Sample introduction laser ablation

Fundamentally, introduction of a gaseous sample is the easiest option for ICP/MS because all of the sample can be passed efficiently along the inlet tube and into the center of the flame. Unfortunately, gases are mainly confined to low-molecular-mass compounds, and many of the samples that need to be examined cannot be vaporized easily. Nevertheless, there are some key analyses that are carried out in this fashion the major one i.s the generation of volatile hydrides. Other methods for volatiles are discussed below. An important method of analysis uses lasers to vaporize nonvolatile samples such as bone or ceramics. With a laser, ablated (vaporized) sample material is swept into the plasma flame before it can condense out again. Similarly, electrically heated filaments or ovens are also used to volatilize solids, the vapor of which is then swept by argon makeup gas into the plasma torch. However, for convenience, the methods of introducing solid samples are discussed fully in Part C (Chapter 17). [Pg.98]

Approximately 70 different elements are routinely determined using ICP-OES. Detection limits are typically in the sub-part-per-billion (sub-ppb) to 0.1 part-per-million (ppm) range. ICP-OES is most commonly used for bulk analysis of liquid samples or solids dissolved in liquids. Special sample introduction techniques, such as spark discharge or laser ablation, allow the analysis of surfaces or thin films. Each element emits a characteristic spectrum in the ultraviolet and visible region. The light intensity at one of the characteristic wavelengths is proportional to the concentration of that element in the sample. [Pg.633]

Ion extraction. The aspirated or laser ablated sample is transported from the sample introduction system into the center of the torch by a 1 1/min flow of Ar carrier gas where it is immediately dissociated and ionized by energy transfer with the hot -6000 K temperature of the surrounding Ar plasma. Ionization efficiencies are >95% for U and Th (Jarvis et al., 1992). For laser ablation sampling, helium may be employed as the carrier... [Pg.41]

Gray AL (1985). Solid sample introduction by laser ablation for inductively coupled plasma source mass spectrometry. Analyst 110 551-556... [Pg.56]

In ICP-AES and ICP-MS, sample mineralisation is the Achilles heel. Sample introduction systems for ICP-AES are numerous gas-phase introduction, pneumatic nebulisation (PN), direct-injection nebulisation (DIN), thermal spray, ultrasonic nebulisation (USN), electrothermal vaporisation (ETV) (furnace, cup, filament), hydride generation, electroerosion, laser ablation and direct sample insertion. Atomisation is an essential process in many fields where a dispersion of liquid particles in a gas is required. Pneumatic nebulisation is most commonly used in conjunction with a spray chamber that serves as a droplet separator, allowing droplets with average diameters of typically <10 xm to pass and enter the ICP. Spray chambers, which reduce solvent load and deal with coarse aerosols, should be as small as possible (micro-nebulisation [177]). Direct injection in the plasma torch is feasible [178]. Ultrasonic atomisers are designed to specifically operate from a vibrational energy source [179]. [Pg.619]

Table 8.36 lists the main classical and newer approaches to solid sampling for elemental analysis. Little work on the introduction of solids into flames has been reported, because of problems of sample delivery and the relatively low source temperature. In arc and spark emission and in laser ablation as a sampling technique, the ablated sample material cannot be determined exactly. The limitations of arc or... [Pg.626]

A particular advantage of ICP-MS derives from its ability to display a complete mass spectrum at one time. Combined with sample introduction by laser ablation it constitutes a very powerful tool for first-look analysis, e g. in geological prospecting or ecological surveys. ICP-MS is applicable to the whole range of areas where minor or trace elements are to be determined. [Pg.308]

There are several sample introduction methods that are used in conjunction with ICP, including nebulization, electrothermal evaporation, gas chromatography, hydride generation, and laser ablation [30]. Laser ablation combined with ICP (LA-ICP) is useful for analysis of solids. In such a source the sample is positioned in a chamber prior to the ICP source, the ablation cell. Argon gas at atmosperic pressure flows through the cell towards the ICP source. The sample is irradiated by a laser beam and... [Pg.22]

The MC-ICP-MS consists of four main parts 1) a sample introduction system that inlets the sample into the instrument as either a liquid (most common), gas, or solid (e.g., laser ablation), 2) an inductively coupled Ar plasma in which the sample is evaporated, vaporized, atomized, and ionized, 3) an ion transfer mechanism (the mass spectrometer interface) that separates the atmospheric pressure of the plasma from the vacuum of the analyzer, and 4) a mass analyzer that deals with the ion kinetic energy spread and produces a mass spectrum with flat topped peaks suitable for isotope ratio measurements. [Pg.118]

Various efficient devices have been utilized for sample introduction into an inductive plasma source, for example the application of several nebulizers, hyphenated techniques, hydride generation, laser ablation and electrothermal vaporization. The role of the solution introduction system in an inductively coupled plasma source is to convert the liquid sample into a suitable form (e.g.,... [Pg.37]

