Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Equipment, calibration

Examine a procedure from Standard Methods for the Analysis of Waters and Wastewaters (or another manual of standard analytical methods), and identify the steps taken to compensate for interferences, to calibrate equipment and instruments, to standardize the method, and to acquire a representative sample. [Pg.52]

The intent is that wherever the calibration is performed, the same standards apply. Calibrating equipment in-house should not absolve you from complying with the same requirements that you would need to impose on an external test house. [Pg.393]

SRM 2390 can be used in several different ways depending on quality assurance requirements. Components of the SRM are designed to provide assurance that each step of the RFLP protocol is functioning properly, but they can also be used for trouble shooting, for calibrating equipment, and for testing the efficacy of new lots of reagents. [Pg.164]

Calibrated equipment is appropriately labelled or otherwise identified to ensure that it is not confused with uncalibrated equipment and to ensure that its calibration status is... [Pg.248]

In conclusion, one must insist on the importance of the main metrological characteristic, the traceabilility (generally of a result), ensuring a clear (unbroken) relationship between the final result and the complete measurement scheme by using appropriate procedures, standards and calibrated equipment. However, for chemical metrology and particularly for on-site measurement, some adaptations are needed for a wider meaning of traceability [66]. [Pg.265]

One or more such solutions may be needed to calibrate equipment or to otherwise compare to or react with the analyte in the sample. [Pg.6]

One or more such solutions are needed to calibrate equipment. [Pg.153]

The plant fire brigade and the local volunteer fire department were notified by the supervisor of the catalyst preparation area by 11 12 A.M. On their arrival to the scene of the fire at 11 15 A.M., the plant fire brigade saw the lead outside operator down about 40 feet from the fire, in between the catalyst preparation area and reactor building No. 1. They also found a seriously burned unknown person about 120 feet from the fire, near the finishing building. (This person was eventually determined to be a service contractor who entered the premises at 10 30 A.M. to calibrate equipment in the instrument house for Reactor No. 1.)... [Pg.370]

AM. Service contractor enters area to calibrate equipment in the Polyethylene Reactor No. 1 instrument house. (Interview)... [Pg.382]

In the above discussion, standard reference materials (SRMs) were mentioned often. A reference material (RM) is a material or substance suitable for use in calibrating equipment or standardizing solutions. A certified reference material (CRM) that a vendor indicates, via a certificate, is an RM. A standard reference material (SRM) is one that is distributed and certified by a certifying body, such as NIST. The SRM is the material to which all calibration and standardization materials should be traceable. A standard material becomes one when it is compared to or prepared from another. Ultimately, it all rests on the SRM — meaning all standard materials are traceable to an SRM (see Figure 5.10). [Pg.35]

As explained in Sections 16.3.4, 6.4.1, and 16.4.2, SEC is a nonabsolute method, which needs calibration. The most popular calibration materials are narrow molar mass distribution polystyrenes (PS). Their molar mass averages are determined by the classical absolute methods—or by SEC applying either the absolute detection or the previously calibrated equipment. The latter approach may bring about the transfer and even the augmentation of errors. Therefore, it is recommended to apply exclusively the certified well-characterized materials for calibrations. These are often called PS calibration standards and are readily available from numerous companies in the molar mass range from about 600 to over 30,000,000g moL. Their prices are reasonable and on average (much) lower than the cost of other narrow MMD polymers. Other available homopolymer calibration materials include various poly(acrylate)s and poly(methacrylate)s. They are, similar to PS, synthesized by anionic polymerization. Some calibration materials are prepared by the methods of preparative fractionation, for example, poly(isobutylene)s and poly(vinylchloride)s. [Pg.491]

One of the key concerns of analytical science is how good are the numbers produced . Even with an adequately developed, optimised and collaboratively tested method which has been carried out on qualified and calibrated equipment the question remains. Recently it has become fashionable to extend the concepts of the physical metrology into analytical measurements and to quantify confidence in terms of the much more negative uncertainty.It is based on the bottom-up principle or the so called error budget approach. This approach is based on the theory that if the variance contributions of all sources of error involved in analytical processes then it is possible to calculate the overall process... [Pg.56]

Testing/calibration equipment identification (e.g., traceable to national standards)... [Pg.602]

The rate of heating of the sample influences volatile matter values and makes it necessary to calibrate equipment to achieve a satisfactory and reproducible heating rate. This calibration can be accomplished by using either a manual or an automatic mechanical device that lowers the sample crucible into the electrically heated furnace at a reproducible rate. [Pg.57]

