Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration system

A basic calibration procedure is required to adjust the spectrophotometer amplifiers to convert the two measured light intensities, I (sample) and /q (optical or electronic reference), into the correct solution of the term log foil), Le., the absorbance as defined by Beer-Lambert s law (see Section 10.2). Mechanical or electronic simulation of an intensity Z = 7o must result in a zero photometer output. The linearity of the absorbance output should be checked by measuring a standard dilution sequence (see Section 10.2.4). [Pg.223]

The system is calibrated by determining the correlation between concentration and spectrophotometer output for a zero sample and a sample of known concentration (standard). [Pg.223]

The concentration of the standard should be just above the expected maximum sample concentration. If intermediate sample concentrations are expected, calibration between a low and high standard may be preferable. The calibration has to be repeated at regular intervals of from one to several hours depending on the overall performance of the analytical system. [Pg.223]

Aging of the peristaltic pump tubes may modify the sample dilution. Some of the analytical channels tend to form precipitates on the cuvette walls, thus changing the spectrophotometer response, especially after a period of analysing high compound concentrations. Temperature effects may alter the equilibrium state of the formation of the spectrophotometri-cally detected compound, and, finally, the stability of some of the reagents is limited. [Pg.223]

Zero water (wash and blanks) and standards should use a seawater matrix identical (or at least similar) to the samples (Section 10.2.2). Analysers employing steady-state mode are calibrated with a low standard (or zero) and a high standard, which are inserted in the same way as samples. A third, medium standard may be added if a linearity test is desired. After [Pg.223]

for intra- and inter-laboratory comparison, it is important to calibrate the intensity axis also. Recommendations from manufacturers vary, and [Pg.118]

laser reference peak verification (cm ) target (instrument) [Pg.119]

dark current acquisition 4. spectrum of intensity standard Postcalibration target (signal) [Pg.119]

Raman spectrum corrected for dark current and white light [Pg.120]


Software for testing of planar welds from both sides of the weld without changing the system calibration... [Pg.776]

Absolute system calibration with a set of narrow distributed nb glucans (pullulans) by means of dual detection of mass and scattering intensity applying a mixed peak position/ broad standard calibration... [Pg.461]

Absolute system calibration is achieved from normalized DRI (mass mass ev)... [Pg.490]

Molecular weight calibration from a monomer to several million daltons can be carried out by a variety of techniques. Because narrow standards of p(methyl methacrylate) (pMMA) are available, these are often used. Narrow standards of p(styrene) (pSty) are also available and can be used. Using the Mark-Houwink-Sakurada equation and the parameters for pSty and pMMA, a system calibrated with pSty can give pMMA-equivalent values, and vice versa. [Pg.540]

Valko et al. [37] developed a fast-gradient RP-HPLC method for the determination of a chromatographic hydrophobicity index (CHI). An octadecylsilane (ODS) column and 50 mM aqueous ammonium acetate (pH 7.4) mobile phase with acetonitrile as an organic modifier (0-100%) were used. The system calibration and quality control were performed periodically by measuring retention for 10 standards unionized at pH 7.4. The CHI could then be used as an independent measure of hydrophobicity. In addition, its correlation with linear free-energy parameters explained some molecular descriptors, including H-bond basicity/ acidity and dipolarity/polarizability. It is noted [27] that there are significant differences between CHI values and octanol-water log D values. [Pg.416]

Sections on matrix algebra, analytic geometry, experimental design, instrument and system calibration, noise, derivatives and their use in data analysis, linearity and nonlinearity are described. Collaborative laboratory studies, using ANOVA, testing for systematic error, ranking tests for collaborative studies, and efficient comparison of two analytical methods are included. Discussion on topics such as the limitations in analytical accuracy and brief introductions to the statistics of spectral searches and the chemometrics of imaging spectroscopy are included. [Pg.556]

Maintain and calibrate the HPLC system periodically. Preventive maintenance performed by a service engineer annually and system calibrations performed every 6 months are recommended. ... [Pg.263]

This chapter reviews the principles and strategies used for HPLC system calibration that includes the pump, the detector, the autosampler, and the column oven. A case study is used to illustrate the development of the calibration procedures for all system modules and the rationale of setting up acceptance criteria that balance productivity and compliance. [Pg.291]

An additional feature of chemometrics that is appealing to process analytical applications is the use of qualitative models to detect and characterize faults in the analyzer system (calibration, instrument, sampling interface, and sampling systems), sample chemistry, and process dynamics. Such faults can be used to trigger preventive maintenance, and to troubleshoot- thus supporting the long-term reliability of the analyzer system. Specihc examples of such fault detection are given in references [15-16]. [Pg.356]

The differences between the ANSI (1983) and DOE (1986) standards become most apparent for photons with energies between 30 and 100 keV. For the same irradiation condition, a monitoring device calibrated according to ANSI (1983) may yield values that differ by up to 20 percent from those obtained with a system calibrated according to DOE (1986) or ANSI (1993). The differences between the revised ANSI standard (ANSI, 1993) and the DOELAP standard are smaller because the specifications for the size and shape of the slab phantom are similar. [Pg.9]

The calibration procedure provides a body of data about how the personal monitor responds to the various irradiation conditions. These data are converted into formulas or algorithms that generate a value for Hp(lO) for the irradiation conditions assumed in the workplace. The formulas or algorithms apply to the personal monitor system calibrated, and do not change unless there is a modification in the design or types of radiation detectors used in the personal monitor. An example of such a body of data for a particular monitoring device is provided by Ehrlich and Soodprasert (1994). [Pg.10]

Data Analysis. All samples are analyzed for radioactive nuclides by a Ge(Li)-pulse height analyzer system calibrated for counting... [Pg.128]


See other pages where Calibration system is mentioned: [Pg.680]    [Pg.329]    [Pg.73]    [Pg.494]    [Pg.201]    [Pg.217]    [Pg.29]    [Pg.329]    [Pg.444]    [Pg.92]    [Pg.240]    [Pg.224]    [Pg.227]    [Pg.441]    [Pg.11]    [Pg.1]    [Pg.7]    [Pg.8]    [Pg.291]    [Pg.292]    [Pg.293]    [Pg.295]    [Pg.297]    [Pg.299]    [Pg.301]    [Pg.316]    [Pg.332]    [Pg.91]    [Pg.27]    [Pg.27]    [Pg.151]    [Pg.51]    [Pg.805]    [Pg.150]   
See also in sourсe #XX -- [ Pg.196 , Pg.227 ]

See also in sourсe #XX -- [ Pg.223 ]




SEARCH



Calibrated system

Calibrated system

Calibration commercial systems

Calibration flow-injection analysis systems

Calibration of the HPLC system

Calibration supply systems

Calibration, 296 data system

HPLC system calibration

Raman spectroscopy system calibration

System calibration program

System equation of a calibration line with drift

© 2024 chempedia.info