Amperometric process

Figure 4.2 — (A) Schematic diagram of an ammonia-N-sensitive probe based on an Ir-MOS capacitor. (Reproduced from [20] with permission of the American Chemical Society). (B) Pneumato-amperometric flow-through cell (a) upper Plexiglas part (b) metallized Gore-Tec membrane (c) auxiliary Gore-Tec membrane (d) polyethylene spacer (e) bottom Plexiglas part (/) carrier stream inlet (g) carrier stream outlet. (C) Schematic representation of the pneumato-amperometric process. The volatile species Y in the carrier stream diffuses through the membrane pores to the porous electrode surface in the electrochemical cell and is oxidized or reduced. (Reproduced from [21] with permission of the American Chemical Society). Figure 4.2 — (A) Schematic diagram of an ammonia-N-sensitive probe based on an Ir-MOS capacitor. (Reproduced from [20] with permission of the American Chemical Society). (B) Pneumato-amperometric flow-through cell (a) upper Plexiglas part (b) metallized Gore-Tec membrane (c) auxiliary Gore-Tec membrane (d) polyethylene spacer (e) bottom Plexiglas part (/) carrier stream inlet (g) carrier stream outlet. (C) Schematic representation of the pneumato-amperometric process. The volatile species Y in the carrier stream diffuses through the membrane pores to the porous electrode surface in the electrochemical cell and is oxidized or reduced. (Reproduced from [21] with permission of the American Chemical Society).
It is used in IC systems when the amperometric process confers selectivity to the determination of the analytes. The operative modes employed in the amperometric techniques for detection in flow systems include those at (1) constant potential, where the current is measured in continuous mode, (2) at pulsed potential with sampling of the current at dehned periods of time (pulsed amperometry, PAD), or (3) at pulsed potential with integration of the current at defined periods of time (integrated pulsed amperometry, IPAD). Amperometric techniques are successfully employed for the determination of carbohydrates, catecholamines, phenols, cyanide, iodide, amines, etc., even if, for optimal detection, it is often required to change the mobile-phase conditions. This is the case of the detection of biogenic amines separated by cation-exchange in acidic eluent and detected by IPAD at the Au electrode after the post-column addition of a pH modiher (NaOH) [262]. [Pg.409]

In the area of consumer products, amperometric glucose sensors hold high potential. Industrially, process monitors for the manufacture of consumer chemicals are under development. However, replacement of defective reference electrodes, which in a laboratory environment may be trivial, may be prohibitively difficult m vivo or in an industrial process environment. [Pg.58]

The continuous methods combine sample collection and the measurement technique in one automated process. The measurement methods used for continuous analyzers include conductometric, colorimetric, coulometric, and amperometric techniques for the determination of SO2 collected in a liquid medium (7). Other continuous methods utilize physicochemical techniques for detection of SO2 in a gas stream. These include flame photometric detection (described earlier) and fluorescence spectroscopy (8). Instruments based on all of these principles are available which meet standard performance specifications. [Pg.201]

Pulsed amperometric detection (PAD), introduced by Johnson and LaCourse (64, 65) has greatly enhanced the scope of liquid chromatography/electrochemistry (66). This detection mode overcomes the problem of loss of activity of noble metal electrodes associated with the fixed-potential detection of compounds such as carbohydrates, alcohols, amino acids, or aldehydes. Pulsed amperometric detection couples tlie process of anodic detection with anodic cleaning and cathodic reactivation of a noble metal electrode, thus assuring a continuously cleaned and active... [Pg.92]

Figure 2.4 Noradrenergic inhibition of Ca " currents and transmitter release in sympathetic neurons and their processes, (a) Inhibition of currents through N-type Ca " channels by external application of noradrenaline (NA) or by over-expression of G-protein P y2 subunits, recorded from the soma and dendrite of a dissociated rat superior cervical sympathetic neuron. Currents were evoked by two successive 10 ms steps from —70 mV to OmV, separated by a prepulse to -1-90 mV. Note that the transient inhibition produced by NA (mediated by the G-protein Go) and the tonic inhibition produced by the G-protein Piy2 subunits were temporarily reversed by the -1-90 mV depolarisation. (Adapted from Fig. 4 in Delmas, P et al. (2000) Nat. Neurosci. 3 670-678. Reproduced with permission), (b) Inhibition of noradrenaline release from neurites of rat superior cervical sympathetic neurons by the 2-adrenoceptor stimulant UK-14,304, recorded amperometrically. Note that pretreatment with Pertussis toxin (PTX), which prevents coupling of the adrenoceptor to Gq, abolished inhibition. (Adapted from Fig. 3 in Koh, D-S and Hille, B (1997) Proc. Natl. Acad. Sci. USA 1506-1511. Reproduced with permission)... Figure 2.4 Noradrenergic inhibition of Ca " currents and transmitter release in sympathetic neurons and their processes, (a) Inhibition of currents through N-type Ca " channels by external application of noradrenaline (NA) or by over-expression of G-protein P y2 subunits, recorded from the soma and dendrite of a dissociated rat superior cervical sympathetic neuron. Currents were evoked by two successive 10 ms steps from —70 mV to OmV, separated by a prepulse to -1-90 mV. Note that the transient inhibition produced by NA (mediated by the G-protein Go) and the tonic inhibition produced by the G-protein Piy2 subunits were temporarily reversed by the -1-90 mV depolarisation. (Adapted from Fig. 4 in Delmas, P et al. (2000) Nat. Neurosci. 3 670-678. Reproduced with permission), (b) Inhibition of noradrenaline release from neurites of rat superior cervical sympathetic neurons by the 2-adrenoceptor stimulant UK-14,304, recorded amperometrically. Note that pretreatment with Pertussis toxin (PTX), which prevents coupling of the adrenoceptor to Gq, abolished inhibition. (Adapted from Fig. 3 in Koh, D-S and Hille, B (1997) Proc. Natl. Acad. Sci. USA 1506-1511. Reproduced with permission)...
Hitherto, there are primarily three ways in which an amperometric electrode can be used to simultaneously induce and monitor interfacial processes. These are illustrated schematically in Fig. 2 for the most general case where diffusion may occur in both of the liquid phases, which comprise the interface of interest. [Pg.291]

