Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detector standard

Caffeine in coffee, tea, and soda is determined by a solid-phase microextraction using an uncoated silica fiber, followed by a GC analysis using a capillary SPB-5 column with an MS detector. Standard solutions are spiked with G3 caffeine as an internal standard. [Pg.612]

Spectrometers. See Monochromators Detectors Standard additions. See Calibration STPF, M T... [Pg.211]

The disadvantages of the canister method are its inability to concentrate pollutants during sample collection and the potential analytical problems associated with the presence of moisture in the sample (Bianchi and Varney 1993). In all methods, however, there is a consensus that after the sample collection and preparation stage, mixture separation and quantitative analysis is best done with GC. coupled with an assortment of detectors. Standardized methods, with slight alterations, also can be used for determining the metabolites of 1.1,1 -trichloroethane (Hallen et al. 1986 Parsons et al. 1985 Vogel and McCarty 1987). [Pg.170]

Chromatographic techniques are very commonly used in the characterisation of natural waters (Aoyama (1996a, 1996b), Huber (unpublished), Shaw et al. (1994), Mori (1988), Becher et al. (1985), Amy et al. (1987)). However, most authors use different techniques, detectors, standards and eluents, making the comparison of results difficult. [Pg.15]

Microsorb phenyl 5-pm column (250 X 4.6 mm) Resolve C-18 column (300 X 3.9 mm, 5 pm) Gradient flow A - methanol, B - water-methanol-acetic acid (90 9 1) containing 0.1% ammonium acetate, C - acetonitrile Photodiode detector Standard compounds Glaucarubolone chaparrinone holacanthone glaucarubinone pitaasin B paraine neoquassin quassin [93]... [Pg.3356]

The hydrocarbons are separated in another column and analyzed by a flame ionization detector, FID. As an example, Figure 3.13 shows the separation obtained for a propane analyzed according to the ISO 7941 standard. Note that certain separations are incomplete as in the case of ethane-ethylene. A better separation could be obtained using an alumina capillary column, for instance. [Pg.72]

The INTROS Flaw Detector is certified by the Russian State Standard Service (GOSSTANDART) as well as approved by the Russian State Mining and Technology Safety Inspection (GOSGORTECHNADZOR). It is used to inspect mining hoist and crane ropes. Fig. 5 illustrates the INTROS use at the mining hoist of an Ural ore mine. The previous model of the instrument, MDK-11 was used to inspect ropes of the air rope ways in Caucasus and Kazakhstan in 1996. Fig. 6 shows the INTROS MDK-11 inspection of 45 mm skyline rope in Almaty, Kazakhstan. [Pg.337]

The Zond VD - 96 portable eddy-current flaw detector-tester is an original Russian development possessing heightened sensitivity for the surface defects and high inspection capacity. (Russia patent Xs 2063025. All-Union state standard certificate of Russian Federation JVa 2846 of 14. 07. 97)... [Pg.342]

The X-ray instrumentation requires a commercial small angle X-ray camera, a standard fine structure X-ray generator and a sample manipulator if scanning is requested. The essential signal is the relative difference between the refraction level Ir and the absorption level Ia. Both levels are measured simultaneously by two scintillation detectors. At fixed angles of deflection this signal depends solely on the inner surface density factor C and thickness d of the sample [2] ... [Pg.558]

Periodical Verification of Ultrasonic Flaw Detectors According to European Standardization. [Pg.700]

The verifications can be performed by the user himself, with electronic measurement equipment described in this project. The consequences of the application of future European standard are very important since is established a mandatory verification of each particular flaw detector, at least once a year. Their verification is to be performed according to a well defined procedure of measurement, including acceptance criteria for each parameter. [Pg.701]

Declare the conformity or non conformity of the particular flaw detector under test, in accordance with the criteria of CEN standards. [Pg.704]

The unit consists of a mechanical scanner (AWS-6), module of scanner and flaw detector control PSP-3. A standard PC is used for processing the UT results. [Pg.790]

The method is based on the international standard ISO 4053/IV. A small amount of the radioactive tracer is injected instantaneously into the flare gas flow through e.g. a valve, representing the only physical interference with the process. Radiation detectors are mounted outside the pipe and the variation of tracer concentration with time is recorded as the tracer moves with the gas stream and passes by the detectors. A control, supply and data registration unit including PC is used for on site data treatment... [Pg.1054]

