Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detectors, calibration

Published refractive index data for the mobile phase, polystyrene, polyacrylonitrile, and the two monomers were used to calculate refractive index detector calibrations for the two homopolymers. The published data were used to determine relationship between refractive index increments of monomer and corresponding homopolymer. Chromatographic refractometer calibrations for the two homopelymers were then calculated from experimentally measured calibration data for the two monomers. [Pg.81]

Seven standards are prepared from the above HEMA/EMA standard solutions typically with every sample set. The preparation of the detector calibration standards, in this way, accounts for the completeness of the HEMA conversion to MEMA. [Pg.355]

The detector calibration standards are made at convenient concentrations of each analyte. All standards should be refrigerated (2-10 °C) in clean amber-glass bottles with foil/Teflon-lined screw-caps. The absolute volume of the standard solutions may be varied at the discretion of the analyst, as long as the correct proportions of the solute and solvent are maintained. [Pg.372]

Dilute each of the detector calibration standards to a final volume of 100 mL with isooctane-ethyl acetate (9 1, v/v). The final concenfration of each deuferafed com-ponenf is 100 igL in all analytical standards. [Pg.373]

The detector calibration standards are made at convenient concentrations of each analyte. Experience has shown that linearity can be obtained over a range from 0.25 to 20.0 pgL with the listed instmmentation. Concentrations other than those shown below also may be prepared and used if necessary. The LC/MS/MS calibration standards are prepared at 0.10, 0.25,0.50, 1.0,2.0, 5.0, 10.0, and 20.0 pgL . ... [Pg.382]

Detector calibration. A calibration curve is generated for every set of samples with a minimum of five standard levels. The standards are interspersed among the analytical samples of each set. The first and last sample in each analytical sample set must be a standard. [Pg.383]

Injection volume Carrier gas flow rate Transfer line temperature Ionization mode Detector calibration Acquisition type Acquisition masses... [Pg.1174]

Concentration assays are often the least demanding, since usually the component to be measured is abundant and minor components scarce. Even if resolution is poor or there is detector noise, accurate measurements of concentration can still be obtained. In concentration assays, the principal requirements are stringency in the precision of sample dilution and measurement of column losses of the major component. Detector calibration, another important issue in concentration assays, has been discussed above. [Pg.27]

ESI-Mass spectra were obtained using JEOL HX/HX110A mass spectrometer equipped with ESI source (Analytica of Branford, Branford, Cl) and channel array detector, calibrated in the FAB mode with alkali metal iodides. Sample solutions were diluted 2000- fold with 50% aqueous methanol (v/v) containing 0.1% acetic acid. [Pg.121]

The reaction now affords investigators another valuable tool because the absolute intensity of the chemiluminescence has been carefully measured. It can thus be used as a standard light source against which other chemiluminescent reactions can be measured without the need of detector calibrations and geometry corrections. This convenience is due largely to the fine work of Fontijn, Meyer, and Schiff,145,147 who used chemical actinometry to measure the emission from a discharge-flow system. In their first report on this reaction, they determined the value of k12 as 1.0 x 104 M x sec-1 for emission in the... [Pg.240]

If the malfunction is in the circuitry and the detector itself is still functional, it may be possible to obtain temperatures by connecting an external bridge circuit to the detector. Resistance readings may then be taken and a corresponding temperature obtained from the detector calibration curves. [Pg.27]

Figure 13 shows a typical pressurizer level system. Pressurizer temperature is held fairly constant during normal operation. The AP detector for level is calibrated with the pressurizer hot, and the effects of density changes do not occur. The pressurizer will not always be hot. It may be cooled down for non-operating maintenance conditions, in which case a second AP detector, calibrated for level measurement at low temperatures, replaces the normal AP detector. The density has not really been compensated for it has actually been aligned out of the instrument by calibration. [Pg.77]

Germanium detector calibration as function of sample volume and gamma-ray energy... [Pg.25]

These ratios are sensitive to detector calibration and should always be determined on a given dual detector system by monitoring the chromatography of standards. [Pg.119]

Breakthrough Experiments. For the breakthrough experiments the sorbents were packed in 100-mg beds, 4 mm in diameter, and inserted into the six ports in the manifold. The effluent ends of the tubes were connected to a common line leading to the flame ionization detector. While the flame ionization detector, calibrated with the challenge atmosphere, monitored the concentration of vinyl acetate in the combined bed effluents, the pump in the detector drew the challenge atmosphere through each bed at approximately 0.2 L/min. The output from the detector, the breakthrough curve, was recorded with a strip-chart recorder. Most of the porous polymer sorbents tested were first washed with acetone and dried. [Pg.174]

The logarithmic dilution method for detector calibration was introduced by Lovelock [6] for the calibration of GC detectors and was modified by Fowlis and Scott [4] for the calibration of LC detectors. The system provides a continuous flow of gas or liquid through the detector that contains a solute, the concentration of which decreases exponentially with time. A diagram of the log dilution apparatus is shown in figure 3. [Pg.28]

The flow and mixing patterns in this system arc studied by dissolving 1500 g of a salt (S) in the first tank, filling each of the other two tanks with pure solvent, and then starting the 40 L/s flow through the system. Each tank outlet stream is monitored with an on-line thermal conductivity detector calibrated to provide instantaneous readings of salt concentration. The data are plotted versus time and the results are compared with the plots that would be expected if the tanks are all perfectly mixed. Your job is to generate the latter plots. [Pg.573]

This can be done, for example, by multiple detection, where the absolute comonomer concentration, Wk, is measured for each analytical fraction after concentration detector calibration according to... [Pg.228]

K. M. Miller and P. Shebell, "Extension of a Generic Ge Detector Calibration Method for In Situ Gamma-Ray Spectrometry" Radioactivity Radiochemistry, Vol. 11, pp. 14 -24, December, 2000. [Pg.192]

X-ray and gamma-ray standards for detector calibration, lAEA-TECDOC-619, 1991. [Pg.258]


See other pages where Detectors, calibration is mentioned: [Pg.586]    [Pg.459]    [Pg.67]    [Pg.355]    [Pg.363]    [Pg.372]    [Pg.374]    [Pg.382]    [Pg.45]    [Pg.271]    [Pg.39]    [Pg.208]    [Pg.222]    [Pg.74]    [Pg.193]    [Pg.73]    [Pg.76]    [Pg.48]    [Pg.3]    [Pg.24]    [Pg.83]    [Pg.193]    [Pg.80]    [Pg.83]    [Pg.84]    [Pg.163]   
See also in sourсe #XX -- [ Pg.67 ]

See also in sourсe #XX -- [ Pg.45 ]

See also in sourсe #XX -- [ Pg.69 ]

See also in sourсe #XX -- [ Pg.253 , Pg.254 , Pg.257 , Pg.268 , Pg.282 , Pg.284 , Pg.286 , Pg.289 ]

See also in sourсe #XX -- [ Pg.69 ]

See also in sourсe #XX -- [ Pg.2566 , Pg.2567 , Pg.2568 , Pg.2569 , Pg.2570 , Pg.2571 , Pg.2572 , Pg.2573 , Pg.2574 ]




SEARCH



Bare-track detectors calibration

Calibrating leak detectors test leaks

Calibration detector response

Calibration of detectors

Detector efficiency calibrations

Detectors cross calibration

High performance liquid chromatography detector calibration

Refractive index detectors universal calibration

Results of the Detector Calibration

Scintillation detectors calibration

Standards for Calibration of Explosive Detectors

Viscosity detectors universal calibration

Viscosity detectors universal calibration plots

© 2024 chempedia.info