Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytical procedures calibration methods

The method enables one to evaluate the basic process parameters, such as the conversion and yield, without a prior knowledge of the composition of the mixture and without calibrating the instrument. The method significantly simplifies the material balance determination, reduces the analysis time and makes it possible to automate the analytical procedure. The method provides for a fuller utilization of computers and considerably improves the accuracy of the analysis. For example, it can help to control the entire process of producing monomers for synthetic rabber, including such analyses as the determination of the concentration of volatile components in solvents and absorbents, and the concentration.of residual monomers and other volatile components in polymers and their aqueous suspensions [176]. [Pg.243]

Although a routine aspect of any analytical procedure, the method of calibration is a major potential source of systematic error. Error in the preparation of calibration standards or their working dilutions is one obvious source of discrepancy. The recovery of an... [Pg.215]

In most analytical procedures, calibration is carried out by means of a calibration curve using com-pound(s) prepared with chemicals of an appropriate purity and verified stoichiometry. Matrix effects must often be taken into account and, consequently, the calibration solutions should be matrix-matched. CRMs of pure compounds may be used for calibration. However, matrix CRMs should in principle not be used for the purpose of calibration unless no other suitable calibrants are available, with the exception of those methods (e.g., spark source mass spectrometry, wavelength-dispersive XRF, etc.) that require calibration with CRMs of a similar, fully characterized matrix (e.g., metal alloys, cements). For such methods, accuracy can only be achieved when certified RMs are used for the calibration. [Pg.4031]

The choice between X-ray fluorescence and the two other methods will be guided by the concentration levels and by the duration of the analytical procedure X-ray fluorescence is usually less sensitive than atomic absorption, but, at least for petroleum products, it requires less preparation after obtaining the calibration curve. Table 2.4 shows the detectable limits and accuracies of the three methods given above for the most commonly analyzed metals in petroleum products. For atomic absorption and plasma, the figures are given for analysis in an organic medium without mineralization. [Pg.38]

In summary, official German analytical methods for pesticide residues are always validated in several laboratories. These inter-laboratory studies avoid the acceptance of methods which cannot readily be reproduced in further laboratories and they do improve the ruggedness of analytical procedures applied. The recently introduced calibration with standards in matrix improves the trueness of the reported recovery data. Other aspects of validation (sample processing, analyte stability, extraction efficiency) are not considered. [Pg.128]

Residue study protocols typically either include quality specifications for analytical procedures or refer to a written analytical method that includes such specifications. The protocol for an LSMBS should also include analytical quality specifications, either directly or by reference to a method. Analytical specifications usually include minimum and maximum recovery of analyte from fortified control samples, minimum number of such fortifications per set of samples, minimum linearity in calibration, minimum stability of response to injection of calibration solutions, and limits of quantitation and of detection. [Pg.239]

Reference and research methods. These are generally more sophisticated procedures that are used by central quality control laboratories or by government agencies with qualified personnel. These methods are often used to verify results obtained by the rapid methods described below or to calibrate the instruments. They may also be used to verify additives declared on the label of a food product or to check for the use of non-permitted additives. This aspect is of increasing importance since legislation may vaiy between different countries, and food products are frequently subject to inter-market exports . The most important characteristics of these analytical procedures are ... [Pg.111]

In order to chemically analyse a sample of com for its protein content, a rather complex analytical procedure (e.g. Kjeldahl analysis) is required, a slow and expensive process. In our example, the PCR/PLS group of methods replaces this procedure with a much faster spectroscopic analysis. First, a mathematical relationship is established from a calibration set, comprising a matrix of NIR-spectra of the collection of samples and the vector of... [Pg.295]

A comprehensive two-volume Handbook of Chemometrics and Qualimetrics has been published by D. L. Massart et al. (1997) and B. G. M. Vandeginste et al. (1998) predecessors of this work and historically interesting are Chemometrics A Textbook (Massart et al. 1988), Evaluation and Optimization of Laboratory Methods and Analytical Procedures (Massart et al. 1978), and The Interpretation of Analytical Chemical Data by the Use of Cluster Analysis (Massart and Kaufmann 1983). A classical reference is still Multivariate Calibration (Martens and Naes 1989). A dictionary with extensive explanations containing about 1700 entries is The Data Analysis Handbook (Frank and Todeschini 1994). [Pg.20]

Two synthetic internal standards are used (nor-DCA and nor-CA) to perform an external standardisation method. Calibration curves for each compound are built, starting from known amounts of standards ranging from 0.02 pg to 0.2 pg. BA standards are prepared and analysed as described in below in section 5.4.5.3 Analytical, Procedure . [Pg.613]

