Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration 654 INDEX

The International Friction Index (IFI) is similar to the SRI discussed above and was developed in the PIARC international experiment to compare and harmonise texture and skid resistance measurements. The index allows for the harmonising of friction measurements with different devices that use smooth tread test tyres to a common calibrated index. The procedure for determining the IFI is covered in ASTM E 1960 (2011). [Pg.731]

Fundamental Limitations to Beers Law Beer s law is a limiting law that is valid only for low concentrations of analyte. There are two contributions to this fundamental limitation to Beer s law. At higher concentrations the individual particles of analyte no longer behave independently of one another. The resulting interaction between particles of analyte may change the value of 8. A second contribution is that the absorptivity, a, and molar absorptivity, 8, depend on the sample s refractive index. Since the refractive index varies with the analyte s concentration, the values of a and 8 will change. For sufficiently low concentrations of analyte, the refractive index remains essentially constant, and the calibration curve is linear. [Pg.386]

Composition The law of mass aclion is expressed as a rate in terms of chemical compositions of the participants, so ultimately the variation of composition with time must be found. The composition is determined in terms of a property that is measured by some instrument and cahbrated in terms of composition. Among the measures that have been used are titration, pressure, refractive index, density, chromatography, spectrometry, polarimetry, conduclimetry, absorbance, and magnetic resonance. In some cases the composition may vary linearly with the observed property, but in every case a calibration is needed. Before kinetic analysis is undertaken, the data are converted to composition as a function of time (C, t), or to composition and temperature as functions of time (C, T, t). In a steady CSTR the rate is observed as a function of residence time. [Pg.707]

Several sources (References 19. 20, 21, and 22) are available for estimating pumps and drivers to check in-house coiTelations or to fill in w here data is not available. Care must be exercised in using construction cost indexes to update the literature data. It would be w ise to calibrate the indexes and literature data by getting vendor prices on a few of the larger, more expensive pumps, and 5% or 10% of the common types of pumps in the project being estimated. [Pg.233]

The documents that define the calibration process themselves are derived documents and therefore will be governed by your control procedure. They do not need to be listed along with all your other control and operating procedures in the index of quality system documents. A separate index of calibration methods should be maintained. Calibration methods are like test and inspection procedures, they are product specific. [Pg.415]

Use of a refractometer to make measurements of the refractive index of liquids. This will often provide a means of identifying a pure compound, and can also be used (in conjunction with a calibration curve) to analyse a mixture of two liquids. [Pg.10]

Direct or indirect methods may be used to determine moisture in dehydrated foods. Indirect methods must be calibrated in terms of direct methods—the most common of which are the oven, distillation, and Fischer methods. Accuracy of the direct methods is difficult to evaluate except by comparison with a chosen reference method. Several reference methods are reviewed, but none can be given an unqualified recommendation as most practical and suitable for all foods. An indirect measure of moisture is the equilibrium vapor pressure of water, which can be measured easily and accurately. Arguments are presented to show that vapor pressure may be a better index of the stability of dehydrated foods than the moisture content, which has been frequently used for this purpose. [Pg.37]

Figure 5. Calibration curve for the determination of the pyrolusite concentration of orthorhombically indexed y - MnO, samples by comparison of the calculated with the observed 4m0) value. Figure 5. Calibration curve for the determination of the pyrolusite concentration of orthorhombically indexed y - MnO, samples by comparison of the calculated with the observed 4m0) value.
Photoelastic measurement is a very useful method for identifying stress in transparent plastics. Quantitative stress measurement is possible with a polarimeter equipped with a calibrated compensator. It makes stresses visible (Fig. 5-2). The optical property of the index of refraction will change with the level of stress (or strain). When the photoelastic... [Pg.302]

Published refractive index data for the mobile phase, polystyrene, polyacrylonitrile, and the two monomers were used to calculate refractive index detector calibrations for the two homopolymers. The published data were used to determine relationship between refractive index increments of monomer and corresponding homopolymer. Chromatographic refractometer calibrations for the two homopelymers were then calculated from experimentally measured calibration data for the two monomers. [Pg.81]

The CHI index is reportedly a relevant parameter in quantitative structure-activity relationship (QSAR) studies [41]. With this approach, log P could be determined in the range -0.45more than 25000 compounds with excellent reproducibility (within 2 index units) and reported in a GlaxoSmithKline database [11]. Two main drawbacks were identified using this approach (i) the assumptions used in Ref [7], i.e. that S is constant for all compounds and that the system dwell volume is excluded in calculations, yield some discrepancies in the resulting log P, and (ii) the set of gradient calibration... [Pg.342]

