Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytical methods weighing

In 1972 Pittam and Pilcher [24] remeasured the enthalpy of combustion of methane using the same method (flow calorimetry) but with a much purer sample and a better analytical method (weighing the CO2). Their six determinations are shown on the right side of Figure 1. This is abetter measurement, but the scatter in the data is similar to that in the study by Rossini methane is very difficult to bum completely in a calorimeter [33]. When the two data sets are considered together neither stands out, and it is no longer certain that the point rejected by Rossini can be dropped. In the new assessment, the recomended value is an average of all twelve points. [Pg.16]

The classical analytical method of deterruination of barium ion is gravimetric, by precipitating and weighing insoluble barium sulfate. Barium chromate, which is more insoluble than strontium chromate in a slightly acidic solution, gives a fairly good separation of the two elements. [Pg.484]

For a set of 10 samples, the analytical method can be completed within 16 laboratory hours from the time of sample weighing to GC injection. [Pg.486]

Organic solvent extraction. Two analytical methods for acetamiprid have been developed One method is for the parent only and the other determines the total residue of the parent and its metabolites (lM-1-2, lM-1-4 and lC-0). Air-dried soil (20-g equivalent dry soil) is weighed into a centrifuge tube and imidacloprid residue is extracted with 100 mL of methanol-0.1M ammonium chloride (4 1, v/v) using a mechanical shaker for about 30 min. After shaking, the tube is centrifuged at 8000 rpm for 2 min. The supernatant is filtered and the analysis of the soil residue is carried out in the same manner as described above for the parent compound. [Pg.1139]

The linearity of the detector can be obtained by diluting the analyte stock solution and measuring the associated responses, while the linearity of the analytical method can be determined by making a series of concentrations of the analyte from independent sample preparations (weighing and spiking) [15]. It is also essential that the basic calibration linear curve be obtained by using independent samples, and not by using samples that have been prepared by dilution and injected into HPLC/GC, or spotted on one TLC plate. [Pg.249]

Instructions on how to prepare samples for thermal analyses can be given only very generally since the variety and complexity of the problems involved makes it impossible to specify strict rules of procedure. In carrying out series of tests on any particular material, it is advisable to make this up in sufficient quantity and also to test it by other analytical methods such as X-ray and IR. All samples should be as far as possible homogeneous. Sample conditioning, i.e. adjustment to a predetermined degree of moisture content, may be carried out either before or after samples are weighed and loaded on the thermobalance. [Pg.87]

A sample of specified weight is normally required in the procedure. An interactive form of weighing is used, in which the display or printing unit of the entry station indicates whether or not the sample has been accepted. Before analysis it is necessary to specify the code number of the analytical method that is to be used, and to store this in the memory of the central control. To indicate where samples are located, it is necessary to identify them before weighing. Optical readers are therefore mounted on the sample-transport mechanism to register each sample. The sample is identified by a unique code placed on the outside of the sample cup. [Pg.43]

After the experiment, the experimental charge is prepared for analysis of the diffusion component or species. The analytical methods include microbeam methods such as electron microprobe, ion microprobe, Rutherford backscatter-ing, and infrared microscope to measure the concentration profile, as well as bulk methods (such as mass spectrometry, infrared spectrometry, or weighing) to determine the total gain or loss of the diffusion component or species. Often, the analysis of the diffusion profile is the most difficult step in obtaining diffusivity. [Pg.285]

The aerosol samples collected by the SFU were analyzed both gravimetrlcally for total suspended particulate mass less than 15pm, and by particle induced x-ray emission (PIXE) for elemental content. The filters were weighed before and after sampling using a Cahn 25 electrobalance sensitive to Ipg. Typical precision of TSP determined by this analytical method is 0.5pg/m for samples collected under conditions of low aerosol concentrations ( ). After weighing, the filters were analyzed for elemental content (elements heavier than Na) using the UC Davis PIXE system. This analysis technique is described in Cahill al ( ). [Pg.329]

