Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Column improvements

Nye Tray, 10-20% increased tray (over sieve or valve) capacity and good efficiency. More capacity from existing column. Improved inlet area for sieve or valve tray with greater area for vapor-liquid disengagement. [Pg.124]

The use of two or more columns improves the probability that the identity of an unknown compound is the same as that of a compound with identical retention times however, these data alone are not conclusive proof. The reliability of the identification depends on the efficiency and polarity of the columns used. With efficient columns, the probability of having two or more components under one peak diminishes and the peaks are generally well resolved. Care must be taken in selecting columns to be certain the columns have different selectivities and not just different names. Thus, the McReynold s constants (Chapter 3) must be compared and should be quite different for each column. For example, in the analysis of pesticides, four different liquid phases might be chosen arbitrarily (e.g., OV-1, UCW-98, SE-30, and DC-200). However, the relative retention in this case will be the same for all four columns since they are all methyl silicones and have essentially the same McReynold s constants (2). [Pg.157]

While various dropping mercury electrodes (DMEs) have long been proposed as LC detectors, it is fair to say that they have never been put to significant practical use for this purpose. Kemula innovated this area in 1952 [16]. As the efficiency of LC columns improved, it became essential to reduce the volume... [Pg.829]

Use fresh sample or standard to confirm sample as source of problem. If some or all peaks are still smaller than expected, replace column. If new column improves analysis, try to restore the old column, following appropriate procedure. If performance does not improve, discard old column. [Pg.127]

Kushi, H., Miki, T., Nakahara, J., Okamoto, K., Kawahara, Y., Saito, T., Tanjoh, K. Hemoperfusion with an immobilized polymyxin B fiber column improves tissue oxygen metabolism. Ther Apher Dial 10 (2006a) 430-435. [Pg.335]

Sato, H., Oshima, K., Kobayashi, K., Yamazaki, H., Suto, Y., Takeyoshi, I. Hemoperfusion with polymyxin B-immobilized fiber column improves liver function after ischemia-reperfusion injury. World J Gastroenterol 15 (2009) 4571-4575. [Pg.337]

Shimizu, T., Hanasawa, K., Sato, K., Umeki, M., Koga, N., Naganuma, T., Sato, S., Shimonishi, T., Ikeda, T., Matsuno, N., Ono, S., Saitoh, H., Satoh, K., Otani, Y., Endo, Y., Eguchi, Y., Tani, T. Direct hemoperfusion with polymyxin-B-immobilized fiber columns improves septic hypotension and reduces inflammatory mediators in septic patients with colorectal perforation. Langenbecks Arch Surg 394 (2009) 303-311. [Pg.337]

With the columns improved as recommended above, it is desirable to repeat many off-line experiments performed earlier. To avoid necessity of extra corrections, it is preferable to simultaneously chromatograph as many elements as possible, and to run experiments with this mixture under various experimental conditions. In particular, because our major concern is the column surface, of great interest would be experiments in the columns made of metals or temperature-resistant fluorinated polymers. The polymeric columns remain unemployed in gas-phase radiochemistry. Their surface will probably not be modified, which can lead to unexpected phenomena or regularities. [Pg.180]

Variation in Plate Height As shown by Equation 30-30, the resolution of a column improves as the square root of the number of plates it contains increases. Example 30-2e reveals, however, that increasing the number of plates is expensive in terms of time unless the increase is achieved by reducing the plate height and not... [Pg.939]

During the 1970s the ECD became firmly established as the most sensitive gas chromatographic detector for some compounds. The kinetic model was described in terms of a numerical solution of the differential equations. This was assisted by the development of the constant current mode of measuring the response and the development of Ni-63 sources for the detector. The purification of the carrier gas and the further development of capillary columns improved the operation of the ECD. In addition, chemical reactions were used to make derivatives with a greater sensitivity in the ECD. Other ion molecule reactions were used to improve the sensitivity of... [Pg.38]

Explain, in terms of Henry s Law coefficients, whether or not increasing the temperature in a stripping column improves the separation at each stage. [Pg.180]

