Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Loop flow analysis

N. Teshima, S. Ohno, T. Sakai, Stopped-in-loop flow analysis of trace vanadium in water, Anal. Sci. 23 (2007) 1. [Pg.288]

The advent of easy access to the satellite-based global positioning system (GPS) and availability of off-the-shelf portable probes and rapid analyzers for a number of water quality determinants have enabled the development of systems that can be carried on small survey vessels to map water quality conditions. Rapid data acquisition is now practical using probes and sondes for measuring temperature, conductivity, turbidity, pH, and dissolved oxygen fluorometric technologies for chlorophyll biomass and phytoplankton composition flow injection and loop flow analysis for some nutrient species and acoustic Doppler-based devices for current profiling. [Pg.25]

The concept of temporal variations in concentration at the flow-through detector explains why pronounced skewed peaks are often observed in flow analysis, especially with loop-based sample introduction. Taylor assumed that dispersion is symmetric in relation to an observer located at the dispersing zone [55,56], but in practice the recorded peaks are usually characterised by a rise time much shorter than the fall time (see also Fig. 1.3e). This skew effect is explained by the fact that the front and trailing portions of the flowing sample, which relate to the rise time and the fall time, respectively, have different residence times in the manifold and are therefore subjected to different extents of dispersion. [Pg.75]

Precise introduction of a well-defined portion of the aqueous sample into the carrier stream is fundamental to flow analysis. This is generally accomplished by continuously pumping the sample into the manifold for a fixed period of time (time-based introduction) or by inserting the sample aliquot via a fixed-volume loop attached to a rotary or sliding bar valve (loop-based introduction). Hydrodynamic [48] and nested [49] sample introduction have also been used. [Pg.217]

A flow system with zone sampling comprises two almost independent subsystems, and the re-sampling loop (L2) is the only component common to both. This architecture permits the time window associated with the tzs value to be more efficiently exploited. Implementation of zone sampling expands the applicability of flow analysis (see Table 7.2) and some applications of this strategy are highlighted in the following sections. [Pg.256]

There are a few ways of linking the techniques of LC-MS and LC-NMR. The most common method is in a parallel mode by splitting the flow, e.g. 50 1, so as to direct the majority of it to the NMR due to its relative insensitivity. This means that the analytes are detected simultaneously by both detectors and possibly also by UV, which may actually be used as the trigger to begin detection by the NMR and MS modules. Alternatively, the rapidly acquired MS data can be used to direct the NMR experiments or vice versa. A second method of interfacing the two techniques is to use the serial mode or stopped flow mode, which enables more sensitive NMR experiments to be carried out. A recent development in stopped flow NMR is the inclusion of in-line solid phase extraction (SPE) after the LC. The SPE acts as a fraction collector for individual compounds. This trapping/ washing step can improve sensitivity several fold. A third method is fraction collection, where samples from the LC are collected in a loop for analysis later, perhaps after certain data have been reviewed. [Pg.119]

Figure 10. Hamster breathing loop for analysis of exhaled Ng The system consists of (a) compressed gas (80/20 He/Oi) and appropriate flow regulators (b) fine flow rate control value (c) rotometer (d) T-joint with gas tight fitting for attaching hamster endotracheal tube (e) heating pallet for hamster (f) chromatograph sampling valve (g) water ana CO absorbent bed (h) chromatograph column (i) thermal conductivity detector. Figure 10. Hamster breathing loop for analysis of exhaled Ng The system consists of (a) compressed gas (80/20 He/Oi) and appropriate flow regulators (b) fine flow rate control value (c) rotometer (d) T-joint with gas tight fitting for attaching hamster endotracheal tube (e) heating pallet for hamster (f) chromatograph sampling valve (g) water ana CO absorbent bed (h) chromatograph column (i) thermal conductivity detector.
Control Flow analyser examines the structure of a procedure to identify all entry and exit points. It also identifies all loops with their entry and exit points. Control flow analysis also reveals more serious errors within the source code such as unreachable code and d5mamic halts. [Pg.93]

The mold cooling analysis and the flow analysis are essentially coupled since the transient cavity wall temperature and heat flux are unknown in both analysis, although in practice one may only need a couple of iterative loops, depending on the required accuracy and efficiency. [Pg.111]

The determination of the temperature control point inside the column is from a combination of closed-loop sensitivity analysis and verification by open-loop sensitivity analysis. For the closed-loop sensitivity analysis, closed-loop simulation runs are made in which both the top and bottom product purities are held at their specifications for different feed compositions. This is to mimic ideal (although not practical) situation with two online composition measurements and two composition control loops setting aqueous reflux flow and reboiler duty. The tray temperature with the least amount of variation for changes in feed composition is selected as the temperature control point. The specific feed composition changes are feed F3 water molar composition +10% changes while keeping the total molar flowrate of F3 the same by adjusting the acetic acid molar composition. [Pg.278]

