Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Assay performance

The relationship between biomarker responses and effects at the population level can be tested in both field experiments and more controlled experiments in mesocosms. It may be possible to define thresholds for biomarker assays performed on indicator species, above which population effects have been shown to occur. Indicator species may be either free living or deployed. The advantage of the latter is... [Pg.315]

Table 4. Kinetic parameters of mung bean PME isoforms. Assays performed at pH 7.6 in 150 mM NaCl values between brackets have been obtained at pH 5.6. Table 4. Kinetic parameters of mung bean PME isoforms. Assays performed at pH 7.6 in 150 mM NaCl values between brackets have been obtained at pH 5.6.
Immunoassays for diuron (Figure 13) are another example of improved assay performance using heterologous assay conditions. One antibody was derived from a hapten that extended the dimethylamine side chain of diuron with methylene groups. [Pg.637]

Hapten density is important for both immunization and assay performance, and hence the extent of conjugation or hapten density should be confirmed by established methods. A characteristic ultraviolet (UV) or visible absorbance spectrum that distinguishes the hapten from the carrier protein or use of a radiolabeled hapten can be used to determine the degree of conjugation. If the hapten has a similar A. iax to the protein, the extent of incorporation can still be estimated when the concentration of the protein and the spectral characteristics of the hapten and protein are known. The difference in absorbance between the conjugate and the starting protein is proportional to... [Pg.643]

Beasley et al. developed a panel of immunoassays to monitor DDT, its metabolites, and structurally related compounds, but they found that milk has a severe effect on the assay performance. They found that when directly utilizing whole milk, color development was completely inhibited. Even when using 1 100 dilutions of whole milk, the assay sensitivity was reduced by 90% (based on the IC50 shift, not simply the dilution factor). A number of procedures were evaluated to eliminate the interferences from the fat-soluble analytes. However, many of the procedures that removed interferences also removed the analytes. Extraction with a mixture of solvents and the use of similarly processed blank milk to prepare the standards ultimately yielded more accurate results. This article demonstrates the difficulties encountered in analyzing lipid-soluble analytes. [Pg.698]

The errors mentioned above represent the reproducibility obtained on the same microtiter plate when the sample molecule is assayed in several different wells. When the reproducibility of Pe measurement is assessed on the basis of assays performed at different times over a long period of time, more systematic sources of errors show up, and the reproducibility can be about 2-3 times worse. Figure 7.56 shows reproducibility of standard compounds taken over a period of about 12 months. Carbamazepine show a long-term reproducibility error of 15%. The other compounds show somewhat higher errors. [Pg.232]

Data was reported as transformation frequency (TF), calculated on the cells that survived after chemical exposure. TF is expressed as a function of the total number of foci per treatment divided by the number of surviving cells estimated from the clonal efficiency observed in the cytotoxicity assay performed in parallel with the transformation test [52],... [Pg.193]

Berkowitz FE, Levin MJ. Use of an enzyme-linked immunosorbent assay performed directly on fixed infected cell monolayers for evaluating drugs against varicella-zoster virus. Antimicrob. Agents Chemother. 1985 28 207-210. [Pg.86]

These biosensors were tested for glucose and lactate measurements in sera, and for lactate measurements in whey solutions. Good agreements were obtained between the present method and reference methods. For glucose analysis in serum, the coefficient of variation for 53 repeated measurements performed over a 10 h period was 4.8% while for lactate analysis, 80 assays performed over a 15 h period gave a coefficient of variation of 6.7%. Thus,... [Pg.170]

Foot and mouth disease virus VP1 epitope Cowpea mosaic virus in cowpea leaf No immunogenicity assays performed. - 79... [Pg.136]

Hepatitis E virus (HEV) ORF2 epitope Tomato leaf and fruit No immunogenicity assays performed. 6... [Pg.145]

Human immunodeficiency virus (HIV) type 1 p24 protein Tomato bushy stunt virus in tobacco leaf No immunogenicity assays performed. 97, 98... [Pg.145]

V. cholerae B subunits of the cholera toxin (CTB) Tobacco leaf chloroplast No immunogenicity assays performed. 119... [Pg.149]

E. coli B subunit of the heat labile enterotoxin (LTB) fused to an immu-nocontraceptive epitope Tomato fruit and leaves No immunogenicity assays performed. 91... [Pg.150]

The intensive research carried out to achieve good stability and easy handling of the BL reagents allowed development of many simple, rapid, highly sensitive and specific assays, many of which are available as commercial kits. More recently application of the newest techniques of molecular biology greatly accelerated the continuous improvement of BL assay performance. [Pg.276]

