Analysis external standard

Since a standard additions calibration curve is constructed in the sample, it cannot be extended to the analysis of another sample. Each sample, therefore, requires its own standard additions calibration curve. This is a serious drawback to the routine application of the method of standard additions, particularly in laboratories that must handle many samples or that require a quick turnaround time. For example, suppose you need to analyze ten samples using a three-point calibration curve. For a normal calibration curve using external standards, only 13 solutions need to be analyzed (3 standards and 10 samples). Using the method of standard additions, however, requires the analysis of 30 solutions, since each of the 10 samples must be analyzed three times (once before spiking and two times after adding successive spikes). [Pg.115]

Standardization—External standards, standard additions, and internal standards are a common feature of many quantitative analyses. Suggested experiments using these standardization methods are found in later chapters. A good project experiment for introducing external standardization, standard additions, and the importance of the sample s matrix is to explore the effect of pH on the quantitative analysis of an acid-base indicator. Using bromothymol blue as an example, external standards can be prepared in a pH 9 buffer and used to analyze samples buffered to different pHs in the range of 6-10. Results can be compared with those obtained using a standard addition. [Pg.130]

The generalized standard addition method (GSAM) extends the analysis of mixtures to situations in which matrix effects prevent the determination of 8x and 8y using external standards.When adding a known concentration of analyte to a solution containing an unknown concentration of analyte, the concentrations usually are not additive (see question 9 in Chapter 5). Conservation of mass, however, is always obeyed. Equation 10.11 can be written in terms of moles, n, by using the relationship... [Pg.402]

When possible, a quantitative analysis is best conducted using external standards. Unfortunately, matrix interferences are a frequent problem, particularly when using electrothermal atomization. Eor this reason the method of standard additions is often used. One limitation to this method of standardization, however, is the requirement that there be a linear relationship between absorbance and concentration. [Pg.420]

Gluodenis describes the use of ICP to analyze samples containing Pb and Ni in brass. The analysis for Pb uses external standards prepared from brass samples containing known amounts of lead. Results are shown in the following table. ... [Pg.456]

Quantitative Analysis Using External Standards To determine the concentration of analyte in a sample, it is necessary to standardize the electrode. If the electrode s response obeys the Nernst equation. [Pg.486]

Eatty acids from commercial fats and oils, such as peanut oil, are extracted with methanolic NaOH and made volatile by derivatizing with a solution of methanol/BE3. Separations are carried out using a capillary 5% phenylmethyl silicone column with MS detection. By searching the associated spectral library students are able to identify the fatty acids present in their sample. Quantitative analysis is by external standards. [Pg.611]

Janusa and co-workers report the determination of chloride by CZE. Analysis of a series of external standards gives the following calibration curve. [Pg.619]

The concentration of Mn in steel can be determined by a neutron activation analysis using the method of external standards. A 1.000-g sample of an unknown steel sample and a 0.950-g sample of a standard steel known to contain 0.463% w/w Mn, are irradiated with neutrons in a nuclear reactor for 10 h. After a 40-min cooling period, the activities for gamma-ray emission were found to be 2542 cpm (counts per minute) for the unknown and 1984 cpm for the standard. What is the %w/w Mn in the unknown steel sample ... [Pg.646]

Standards used to constmct a cahbration curve must be prepared such that the matrix of the standard is identical to the sample s matrix because the values of the parameters k and b associated with a linear cahbration curve are matrix dependent. Many areas of chemical analysis are plagued by matrix effects, and it is often difficult to duphcate the sample matrix when preparing external standards. Because it is desirable to eliminate matrix effects, cahbration in the sample matrix itself can be performed. This approach is called the standard addition method (SAM) (14). In this method, the standards are added to the sample matrix and the response of the analyte plus the standard is monitored as a function of the added amount of the standard. The initial response is assumed to be Rq, and the relationship between the response and the concentration of the analyte is... [Pg.427]

Purity. Gas chromatographic analysis is performed utilizing a wide-bore capillary column (DB-1, 60 m x 0.32 mm ID x 1.0 //m film) and a flame ionization detector in an instmment such as a Hewlett-Packard 5890 gas chromatograph. A caUbration standard is used to determine response factors for all significant impurities, and external standard calculation techniques are used to estimate the impurity concentrations. AHyl chloride purity is deterrnined by difference. [Pg.35]

At X-ray fluorescence analysis (XRF) of samples of the limited weight is perspective to prepare for specimens as polymeric films on a basis of methylcellulose [1]. By the example of definition of heavy metals in film specimens have studied dependence of intensity of X-ray radiation from their chemical compound, surface density (P ) and the size (D) particles of the powder introduced to polymer. Have theoretically established, that the basic source of an error of results XRF is dependence of intensity (F) analytical lines of determined elements from a specimen. Thus the best account of variations P provides a method of the internal standard at change P from 2 up to 6 mg/sm the coefficient of variation describing an error of definition Mo, Zn, Cu, Co, Fe and Mn in a method of the direct external standard, reaches 40 %, and at use of a method of the internal standard (an element of comparison Ga) value does not exceed 2,2 %. Experiment within the limits of a casual error (V changes from 2,9 up to 7,4 %) has confirmed theoretical conclusions. [Pg.104]

