Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Addition calibration

Table 14 can be regarded as providing a reasonable overall picture, even if the results cannot applied to any particular case. However, if the underlying principle is accepted, it becomes clear that improvements in a single stage, for example the reduction of instrument variation, has a negligible beneficial effect (if this variation was not outside the normal range ). Even if the contribution of repeatability is re-duced to zero, the cumulative uncertainty is reduced by 10% only, i.e. from 2.2 to y(0.0)2 (0.8)2 (1.0)2 + (1.5)2 = 2.0. This statistical view of errors should help to avoid some unnecessary efforts to improve, e.g., calibration. Additionally, this broad view on all sources of error may help to detect the most important ones. Consequently, without participation in proficiency tests, any method validation will remain incomplete. [Pg.131]

Classical calibration procedure can only be applied when all the species that contribute to the form of the spectra are known and can be included into the calibration. Additionally, there is the constraint that no interactions between the analytes and other species (e.g. solvent) or effects (e.g. of temperature) should occur. [Pg.184]

Protein based assays are influenced by the relative amount of protein in the cells, the so-called expression level. Expression of a specific protein varies according to different parameters, such as climate, soil, drought conditions, etc. If the expression level in a given sample is different from the expression level of the materials used as calibrators, additional uncertainty will impact the result expressed in mass fractions. While again, for individual and intact kernels other analytical approaches (pooled sampling procedures from GIPSA, see above) are possible that would not be affected by this phenomenon, this would not be applicable for processed samples. [Pg.137]

The checkers obtained n-butyllithium from the Foote Mineral Co., Exton, Pennsylvania. The concentration of n-butyllithium was determined by the method of Gilman and Haubein.2 This reagent is conveniently transferred to a previously calibrated addition funnel under nitrogen pressure through a short length of inert (e.g., Teflon) tubing. [Pg.105]

The unit used to compare the computer line s pressure measurement readings with equipment air pressure measurements has not been calibrated. Additionally there has been no periodic maintenance to assure that the unit is operating appropriately. The issue becomes even more critical because the. .. computer system is not validated. It is essential that this unit. .. be accurate and reliable. [FDA Warning Letter, 1999]... [Pg.681]

Cahbrations are commonly performed with phantoms simulating the neck. Due to the mentioned parameters a neck phantom alone may be inadequate as a means of calibration. For individual calibrations, additional mathematical corrections are always required and, even then, there will be uncertainties in the results due to the namre of radiation interaction, which makes it difficult to completely correct for all scattering and absorption events. [Pg.34]

This discussion on camera calibration is not meant to be comprehensive. However, it does provide the basic background for understanding how and why cameras ate calibrated. Additional terms can be added to the basic 11-parameter DLT model to correct for symmetrical and asymmetrical lens distortions. These errors can be treated, in part, during camera calibration, and may also be accounted for by using lens correction maps provided by the manufacturer. [Pg.123]

All the above-mentioned experiments were performed with the use of very sensitive fluorescence technique by means of commercial fluorimeter and of modified quartz plates in a single fast measurement. The covalent immobilization of the label enables multiple use of the same plate without the need to prepare and calibrate additional solutions - an important advantage for possible industrial use. [Pg.295]

Raman Spectroscopy Historically, Raman spectroscopy was never considered a sensitive technique because only 1 in 10 photons emitted from a molecule is collected. However, Raman systems have improved tremendously in the last several years. It is no longer deemed an insensitive, irreproducible, fluorescence-dominated technique. Raman is a versatile technique capable of providing information on several parameters simultaneously, such as monomer concentration and particle size. Raman is especially amenable for monomer detection in water-soluble polymers because symmetric vinylic monomer structures are good Raman scatterers and water has a weak signal. To that end, Raman is a complementary technique to FTIR and can be used to monitor monomer concentration and conversion. By employing a near-IR laser (785 nm) which removes most of the fluorescence, along with sharp monomer and polymer peaks that are often separated, monomer concentrations may be determined with univariate calibration. Additionally, since Raman is sensitive to the local molecular environment, it may be used to provide particle size information. [Pg.392]

Fig. B.9 Polagram of the zinc standard addition experiment originally from Ref. [15] where the figure caption reads Determination of Zinc by calibration (standard) addition, a unknown and supporting electrolyte, b same solution after calibration addition. Reproduced with permission from Ref. [12]. Copyright Springer Science and Business Media... Fig. B.9 Polagram of the zinc standard addition experiment originally from Ref. [15] where the figure caption reads Determination of Zinc by calibration (standard) addition, a unknown and supporting electrolyte, b same solution after calibration addition. Reproduced with permission from Ref. [12]. Copyright Springer Science and Business Media...
In multivariate calibration, additional preprocessing with so-called scatter correction is common to remove variation in X that is due to light scattering and other sources unrelated to Y. ... [Pg.2010]

The calibration curve of each rosetta strain gauge was so obtained and ftg.5 shows the sum of the principal stresses at the measuring points versus pressure inside the vessel. Further tests were carried out to obtain the calibration factor and to check that it remained constant on the whole scan area of the test surface. This was achieved through additional measurements using the SPATE system on fixed points on the surface located very close to the applied rosetta strain gauges. This procedure gave the following results ... [Pg.411]