Valko et al. [37] developed a fast-gradient RP-HPLC method for the determination of a chromatographic hydrophobicity index (CHI). An octadecylsilane (ODS) column and 50 mM aqueous ammonium acetate (pH 7.4) mobile phase with acetonitrile as an organic modifier (0-100%) were used. The system calibration and quality control were performed periodically by measuring retention for 10 standards unionized at pH 7.4. The CHI could then be used as an independent measure of hydrophobicity. In addition, its correlation with linear free-energy parameters explained some molecular descriptors, including H-bond basicity/ acidity and dipolarity/polarizability. It is noted [27] that there are significant differences between CHI values and octanol-water log D values. [Pg.416]

Many other subjects are important to achieve successful pattern recognition. To name only two, it should be investigated to what extent outliers are present, because these can have a profound influence on the quality of a model and to what extent clusters occur in a class (e.g. using the index of clustering tendency of Section 30.4.1). When clusters occur, we must wonder whether we should not consider two (or more) classes instead of a single class. These problems also affect multivariate calibration (Chapter 36) and we have discussed them to a somewhat greater extent in that chapter. [Pg.239]

An organized document filing system must be maintained. This could be a paper file, an electronic document file, or a mixture of both. The equipment inventory system contains key information on the components of each system, their performance criteria and maintenance and calibration status. All documents including installation and performance documentation, as with other documents necessary to demonstrate the quality of the data, include SOPs for the qualification procedures, calibration, maintenance, personnel training, etc. If the manufacturer s operating, service, or maintenance manuals are used or cited in the operating procedures, copies of these manuals should be maintained. To facilitate retrieval, documentation should be stored in a central location and be indexed for easy retrieval. [Pg.1042]

ISO/IEC Guide 14025, General Requirements for the Competence of Testing and Calibration in Laboratories, International Organization of Standardization, Geneva (2000). Also available on the World Wide Web http //www.iso.ch/iso/en/iso9000-14000/index.html. [Pg.1085]

Sample preparation, injection, calibration, and data collection, must be automated for process analysis. Methods used for flow injection analysis (FLA) are also useful for reliable sampling for process LC systems.1 Dynamic dilution is a technique that is used extensively in FIA.13 In this technique, sample from a loop or slot of a valve is diluted as it is transferred to a HPLC injection valve for analysis. As the diluted sample plug passes through the HPLC valve it is switched and the sample is injected onto the HPLC column for separation. The sample transfer time typically is determined with a refractive index detector and valve switching, which can be controlled by an integrator or computer. The transfer time is very reproducible. Calibration is typically done by external standardization using normalization by response factor. Internal standardization has also been used. To detect upsets or for process optimization, absolute numbers are not always needed. An alternative to... [Pg.76]

The indexed relative sensitivity factor approach obviates the necessity of measuring the relative sensitivity factors from all possible matrices, by transferring relative sensitivity factors for elements between different matrices by using the matrix-dependence of characteristic intensity ratios in the spectra. Calibration curves are constructed relating RSFs for an element in a matrix to the matrix ion species ratio (e.g. M2+/M+ for element M) generated from a single standard. [Pg.79]

Products were analyzed via Waters Model 515 HPLC Pump fitted with a Waters model 2410 refractive index detector. Separations was performed via an Aminex HP-87H 300mm column at 65°C using 0.005M H2SO4 as the mobile phase. Compounds calibrated for this work included xylitol, arabitol, erythritol, threitol, PG, EG, glycerol, lactate, 1-propanol, 2-propanol, ethanol, methanol, and the butanetriol isomers. Any compounds not visible by RID were not quantified in this work. [Pg.168]


See other pages where Calibration 654 INDEX is mentioned: [Pg.355]    [Pg.11]    [Pg.644]    [Pg.692]    [Pg.190]    [Pg.214]    [Pg.59]    [Pg.469]    [Pg.560]    [Pg.71]    [Pg.556]    [Pg.91]    [Pg.115]    [Pg.119]    [Pg.231]    [Pg.290]    [Pg.302]    [Pg.324]    [Pg.347]    [Pg.357]    [Pg.260]    [Pg.645]    [Pg.263]    [Pg.146]    [Pg.271]    [Pg.185]    [Pg.165]    [Pg.125]    [Pg.141]    [Pg.234]    [Pg.445]   


SEARCH



© 2024 chempedia.info