Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Thermal conduction calorimeters

Passive diathermal calorimeters are those in which good heat-exchange between the system S and the surrounding thermostat T is achieved by good thermal conduction. The sample temperature passively follows, here, the thermostat temperature and, except in transient situations, there is no heat stored in the system S, These calorimeters can also be called, quite correctly, thermal conduction calorimeters . [Pg.34]

When heat is liberated or absorbed in the calorimeter vessel, a thermal flux is established in the heat conductor and heat flows until the thermal equilibrium of the calorimetric system is restored. The heat capacity of the surrounding medium (heat sink) is supposed to be infinitely large and its temperature is not modified by the amount of heat flowing in or out. The quantity of heat flowing along the heat conductor is evaluated, as a function of time, from the intensity of a physical modification produced in the conductor by the heat flux. Usually, the temperature difference 0 between the ends of the conductor is measured. Since heat is transferred by conduction along the heat conductor, calorimeters of this type are often also named conduction calorimeters (20a). [Pg.195]

An extremely simplified scheme of a calorimeter (composite thermal detector) is shown in Fig. 15.6. The temperature of an absorber A (TA) is measured by a thermometer T. A thermal conductance G forms a thermal link with the heat sink B at the temperature Ts. In the ideal adiabatic situation (G = 0), an absorption of an energy AE produces an absorber temperature increase ... [Pg.331]

As in the case of calorimeters, a bolometer consists of an absorbing element with heat capacity C, which converts the impinging electromagnetic radiation to heat, and which is linked to a heat sink at temperature Ts via a thermal conductance G. The temperature TA of the absorber is measured by a thermometer in thermal contact with the absorber. [Pg.336]

Another problem related to the validity of equation 9.9 is that equation 9.6 applies only to heat conduction. If T — 12 is large, some significant fraction of heat will be transferred by convection and radiation and thus will not be monitored by the thermopile. Consequently, the use of partial compensating Peltier or Joule effects was essential in the experiments involving Calvet s calorimeter, whose thermopiles had a fairly low thermal conductivity. [Pg.140]

The problems associated with direct reaction calorimetry are mainly associated with (1) the temperature at which reaction can occur (2) reaction of the sample with its surroundings and (3) the rate of reaction which usually takes place in an uncontrolled matmer. For low melting elements such as Zn, Pb, etc., reaction may take place quite readily below S00°C. Therefore, the materials used to construct the calorimeter are not subjected to particularly high temperatures and it is easy to select a suitably non-reactive metal to encase the sample. However, for materials such as carbides, borides and many intermetallic compounds these temperatures are insufficient to instigate reaction between the components of the compound and the materials of construction must be able to withstand high temperatures. It seems simple to construct the calorimeter from some refractory material. However, problems may arise if its thermal conductivity is very low. It is then difficult to control the heat flow within the calorimeter if some form of adiabatic or isothermal condition needs to be maintained, which is further exacerbated if the reaction rates are fast. [Pg.82]

Adiabatic calorimeters have also been used for direct-reaction calorimetry. Kubaschewski and Walter (1939) designed a calorimeter to study intermetallic compoimds up to 700°C. The procedure involved dropping compressed powders of two metals into the calorimeter and maintaining an equal temperature between the main calorimetric block and a surrounding jacket of refractory alloy. Any rise in temperature due to the reaction of the metal powders in the calorimeter was compensated by electrically heating the surrounding jacket so that its temperature remained the same as the calorimeter. The heat of reaction was then directly a function of the electrical energy needed to maintain the jacket at the same temperature as the calorimeter. One of the main problems with this calorimeter was the low thermal conductivity of the refractory alloy which meant that it was very difficult to maintain true adiabatic conditions. [Pg.83]

Thermal conductivity data are even more difficult to obtain. In the case of calorimetric data of heat capacity and heats of dissociation, the measurements though still reasonably challenging are aided by significant improvements in commercial calorimeters that can operate at high pressures. Thermal property data are presented in Section 6.3.2. [Pg.320]

Figure 4. Potential-time curves from experiments with a thermopile heat conduction calorimeter. A A short heat pulse released at time t,. B A constant thermal power released between time t, and t2. The steady-state potential value, USI is proportional to the released thermal power. Figure 4. Potential-time curves from experiments with a thermopile heat conduction calorimeter. A A short heat pulse released at time t,. B A constant thermal power released between time t, and t2. The steady-state potential value, USI is proportional to the released thermal power.
The thermal inertia of a heat conduction calorimeter is described by its time constant x (equation (18)). In practice, X is given by... [Pg.281]