Anothaf option is the use of buffies. The use of ha flies2 in continuous flow foam columns improves removal efficiency and increases maximum hydraulic loading rates by reducing channeling in the fonm and other types of axial dispersion. [Pg.818]

The feed stage location of the column is optimized first in the simulation prior to the start of the column thermal analysis since the feed stage may strongly interact with the other options for column improvements. This can be carried out by trying alternate feed stage locations in simulation and evaluating its impact on reboiling duty. [Pg.301]

Peaks characteristic of different solutes grow further apart linearly in retention time, while the peak width grows as the square root of the retention time. Thus, the selectivity of a column improves with column length. In the analytical mode, individual peaks can be detected in a flame or IR detector. In the preparative mode, different eluted solutes can... [Pg.23]

LC with electrochemical detection has now reached the stage where it is becoming the assay of choice for brain amines and their metabolites. However, it is possible to expect significant advances in the methods in the future especially as the reliability of the chromatographic columns improves and various technological ideas are adopted. Mention has been... [Pg.38]

The objective here is to focus the partially resolved peaks of interest onto a concentrator column in the second stage. This is accomplished by valve switching. The second column is of smaller diameter and is operated at a lower flow rate relative to the first column. Resolution of the peaks of interest is completed on the second column. Improved resolution is frequently possible by using eluents with different chemical compositions in the two columns. [Pg.33]

This means that a supercritical fluid mobile phase results in lower column backpressure than a liquid. Due to the higher diffusion rates in a supercritical fluid than in a liquid, the column effldency should, in principle, be higher in supercritical fluid chromatography (SFC) than in liquid chromatography (LC), but lower than that in gas chromatography (GC). However, for long columns with a density gradient over the column, improved efficiency compared to HPLC is not necessarily the case. [Pg.115]

Adding a ratio block between feed and heat input to the low-pressure column improves the response of the system. Figure 6.23 shows one way to implement this structure using an additive feedforward scheme. The output signal of the multiplier block ratioQtot/F is the calculated total heat input to the reboiler of the low-pressure column. This is fed... [Pg.185]

The use of two or more columns improves the probability that the identity of an unknown compound is the same as that of a compound with identical retention times. However, these data alone are not conclusive proof. The reliability of the identification depends on the efficiency and polarities of the column used. With... [Pg.409]

Programmed-temperature vaporizer electronic pressure-controlled sample inlet systems -ms grade columns, solid-phase microextraction sampling techniques, large-volume injectors, integrated guard columns, multidimensional GC, sol-gel columns, improvements in silphenylene phases, advances in GC-MS, more affordable benchtop GC-MS systems... [Pg.97]

The above indicates that, while heat removal in the upper sections of multiproduct columns improves fractionation, the means for heat removal could undermine it. The net outcome depends... [Pg.243]

The majority of LC separations of fat-soluble vitamins are based on RP chromatography on C18 columns, but the use of a triacontyl stationary phase (C30) is becoming more common, due to its superior shape selectivity [90,91]. Shape discrimination with C30 columns improves at subambient temperature, permitting resolution of geometric and positional isomers partly or fully, depending on the mobile phase composition. Nevertheless, C30 columns are less efficient and the chromatographic peaks broader than those obtained with C18 columns. [Pg.491]

Column 2, 30 m, 0.53 mm ID, 5pm film thickness methyl silicone. This column improves the separation of methyl acetylene, wo-pentane, and n-pentane. [Pg.1035]


See other pages where Column improvements is mentioned: [Pg.76]    [Pg.901]    [Pg.254]    [Pg.109]    [Pg.48]    [Pg.74]    [Pg.464]    [Pg.130]    [Pg.471]    [Pg.87]    [Pg.314]    [Pg.616]    [Pg.116]    [Pg.73]    [Pg.26]    [Pg.47]    [Pg.300]    [Pg.313]    [Pg.1069]    [Pg.10]    [Pg.720]    [Pg.111]   
See also in sourсe #XX -- [ Pg.428 ]




SEARCH



Column efficiency improvement

Improving Separation by Increasing the Column Efficiency

Separation Based on an Improved Column Efficiency

© 2024 chempedia.info