Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Stability chambers calibration

All stability studies on clinical trial materials must be carried out in full accordance with cGMPs, even if a research department carries out the studies. All studies must be carried out by adequately trained personnel under adequate work conditions. The personnel must use properly qualified and calibrated stability chambers, instruments, reagents, and standards. They must follow validated analytical methods and approved written procedures, and they must properly document all work. There must be proper sample and data traceability, change control, and go on. [Pg.191]

Use of validated, stability-indicating methods Evaluation of forced degradation products Properly executed method transfer protocols Use of qualified and calibrated equipment by trained analyst Use of qualified and calibrated stability chambers or rooms Handling of OOS and OOT results Timely implementation of corrective actions Stability chambers/rooms... [Pg.219]

Because the stability chambers are an integral part of the stability program and require continuous performance to specifications for long-term studies, all aspects of the chambers must be described in detail in an SOP. The SOP should include the procedures and the schedule for calibration of the chambers, the description and operating parameters, a routine maintenance schedule, inventory system, IQ/OQ procedures and a monitoring system (PQ), and emergency procedures for malfunctions or unusual occurrences. [Pg.452]

Abstract An important aspect of all stability studies is the stability chambers themselves. This chapter is intended to provide a description of the different sizes and types of chambers that are available, the chamber tolerances required, and to provide some practical information for qualification, calibration, maintenance and monitoring of the chambers. Temperature, humidity and photo-stability chambers are included. Also included are guidelines on how to handle chamber excursions. [Pg.285]

Chapter 14 introduces stability chambers. It also discusses factors to be considered for chamber validation, calibration, and maintenance. This chapter also elaborates on ICH QIB guideline, which established the requirements for photostability condition. [Pg.369]

Resistors of 100-200 kO, realized with a heavily doped meander implanted on a 1 mm3 silicon chip, are attached to each absorber in order to calibrate and stabilize the gain of the bolometer over long running periods. The towers will be mechanically connected to an OFHC copper top plate, elastically suspended from the DR mixing chamber. [Pg.361]

The instrument is calibrated by solution standards, which contain the elements of interest in an appropriate matrix. The concentration range covered for ICP-AES may be several orders of magnitude. Standards, blanks, and samples are analyzed in a sequence appropriate with the instrumental stability and precision desired. During nebullzation, approximately fifteen seconds is required to obtain a steady-state signal another ten seconds is required to integrate the signal and a thirty second blank rinse is required to clean out the spray chamber. The actual time intervals will vary from system to system. [Pg.114]

Firstly, the calibration of the individual chambers has been verified with muon runs. The charge yield obtained agrees well with the cosmic-ray muon calibrations performed several years ago. This is an indication for the longterm signal stability of these detectors. [Pg.385]

The air circulation system for the OPC drying system was plugged to stabilize the airflow in the chamber. The reproducibility of dcmemeasurement was tested using (NH4)2S04 particles. The random errors associated with dc me measurement were 0.1 and 0.3 nm for AT 1 and AT 2, respectively. The influence of this error on the critical S was calculated to be negligibly small (less than 0.001%, absolute). Although, we employed the CCNC manufactured by DMT, the present result is applicable for other types of CCNCs, as they are calibrated similarly (e.g., Snider et al., 2006 Frank et al., 2007). [Pg.246]

Figure 4 - Preparation and insertion of pH probe. 1) The residual 4M KCL solution in the reference chamber is aspirated and replaced with fresh solution. 2 and 3) The sterile probe is calibrated in low (6.86) and high (7.41) standard pH buffers. 4) A 2 cm. incision is made down to the fascia lata in which a 1 cm, incision is made. 5) A subfascial tunnel which will accept the distal 2 cm. of the probe is made with blunt dissection. 6) The probe, the cable of which is attached to a pH meter, is stabilized by closing the fascia and skin over the probe and taping the exposed portion of the probe to the skin. 7) Final position of the probe on the muscle surface. The right angle extension on the exposed plastic cap is useful for securing the probe to the limb and orienting the sensing tip. Figure 4 - Preparation and insertion of pH probe. 1) The residual 4M KCL solution in the reference chamber is aspirated and replaced with fresh solution. 2 and 3) The sterile probe is calibrated in low (6.86) and high (7.41) standard pH buffers. 4) A 2 cm. incision is made down to the fascia lata in which a 1 cm, incision is made. 5) A subfascial tunnel which will accept the distal 2 cm. of the probe is made with blunt dissection. 6) The probe, the cable of which is attached to a pH meter, is stabilized by closing the fascia and skin over the probe and taping the exposed portion of the probe to the skin. 7) Final position of the probe on the muscle surface. The right angle extension on the exposed plastic cap is useful for securing the probe to the limb and orienting the sensing tip.
Those users of LC/MS with some experience in mass spectrometry tend to polish the ion source and ion optics before introducing an important series of samples. Of course, maximum signal intensities in API LC/MS ionization techniques will be achieved when all residues are completely eliminated from the ion source surfaces. The electrical fields that are necessary to form and/or transport the ionic analytes before they reach the mass analyzer are influenced by the degree of contamination with non-volatile sample constituents. On the other hand, experienced LC/MS users know that the long-term stability of an LC/MS system will be increased if the spray chamber is intentionally slightly contaminated. This is of particular importance when one wishes or has to quantify by means of external calibration. With the most commonly used ion sources it is sufficient to inject the sample matrix 10-20 times in order to achieve an almost constant analyte response after an initial exponential drop. [Pg.544]

Chambers used in the formal stability program must go through qualification and calibration procedures. The chambers should be qualified at the time of purchase to assure a proper installation, a process denoted installation qualification (IQ). This procedure documents that the chamber has been installed properly and that it meets the manufacturer s specifications. The chamber should also be... [Pg.467]


See other pages where Stability chambers calibration is mentioned: [Pg.253]    [Pg.39]    [Pg.208]    [Pg.253]    [Pg.285]    [Pg.287]    [Pg.289]    [Pg.291]    [Pg.293]    [Pg.295]    [Pg.297]    [Pg.299]    [Pg.301]    [Pg.467]    [Pg.169]    [Pg.146]    [Pg.105]    [Pg.11]    [Pg.180]    [Pg.215]    [Pg.88]    [Pg.328]    [Pg.261]    [Pg.317]    [Pg.69]    [Pg.109]    [Pg.732]    [Pg.106]    [Pg.199]    [Pg.140]   
See also in sourсe #XX -- [ Pg.293 , Pg.299 ]




SEARCH



Stability chambers

Stability chambers chamber

© 2024 chempedia.info