Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration defined

Physical and chemical effects can be combined for identification as sample matrix effects. Matrix effects alter the slope of calibration curves, while spectral interferences cause parallel shifts in the calibration curve. The water-methanol data set contains matrix effects stemming from chemical interferences. As already noted in Section 5.2, using the univariate calibration defined in Equation 5.4 requires an interference-free wavelength. Going to multivariate models can correct for spectral interferences and some matrix effects. The standard addition method described in Section 5.7 can be used in some cases to correct for matrix effects. Severe matrix effects can cause nonlinear responses requiring a nonlinear modeling method. [Pg.135]

Fig. 15. Straight line calibration with confidence interval of calibration defined in y direction a and confidence intervals of the analytical result defined in x-direction b applying variing amounts of repeated measurements in the analytical procedure S8>... Fig. 15. Straight line calibration with confidence interval of calibration defined in y direction a and confidence intervals of the analytical result defined in x-direction b applying variing amounts of repeated measurements in the analytical procedure S8>...
Camera Calibration. Each camera must be calibrated before it can contribute to locating a target in object-space. Camera calibration defines a mapping from three-dimensional object-space into the two-dimensional u, v coordinates of the camera. This mapping is expressed in Eq. (5.2) in terms of homogeneous coordinates ... [Pg.121]

The ATa calibration defined for a tensile loading (i.e. same A, ) values listed in Table 1 and with characteristic length = h/2) also applies when the adhesive undergoes a uniform temperature change, AT, if the characteristic stress is defined as... [Pg.155]

The viscosity is determined by measuring the time it takes for a crude to flow through a capillary tube of a given length at a precise temperature. This is called the kinematic viscosity, expressed in mm /s. It is defined by the standards, NF T 60-100 or ASTM D 445. Viscosity can also be determined by measuring the time it takes for the oil to flow through a calibrated orifice standard ASTM D 88. It is expressed in Saybolt seconds (SSU). [Pg.318]

An experimental activity on the stress measurement of a pressure vessel using the SPATE technique was carried out. It was demontrated that this approach allows to define the distribution of stress level on the vessel surface with a quite good accuracy. The most significant advantage in using this technique rather than others is to provide a true fine map of stresses in a short time even if a preliminary meticolous calibration of the equipment has to be performed. [Pg.413]

With most non-isothemial calorimeters, it is necessary to relate the temperature rise to the quantity of energy released in the process by determining the calorimeter constant, which is the amount of energy required to increase the temperature of the calorimeter by one degree. This value can be detemiined by electrical calibration using a resistance heater or by measurements on well-defined reference materials [1], For example, in bomb calorimetry, the calorimeter constant is often detemiined from the temperature rise that occurs when a known mass of a highly pure standard sample of, for example, benzoic acid is burnt in oxygen. [Pg.1902]

Calibrating the electrode presents a third complication since a standard with an accurately known activity for H+ needs to be used. Unfortunately, it is not possible to calculate rigorously the activity of a single ion. For this reason pH electrodes are calibrated using a standard buffer whose composition is chosen such that the defined pH is as close as possible to that given by equation 11.18. Table 11.6 gives pH values for several primary standard buffer solutions accepted by the National Institute of Standards and Technology. [Pg.492]

The components of a quality assurance program are designed to serve the two functions just mentioned—control and assessment. Quality control operations are defined by operational procedures, specifications, calibration procedures, and standards and contain the following components ... [Pg.223]

Hartree-Fock MO approach, the minimization of energy should provide the most accurate description of the electronic field. The mathematical problem is to define each of the terms, with being the most challenging. The formulation carmot be done exactly, but various approaches have been developed and calibrated by comparison with experimental data. The methods used most frequently by chemists were developed by A. D. Becke. " This approach is often called the B3LYP method. The computations can be done with... [Pg.59]

Systematic error, as stated above, can be eliminated— not totally, but usually to a sufficient degree. This elimination process is called calibration. Calibration is simply a procedure where the result of measurement recorded by an instrument is compared with the measurement result of a standard. A standard is a measuring device intended to define, to represent physically, to conserve, or to reproduce the unit of measurement in order to transmit it to other measuring instruments by comparison. There are several categories of standards, but, simplifying a little, a standard is an instrument with a very high accuracy and can for that reason be... [Pg.1124]

The bias error is a quantity that gives the total systematic error of a measuring instrument under defined conditions. As mentioned earlier, the bias should be minimized by calibration. The repeatability error consists of the confidence limits of a single measurement under certain conditions. The mac-curacy or error of indication is the total error of the instrument, including the... [Pg.1130]

