactant. The initial rate method is extremely convenient, and the preponderance of enzyme rate data has been obtained by initial rate measurements. Finally, one should note that the initial rate method can yield erroneous results if the initial reactant concentration is in doubt. This is not true for the plots of ln ([Ao] - [At]/([Ao] -[Aoo]) versus reaction time because one is considering the fraction of reactant A remaining. [Pg.135]

Dzubay, T.G. Stevens, R.K. "Ambient Air Analysis with Dichotomous Sampler and X-ray Fluorescence Spectrometer" Environ. Sci. Technol., 1975, 9, 663. [Pg.285]

Elemental Analysis. A Phillips PW 1410/70 X-ray fluorescence spectrometer with Cr radiation was used to measure the relative quantities of Br in the molded polymer samples. Extractable bromide and chloride ions were detected with specific ion electrodes after a 48-hour, 120 C steam bomb extraction. [Pg.214]

Atomic fluorescence spectrometer PS Analytical Excalibur atomic fluorescence spectrometer (AFS). [Pg.174]


See other pages where Fluorescence Spectrometers is mentioned: [Pg.427]    [Pg.58]    [Pg.170]    [Pg.319]    [Pg.78]    [Pg.22]    [Pg.70]    [Pg.338]    [Pg.123]    [Pg.86]    [Pg.344]    [Pg.476]    [Pg.134]    [Pg.233]    [Pg.379]    [Pg.143]    [Pg.29]    [Pg.285]   
See also in sourсe #XX -- [ Pg.29 , Pg.30 ]




SEARCH



Application of field-portable x-ray fluorescence spectrometers in mineral exploration, with examples from the Abitibi Greenstone Belt

Fiber optic fluorescence spectrometers

Fluorescence lifetime spectrometers

Fluorescence spectrometer for

Fluorescence spectrometer, Perkin-Elmer

Microscope fluorescence spectrometer

X-ray fluorescence spectrometer

© 2024 chempedia.info