A calibration curve shows the response of an analytical method to known quantities of analyte.8 Table 4-7 gives real data from a protein analysis that produces a colored product. A spectrophotometer measures the absorbance of light, which is proportional to the quantity of protein analyzed. Solutions containing known concentrations of analyte are called standard solutions. Solutions containing all the reagents and solvents used in the analysis, but no deliberately added analyte, are called blank solutions. Blanks measure the response of the analytical procedure to impurities or interfering species in the reagents. [Pg.69]

In principle, all performance measures of an analytical procedure mentioned in the title of this section can be derived from a certain critical signal value, ycrit. These performance measures are of special interest in trace analysis. The approaches to estimation of these measures may be subdivided into methods of blank statistics , which use only blank measurement statistics, and methods of calibration statistics , which in addition take into account calibration confidence band statistics. [Pg.66]

This group of elements contains a large volume of information on analytical laboratory method requirements and procedures. Therefore, the laboratory that will conduct the analysis should provide this information to the project team for the incorporation into the QAPP. For example, the selection of analytical laboratory methods and QC requirements definitely needs input from the analytical laboratory, particularly if low detection limits or non-routine analyses are concerned. Analytical instrument calibration and maintenance requirements should also be developed as a cooperative effort with the analytical laboratory or by the individuals who are well-versed in laboratory practices and procedures. [Pg.79]

RMs Considering the limitations of available primary methods, emphasis is placed by the CCQM on the elaboration of synthetic RMs derived from pure materials. These would then be used for calibration of instruments and hence, could help in the metrological step of analytical procedures. As appears from the cited examples, such RMs are hardly suitable as reference samples in many applications of analytical chemistry. [Pg.42]

Apart from the standard Shewart charts, the analyst can also apply X-charts, on which the mean of several replicate measurements is plotted, or R-charts, where the difference between two replicate measurements is plotted. X- and R-charts give an indication of the reproducibility of the method. Drift in analytical procedure, for example, slows changes in the system caused by the aging of parts of instruments, decalibration in wavelength, or the aging of calibration stock solutions, can be detected early when a Cusum chart (cumulative sum) is applied. In Cusum charts, the analyst reports the cumulative sum of the differences between delivered and reference values. If this reference value is certified (CRM), the Cusum chart allows the accuracy of the determination to be monitored. [Pg.395]

From this it can be concluded that the curvature of IDMS calibration curves can be described very accurately by means of higher order polynomials. The ability to check different models allows one to adapt the same calculation procedure regardless of the actual analytical situation. This is especially important in cases where, due to low efficiency of the synthesis, a large amount of unlabeled or partially labeled material is present in the IS. The setup of analyses with an IS of low mass increment is also facilitated (Jonckheere et al., 1982). In contrast to other calibration methods, no initial estimates of the amount of unlabeled product and/or influence of naturally... [Pg.138]

The resulting metal, in the salt state, is dissolved in nitric acid and diluted to a known volume with double distilled water. The solution is analyzed by Atomic Absorption for the metals of interest. Metals such as Cadmium, Berylium and Nickel have very low threshold limit values which are set by OSHA and are frequently present in the sample in very low quantities. Many Atomic Absorption units are equipped with a hollow graphite tube atomizer which increases the sensitivity dramatically making it easier for the analyst to obtain reliable results for species present in the sample in very low concentrations. The validity of the metal fume data depends on sampling train calibration and the precision and accuracy of the analytical procedure. NIOSH has reported a 2% relative standard deviation in the analytical method which has been collaboratively tested. [Pg.189]

In the method described, the procedural blank determination is essential for the analytical procedure since it allows the estimation of all kinds of contaminations. To assess the matrix effect for each sample, recovery rates were determined. The checking of the half range of the calibration graph for each 10 runs of samples allowed the evaluation of calibration graph deviation. [Pg.16]

In the analytical procedures of ferrous metallurgy calibration methods have quickly proved useful, interferences being compensated for from the outset although they are not known individually. [Pg.242]


See other pages where Analytical procedures calibration methods is mentioned: [Pg.62]    [Pg.56]    [Pg.398]    [Pg.157]    [Pg.38]    [Pg.518]    [Pg.350]    [Pg.417]    [Pg.570]    [Pg.589]    [Pg.38]    [Pg.16]    [Pg.137]    [Pg.251]    [Pg.73]    [Pg.131]    [Pg.387]    [Pg.17]    [Pg.750]    [Pg.57]    [Pg.196]    [Pg.272]    [Pg.380]    [Pg.543]    [Pg.543]    [Pg.432]    [Pg.31]    [Pg.545]    [Pg.234]    [Pg.128]    [Pg.196]   
See also in sourсe #XX -- [ Pg.8 , Pg.14 ]




SEARCH



Analytic Procedures

Analytical Calibration

Analytical methods procedures

Analytical methods, calibration

Analytical procedures

Calibration procedures method

Method calibration

Method procedure

Methodical procedures

© 2024 chempedia.info