Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration mass spectrometry

To study catalytic processes of clusters on surfaces with high sensitivity, a newly designed pulsed valve with excellent pulse-to-pulse stability, in combination with absolutely calibrated mass spectrometry, was used to determine turn-over frequencies [75]. [Pg.578]

MS, mass spectrometry CMS, calibrated mass spectrometry KE, Knudsen effusion KEMS, Knudsen effusion mass spectrometry. Given only in the form of the Clausius-Clapeyron equation. Run 3 of Habermann and Daane (1964) given in detail. ... [Pg.358]

Quantitative mass spectrometry, also used for pharmaceutical appHcations, involves the use of isotopicaHy labeled internal standards for method calibration and the calculation of percent recoveries (9). Maximum sensitivity is obtained when the mass spectrometer is set to monitor only a few ions, which are characteristic of the target compounds to be quantified, a procedure known as the selected ion monitoring mode (sim). When chlorinated species are to be detected, then two ions from the isotopic envelope can be monitored, and confirmation of the target compound can be based not only on the gc retention time and the mass, but on the ratio of the two ion abundances being close to the theoretically expected value. The spectrometer cycles through the ions in the shortest possible time. This avoids compromising the chromatographic resolution of the gc, because even after extraction the sample contains many compounds in addition to the analyte. To increase sensitivity, some methods use sample concentration techniques. [Pg.548]

In a separate set of experiments designed to follow the gas phase reactions of CHj-radicals with NO, CHj- radicals were generated by the thermal decomposition of azomethane, CHjN NCHj, at 980 °C. The CH3- radicals were subsequently allowed to react with themselves and with NO in a Knudsen cell that has been described previously [12]. Analysis of intermediates and products was again done by mass spectrometry, using the VIEMS. Calibration of the mass spectrometer with respect to CH,- radicals was carried out by introducing the products of azomethane decomposition directly into the high vacuum region of the instrument. [Pg.713]

Residues of isoxaflutole, RPA 202248 and RPA 203328 are extracted from surface water or groundwater on to an RP-102 resin solid-phase extraction (SPE) cartridge, then eluted with an acetonitrile-methanol solvent mixture. Residues are determined by liquid chromatography/tandem mass spectrometry (LC/MS/MS) on a Cg column. Quantitation of results is based on a comparison of the ratio of analyte response to isotopically labeled internal standard response versus analyte response to internal standard response for calibration standards. [Pg.510]

Conventional calibration MDRD equation [used only with those creatinine methods that have not been recalibrated to be traceable to isotope dilution mass spectrometry (IDMS)]... [Pg.1543]

Principles and Characteristics Mass spectrometry can provide the accurate mass determination in a direct measurement mode. For a properly calibrated mass spectrometer the mass accuracy should be expected to be good to at least 0.1 Da. Accurate mass measurements can be made at any resolution (resolution matters only when separating masses). For polymer/additive deformulation the nominal molecular weight of an analyte, as determined with an accuracy of 0.1 Da from the mass spectrum, is generally insufficient to characterise the sample, in view of the small mass differences in commercial additives. With the thousands of additives, it is obvious that the same nominal mass often corresponds to quite a number of possible additive types, e.g. NPG dibenzoate, Tinuvin 312, Uvistat 247, Flexricin P-1, isobutylpalmitate and fumaric acid for m = 312 Da see also Table 6.7 for m = 268 Da. Accurate mass measurements are most often made in El mode, since the sensitivity is high, and reference mass peaks are readily available (using various fluorinated reference materials). Accurate mass measurements can also be made in Cl... [Pg.355]

Resolution does not affect the accuracy of the individual accurate mass measurements when no separation problem exists. When performing accurate mass measurements on a given component in a mixture, it may be necessary to raise the resolution of the mass spectrometer wherever possible. Atomic composition mass spectrometry (AC-MS) is a powerful technique for chemical structure identification or confirmation, which requires double-focusing magnetic, Fourier-transform ion-cyclotron resonance (FTICR) or else ToF-MS spectrometers, and use of a suitable reference material. The most common reference materials for accurate mass measurements are perfluorokerosene (PFK), perfluorotetrabutylamine (PFTBA) and decafluorotriph-enylphosphine (DFTPP). One of the difficulties of high-mass MS is the lack of suitable calibration standards. Reference inlets to the ion source facilitate exact mass measurement. When appropriately calibrated, ToF mass... [Pg.356]

Mass spectrometers must be regularly tuned or calibrated against a known standard, e.g. perfluorotributy-lamine (PFTBA). The trend is towards miniaturisation (10 x 24 x 14 in.). A concept for a micro mass spectrometer, with potential applications in process monitoring, has been presented [167]. Mass-spectrometry instrumentation (1997) has been reviewed [166]. [Pg.387]

SEC-ESI-FTMS combines the size separation based technique of SEC with one of the most powerful mass spectrometric techniques of FTMS offering high mass accuracy (ppm), ultrahigh resolving power (>10(i) 6) and the capability to perform tandem mass spectrometry. The technique enables generation of oligomer elution profiles, which can be used for accurate calibration of standard SEC data. Coupling of SEC to ESI-MS is further described in ref. [710],... [Pg.529]