In scanning electrochemical microscopy (SECM) a microelectrode probe (tip) is used to examine solid-liquid and liquid-liquid interfaces. SECM can provide information about the chemical nature, reactivity, and topography of phase boundaries. The earlier SECM experiments employed microdisk metal electrodes as amperometric probes [29]. This limited the applicability of the SECM to studies of processes involving electroactive (i.e., either oxidizable or reducible) species. One can apply SECM to studies of processes involving electroinactive species by using potentiometric tips [36]. However, potentio-metric tips are suitable only for collection mode measurements, whereas the amperometric feedback mode has been used for most quantitative SECM applications. [Pg.397]

Because process mixtures are complex, specialized detectors may substitute for separation efficiency. One specialized detector is the array amperometric detector, which allows selective detection of electrochemically active compounds.23 Electrochemical array detectors are discussed in greater detail in Chapter 5. Many pharmaceutical compounds are chiral, so a detector capable of determining optical purity would be extremely useful in monitoring synthetic reactions. A double-beam circular dichroism detector using a laser as the source was used for the selective detection of chiral cobalt compounds.24 The double-beam, single-source construction reduces the limitations of flicker noise. Chemiluminescence of an ozonized mixture was used as the principle for a sulfur-selective detector used to analyze pesticides, proteins, and blood thiols from rat plasma.25 Chemiluminescence using bis (2,4, 6-trichlorophenyl) oxalate was used for the selective detection of catalytically reduced nitrated polycyclic aromatic hydrocarbons from diesel exhaust.26... [Pg.93]

The above system of directly sensing a process stream without more is often not sufficiently accurate for process control so, robot titration is preferred in that case by means of for instance the microcomputerized (64K) Titro-Analyzer ADI 2015 (see Fig. 5.28) or its more flexible type ADI 2020 (handling even four sample streams) recently developed by Applikon Dependable Instruments20. These analyzers take a sample directly from process line(s), size it, run the complete analysis and transmit the calculated result(s) to process operation (or control) they allow for a wide range of analyses (potentiometric, amperometric and colorimetric) by means of titrations to a fixed end-point or to a full curve with either single or multiple equivalent points direct measurements with or without (standard) addition of auxiliary reagents can be presented in any units (pH, mV, temperature, etc.) required. [Pg.374]

M.T. Perez Pita, A.J. Reviejo, F.J.M. de Villena, and J.M. Pingarron, Amperometric selective biosensing of dimethyl- and diethyldithiocarbamates based on inhibition processes in a medium of reversed micelles. Anal. Chim. Acta 340, 89-97 (1997). [Pg.74]

An amperometric method or amperometry is concerned with the measurement of current under a constant applied voltage and under such experimental parameters the concentration of the analyte exclusively determines the quantum and magnitude of the current. Hence, these measurements may be employed effectively to record the alteration in concentration of an ion in question in the course of a titration, and ultimately the end-point is established. This specific process is commonly referred to as amperometric method or amperometry. [Pg.253]

Figure 3.8 Amperometric detectors (a) measure the current that flows between the working electrode, usually a glassy carbon electrode, and a reference electrode, at a fixed voltage, usually close to the discharge potential for the compound. Coulometric detectors (b) are less common and are designed with a porous carbon flow cell so that all the analyte reacts in the cell, the amount of current consumed during the process being proportional to the amount of the substance. Figure 3.8 Amperometric detectors (a) measure the current that flows between the working electrode, usually a glassy carbon electrode, and a reference electrode, at a fixed voltage, usually close to the discharge potential for the compound. Coulometric detectors (b) are less common and are designed with a porous carbon flow cell so that all the analyte reacts in the cell, the amount of current consumed during the process being proportional to the amount of the substance.
The potentialities offered by chemical processes catalyzed by enzymes immobilized in a polymeric matrix are obvious and are now successfully utilized in various ways [6, 13, 16, 17, 40, 43,44, 50, 58, 61], This idea was introduced into electroanalytical chemistry by Clark and Lyons [9], who proposed a glucose electrode, with glucose oxidase immobilized between cuprophane membranes and with amperometric determination of the hydrogen peroxide formed by the reaction... [Pg.202]


See other pages where Amperometric process is mentioned: [Pg.226]    [Pg.226]    [Pg.485]    [Pg.494]    [Pg.172]    [Pg.173]    [Pg.190]    [Pg.112]    [Pg.105]    [Pg.295]    [Pg.397]    [Pg.401]    [Pg.593]    [Pg.812]    [Pg.495]    [Pg.41]    [Pg.457]    [Pg.467]    [Pg.261]    [Pg.269]    [Pg.370]    [Pg.379]    [Pg.540]    [Pg.582]    [Pg.513]    [Pg.105]    [Pg.145]    [Pg.2]    [Pg.154]    [Pg.331]    [Pg.1073]    [Pg.148]    [Pg.535]    [Pg.385]    [Pg.392]    [Pg.410]    [Pg.650]   
See also in sourсe #XX -- [ Pg.226 ]