The external heat flux (q"x) from the cone heater does not exclusively determine the heat flux important for samples pyrolysis in the cone calorimeter, since the reradiation from the hot sample surface (q"eTad), the loss by thermal conductivity into the specimen and the surroundings ( I SS), and the heat flux from the flame (q Lmt) are also of the same order of magnitude.82 85 Thus, the heat flux effective with respect to pyrolysis during a cone calorimeter run (qeii) is the result of the external heat flux and the material s response (qeB = q L + < L - gCad - qLs). [Pg.397]

The calorimeter method is an older technique which is a direct measurement of Fourier s law. It is one of the ASTM [2] standard tests for thermal conductivity, designation C201. The experimental configuration is shown in Figure 9.3. A SiC slab... [Pg.231]

Figure 9.3 Schematic of the calorimeter method of measuring thermal conductivity [2]. Specimen sizes are approximately three bricks of dimensions 23 x 11.4 x 6.4 cm3. Figure 9.3 Schematic of the calorimeter method of measuring thermal conductivity [2]. Specimen sizes are approximately three bricks of dimensions 23 x 11.4 x 6.4 cm3.
The back-up insulation between the base of the test specimen and the calorimeter is optional. Under conditions of steady state heat flow, introduction of back-up insulation diminishes the heat flow to the calorimeter as well as the temperature drop across the specimen the thermal conductivity, as... [Pg.232]

Figure 9.4 Water cooling system specified in ASTM thermal conductivity standard C201. The center-most series of cooling coils makes up the calorimeter . Outside of the calorimeter are the inside guard cooling coils, which in turn are surrounded by the outside guard coils [2],... Figure 9.4 Water cooling system specified in ASTM thermal conductivity standard C201. The center-most series of cooling coils makes up the calorimeter . Outside of the calorimeter are the inside guard cooling coils, which in turn are surrounded by the outside guard coils [2],...
Flow adsorption microcalorimetry has been used to measure the heats of adsorption of ammonia in a nitrogen carrier on the H and Na forms of a Y zeolite [21]. The calorimeter was linked to a thermal conductivity detector in which the rates of adsorption and desorption and the associated rates of heat evolution or absorption were measured simultaneously at atmospheric pressure. The authors found that, as surface coverage increased, the sites covered first were not necessarily those with the highest molar heats of adsorption. [Pg.426]

The Seebeck coefficient were calculated from measurement of electromotive force with temperature difference of lOK. The electrical resistivity and Hall measurement were performed by van der Pauw method. The thermal conductivity were calculated from the thermal diffusivity, the specific heat and the density. The thermal diffusivity and the specific heat were measured by laser flash method and differential scanning calorimeter (DSC), respectively. [Pg.534]

Electrical resistivity measurement adopted conventional four probes method. Seebeck coefficient was measured by the standard DC method. Thermal conductivity k was calculated from density, specific heat, and thermal diffiisivity. Specific heat measurement was carried out by use of a differential scanning calorimeter (DSC model 8230, Rigaku, Japan) compared with a standard material of a -AI2O3. The values of thermal diffiisivity obtained from a differential phase analysis of photo-pyroelectric signal (AL- A 0 analysis) [9]. All measiu ements were done at room temperature. [Pg.613]

The copper calorimeter works on the principle of the mixture calorimeter in place of the calorimetric liquid there is a copper block with good thermal insulation, provided with an axial cavity for receiving the warmer or cooler substance under investigation. The change in the temperature of the block is measured with the aid of thermocouples the good thermal conductivity of copper ensures that all parts of it are always at practically the same temperature. With a view to the best thermal in-FlG 3- sulation, it is contained in a double-walled vessel which is evacuated and silvered. [Pg.26]

As calorimeter vessels the three forms illustrated below were employed. Metals, on account of their high thermal conductivity, could be used without any enclosure simply in the form of a block this was of cylindrical shape, and was provided with a hole in which was inserted a core made of the same metal and wound with platinum wire. Thin waxed paper was used for insulation the very small space between the core and the block was filled up with melted paraffin wax. The upper part of the core was... [Pg.30]


See other pages where Thermal conduction calorimeters is mentioned: [Pg.216]    [Pg.229]    [Pg.311]    [Pg.79]    [Pg.84]    [Pg.137]    [Pg.140]    [Pg.189]    [Pg.170]    [Pg.548]    [Pg.2379]    [Pg.282]    [Pg.285]    [Pg.397]    [Pg.574]    [Pg.234]    [Pg.239]    [Pg.240]    [Pg.149]    [Pg.311]    [Pg.221]    [Pg.6]    [Pg.10]    [Pg.2293]    [Pg.19]    [Pg.266]    [Pg.311]   
See also in sourсe #XX -- [ Pg.34 ]




SEARCH



Calorimeters

Calorimeters conduction

© 2024 chempedia.info