Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analysis flow injector

A. F. Kapauan and M. C. Magno, Continuous-Flow Injector for Flow Injection Analysis. Anal. Ghent., 58 (1986) 509. [Pg.450]

The syringe is removed, the port is closed and the eluent stream is started again. An example of such a stopped flow injector, which is commercially available (Varian), is shown in Fig. 7. For precise retention time measurements, stopped flow injection is less preferable because of the time needed to pressurize the system, which might be of the order of 3-2D sec depending on the capacity of the pumping device. On the other hand, for routine analysis this is less important and stopped flow injection might be a very valuable, cheap and reliable sample introduction technique. [Pg.69]

Figure 4.11 provides a schematic illustration of a lab-on-a-chip device for peptide analysis. The actual enzymatic cleavage reaction takes place in the sample reservoir from where the fission products are funneled to the cross-flow injector using 50 mM Tris-HCl as separation path buffer medium. The separation principle consists in different migration rates of the peptide fragments in an electric field that are detected by fluorescence. The methodology has primarily been developed for bioanalysis it has been reasoned to use it for preparative biocatalytic chemistry in terms of a combinatorial chemistry approach, but with no application in sight so far. [Pg.100]

Example of a single-channel manifold for use in flow injection analysis where R1 is a reagent reservoir P is the pump S is the sample I is the injector B is a bypass loop ... [Pg.652]

Two examples of dual-channel manifolds for use In flow Injection analysis where R1 and R2 are reagent reservoirs P Is the pump S Is the sample I Is the Injector B Is a bypass loop W Is waste C Is the mixing and reaction coll and D Is the detector. [Pg.653]

Analysis and purification of the product solution is best accomplished by gas chromatography. The submitters used a 500 cm. by 0.6 cm. aluminum or polyethylene column packed with 21% oxydipropionitrile on Chromosorb P with column, injector and detector operated at 25° and a flow rate of 50 ml./minute. Under these conditions the retention times of bicyclopentene and cyclopentadiene were 3 and 5 minutes, respectively, beyond that of the coinjected air. Since bioyclo-pentene is extremely labile with respect to acid catalysis any contact with water, hydroxylic solvents, and nonprotic acids should be avoided (Note 11). Bicyclopentene stored at —78° in anhydrous tetrahydro-furan is stable indefinitely. [Pg.18]

GC-TEA Analysis. A Bendix model 2200 GC and Thermo Electron model 502 TEA were used. The GC injector temperature was 210 C. The TEA pyrolysis furnace was operated at 450 C and the cold trap was held at -150 C in isopentane slush. Oxygen flow to the ozonator was 20 cc/min and indicated pressure was 1.5 torr at a helium flow rate of 20 cc/min. TEA output was processed by a digital integrator (Spectra Physics System I). [Pg.336]

Principles and Characteristics As mentioned already (Section 3.5.2) solid-phase microextraction involves the use of a micro-fibre which is exposed to the analyte(s) for a prespecified time. GC-MS is an ideal detector after SPME extraction/injection for both qualitative and quantitative analysis. For SPME-GC analysis, the fibre is forced into the chromatography capillary injector, where the entire extraction is desorbed. A high linear flow-rate of the carrier gas along the fibre is essential to ensure complete desorption of the analytes. Because no solvent is injected, and the analytes are rapidly desorbed on to the column, minimum detection limits are improved and resolution is maintained. Online coupling of conventional fibre-based SPME coupled with GC is now becoming routine. Automated SPME takes the sample directly from bottle to gas chromatograph. Split/splitless, on-column and PTV injection are compatible with SPME. SPME can also be used very effectively for sample introduction to fast GC systems, provided that a dedicated injector is used for this purpose [69,70],... [Pg.437]

Flow injection analysis is another technique that has been applied to the determination of nitrate and nitrite in seawater. Anderson [ 126] used flow injection analysis to automate the determination of nitrate and nitrite in seawater. The detection limit of his method was 0.1 imol/l. However, the sampling rate was only 30 per hour which is low for flow injection analysis. Reactions seldom go to completion in a determination by flow injection analysis [127,128] because of the short residence time of the sample in the reaction manifold. Anderson selected a relatively long residence time so that the extent of formation of the azo dye was adequate to give a detection limit of 0.1 pmol/l. This reduced the sampling rate because only one sample is present at a time in the post-injector column in flow injection analysis. Any increase in reaction time causes a corresponding increase in the time needed to analyse one sample. [Pg.87]

FIA has also found wide application in pharmaceutical analysis.214,215 Direct UV detection of active ingredients is the most popular pharmaceutical analysis application of FIA. For single component analysis of samples with little matrix interference such as dissolution and content uniformity of conventional dosage forms, many pharmaceutical chemists simply replace a column with suitable tubing between the injector and the detector to run FIA on standard HPLC instrumentation. When direct UV detection offers inadequate selectivity, simple online reaction schemes with more specific reagents including chemical, photochemical, and enzymatic reactions of derivatization are applied for flow injection determination of pharmaceuticals.216... [Pg.269]

Chromatographic Conditions. GC/MS-MS analyses were performed on a Varian 3800 gas chromatograph (Varian Chromatography Systems, Walnut Creek, CA) equipped with a 1079 split/splitless injector and a ion trap spectrometer (Varian Saturn 2000, Varian Chromatography Systems) with a waveboard for MS-MS analysis. The system was operated by Saturn GC/MS Workstation v5.4 software. The MS-MS detection method was adapted from reference. PCBs were separated on a 25 m length x 0.32 mm i.d., CPSil-8 column coated with a 0.25-pm film. The GC oven temperature program was as follows 90 °C hold 2 min, ramp 30 °C/min to 170 °C, hold for 10 min, rate 3 °C/ min to 250 °C, rate 20 °C/min to a final temperature of 280 °C, and hold for 5 min. Helium was employed as the carrier gas, with a constant column flow of 1.0 mL/min. [Pg.94]

With commercially available automatic sampling devices, large numbers of samples can be routinely analyzed by LC without operator intervention. Such equipment is popular for the analysis of routine samples (e.g., quality control of drugs), particularly when coupled with automatic data-handling systems. Automatic injectors are indispensable in unattended searching (e.g., overnight) for chromatographic parameters such as solvent selectivity, flow rate, and temperature optimization. [Pg.8]


See other pages where Analysis flow injector is mentioned: [Pg.35]    [Pg.3336]    [Pg.2086]    [Pg.236]    [Pg.182]    [Pg.200]    [Pg.25]    [Pg.126]    [Pg.131]    [Pg.645]    [Pg.834]    [Pg.185]    [Pg.469]    [Pg.470]    [Pg.192]    [Pg.332]    [Pg.370]    [Pg.146]    [Pg.282]    [Pg.9]    [Pg.351]    [Pg.26]    [Pg.166]    [Pg.98]    [Pg.74]    [Pg.420]    [Pg.438]    [Pg.156]    [Pg.149]    [Pg.69]    [Pg.37]    [Pg.6]    [Pg.185]    [Pg.206]    [Pg.136]    [Pg.216]    [Pg.645]   
See also in sourсe #XX -- [ Pg.305 ]




SEARCH



Injector

© 2024 chempedia.info