The GLP regulations boil down to this if you submit a study to a regulatory agency, then this study should have been conducted in a proper facility by qualified personnel, using properly maintained and calibrated equipment, following written standard procedures and checked routinely by an independent and qualified person. All the original data should be archived and it should be possible to validate the final report of the study by an audit of raw data. [Pg.15]

Third, using properly maintained and calibrated equipment. There is nothing particularly new about this. We expect to see records of calibration of equipment, either done as a separate routine or as part of the analytical sequence. Moreover, we expect to see a written record of these calibrations and the record should be unique to that piece of equipment. The equipment should be properly maintained and there is to be a record of this maintenance. You maintain and service it, the dealer does it, the factory does it, whatever. Someplace there is a log that shows when the equipment was taken out of service, what was wrong, when it was fixed and recalibrated, and when it was put back into service. Why - because the... [Pg.19]

In practice, many of the conditions of measurement in practical chemical measurement are specified in terms of physical measurements. So, too, are many of the inputs to a given calculation. Though the establishment of traceability in these fields has been far from trivial, it is now essentially a routine matter for laboratories to obtain suitable calibrated equipment for measuring quantities such as length, volume, mass, temperature and time. The problem for most laboratories is related to their chemical reference values for amount of substance measurements. [Pg.291]

Qualified and calibrated equipment, including robotic methods Validated LIMS, computer programs for calculation, and other data-handling systems... [Pg.215]

Use of validated, stability-indicating methods Evaluation of forced degradation products Properly executed method transfer protocols Use of qualified and calibrated equipment by trained analyst Use of qualified and calibrated stability chambers or rooms Handling of OOS and OOT results Timely implementation of corrective actions Stability chambers/rooms... [Pg.219]

Sometimes a measurement involves a single piece of calibrated equipment with a known measurement uncertainty value o, and then confidence limits can be calculated just as with the coin tosses. Usually, however, we do not know o in advance it needs to be determined from the spread in the measurements themselves. For example, suppose we made 1000 measurements of some observable, such as the salt concentration C in a series of bottles labeled 100 mM NaCl. Further, let us assume that the deviations are all due to random errors in the preparation process. The distribution of all of the measurements (a histogram) would then look much like a Gaussian, centered around the ideal value. Figure 4.2 shows a realistic simulated data set. Note that with this many data points, the near-Gaussian nature of the distribution is apparent to the eye. [Pg.69]

The other issue to consider is cost, both capital and operating. Capital is generally lower for the antisealant feed system than for a softener. An antisealant feed system can be had for as little as about 2,500 for the pump, make-down calibration equipment, and a day... [Pg.309]

Further barriers to a broader use of Raman spectroscopy are problems with intensity calibration and benchmarking, lack of reference materials, and the expense of the equipment. The calibration equipment, which is typically supplied with the instruments by the manufacturers, varies for instance neon or mercury lamps are delivered with dispersive instruments and HeNe lasers with FT spectrometers. [Pg.54]

Closely related to preventive action is the process audit. Such an audit is performed as part of a laboratory assessment for each test on the proposed scope of accreditation. Performance of a test procedure is observed relative not only to the requirements of the ISO 17025 standard (or ISO Guide 25) but also with respect to performing the ASTM or ISO test method properly. As a minimum, the laboratory must have properly calibrated equipment, the current test method, and trained testing personnel. In addition, a representative number of test demonstrations will reveal whether good laboratory practices are observed in sample preparation, using the correct test parameters and calculating the results. When operators and technicians accept responsibility through such demonstrations, they tend to commit to the quality process. Real improvement occurs with involvement at all levels. [Pg.84]


See other pages where Equipment, calibration is mentioned: [Pg.45]    [Pg.50]    [Pg.815]    [Pg.102]    [Pg.1152]    [Pg.518]    [Pg.175]    [Pg.181]    [Pg.993]    [Pg.298]    [Pg.6]    [Pg.150]    [Pg.153]    [Pg.338]    [Pg.369]    [Pg.26]    [Pg.400]    [Pg.312]    [Pg.330]    [Pg.19]    [Pg.19]    [Pg.135]    [Pg.113]    [Pg.127]    [Pg.434]    [Pg.20]    [Pg.547]   
See also in sourсe #XX -- [ Pg.15 , Pg.122 , Pg.136 ]




SEARCH



Calibration and Validation Equipment

Calibration equipment and

Calibration of equipment

Die and calibration equipment

Equipment Calibration and Qualification

Equipment enthalpy calibration

Equipment maintenance and calibration

Equipment temperature calibration

Feed Equipment Calibration or Verification

Initial preparation and calibration of equipment

Maintenance and Calibration of Equipment

Measuring Equipment and Calibration

Volumetric equipment, calibration

© 2024 chempedia.info