Figure 5.1 Main parts of an inductively coupled plasma mass spectrometer sample introduction systems (left column), e.g., Meinhard or MicroMist nebulizer with cyclonic spray chamber, ultrasonic nebulizer, microconcentric nebulizer and laser ablation system (all from CETAC Technologies), ion source (middle column) and several types of mass spectrometers, (a) Agilent 7500 from Agilent, (b) Platform from CV Instruments, or (c) Element from Thermo Fisher Scientific. (Parts of this figure were reproduced with permission from CETAC Technologies, Agilent, CV Instruments and Thermo Tisher Scientific, respectively.)... Figure 5.1 Main parts of an inductively coupled plasma mass spectrometer sample introduction systems (left column), e.g., Meinhard or MicroMist nebulizer with cyclonic spray chamber, ultrasonic nebulizer, microconcentric nebulizer and laser ablation system (all from CETAC Technologies), ion source (middle column) and several types of mass spectrometers, (a) Agilent 7500 from Agilent, (b) Platform from CV Instruments, or (c) Element from Thermo Fisher Scientific. (Parts of this figure were reproduced with permission from CETAC Technologies, Agilent, CV Instruments and Thermo Tisher Scientific, respectively.)...
Calibration and quantification procedures are easier in LA-ICP-MS compared to other solid-state mass spectrometric techniques because the laser ablation and the ICP ion source operate at normal pressure and the laser ablation of solid samples and ionization of analytes are separated in space and time. Therefore the advantage of solution calibration in ICP-MS can be applied in this solid-state analytical technique. The introduction of solution based calibration, which is only possible in LA-ICP-MS, was an innovative step in the development of this sensitive mass spectrometric technique. A number of different calibration approaches using aqueous standard solutions in the dual gas flow technique have been discussed by various authors.74 75 In the dual gas flow injection technique , the nebulized standard solution and the laser ablated sample material are mixed in the -piece and the two gas flows from the nebulizer (e.g. ultrasonic nebulizer) and laser ablation chamber are added. Using solution based calibration with the addition of a standard solution, Leach et alP determined minor elements in steel reference materials with a relative accuracy of a few %. In comparison to the so-called dual gas flow technique proposed in the literature, where the argon flow rates through the nebulizer and ablation cell add up to 11 min-1 (e.g. 0.451 min-1 and... [Pg.201]

On line additions of aqueous standard solutions for the calibration of LA-ICP-MS including a comparison of wet and dry plasma conditions are discussed by O Connor et al.ls For solution calibration of standard solutions the authors used a 100 (xl PFA nebulizer together with a cyclonic spray chamber or a MCN-6000 sample introduction system with desolvator, to study the wet and dry plasma, respectively. A polypropylene Y piece was applied to mix the laser ablated material and the nebulized standard solutions. The authors found that the on line addition of water is the preferred mode of operation for quantification by LA-ICP-MS, i.e., wet plasma is more stable (improved standard deviation of sensitivity ratios). [Pg.204]

Because ICP-MS with different instrumentations and sample introduction systems (besides solution nebulization, also laser ablation or hyphenated methods, such as HPLC, CE, SPME) is today the most frequently used analytical technique for precise and accurate isotope ratio measurements, the following section will mainly focus on this form of mass spectrometry with an inductively coupled plasma source. [Pg.228]

Use of inductively coupled plasma-mass spectrometry (1CP-MS) coupled to a laser-ablation sample introduction system (LA-ICP-MS) as a minimally destructive method for chemical characterization of archaeological materials has gained favor during the past few years. Although still a relatively new analytical technique in archaeology, LA-ICP-MS has been demonstrated to be a productive avenue of research for chemical characterization of obsidian, chert, pottery, painted and glazed surfaces, and human bone and teeth. Archaeological applications of LA-ICP-MS and comparisons with other analytical methods are described. [Pg.275]

Sr). Over the past 30 years, lead and strontium isotope ratios have been measured with thermal ionization mass spectrometry (TIMS). Elemental salts are deposited on a filament heated to produce ionized particles, which are then sent into a mass spectrometer where they are detected by multiple Faraday cups arrayed such that ions of several masses are collected simultaneously. TIMS is capable of high precision isotope discrimination, but the instruments tend to be large and expensive, and extensive sample preparation is required prior to sample introduction. Newer ICP-MS-based technologies like multi-collector ICP-MS (especially laser ablation) circumvent some of the sample preparation issues while exploiting the precision of simultaneous mass discrimination, but they are still limited by the number and configuration of ion collectors. [Pg.299]

To test for the possibility of laser-induced fractionation, we compared NIST standards in the TOF with both laser and nebulizer aspirated solutions (Figure 1). Our results indicate that while a small amount of bias correction difference exists between laser and aspirated sample introduction at lower signal intensities, no significant difference in isotopic ratios between ablated and aspirated samples exists when signal intensities are maintained at 300,000 cps and higher. [Pg.300]

The role of the sample introduction system is to convert a sample into a form that can be effectively vaporized into free atoms and ions in the ICP. A peristaltic pump is typically used to deliver a constant flow or sample solution (independent of variations in solution viscosity) to the nebulizer. Several different kinds of nebulizers are available to generate the sample aerosol, and several different spray chamber designs have been used to modify the aerosol before it enters the ICP Gases can be directly introduced into the plasma, for example, after hydride generation. Solids can be introduced by using electrothermal vaporization or laser ablation. [Pg.73]


See other pages where Sample introduction laser ablation is mentioned: [Pg.1555]    [Pg.41]    [Pg.626]    [Pg.657]    [Pg.158]    [Pg.154]    [Pg.28]    [Pg.38]    [Pg.43]    [Pg.120]    [Pg.264]    [Pg.395]    [Pg.395]    [Pg.429]    [Pg.512]    [Pg.512]    [Pg.513]    [Pg.513]    [Pg.276]    [Pg.297]    [Pg.298]    [Pg.298]    [Pg.299]    [Pg.300]    [Pg.304]    [Pg.305]    [Pg.308]    [Pg.562]    [Pg.85]   
See also in sourсe #XX -- [ Pg.36 , Pg.54 , Pg.62 , Pg.77 , Pg.81 , Pg.82 , Pg.83 , Pg.85 , Pg.86 , Pg.106 , Pg.129 ]




SEARCH



Ablate

Ablation

Ablator

Ablators

Laser ablation

Laser sampling

Sample introduction

Sample introduction systems laser ablation method

© 2024 chempedia.info