In the future, it is likely that many labor-intensive functional immunological assays performed in vivo or ex vivo will be replaced by high throughput multiparameter cytometric-based assays. The challenges faced both today and in the future have been recently described.56 For example, translational aspects of the nonclinical data to the clinical setting are difficult because despite much similarity, there are still species differences in some of the very basic aspects of the immune system. Additionally, reagents... [Pg.117]

Matrix effect — To demonstrate that the assay performance was independent from the sample matrix, QC samples were prepared using two different lots of matrix. The QC samples were evaluated using the same calibration curve. With regard to analytical recovery, no significant difference was observed for the QCs prepared in two lots of plasma. [Pg.82]

Standards performance — The assay performed well. Mean bias at LLOQ was -0.2% and for other standards varied from -1.2 to 1.0%. Coefficient of determination (r2) was greater than 0.989372 for all batches run. [Pg.82]

Many other types of solid phase adsorbents, including those based on conventional and specialty materials like restricted access media (RAM), can increase analysis speed and improve assay performance. These types of materials, also known as internal reversed-phase packings, are especially useful for assaying target compounds in biological samples such as serum and plasma. They are chemically modified porous silicas that have hydrophilic external surfaces and restricted-access hydrophobic internal surfaces. The ratio of interior to external surface areas is large. Macromolecules such as proteins cannot enter the pores of the RAM (they are excluded from the hydrophobic internal surface) and they elute quickly through the column. However, the smaller analyte molecules that can enter the pores are retained via interactions with the hydrophobic bonded phase within... [Pg.350]

The results from all biological assays performed showed that chemically synthesized E. coli lipid A (compound 506 or LA-15-PP) expresses, with similar doses, the same spectrum of endotoxic effects as bacterial (E. coli) free lipid A (5,234-237). Thus, lipid A constitutes the lethal, pyrogenic, leukopenic, and mediator-inducing, that is, the endotoxically essential region of LPS, its endotoxic properties being embedded in a molecule having the structure shown in Fig. 2. [Pg.257]

The degradative bacterial activity is increased with water temperature [2,4,5,7,22], as it increases all the metabolic processes. Consequently, an increase in this factor encourages both the extent and the kinetics of the process (see Table 5.3.2). In assays performed at high concentrations... [Pg.630]

Fig. 39 (a) Design and principle of operation of the BL-QD system and (b) spectral manifestation of the assay, (c) Two assays performed under identical conditions in mouse serum green) and mouse whole blood red). (Reprinted with permission from [222]. Copyright 2006 Macmillan Publishers Ltd)... [Pg.94]

Although most assays perform well with regard to specificity and reproducibility, the major problem remains their standardization (A9, Dl, K30, L4). There is currently no internationally accepted standard, and the selection of a reference material raised many problems (A8, G5, K30, L4). A number of questions have not been solved Should the standard consist of several apo(a) isoforms Can the reference material be lyophilized Should results be expressed as mass or as moles of apoprotein or lipoprotein How should the protein mass of the primary standard be determined What are optimal storage conditions for the secondary standard Which method can be used as a reference method Can recombinant apo(a) represent an alternative for a primary standard These problems came to light in the course of the international surveys whose results were presented at the Lp(a) Workshop in New Orleans (1992) (L4). [Pg.109]

Reliable statistical calculation of the data is performed by the cytometer s software, resulting in a simple evaluation of the association assays performed to characterize the interaction between targeted systems and the cells. The individual cell populations are usually described by the relative mean, either arithmetic or geometric, and the coefficient of variation of a particular parameter, as well as the number of cells within the respective population. [Pg.655]


See other pages where Assay performance is mentioned: [Pg.244]    [Pg.395]    [Pg.57]    [Pg.584]    [Pg.160]    [Pg.762]    [Pg.637]    [Pg.55]    [Pg.57]    [Pg.287]    [Pg.94]    [Pg.65]    [Pg.70]    [Pg.123]    [Pg.313]    [Pg.102]    [Pg.73]    [Pg.73]    [Pg.82]    [Pg.87]    [Pg.663]    [Pg.382]    [Pg.439]   
See also in sourсe #XX -- [ Pg.48 ]




SEARCH



Analytical performance assays, calibration curves

Characterization and Performance Assays

Enzyme-linked immunosorbent assays performance/limitations

Estimating Assay Performance Characteristics

High performance liquid chromatography assay preparation

High-performance liquid chromatography assays

Performance characteristics, ligand binding assays

Psychomotor Vigilance Performance Neurocognitive Assay Sensitive to Sleep Loss

Quality assurance assay performance

Reverse-phase high-performance liquid assay

The importance of assay performance in aquatic disease diagnosis

© 2024 chempedia.info