The low MW power levels conuuonly employed in TREPR spectroscopy do not require any precautions to avoid detector overload and, therefore, the fiill time development of the transient magnetization is obtained undiminished by any MW detection deadtime. (3) Standard CW EPR equipment can be used for TREPR requiring only moderate efforts to adapt the MW detection part of the spectrometer for the observation of the transient response to a pulsed light excitation with high time resolution. (4) TREPR spectroscopy proved to be a suitable teclmique for observing a variety of spin coherence phenomena, such as transient nutations [16], quantum beats [17] and nuclear modulations [18], that have been usefi.il to interpret EPR data on light-mduced spm-correlated radical pairs. [Pg.1566]

In TOF-SARS [9], a low-keV, monoenergetic, mass-selected, pulsed noble gas ion beam is focused onto a sample surface. The velocity distributions of scattered and recoiled particles are measured by standard TOF methods. A chaimel electron multiplier is used to detect fast (>800 eV) neutrals and ions. This type of detector has a small acceptance solid angle. A fixed angle is used between the pulsed ion beam and detector directions with respect to the sample as shown in figure Bl.23.4. The sample has to be rotated to measure ion scattering... [Pg.1805]

Another approach to multielemental analysis is to use a multichannel instrument that allows for the simultaneous monitoring of many analytes. A simple design for a multichannel spectrometer consists of a standard diffraction grating and 48-60 separate exit slits and detectors positioned in a semicircular array around the diffraction grating at positions corresponding to the desired wavelengths (Figure 10.50). [Pg.436]

Calibration curves are usually constructed by analyzing a series of external standards and plotting the detector s signal as a function of their known concentrations. As long as the injection volume is identical for every standard and sample, calibration curves prepared in this fashion give both accurate and precise results. Unfortunately, even under the best of conditions, replicate injections may have volumes that differ by as much as 5% and often may be substantially worse. For this... [Pg.573]

With conventional nonspectroscopic detectors, other methods must be used to identify the solutes. One approach is to spike the sample by adding an aliquot of a suspected analyte and looking for an increase in peak height. Retention times also can be compared with values measured for standards, provided that the operating conditions are identical. Because of the difficulty of exactly matching such conditions, tables of retention times are of limited utility. [Pg.575]

Precision The precision of a gas chromatographic analysis includes contributions from sampling, sample preparation, and the instrument. The relative standard deviation due to the gas chromatographic portion of the analysis is typically 1-5%, although it can be significantly higher. The principal limitations to precision are detector noise and the reproducibility of injection volumes. In quantitative work, the use of an internal standard compensates for any variability in injection volumes. [Pg.577]

An internal standard of 1-butanol is used to determine the concentrations of one or more of the following impurities commonly found in whiskey acetaldehyde, methanol, ethyl acetate, 1-propanol, 2-methyl-1-propanol, acetic acid, 2-methyl-1-butanol and 3-methyl-1-butanol. A packed column using 5% Garbowax 20m on 80/120 Garbopak B and an EID detector were used. [Pg.611]

Caffeine in tea and coffee is determined by CZE using nicotine as an internal standard. The buffer solution is 50 mM sodium borate adjusted to pH 8.5 with H3PO4. A UV detector set to 214 nm is used to record the electropherograms. [Pg.614]

Zhou and colleagues determined the %w/w H2O in methanol by GG, using a capillary column coated with a nonpolar stationary phase and a thermal conductivity detector. A series of calibration standards gave the following results. [Pg.616]


See other pages where Detector standard is mentioned: [Pg.1574]    [Pg.238]    [Pg.10]    [Pg.348]    [Pg.29]    [Pg.225]    [Pg.213]    [Pg.1574]    [Pg.142]    [Pg.375]    [Pg.328]    [Pg.1574]    [Pg.238]    [Pg.10]    [Pg.348]    [Pg.29]    [Pg.225]    [Pg.213]    [Pg.1574]    [Pg.142]    [Pg.375]    [Pg.328]    [Pg.81]    [Pg.196]    [Pg.272]    [Pg.509]    [Pg.580]    [Pg.856]    [Pg.1122]    [Pg.1215]    [Pg.1420]    [Pg.1428]    [Pg.1979]    [Pg.411]    [Pg.586]    [Pg.589]    [Pg.597]   
See also in sourсe #XX -- [ Pg.969 , Pg.971 , Pg.972 , Pg.973 , Pg.974 , Pg.975 , Pg.978 , Pg.980 , Pg.981 , Pg.982 , Pg.983 , Pg.986 , Pg.987 ]




SEARCH



Standards for Calibration of Explosive Detectors

© 2024 chempedia.info