Applications Table 8.58 shows the main fields of application of inorganic mass spectrometry. Mass-spectrometric techniques find wide application in inorganic analysis, and are being used for the determination of elemental concentrations and of isotopic abundances for speciation and surface characterisation for imaging and depth profiling. Solid-state mass spectrometry is usable as a quantitative method only after calibration by standard samples. [Pg.650]

Applications Real applications of spark-source MS started on an empirical basis before fundamental insights were available. SSMS is now considered obsolete in many areas, but various unique applications for a variety of biological substances and metals are reported. Usually, each application requires specific sample preparation, sparking procedure and ion detection. SSMS is now used only in a few laboratories worldwide. Spark-source mass spectrometry is still attractive for certain applications (e.g. in the microelectronics industry). This is especially so when a multi-element survey analysis is required, for which the accuracy of the technique is sufficient (generally 15-30% with calibration or within an order of magnitude without). SSMS is considered to be a... [Pg.651]

Isotope dilution mass spectrometry (IDMS) can be applied with most of the ionisation methods used in mass spectrometry to determine isotope ratios with greater or lesser accuracy. For calibration by means of isotope dilution, an exactly known amount of a spike solution, enriched in an isotope of the element(s) to be determined, is added to an exactly known amount of sample. After isotopic equilibration, the isotope ratio for the mixture is determined mass spectrometrically. The attraction of IDMS is its potential simplicity it relies only on the measurement of ratios. The... [Pg.659]

In isotope dilution inductively coupled plasma-mass spectrometry (ID-ICP-MS) the spike, the unspiked and a spiked sample are measured by ICP-MS in order to determine the isotope ratio. Using this technique, more precise and accurate results can be obtained than by using a calibration graph or by standard addition. This is due to elimination of various systematic errors. Isotopes behave identically in most chemical and physical processes. Signal suppression and enhancement due to the matrix in ICP-MS affects both isotopes equally. The same holds for most long-term instrumental fluctuations and drift. Accuracy and precision obtained with ID-ICP-QMS are better than with other ICP-QMS calibration... [Pg.660]

Applications The application of the isotope dilution technique is especially useful in carrying out precise and accurate micro and trace analyses. The most accurate results in mass spectrometry are obtained if the isotope dilution technique is applied (RSDs better than 1 % in trace analysis). Therefore, application of IDMS is especially recommended for calibration of other analytical data, and for certification of standard reference materials. The technique also finds application in the field of isotope geology, and is used in the nuclear industry for quantitative isotope analysis. [Pg.661]

While it is certainly possible to replace the constant with the appropriate values for constituent parameters ( 2eVID2), it is much more practical to determine the value empirically. A common calibration equation for TOF mass spectrometry is slightly more complicated (with the addition of an offset m/z value, b) and is given by... [Pg.44]

Goodacre, R. Trew, S. Wrigley-Jones, C. Neal, M. J. Maddock, J. Ottley, T. W. Porter, N. Kell, D. B. Rapid screening for metabolite overproduction in fermentor broths using pyrolysis mass spectrometry with multivariate calibration and artificial neural networks. Biotechnol. Bioengin. 1994, 44,1205-1216. [Pg.124]

Ledford, E. B., Jr. Rempel, D. L. Gross, M. L. Space charge effects in Fourier transform mass spectrometry Mass calibration. Anal. Chem. 1984,56, 2744-2748. [Pg.297]

In our particular application, we do not have an absolute method of calibration because the alkyl chain length affects the EO distribution retention. However, mass spectrometry would be an ideal third dimension. The automated combination of two-dimensional chromatography and mass spectrometry is the next step toward the future of simultaneous separation and identification of very complicated samples. [Pg.443]

Merritt, D. A., Brand, W. A. and Hayes, J. M. (1994) Isotope ratio monitoring gas chromatography mass spectrometry methods for isotopic calibration. Organic Geochemistry 21, 573 583. [Pg.429]

O. Vorm and P. Roepstorff. Detector Bias Gating for Improved Detector Response and Calibration in Matrix-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry. J. Mass Spectrom., 31(1996) 351-356. [Pg.86]


See other pages where Calibration mass spectrometry is mentioned: [Pg.1828]    [Pg.446]    [Pg.33]    [Pg.106]    [Pg.14]    [Pg.138]    [Pg.85]    [Pg.696]    [Pg.486]    [Pg.342]    [Pg.496]    [Pg.370]    [Pg.397]    [Pg.410]    [Pg.589]    [Pg.431]    [Pg.347]    [Pg.83]    [Pg.259]    [Pg.134]    [Pg.8]    [Pg.19]    [Pg.350]    [Pg.1]    [Pg.1]    [Pg.224]   
See also in sourсe #XX -- [ Pg.610 ]




SEARCH



Isotope-dilution mass spectrometry calibration

Isotope-dilution mass spectrometry calibration standards

Mass calibration

Quantification of Analytical Data via Calibration Curves in Mass Spectrometry Using Certified Reference Materials or Defined Standard Solutions

Tandem mass spectrometry calibration

© 2024 chempedia.info