Constriction measurement devices constructed to standards do not necessarily require calibration. One idea of strict standardization is to define the manufacturing, tolerances, and other features in such a way that the instruments made according to these rules require no calibration. The properties are so well known that a certain accuracy can be guaranteed. If the accuracy specified in the standard is inadequate, additional calibration procedures are required. The same applies to Pitot-static tubes made according to standard specifications." ... [Pg.1168]

Some electronic equipment has self-calibration routines built in to the start-up sequence. This should be taken as an indication of serviceability and not of absolute calibration. The device should still be subject to independent calibration at a defined frequency. [Pg.403]

The standard requires the supplier to define the process employed for the calibration of inspection, measuring, and test equipment, including details of equipment type, unique identification, location, frequency of checks, check method, acceptance criteria, and the action to be taken when results are unsatisfactory. [Pg.414]

The standard requires you to define the calibration process rather than the calibration procedures. Calibration procedures will define how you control all calibration activities. The calibration process required only applies to the process of calibrating a particular device. This does not mean that you need to define the process of calibrating every device. The process of calibration may be ... [Pg.414]

Whichever method applies you should define it within your quality system. You should maintain lists of all the devices that require calibration and indicate in these lists ... [Pg.415]

The documents that define the calibration process themselves are derived documents and therefore will be governed by your control procedure. They do not need to be listed along with all your other control and operating procedures in the index of quality system documents. A separate index of calibration methods should be maintained. Calibration methods are like test and inspection procedures, they are product specific. [Pg.415]

In what documents is the process employed for the calibration of inspection, measuring, and test equipment defined ... [Pg.424]

In between come the semi-empirical models such as PPP or ZINDO. Here we start with a rigorous Hamiltonian and perhaps a well-defined basis set, but we then calibrate the difficult integrals against experiment. [Pg.173]

The steel housing rarely exceeds 0.5 in. (12 mm) and a calibration is done in terms of API units, arbitrary units defined in a standard calibration pit located at the University of Houston. [Pg.972]

To define the position of an absorption, the NMR chart is calibrated and a reference point is used. In practice, a small amount of tetramethylsilane [TMS (CH )4Si] Is added to the sample so that a reference absorption peak is produced when the spectrum is run. TMS is used as reference for both l H and 13C measurements because it produces in both a single peak that occurs upfield of other absorptions normally found in organic compounds. The ]H and 13C spectra of methyl acetate in Figure 13.3 have the l MS reference peak indicated. [Pg.445]

For scientific work the fundamental standard of mass is the international prototype kilogram, which is a mass of platinum-iridium alloy made in 1887 and deposited in the International Bureau of Weights and Measures near Paris. Authentic copies of the standard are kept by the appropriate responsible authorities in the various countries of the world these copies are employed for the comparison of secondary standards, which are used in the calibration of weights for scientific work. The unit of mass that is almost universally employed in laboratory work, however, is the gram, which may be defined as the one-thousandth part of the mass of the international prototype kilogram. [Pg.75]

The capacity of a glass vessel varies with the temperature, and it is therefore necessary to define the temperature at which its capacity is intended to be correct in the UK a temperature of 20 °C has been adopted. A subsidiary standard temperature of 27 °C is accepted by the British Standards Institution, for use in tropical climates where the ambient temperature is consistently above 20 °C. The US Bureau of Standards, Washington, in compliance with the view held by some chemists that 25 °C more nearly approximates to the average laboratory temperature in the United States, will calibrate glass volumetric apparatus marked either 20 °C or 25 °C. [Pg.80]


See other pages where Calibration defined is mentioned: [Pg.150]    [Pg.150]    [Pg.739]    [Pg.1121]    [Pg.1828]    [Pg.47]    [Pg.127]    [Pg.411]    [Pg.444]    [Pg.2547]    [Pg.2]    [Pg.66]    [Pg.435]    [Pg.363]    [Pg.1125]    [Pg.1140]    [Pg.79]    [Pg.392]    [Pg.148]    [Pg.858]    [Pg.643]    [Pg.232]    [Pg.313]    [Pg.240]    [Pg.565]   
See also in sourсe #XX -- [ Pg.11 ]




SEARCH



Calibration curve defined

Defining an SEC calibration

Quantification of Analytical Data via Calibration Curves in Mass Spectrometry Using Certified Reference Materials or Defined Standard Solutions

© 2024 chempedia.info