To analyze pesticides from the sample, several GC techniques were used GC with FID and EC detectors and GC/MS with external standards. Pesticides are mostly analyzed using split/splitless technique where higher amount of injected solution exits the gas chromatograph without decomposing therefore by quantification of the several pesticides in the filter, we found out how harmful is exposing analysts to pesticide compounds during the GC analysis. [Pg.192]

One of the important advantages of ICPMS in problem solving is the ability to obtain a semiquantitative analysis of most elements in the periodic table in a few minutes. In addition, sub-ppb detection limits may be achieved using only a small amount of sample. This is possible because the response curve of the mass spectrometer over the relatively small mass range required for elemental analysis may be determined easily under a given set of matrix and instrument conditions. This curve can be used in conjunction with an internal or external standard to quantily within the sample. A recent study has found accuracies of 5—20% for this type of analysis. The shape of the response curve is affected by several factors. These include matrix (particularly organic components), voltages within the ion optics, and the temperature of the interffice. [Pg.630]

Ethanol concentration in the fermentation broth is determined by using gas chromatography (HP 5890 series II with HP Chemstation data processing software, Hewlett-Packard, Avondale, PA) with a Poropak Q Column, and a Hewlett-Packard model 3380A integrator. A flame ionisation detector (FID) is used to determine ethanol. The oven temperature is maintained at 180 °C, and the injector and detector temperature are maintained at 240 °C. The sample taken from the fermentation media has to be filtered and any internal standard must be added for analysis based on internal standard methods otherwise, the area under the peak must be compared with known standard samples for calculation based on external standard methods. [Pg.257]

Table 5.17 Quantitative results obtained for the determination of four diarrhetic shellfish poisons (DSPs) using external standards and the method of standard additions. Reprinted from J. Chromatogr., A, 943, Matrix effect and correction by standard addition in quantitative liquid chromatographic-mass spectrometric analysis of diarrhetic shellfish poisoning toxins , Ito, S. and Tsukada, K., 39-46, Copyright (2002), with permission from Elsevier Science... Table 5.17 Quantitative results obtained for the determination of four diarrhetic shellfish poisons (DSPs) using external standards and the method of standard additions. Reprinted from J. Chromatogr., A, 943, Matrix effect and correction by standard addition in quantitative liquid chromatographic-mass spectrometric analysis of diarrhetic shellfish poisoning toxins , Ito, S. and Tsukada, K., 39-46, Copyright (2002), with permission from Elsevier Science...
Internal rather than external standards are recommended for quantitative analysis. If standard additions are used for calibration all the parameters listed in this table must be held constant. [Pg.128]

Sample preparation, injection, calibration, and data collection, must be automated for process analysis. Methods used for flow injection analysis (FLA) are also useful for reliable sampling for process LC systems.1 Dynamic dilution is a technique that is used extensively in FIA.13 In this technique, sample from a loop or slot of a valve is diluted as it is transferred to a HPLC injection valve for analysis. As the diluted sample plug passes through the HPLC valve it is switched and the sample is injected onto the HPLC column for separation. The sample transfer time typically is determined with a refractive index detector and valve switching, which can be controlled by an integrator or computer. The transfer time is very reproducible. Calibration is typically done by external standardization using normalization by response factor. Internal standardization has also been used. To detect upsets or for process optimization, absolute numbers are not always needed. An alternative to... [Pg.76]

The determination of the relationship between detector response and the sample concentration is termed the calibration of the method. There are two types of methods in use for the quantitative analysis of a sample, i.e., the external standard and the internal standard method. An external standard method is a direct comparison of the detector response of a pure compound (standard) to a sample.2 The calibration of the method is performed by preparing standards of varying concentration and analyzing them by a developed method. Method 1 (below) was developed for toluene, and standards of varying concentration were prepared and analyzed. The results obtained are summarized in Table 2 see Figure 3. [Pg.156]

Quantitative analysis using FAB is not straightforward, as with all ionisation techniques that use a direct insertion probe. While the goal of the exercise is to determine the bulk concentration of the analyte in the FAB matrix, FAB is instead measuring the concentration of the analyte in the surface of the matrix. The analyte surface concentration is not only a function of bulk analyte concentration, but is also affected by such factors as temperature, pressure, ionic strength, pH, FAB matrix, and sample matrix. With FAB and FTB/LSIMS the sample signal often dies away when the matrix, rather than the sample, is consumed therefore, one cannot be sure that the ion signal obtained represents the entire sample. External standard FAB quantitation methods are of questionable accuracy, and even simple internal standard methods can be trusted only where the analyte is found in a well-controlled sample matrix or is separated from its sample matrix prior to FAB analysis. Therefore, labelled internal standards and isotope dilution methods have become the norm for FAB quantitation. [Pg.369]


See other pages where Analysis external standard is mentioned: [Pg.84]    [Pg.84]    [Pg.117]    [Pg.123]    [Pg.130]    [Pg.451]    [Pg.451]    [Pg.489]    [Pg.586]    [Pg.811]    [Pg.811]    [Pg.813]    [Pg.445]    [Pg.780]    [Pg.833]    [Pg.49]    [Pg.51]    [Pg.131]    [Pg.570]    [Pg.359]    [Pg.111]    [Pg.198]    [Pg.202]    [Pg.238]    [Pg.243]    [Pg.371]    [Pg.515]    [Pg.609]    [Pg.618]   
See also in sourсe #XX -- [ Pg.97 ]