Potentiostat performance

A potentiostatic performance, e.g. with the circuit of Fig. 9, is preferable, because then it is possible to superimpose a small-amplitude a.c. voltage on the mean d.c. potential fed to the potentiostat. Consequently, an a.c. current will flow through the cell. In the first-order approximation its current density, Ay will be of the form... [Pg.240]

Standard ASTM potentiodynamic anodic polarization plot for certification of potentiostat performance. Type 430 stainless steel in 1 N H2S04at 30 °C. Scan rate of 600 mV/h. Test curve is to lie within the shaded region. Redrawn from Ref 27... [Pg.211]

The upper limit to the electron injection rate constants that can be measured by IMPS is determined by Xceii and potentiostat performance. If the space charge capacitance is high, it may be impossible to determine kinj. This is the case, for example, for formic acid oxidation on Ti02, where the IMPS response is dominated by the combination of the space charge and Helmholtz capacitances [84]. It would be interesting to explore the possibilities of using intensity modulated photopotential to overcome this limitation. [Pg.131]

Potentiodynamic polarisation The characteristics of passive/active conditions for metals can be readily defined using this technique ". Details for laboratory application can be found in ASTM Standard G5 (latest revision). Application in plant is easily performed as portable equipment (potentiostat) is available from several manufacturers, with some models incorporating built-in computer facilities. [Pg.1138]

The basic instrumentation required for controlled-potential experiments is relatively inexpensive and readily available commercially. The basic necessities include a cell (with a three-electrode system), a voltammetric analyzer (consisting of a potentiostatic circuitry and a voltage ramp generator), and an X-Y-t recorder (or plotter). Modem voltammetric analyzers are versatile enough to perform many modes of operation. Depending upon the specific experiment, other components may be required. For example, a faradaic cage is desired for work with ultramicroelectrodes. The system should be located in a room free from major electrical interferences, vibrations, and drastic fluctuations in temperature. [Pg.100]

Cu9ln4 and Cu2Se. They performed electrodeposition potentiostatically at room temperature on Ti or Ni rotating disk electrodes from acidic, citrate-buffered solutions. It was shown that the formation of crystalline definite compounds is correlated with a slow surface process, which induced a plateau on the polarization curves. The use of citrate ions was found to shift the copper deposition potential in the negative direction, lower the plateau current, and slow down the interfacial reactions. [Pg.117]

For coulometric analysis, the substance being examined must react in 100% current yields [i.e., other (secondary) reactions must be entirely absent]. In efforts to avoid side reactions, coulometry most often is performed potentiostatically (amperometrically) (i.e., the electrode potential is kept constant during the experiment), and the current consumed at the electrode is measured. The current is highest at the start of the... [Pg.388]

Electrodeposition of metals can be performed under different electrochemical modes. In the work mentioned in Ref. [18], it was performed in potentiostatic mode. The potential value for formation of platinum nanoparticles is —25 mV vs. SCE the deposition is performed from 2.5 mM solution of H2[PtCl6] in 50 mM KCl. The size of nanoparticles formed depends on the reduction charge. Continuous monitoring of the charge in potentiostatic mode is provided by different potentiostats, for example, by Autolab-PG-stat (EcoChemie, The Netherlands). Conditions for deposition of other metals should be selected according to their electrochemical properties. [Pg.323]

Electrochemical performance of the modified carbons was examined by floating gas diffusion electrode [19] in an electrochemical cell with separated cathode and anode chambers at room temperature with PI-50-1.1 -potentiostat in IN KOH aqueous solution. [Pg.347]

Figure 2.15 Schematic representation of the equipment necessary to perform linear sweep voltammetry LSV) or cyclic voltammetry CV). WFG waveform generator, P potentiostat, CR chart recorder, EC electrochemical cell, WE working electrode, CE counter electrode, RE... Figure 2.15 Schematic representation of the equipment necessary to perform linear sweep voltammetry LSV) or cyclic voltammetry CV). WFG waveform generator, P potentiostat, CR chart recorder, EC electrochemical cell, WE working electrode, CE counter electrode, RE...
Electrochemical Equipment. Electrochemical experiments were performed using either a PAR Model 175 universal programmer and a PAR Model 363 potentiostat/galvanostat, or a Pine Instruments RDE-4 bipotentiostat, coupled with a Kipp and Zonen BD 91 X-y-y recorder. The current-time response for the chronoamperometry experiments was recorded with a Nicolet 4094 digital oscilloscope. All potentials were measured vs. a Ag/10"2 M Ag+ reference electrode. [Pg.411]

