Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sampling controller

Selection of film systems for the random sample control measurements... [Pg.554]

But this is deceptive, for the simple reason that in order to sample the 5 microliters, one requires 100 microliters or more to be present in the cup. Therefore, the sample size required is not the size that is being sampled, but the size that is needed in the sampling cup. This problem may be approached in several ways. The problem of sample control will now be discussed. [Pg.105]

Sample Control. In figure 7, the technician is sampling with a sampler diluter from the capillary. Here we see what is called "sample control". None of the specimen is being wasted on the walls of a cup. All of it is being taken eventually into the capillary and then dispensed. With a setup as shown in Figure 7, synchronized with this diluter, is a multiple dispenser, which can dispense several reagents at the same time that the specimen is added to the tube. [Pg.105]

Procedure for instrumental analysis of samples, controls, and standards. [Pg.87]

Cold split Column independent 100-fold increase in sensitivity with respect to conventional GC injection Reproducible and accurate sampling Controlled evaporation No discrimination on basis of b.p. Rapid transfer of sample to column Low volume of CIS liners (2-3 xL)... [Pg.189]

The strategy depends on the situation and how we measure the concentration. If we can rely on pH or absorbance (UV, visible, or Infrared spectrometer), the sensor response time can be reasonably fast, and we can make our decision based on the actual process dynamics. Most likely we would be thinking along the lines of PI or PID controllers. If we can only use gas chromatography (GC) or other slow analytical methods to measure concentration, we must consider discrete data sampling control. Indeed, prevalent time delay makes chemical process control unique and, in a sense, more difficult than many mechanical or electrical systems. [Pg.102]

The turbidity gradually increased as shown in Table 9.19, with an increase in the volume of solution or amount of CH3COONa in all samples (control, sonicated and boiled). But the increase in precipitation was not in proportion to the reagent added to the A12(S04)3 solution. The turbidity in the first solution for unsonicated, sonicated and boiled solution was roughly in the ratio 1 lVi 3. Turbidity in sonicated and boiled samples was almost equal in solution containing 10 ml of aluminium sulphate and 5.0 ml of sodium acetate. Turbidity in sonicated samples (10 + 10) increased marginally compared to the boiled sample. This trend of sonicated samples, as seen in Table 9.19, indicated the role of ultrasonic power to be important. When 20 ml of A12(S04)3 solution was mixed with 5 ml of sodium acetate and sonicated, the amount of precipitate formed was negligible compared to the precipitation in a mixture of 10 ml of A12(S04)3 and 2.5 ml of sodium acetate, where the intensity of the ultrasonic power was almost double. [Pg.256]

The first STM experiments were performed under UHV conditions, and so the bias potential was simply applied as a difference across the tip and sample. However, introducing an electrolyte above the sample brought with it some particular problems. It is no longer sufficient simply to apply a bias voltage equal to the potential difference between tip and sample as this means that the potentials of the tip and sample are undefined with respect to any fixed reference, a wholly undesirable situation. Consequently, modern electrochemical STM systems operate under bipotentiostatic control with the tip and sample controlled and monitored independently with respect to the reference electrode. The bias potential is then still given by (Fs — FT), but VT and Fs are now potentials with respect to the reference electrode. [Pg.79]

Itaya and colleagues decided to employ STM to see if any correlation existed between the surface topography and the electrochemistry in Figures 2.27(a) and (b). The images were obtained at constant current with the tip and sample controlled independently. [Pg.82]

Fig. 3 Variation in the sample flow rate with the angle of rotation of the sample control valve. The integers within the graph refer to the number of rotations. The y axis is shown on a log scale. The parallel, dashed lines, show the eleven flow rates corresponding to the eleven dilutions... Fig. 3 Variation in the sample flow rate with the angle of rotation of the sample control valve. The integers within the graph refer to the number of rotations. The y axis is shown on a log scale. The parallel, dashed lines, show the eleven flow rates corresponding to the eleven dilutions...
Antipsychotics have long since replaced ECT for the treatment of schizophrenia. Several studies, however, have found ECT equal in efficacy to these agents, while one large-sample, controlled trial found it less effective than drugs, but more effective than psychotherapy ( 406). Some clinicians believe that selected patients may benefit when ECT is given concurrently with an antipsychotic. One controlled study, for example, found that ECT in combination with a phenothiazine led to a more rapid remission than the phenothiazine alone ( 407). Clinical experience has clearly documented an important role for ECT in catatonic excitement or withdrawal, as well as for other severe, life-endangering psychotic states. More recently, ECT combined with novel antipsychotics has been reported to benefit previously poorly responsive psychotic patients and was well tolerated (106, 408, 409). [Pg.80]

Fig. 6.2.2a-bc High-performance liquid chromatography (HPLC) with electrochemical (EC) detection of neurotransmitter metabolites, a standard mixture b cerebrospinal fluid (CSF) sample - control c CSF sample - aromatic amino acid decarboxylase (AADC) deficiency. Peak identification 1 = 5HIAA (7.7 min), 2 = 3-MD (9.6 min), 3 = HVA (11.7 min)... [Pg.708]