Figure 9.25 shows the deviation of temperature under this ideal control situation. Temperature at the fourth stage is selected as the control point, since it shows the least deviation from base-case condition. Therefore, by holding the temperature at this control point, both bottom and top product purities should be maintained near their specifications. The other two free manipulated variables (reboiler heat duty and isobutyl acetate makeup flow) are fixed and only adjusted according to the total molar feed fiowrate using a ratio scheme. Notice that the selected temperature control point should be verified by open-loop sensitivity analysis similar to those shown in Figure 9.9, so that enough open-loop sensitivity between the manipulated variable to the controller variable is confirmed. [Pg.278]

A modified overall control strategy is to ratio both reflux flowrates to the fresh feed flow-rate. This modified overall control strategy was recommended in Grassi and is shown in Figure 10.26. Notice that the temperature control points for both columns are the same as with the other reflux-ratio structure because there was httle change in the open-loop sensitivity analysis from those shown in Figures 10.22 and 10.23. [Pg.322]

Technologies to ensure reliable core operation Corrosion test loop will be built to define and validate a technology for the protection of stractural materials from corrosion / erosion in a flow of Pb-Bi coolant experimental facility is under construction. Demonstration loop scaling analysis completed, ongoing experimental programme. [Pg.662]

The ECS phase analysis is based on a series of calculations of the primary system tank and piping with air ingress caused by the inability of the ECS to fully make up the coolant lost out broken pipe. Air enters the system through the vents and plenum inlet line and markedly degrades system loop flow. [Pg.555]

The analysis of the ECS phase of the accident is based on a combination of bounding best estimate for loop flows and a mechanistic modeling and engineering approach for phenomena within the fuel assembly. Because of the nature of the uncertainties and the lack of prototype experimental data to perform a detailed statistical analysis comparable to that performed for FI, ECS uncertainties have been classified as biases and random uncertainties and treated in a conventional manner. Thus, the methodology is still evolving in... [Pg.565]

An analysis of the subsequent ruptures of the main pressure pipe and the safeguard vessel was also performed. This analysis has shown that such an event results in an increase of the pressure in the compartment below the middle plate of the support belt of the emergency loop. The analysis showed that the pressure increase could lead to fuel assemblies (FAs) and/or support belt being pushed to the coolant surface. To avoid the risk of FAs or support belt being pushed to the surface, the compartments of the support belt at the middle plate have two safety valves that provide a by-pass and flow supply to the suction area of the main circulation pump (see Fig. 5). Even with only one open valve, there are no FAs being pushed to the surface. [Pg.132]

In addition to Lt/Lh other length scales also affect the stability behaviour. This can be established by carrying out a linear stability analysis. In this method, the loop flow rate and temperature are perturbed as... [Pg.144]

Example of a single-channel manifold for use in flow injection analysis where R1 is a reagent reservoir P is the pump S is the sample I is the injector B is a bypass loop ... [Pg.652]

Two examples of dual-channel manifolds for use In flow Injection analysis where R1 and R2 are reagent reservoirs P Is the pump S Is the sample I Is the Injector B Is a bypass loop W Is waste C Is the mixing and reaction coll and D Is the detector. [Pg.653]

Time-Delay Compensation Time delays are a common occurrence in the process industries because of the presence of recycle loops, fluid-flow distance lags, and dead time in composition measurements resulting from use of chromatographic analysis. The presence of a time delay in a process severely hmits the performance of a conventional PID control system, reducing the stability margin of the closed-loop control system. Consequently, the controller gain must be reduced below that which could be used for a process without delay. Thus, the response of the closed-loop system will be sluggish compared to that of the system with no time delay. [Pg.733]

Operational sequence diagrams are flcw-charting techniques that represent any sequence of control movements and information collection activities that are executed in order to perform a task. Various activities in the diagram are represented with a symbolic notation, supported where necessary by a text description. For the majority of simple applications, OSDs assume a linear flow drawn from top to bottom with a limited degree of branching and looping. The symbols used are usually tailored to fit the type of task being studied and its level of analysis. [Pg.172]

In 1990, Bushey and Jorgenson developed the first automated system that eoupled HPLC with CZE (19). This orthogonal separation teehnique used differenees in hydrophobieity in the first dimension and moleeular eharge in the seeond dimension for the analysis of peptide mixtures. The LC separation employed a gradient at 20 p.L/min volumetrie flow rate, with a eolumn of 1.0 mm ID. The effluent from the ehromatographie eolumn filled a 10 p.L loop on a eomputer-eontrolled, six-port miero valve. At fixed intervals, the loop material was flushed over the anode end of the CZE eapillary, allowing eleetrokinetie injeetions to be made into the seeond dimension from the first. [Pg.204]


See other pages where Loop flow analysis is mentioned: [Pg.271]    [Pg.271]    [Pg.521]    [Pg.369]    [Pg.521]    [Pg.524]    [Pg.699]    [Pg.30]    [Pg.220]    [Pg.300]    [Pg.390]    [Pg.268]    [Pg.316]    [Pg.28]    [Pg.179]    [Pg.15]    [Pg.14]    [Pg.87]    [Pg.81]    [Pg.88]    [Pg.579]    [Pg.88]    [Pg.177]    [Pg.362]    [Pg.622]    [Pg.238]   
See also in sourсe #XX -- [ Pg.25 ]




SEARCH



Loop analysis

© 2024 chempedia.info