Isotopic ratio

It is interesting to note that recent evidence shows that even extra-terrestrially formed hydrocarbons can reach the Earth. The Earth continues to receive some 40,000 tons of interplanetary dust every year. Mass-spectrometric analysis has revealed the presence of hydrocarbons attached to these dust particles, including polycyclic aromatics such as phenanthrene, chrysene, pyrene, benzopyrene, and pentacene of extraterrestrial origin indicated by anomalous isotopic ratios. [Pg.128]

Lasers can be used in either pulsed or continuous mode to desorb material from a sample, which can then be examined as such or mixed or dissolved in a matrix. The desorbed (ablated) material contains few or sometimes even no ions, and a second ionization step is frequently needed to improve the yield of ions. The most common methods of providing the second ionization use MALDI to give protonated molecular ions or a plasma torch to give atomic ions for isotope ratio measurement. By adjusting the laser s focus and power, laser desorption can be used for either depth or surface profiling. [Pg.12]

A further important use of El mass spectrometry lies in measuring isotope ratios, which can be used in estimating the ages of artifacts, rocks, or fossils. Electron ionization affects the isotopes of any one element equally, so that the true isotope ratio is not distorted by the ionization step. Further information on isotopes can be found in Chapter 46. [Pg.16]

The ion current resulting from collection of the mass-separated ions provides a measure of the numbers of ions at each m/z value (the ion abundances). Note that for this ionization method, all ions have only a single positive charge, z = 1, so that m/z = m, which means that masses are obtained directly from the measured m/z values. Thus, after the thermal ionization process, m/z values and abundances of ions are measured. The accurate measurement of relative ion abundances provides highly accurate isotope ratios. This aspect is developed more fully below. [Pg.46]

Figure 7.9 shows a schematic representation of this effect, in which the ratio of the two isotopes changes with time. To obtain an accurate estimate of the ratio of ion abundances, it is better if the relative ion yields decrease linearly (Figure 7.9) which can be achieved by adjusting the filament temperature continuously to obtain the desired linear response. An almost constant response for the isotope ratio can be obtained by slow evaporation of the sample, viz., by keeping the filament temperature as low as is consistent with sufficient sensitivity of detection (Figure 7.9). [Pg.52]

The previous discussion demonstrates that measurement of precise isotope ratios requires a substantial amount of operator experience, particularly with samples that have not been examined previously. A choice of filament metal must be made, the preparation of the sample on the filament surface is important (particularly when activators are used), and the rate of evaporation (and therefore temperature control) may be crucial. Despite these challenges, this method of surface ionization is a useful technique for measuring precise isotope ratios for multiple isotopes. Other chapters in this book discuss practical details and applications. [Pg.52]

Precise measurement of isotope ratios can be obtained by comparing the yields of isotopic ions desorbing from a sample placed on a strongly heated filament that is generally made from platinum, tantalum, rhenium, or tungsten. [Pg.53]

Gases and vapors of volatile liquids can be introduced directly into a plasma flame for elemental analysis or for isotope ratio measurements. Some elements can be examined by first converting them chemically into volatile forms, as with the formation of hydrides of arsenic and tellurium. It is important that not too much analyte pass into the flame, as the extra material introduced into the plasma can cause it to become unstable or even to go out altogether, thereby compromising accuracy or continuity of measurement. [Pg.102]

El = electron ionization Cl = chemical ionization ES = electrospray APCI = atmospheric-pressure chemical ionization MALDI = matrix-assisted laser desorption ionization PT = plasma torch (isotope ratios) TI = thermal (surface) ionization (isotope ratios). [Pg.280]

The previous discussion has centered on how to obtain as much molecular mass and chemical structure information as possible from a given sample. However, there are many uses of mass spectrometry where precise isotope ratios are needed and total molecular mass information is unimportant. For accurate measurement of isotope ratio, the sample can be vaporized and then directed into a plasma torch. The sample can be a gas or a solution that is vaporized to form an aerosol, or it can be a solid that is vaporized to an aerosol by laser ablation. Whatever method is used to vaporize the sample, it is then swept into the flame of a plasma torch. Operating at temperatures of about 5000 K and containing large numbers of gas ions and electrons, the plasma completely fragments all substances into ionized atoms within a few milliseconds. The ionized atoms are then passed into a mass analyzer for measurement of their atomic mass and abundance of isotopes. Even intractable substances such as glass, ceramics, rock, and bone can be examined directly by this technique. [Pg.284]