Fig. 18b.1. Electrochemical cells and representative cell configurations, (a) Schematic diagram of a cell-potentiostat system, (b) Typical laboratory cell with Hg-drop electrode and drop knocker, (c) Voltammetric cell as detector at the end of a high-performance liquid chromatographic column, (d) A two-electrode (graphite) chip cell for biosensor development, (e) Three-electrode chip cells on a ceramic substrate for bioanalytical work. Fig. 18b.1. Electrochemical cells and representative cell configurations, (a) Schematic diagram of a cell-potentiostat system, (b) Typical laboratory cell with Hg-drop electrode and drop knocker, (c) Voltammetric cell as detector at the end of a high-performance liquid chromatographic column, (d) A two-electrode (graphite) chip cell for biosensor development, (e) Three-electrode chip cells on a ceramic substrate for bioanalytical work.
Electrochemical measurements were performed in an electrochemical cell equipped with quartz windows which fit into the sample compartment of a Cary 14 spectrometer. The cell (CHjCN, 0.1N TEAP vs S.C.E.) employed three electrode (Pt auxiliary electrode) potentiostatic control. A Tacussell PRT Potentiostat and PAR model 175 signal generator were used for the measurements. [Pg.439]

Typical anodization curves of silicon electrodes in aqueous electrolytes are shown in Fig. 5.1 [Pa9]. The oxidation can be performed under potential control or under current control. For the potentiostatic case the current density in the first few seconds of anodization is only limited by the electrolyte conductivity [Ba2]. In this respect the oxide formation in this time interval is not truly under potentiostatic control, which may cause irreproducible results [Ba7]. In aqueous electrolytes of low resistivity the potentiostatic characteristic shows a sharp current peak when the potential is switched to a positive value at t=0. After this first current peak a second broader one is observed for potentials of 16 V and higher, as shown in Fig. 5.1a. The first sharp peak due to anodic oxidation is also observed in low concentrated HF, as shown in Fig. 4.14. In order to avoid the initial current peak, the oxidation can be performed under potentiodynamic conditions (V/f =const), as shown in Fig. 5.1b. In this case the current increases slowly near t=0, but shows a pronounced first maximum at a constant bias of about 19 V, independently of scan rate. The charge consumed between t=0 and this first maximum is in the order of 0.2 mAs cnT2. After this first maximum several other maxima at different bias are observed. [Pg.79]

Occasionally, we prepare a different type of electrode - we screen-print them. With modem screen-printing technology, a complicated arrangement of precisely arranged and well-defined electrodes can be prepared on the solid substrate of choice. In essence, these electrodes are simply dots of conductive material, with suitable electrical contacts to ensure that each can be addressed correctly. As an example, each electrode can be connected to a separate channel of a frequency analyser or potentiostat. In this way, a great many analyses can be performed simultaneously. [Pg.287]


See other pages where Potentiostat performance is mentioned: [Pg.217]    [Pg.217]    [Pg.1004]    [Pg.1041]    [Pg.1113]    [Pg.532]    [Pg.473]    [Pg.104]    [Pg.115]    [Pg.216]    [Pg.221]    [Pg.613]    [Pg.16]    [Pg.417]    [Pg.423]    [Pg.334]    [Pg.11]    [Pg.38]    [Pg.592]    [Pg.66]    [Pg.111]    [Pg.333]    [Pg.284]    [Pg.114]    [Pg.52]    [Pg.234]    [Pg.432]    [Pg.520]    [Pg.331]    [Pg.669]    [Pg.171]    [Pg.246]    [Pg.116]    [Pg.293]    [Pg.535]    [Pg.136]    [Pg.137]   
See also in sourсe #XX -- [ Pg.375 ]