The choice of solvent cannot usually be made on the basis of theoretical considerations alone (see below), but must be experimentally determined, if no information is already available. About 0 -1 g. of the powdered substance is placed in a small test-tube (75 X 11 or 110 X 12 mm.) and the solvent is added a drop at a time (best with a calibrated dropper. Fig. 11, 27, 1) with continuous shaking of the test-tube. After about 1 ml. of the solvent has been added, the mixture is heated to boiling, due precautions being taken if the solvent is inflammable. If the sample dissolves easily in 1 ml. of cold solvent or upon gentle warming, the solvent is unsuitable. If aU the solid does not dissolve, more 11,27,1. solvent is added in 0-5 ml. portions, and again heated to boiling after each addition. If 3 ml. of solvent is added and the substance... [Pg.124]

In addition to the orthodox method, just described, for the determination of the boiling points of liquids, the student should determine the boiling points of small volumes (ca. 0 5 ml.) by Siwolobofifs method. Full details are given iri Section 11,12. Determine the boiling points of the pure liquids listed in the previous paragraph. Observe the atmospheric pressure and if this differs by more than 5 mm. from 760 mm., correct the boiling point with the aid of Table II,9,B. Compare the observed boiling points with the accepted values, and draw a calibration curve for the thermometer. [Pg.231]

To prepare the solution we measure out exactly 0.1500 g of Cu into a small beaker. To dissolve the Cu we add a small portion of concentrated HNO3 and gently heat until it completely dissolves. The resulting solution is poured into a 1-L volumetric flask. The beaker is rinsed repeatedly with small portions of water, which are added to the volumetric flask. This process, which is called a quantitative transfer, ensures that the Cu is completely transferred to the volumetric flask. Finally, additional water is added to the volumetric flask s calibration mark. [Pg.30]

Colorplate 2 shows an example of a set of standard additions and their corresponding standard additions calibration curve. [Pg.113]

Examples of calibration curves for the method of standard additions. In (a) the signal is plotted versus the volume of the added standard, and in (b) the signal is plotted versus the concentration of the added standard after dilution. [Pg.114]

A fifth spectrophotometric method for the quantitative determination of the concentration of Pb + in blood uses a multiple-point standard addition based on equation 5.6. The original blood sample has a volume of 1.00 mb, and the standard used for spiking the sample has a concentration of 1560 ppb Pb +. All samples were diluted to 5.00 mb before measuring the signal. A calibration curve of Sjpike versus Vj is described by... [Pg.114]

Figure 5.7(b) shows the relevant relationships when Sspike is plotted versus the concentrations of the spiked standards after dilution. Standard addition calibration curves based on equation 5.8 are also possible. [Pg.115]

Since a standard additions calibration curve is constructed in the sample, it cannot be extended to the analysis of another sample. Each sample, therefore, requires its own standard additions calibration curve. This is a serious drawback to the routine application of the method of standard additions, particularly in laboratories that must handle many samples or that require a quick turnaround time. For example, suppose you need to analyze ten samples using a three-point calibration curve. For a normal calibration curve using external standards, only 13 solutions need to be analyzed (3 standards and 10 samples). Using the method of standard additions, however, requires the analysis of 30 solutions, since each of the 10 samples must be analyzed three times (once before spiking and two times after adding successive spikes). [Pg.115]

The method of standard additions can be used to check the validity of an external standardization when matrix matching is not feasible. To do this, a normal calibration curve of Sjtand versus Cs is constructed, and the value of k is determined from its slope. A standard additions calibration curve is then constructed using equation 5.6, plotting the data as shown in Figure 5.7(b). The slope of this standard additions calibration curve gives an independent determination of k. If the two values of k are identical, then any difference between the sample s matrix and that of the external standards can be ignored. When the values of k are different, a proportional determinate error is introduced if the normal calibration curve is used. [Pg.115]

In a standard addition the analyte s concentration is determined by extrapolating the calibration curve to find the x-intercept. In this case the value of X is... [Pg.124]


See other pages where Addition calibration is mentioned: [Pg.38]    [Pg.188]    [Pg.425]    [Pg.197]    [Pg.772]    [Pg.1335]    [Pg.2753]    [Pg.310]    [Pg.50]    [Pg.144]    [Pg.276]    [Pg.38]    [Pg.188]    [Pg.425]    [Pg.197]    [Pg.772]    [Pg.1335]    [Pg.2753]    [Pg.310]    [Pg.50]    [Pg.144]    [Pg.276]    [Pg.10]    [Pg.159]    [Pg.286]    [Pg.484]    [Pg.694]    [Pg.999]    [Pg.1289]    [Pg.1908]    [Pg.605]    [Pg.257]    [Pg.429]    [Pg.88]    [Pg.72]    [Pg.27]    [Pg.113]    [Pg.124]   


SEARCH



Addition-calibration method

Calibration by standard addition

Calibration graphs standard addition method used

Calibration standard additions

Calibration standard-additions method

Inductively coupled plasma mass standard addition calibration

Standard addition method calibration graphs using

Standard addition mode of instrument calibration

Standard-additions method, calibration disadvantages

Standard-additions method, calibration effects

© 2024 chempedia.info