Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Column deterioration

To date most of the work which has been done with supercritical fluid extraction has concentrated on the extraction of analytes from solid matrices or liquids supported on an inert solid carrier matrix. The extraction of aqueous matrices presents particular problems [276-278]. The co-extraction of water causes problems with restrictor plugging, column deterioration, and phase separation if a nonpolar solvent is used for sample collection. Also, carbon dioxide isay have limited extraction efficiency for many water soluble compounds. [Pg.411]

Solvent removal.. Lzurge amounts of solvents and excess derivatizing reagents, etc., can be excluded from the main separation column. Prevents column deterioration and improves the performance of sensitive, selective detectors. [Pg.909]

To improve chromatographic separation, another analytical column could be used in addition to the monolith (Xu et al. 2006). The monolith column served as an extraction column only. Hsieh et al. (2000, 2002) utilized a polymer-coated mixed function (PCMF) Capcell C8 column (4.6 x 50 mm, Phenomenex) to provide dual functions—online plasma extraction and analyte separation. The silica was coated with a polymer containing both hydrophilic polyoxythylene and hydrophobic groups. The diluted plasma samples (1 1 to 1 3) were injected directly. No column deterioration was observed after 200 injections. [Pg.285]

The bioanalyst can be required to analyse most biofluids although the most common are urine and the aqueous phase of blood, i.e. plasma or serum. Other samples may be cell and tissue extracts, synovial fluid, cerebrospinal fluid (CSF) and saliva. In the case of urine and CSF with their very low protein content it might be possible to directly inject the sample into an HPLC column. With most silica-based packing materials, direct injection of blood proteins will rapidly lead to column deterioration. HPLC columns are expensive and their efficiency is easily lost so correct preparation of samples will not only improve column life but also improve the results. At its simplest it is only necessary to remove particulate matter from samples to prevent clogging of the column and frits. Modern HPLC packings are very susceptible to contamination by proteins, fats and other macromolecules from biological samples and it is necessary to remove these (except of course for protein analysis). [Pg.210]

There are two basic disadvantages to the coated capillary column. First, the limited solute retention that results from the small quantity of stationary phase in the column. Second, if a thick film is coated on the column to compensate for this low retention, the film becomes unstable resulting in rapid column deterioration. Initially, attempts were made to increase the stationary-phase loading by increasing the internal surface area of the column. Attempts were first made to etch the internal column surface, which produced very little increase in surface area and very scant improvement. Attempts were then made to coat the internal surface with di-atomaceous earth, to form a hybrid between a packed column and coated capillary. None of the techniques were particularly successful and the work was suddenly eclipsed by the production of immobilize films firmly attached to the tube walls. This solved both the problem of loading, because thick films could be immobilized on the tube surface, and that of phase stability. As a consequence, porous-layer open-tubular (PLOT) columns are not extensively used. The PLOT column, however, has been found to be an attractive alternative to the packed column for gas-solid chromatography (GSC) and effective methods for depositing adsorbents on the tube surface have been developed. [Pg.1067]

The most common problem associated with analytical columns is column deterioration. Deterioration may appear as poor peak shapes, split peaks, shoulders, loss of resolution, decreased retention times, and high backpressure. These symptoms indicate contaminants that have accumulated on the frit or column inlet, or there are voids, channels, or a depression in the packing bed. Deterioration is more evident in higher efficiency columns. For example, a column with 3-pm packing is more susceptible to plugging than one with 5- or 10-pm packing. Proper column protection and sample preparation are essential to prolong a column s life and obtain its best performance. [Pg.1664]

Figure 2.16. Example of column deterioration due to aging. The 0.25 mm I. D. fused silica columns are 15 m (A) and 30 m (B) coated with 1.0 xm DB-5 stationary phase. Column activity is determined using the Grob test method (Table 2.16). Column (A) is an old column exhibiting strong activity. Column (B) is a new column. The reduced peak heights of nonanal and 2,6-dimethylaniline on both columns also indicate change in relative composition of the test mixture due to storage time. Figure 2.16. Example of column deterioration due to aging. The 0.25 mm I. D. fused silica columns are 15 m (A) and 30 m (B) coated with 1.0 xm DB-5 stationary phase. Column activity is determined using the Grob test method (Table 2.16). Column (A) is an old column exhibiting strong activity. Column (B) is a new column. The reduced peak heights of nonanal and 2,6-dimethylaniline on both columns also indicate change in relative composition of the test mixture due to storage time.
Good columns eventually deteriorate over time due to modification of the stationary phase film or glass surface induced by moisture or matrix components introduced with the samples or by the carrier gas, thermal stress, attack by oxygen and aging. Column efficiency for inert compounds may remain acceptable but increasing adsorptive and catalytic activity eventually precludes the separation of polar compounds. Figure 2.16 shows an example of the influence of column deterioration on the separation of several... [Pg.154]