This analytical method measures two elements, K and Th, which were also determined via NAA. However, the neutron activation procedure is limited to samples weighing a few hundred milligrams. Thus, another set of internal standards was included in the analysis to insure not only consistent and accurate results, but also homogeneous samples for these elements over a sample size range of 1000-fold. [Pg.137]

Method validation is important to ensure that the analytical method is in statistical control. A method may be validated by the so-called method evaluation function (MEF) (Christensen et al., 1993), which is obtained by linear regression analysis of the measured concentrations versus the true concentrations. A true concentration in a solution can be obtained by use of a high purity standard obtained from another manufacturer or batch than the one used for calibration. Both the high purity standard and the solvent are weighed using a traceable calibrated balance. If certified reference material is available this is preferred. The method evaluation includes the most important characteristics of the method as the following elements (see Figure 2.7) ... [Pg.37]

Gravimetric analysis is one of classical analytical methods. It is based on chemical transformation of the sample using excess of a reagent to a substance, which is weighed after processing. The weight of the substance obtained serves as a base for calculation of amount of substance. [Pg.94]

Gravimetric methods are based on the quantitative measurement of an analyte by weighing a pure and insoluble compound from the analyte. Analytical balances, which provide accurate and precise data, are used for this purpose. Detailed information on gravimetric methods and their applications can be found in the literature.3... [Pg.262]

Samples rarely come in a form that can be injected directly into the instrument some form of sample preparation usually is required. Sample preparation includes any manipulation of the sample prior to analysis, including techniques such as weighing, dilution, concentration, filtration, centrifugation, and liquid- or solid-phase extraction. Sample preparation can be performed either on-line or off-line, but it is usually performed offline. Off-line preparation can be time-consuming and tedious, and the more steps that are required, the more susceptible the analytical method is to operator error and irreproducibility. [Pg.109]

Consider an analytical method involving the titration of hydrochloric acid with anhydrous sodium carbonate to determine the concentration of the acid. The measurements made are mass (weighing out a chemical to make up a solution of known concentration) and volume (dispensing liquids with pipettes and burettes). The reaction between the two chemicals is based on amount of substance - one mole of sodium carbonate reacts with two moles of hydrochloric acid - and the mass of a mole is known (e.g. the formula weight in grams of one mole of sodium carbonate is 105.99). All the measurements are based on either length or mass and are traceable to SI units, so the method is a primary method. [Pg.28]

A measured value is complete only when it is accompanied by a statement of its uncertainty and is required in order to decide whether or not the result is adequate for its intended purpose. The uncertainty value must be suitably small to show that the reported results can be accepted with confidence and to ascertain whether or not it is consistent with similar results. There is an uncertainty in the concentration of the calibration samples used both in synthetic calibration samples and calibrations of standard addition. Weighing and volumes, which are a must in most analytical methods, must include weighing errors volumes must include volume errors to take into account uncertainties associated with these steps of the analysis. These and others must also be included in the overall calculation of the analytical error. [Pg.102]

Standard operating procedures (SOPs) are detailed descriptions of analytical methods (or of other laboratory tasks such as weighing or document archiving). [Pg.314]


See other pages where Analytical methods weighing is mentioned: [Pg.201]    [Pg.152]    [Pg.137]    [Pg.9]    [Pg.610]    [Pg.217]    [Pg.41]    [Pg.69]    [Pg.64]    [Pg.171]    [Pg.94]    [Pg.201]    [Pg.277]    [Pg.217]    [Pg.142]    [Pg.1472]    [Pg.238]    [Pg.306]    [Pg.23]    [Pg.282]    [Pg.239]    [Pg.4468]    [Pg.320]    [Pg.314]    [Pg.81]    [Pg.130]    [Pg.56]    [Pg.443]    [Pg.284]    [Pg.239]    [Pg.271]    [Pg.11]   
See also in sourсe #XX -- [ Pg.18 , Pg.22 , Pg.175 , Pg.201 , Pg.215 ]




SEARCH



Weighing

Weighing methods

© 2024 chempedia.info