Ionization Method Type of Molecular Ion Formed Good Molecular Mass Information Abundant Fragment Ions MS/MS Needed for Structural Information Accurate Values for Isotope Ratios... [Pg.285]

Plasma torches and thermal ionization sources break down the substances into atoms and ionized atoms. Both are used for measurement of accurate isotope ratios. In the breakdown process, all structural information is lost, other than an identification of elements present (e.g., as in inductively coupled mass spectrometry, ICP/MS). [Pg.285]

Since detailed chemical structure information is not usually required from isotope ratio measurements, it is possible to vaporize samples by simply pyrolyzing them. For this purpose, the sample can be placed on a tungsten, rhenium, or platinum wire and heated strongly in vacuum by passing an electric current through the wire. This is thermal or surface ionization (TI). Alternatively, a small electric furnace can be used when removal of solvent from a dilute solution is desirable before vaporization of residual solute. Again, a wide variety of mass analyzers can be used to measure m/z values of atomic ions and their relative abundances. [Pg.285]

This example can be used in reverse to show the usefulness of looking for such isotopes. Suppose there were an unknown sample that had two molecular ion peaks in the ratio of 3 1 that were two mass units apart then it could reasonably be deduced that it was highly likely the unknown contained chlorine. In this case, the isotope ratio has been used to identify a chlorine-containing compound. This use of mass spectrometry is widespread in general analysis of materials, and it... [Pg.339]

The use of accurate isotope ratio measurement is exemplified here by a method used to determine the temperature of the Mediterranean Sea 10,000 years ago. It is known that the relative solubility of the two isotopic forms of carbon dioxide COj) in sea water depends on temperature... [Pg.340]

This accurate measurement of the ratio of abundances of isotopes is used for geological dating, estimation of the ages of antiquities, testing athletes for the use of banned steroids, examining fine details of chemical reaction pathways, and so on. These uses are discussed in this book under various headings concerned with isotope ratio mass spectrometry (see Chapters 7, 14, 15, 16, 17, 47, and 48). [Pg.341]

For other elements that occur with major relative abundances of more than one isotope in the natural state, the isotope pattern becomes much more complex. For example, with chlorine and bromine, the presence of these elements is clearly apparent from the isotopes Cl and for chlorine and Br and Br for bromine. Figure 47.2a shows the molecular ion region for the compound chlorodecane. Now, there are new situations in that C, C, C1, and Cl isotopes all have probabilities of occurring together. Thus, there are molecular ion peaks for + Cl, C + Cl, + Cl, and so on. Even so, the isotopic ratio of 3 1 for Cl to Cl is very clear... [Pg.348]

Partial mass spectra showing the isotope patterns in the molecular ion regions for ions containing carbon and (a) only one chlorine atom, (b) only one bromine atom, and (c) one chlorine and one bromine atom. The isotope patterns are quite different from each other. Note how the halogen isotope ratios appear very clearly as 3 1 for chlorine in (a), 1 1 for bromine in (b), and 3 4 1 for chlorine and bromine in (c). If the numbers of halogens were not known, the pattern could be used in a reverse sense to decide their number. [Pg.349]

Special instruments (isotope ratio mass spectrometers) are used to determine isotope ratios, when needed, to better than about 3%. Such special instruments are described in Chapters 6, 7, and 48. The methods of ionization and analysis for such precise measurements are not described here. [Pg.349]

In a similar vein, mean seawater temperatures can be estimated from the ratio of 0 to 0 in limestone. The latter rock is composed of calcium carbonate, laid down from shells of countless small sea creatures as they die and fall to the bottom of the ocean. The ratio of the oxygen isotopes locked up as carbon dioxide varies with the temperature of sea water. Any organisms building shells will fix the ratio in the calcium carbonate of their shells. As the limestone deposits form, the layers represent a chronological description of the mean sea temperature. To assess mean sea temperatures from thousands or millions of years ago, it is necessary only to measure accurately the ratio and use a precalibrated graph that relates temperatures to isotope ratios in sea water. [Pg.351]