Statement on DNA Sampling Control and Access. Eubios Journal of Asian and... [Pg.212]

Add 80 pi m-hydroxydiphenyl solution to 2 tubes of each sample and the 2 reagent control tubes. Add to the third tube of each sample 80 p.1 of 0.5% NaOH (this is the sample control). Vortex the contents of the tubes three times ensure they are mixed well. [Pg.736]

Between 10 min and 1 hr after complete mixture, read the absorbances at 525 nm against the reagent control. Subtract the values for the sample controls from their corresponding sample absorbances. [Pg.736]

Step 9 Method maintenance Incorporate the new method or analyzer into the existing method maintenance systems for the site, to ensure that the method as practiced continues to meet the technical requirements for as long as it is in use. This is done by the receiver. Method maintenance systems may include check sample control-charting, intra-and/or inter-lab uniformity testing, on-site auditing, instrument preventive maintenance (PM) scheduling, control-charting the method precision and/or accuracy, etc. [Pg.396]

Another possibility for reducing expenditure in sampling and subsequent analysis is transition from 100 percent inspection to the method of sample control according to FELIX and LEMARIE [1964],... [Pg.360]

A wide range of chemical changes are possible. For inorganic samples, controlling the pH can be useful in preventing chemical reactions. For example, metal ions may oxidize to form insoluble oxides or hydroxides. The sample is often acidified with HNO3 to a pH below 2, as most nitrates are soluble, and excess nitrate prevents precipitation. Other ions, such as sulfides and cyanides, are also preserved by pH control. Samples collected for NH3 analysis are acidified with sulfuric acid to stabilize the NH3 as NH4SO4. [Pg.20]

Antidepressant stability after three freeze-thaw cycles was evaluated in triplicate at low and high concentrations. Calculated concentrations in the samples subjected to these conditions (stability samples) were compared to those obtained in freshly prepared samples (control samples). Stability and control samples were quantified with a calibration curve prepared on the day of the analysis. All analytes were stable under these conditions, except sertraline, for which a slight signal decrease was observed at 250 ng/mL in oral fluid (MRE = -33.4% %CV = 6.0%). [Pg.167]


See other pages where Sampling controller is mentioned: [Pg.2551]    [Pg.768]    [Pg.434]    [Pg.107]    [Pg.811]    [Pg.52]    [Pg.78]    [Pg.7]    [Pg.144]    [Pg.497]    [Pg.39]    [Pg.173]    [Pg.429]    [Pg.149]    [Pg.111]    [Pg.909]    [Pg.911]    [Pg.1150]    [Pg.341]    [Pg.341]    [Pg.78]    [Pg.85]    [Pg.63]    [Pg.58]    [Pg.200]    [Pg.260]    [Pg.13]   
See also in sourсe #XX -- [ Pg.114 , Pg.115 ]




SEARCH



Algorithm variable sample-time control

Assays sample control

Biomarkers sample quality controls

Blanks, field quality control sampling

Control sample

Control sample

Control sample comparison

Control sample comparison software

Control samples, use

Controller, with complementary sampling

Controlling the Sample Environment

Damping with sampling controller

Event-Controlled Sampling Surface Water Runoff from Agricultural Land

Event-controlled sampling

External control samples

Feedback control sampled data systems

Groundwater Sampling and Quality Control of Water Analyses

Hand wash samples control

In-process control sampling

Laboratory control sample

Laboratory control sample duplicate

Linear controllers sampling

Mass spectrometry quality control samples

Method Validation and Sample Analysis in a Controlled Laboratory Environment

Molecular speciation, control sample

Process control and sampling

Process control sample size selection

Process control sampling errors

Process control, automatic sampled data

Pump, sampling control

Quality Control (QC) Samples

Quality Control samples

Quality control field fortified sample

Quality control protocol design sample

Quality control protocol design samples used

Quality control sample analysis

Quality control sample preparation

Quality control sample stability

Quality control sample types

Quality control sampling activities

Quality control, sampling

Results for control samples

Sample Hazard Control Policy Statement

Sample Hazard Control Status Report

Sample Points for Bottom Product Analyzer Control

Sample Points for Top Product Analyzer Control

Sample compartment, control

Sample control, field studies

Sample controlled TA

Sample controlled thermal analysis

Sample dimensions, control

Sample preparation, generally contamination controls

Sample preparation, generally quality control

Sample temperature control

Sample-time control algorithm

Sampled Data or Discrete Control Systems

Sampled-Data Control of Distillation Columns

Sampling controller direct-digital

Sampling for quality control

Sampling point, process control

Sampling quality control issues

Statistical Analyses and Plotting of Control Sample Data

Temperature Control of the Sample

The Control of Sample Size for Normal Preparative Column Operation

Tip-Sample Distance Control

Tip-Sample Distance Control and Image Recording

Validation Samples, Quality Controls, and Assay Range

What factors control the extent of random sampling error

© 2024 chempedia.info