Adamsite 1 is the most difficult of the organoarsenicals to analyze. It has very high thermal stability and does not hydrolyze readily in the environment. Like most arsenicals, attempts to assess Adamsite by gas chromatography (GC) lead to rapid column deterioration and Adamsite carmot be derivatized with thiol derivatives <2003JCH(1000)253>. Schoene etal. developed two derivatization reactions for Adamsite. In the first, it was reacted with bromine in AcOH to give 2,2, 4,4, 6,6 -hexabromodiphenylamine 37. The alternative derivatization involved pyrolytic ethylation with dimethylformamide diethyl acetal (DMFDEA) and pyridine to yield 10-ethyl-5,10-dihydrophenarsazinine 38. The reaction mixture, which is Adamsite, EtOH, pyridine, and DMFDEA 39, was stored for 3 days in a closed vial at 90 °C. On column injection of 1 pi from this solution, no ethyl derivative could be detected, whereas splitless injection at 290°C injector temperature afforded EtPA 38 in the expected amount (Scheme 14) <1996JCH(719)401>. [Pg.881]

When properly cared for, a column can last for a long time. Since the major cause for column deterioration is contamination, the protection of the column from contamination is a primary task of colunm care. Column lifetimes of over 10,000 analyses have been achieved with columns protected by guard columns. Under the identical set of conditions, but without this protection, column life was limited to about 2000 injections. [Pg.383]

Columns deteriorate with use over time. Extremes of pH and constant use can cause the packing material to be stripped away from the silica backbone. This causes silanol sites, which interact with components within the... [Pg.195]

It is seen that the separation is completed in about 30 min., a much faster analysis than that of Gil-Av. It was found that the minimal operating temperature was 90°C, but the column could be programmed up to 175°C without column deterioration. As a result of the introduction of the chiral polysiloxane stationary phases by Frank et al, the development of chiral GC gained momentum. [Pg.111]

This type of chiral stationary, however, proved to be very effective. The separation of the enantiomers of hexabarbital on this stationary phase by the direct injection of blood serum is shown in figure 11.27. Chromatogram A was obtained after 20 injections of serum and chromatogram B after 60 consecutive injections of blood serum. It is seen that here is very little column deterioration and that, although the tail of the major peak has become a little extended after 60 injections, the column could still be used very effectively for the analysis (it should be noted, that it is essential to employ titanium frits in the column, to prevent protein precipitation, which could occur on stainless steel frits). [Pg.362]

Since the intrinsic viscosity measurements have been the cornerstone of industrial characterization measurements, it is expected that the measurement of this parameter and the other associated measurements, such as intrinsic viscosity distribution by GPC, will become more important from a quality control standpoint than the traditional measurement of molecular weight distribution, which has not previously been related to intrinsic viscosity. Of particular importance for quality control are the claims made for the technique, namely that IVD measurement is much less sensitive to column deterioration, flow rate variations and column loading. [Pg.75]

It is believed that when micellar mobile phases are used, the chromatographic columns deteriorate very easily. However, with an adequate experimental methodology (see Chapter 4), no apparent degradation of the chromatographic performance will occur. Hundreds of injections can be made without modification of the retention of the confounds or pressure buildup in the chromatographic system. In our laboratories, we have used the same columns for a year or even longer (at least 600 injections), without any apparent deterioration. [Pg.344]

The solvent or solution used for the sample to be introduced into the separation column has to be compatible with the aqueous mobile phase. If it is too strong an eluent and injected in relatively large volumes, peak deformation may occur as in other kinds of chromatography. The addition to the sample solution of a hydrophilic counterion or a strongly competing ionic component of the same charge as the ionic solute may give severe peak distortion, but peak compression under optimized conditions. Column deterioration is probably no more pronounced in ion pair LC than in other systems with bonded stationary phases and mixtures of aqueous buffer solutions and... [Pg.2584]

Thalidomide levels in raw material and finished products was detomined using a C]g column (2 =237nm) and an 85/15 water/acetonitrile mobile phase [1527]. Elution was complete in 8 min. The authors note that past mobile phases used a 0.1 % H3PO4 additive that caused column deterioration much fiister than the mobile phase described here. The effect of changing percent acetonitrile on retention time was also reported. A linear curve was established between 10 and 200pg/mL. [Pg.524]

Stripping off or exchange of the fixed cation. Regeneration of the cations with their respective nitrates should take place regularly to prevent and correct column deterioration. [Pg.157]

Retention times will migrate with column deterioration, and the life of the column is relatively short, especially for samples with high TDS. [Pg.709]

The biggest problem in using NP-HPLC is its dramatic sensitivity to water. Even water traces (in the mobile phase or from the sample) may bind to the column, deteriorate its performance, and cause irreproducibility. In addition, particular care must be taken to ensure accurate pH, as in NP-HPLC, retention is very sensitive to the charge state of the analyte. Owing to these practical problems NP-HPLC is relatively rarely used. Its main application fields are separation of polyaromatic hydrocarbons, sterols, vitamins, chlorophylls, ceramides, and other Upid extracts. [Pg.81]


See other pages where Column deterioration is mentioned: [Pg.434]    [Pg.425]    [Pg.324]    [Pg.70]    [Pg.104]    [Pg.90]    [Pg.20]    [Pg.75]    [Pg.172]    [Pg.293]    [Pg.232]    [Pg.185]    [Pg.210]    [Pg.129]    [Pg.243]    [Pg.178]    [Pg.4]    [Pg.192]    [Pg.287]    [Pg.288]    [Pg.273]    [Pg.690]    [Pg.97]    [Pg.134]    [Pg.52]    [Pg.172]    [Pg.201]   
See also in sourсe #XX -- [ Pg.135 ]




SEARCH



Deterioration

© 2024 chempedia.info