Elemental isotopic compositions (isotope ratios) can be used mass spectrometrically in a routine sense to monitor a substance for the presence of different kinds of elements, as with chlorine or platinum. It can also be used in a precise sense to examine tiny variations in these ratios, from which important deductions can be made in a wide variety of disciplines. [Pg.352]

One of the most significant sources of change in isotope ratios is caused by the small mass differences between isotopes and their effects on the physical properties of elements and compounds. For example, ordinary water (mostly Ej O) has a lower density, lower boiling point, and higher vapor pressure than does heavy water (mostly H2 0). Other major changes can occur through exchange processes. Such physical and kinetic differences lead to natural local fractionation of isotopes. Artificial fractionation (enrichment or depletion) of uranium isotopes is the basis for construction of atomic bombs, nuclear power reactors, and depleted uranium weapons. [Pg.353]

Few of the naturally occurring elements have significant amounts of radioactive isotopes, but there are many artificially produced radioactive species. Mass spectrometry can measure both radioactive and nonradioactive isotope ratios, but there are health and safety issues for the radioactive ones. However, modem isotope instmments are becoming so sensitive that only very small amounts of sample are needed. Where radioactive isotopes are a serious issue, the radioactive hazards can be minimized by using special inlet systems and ion pumps in place of rotary pumps for maintaining a vacuum. For example, mass spectrometry is now used in the analysis of Pu/ Pu ratios. [Pg.354]

Because variations in accurate isotope ratio measurements typically concern only a few parts per 1000 by mass and there are no universal absolute ratios, it is necessary to define some standards. For this purpose, samples of standard substances are produced and made available at two major centers IAEA (International Atomic Energy Authority, U.K.) and NIST (National Institute for Standards and Technology, U.S.). Standards from other sources are also available. These primary standards can be used as such, or alternative standards can be employed if the primary ones are not available. However, any alternative standards need to be related accurately to the primary ones (see formulae below). For example, the material PDB (PeeDee belemnite), used particularly as a standard for the ratio of isotopes, is no longer readily available, and a new standard, VPDB,... [Pg.354]

Before measurement it must be decided exactly which isotopes are to be compared. For oxygen, it is usually the ratio of 0 to 0, and for hydrogen it is H to H. Such isotope ratios are measured by the mass spectrometer. For example, examination of a sample of a carbonaceous compound provides abundances of ions at two m/z values, one related to C and one to C (it could be at m/z 45 and COj at m/z 44). By convention, the heavier isotope is always compared with the lighter isotope. The ratio of isotopes is given the symbol R (Figure 48.1). [Pg.354]

For example, if a carbonaceous sample (S) is examined mass spectrometrically, the ratio of abundances for the carbon isotopes C, in the sample is Rg. This ratio by itself is of little significance and needs to be related to a reference standard of some sort. The same isotope ratio measured for a reference sample is then R. The reference ratio also serves to check the performance of the mass spectrometer. If two ratios are measured, it is natural to assess them against each other as, for example, the sample versus the reference material. This assessment is defined by another ratio, a (the fractionation factor Figure 48.2). [Pg.354]


See other pages where Isotopic ratio is mentioned: [Pg.39]    [Pg.45]    [Pg.94]    [Pg.98]    [Pg.104]    [Pg.110]    [Pg.113]    [Pg.339]    [Pg.343]    [Pg.343]    [Pg.343]    [Pg.345]    [Pg.347]    [Pg.349]    [Pg.349]    [Pg.350]    [Pg.350]    [Pg.351]    [Pg.351]    [Pg.351]    [Pg.353]    [Pg.353]    [Pg.353]    [Pg.354]    [Pg.354]   
See also in sourсe #XX -- [ Pg.285 ]

See also in sourсe #XX -- [ Pg.409 ]

See also in sourсe #XX -- [ Pg.277 ]