Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

The test mixture

As the column must never be overloaded, not more than 1 J,g of each component per [Pg.146]

The test mixture must contain a compound by means of which the breakthrough [Pg.146]

Mobile phase hexane-tcrt-butylmethyl ether, appropriate mixture as to yield suitable k values. [Pg.147]

Methanol-water or acetonitrile-water with a different (weaker) composition than the mobile phase (nonretained, solvent for mixture) [Pg.148]

A specific test mixture may be prepared for working in a special area, e.g. an aflatoxin mixture for aflatoxin determination. (However, a test mixture for the determination of the plate number should not include any peptides or proteins. These compounds need to be eluted by gradient, see Fig. 18.7, whereas the plate [Pg.135]

As the column must never be overloaded, not more than 1 pg of each component per gram of stationary phase should be injected the amount permitted in analytical columns is so small as to make refractive index detection critical or even impossible. [Pg.136]

The test mixture must contain a compound by means of which the breakthrough time can be measured accurately it must be neither retained nor excluded. These conditions are satisfied when the product has a low molar mass and is as closely related to the eluent as possible, e.g. pentane for elution with hexane. Pentane gives no UV absorption but just a small refractive index peak. The plate number cannot be determined from this signal, so another test compound such as toluene or xylene which is rapidly eluted (k 0.2) and which produces a true UV absorption peak is recommended. The theoretical plate number of a little or non-retained substance reflects directly on the packing quality and the extra-column volumes, whereas mass-transfer properties are equally important in peaks that are eluted later. In reversed-phase systems, to can be determined with uracil. The test mixture should also contain compounds for which w 1 and 3-5. [Pg.136]


MESG is defined in terms of die precise test mediod and apparatus used, of which there are three variants British, lEC, and Underwriters Laboratories, Inc. Each apparatus consists of two combusdon chambers connected by a slot of specified size and variable widdi. The separate chambers are filled with the test mixture. The MESG is die maximum slot widdi that prevents flame propagadoii between die chambers for all composi-doiis of die test gas in air under the specified test coiididoiis. Phillips (1987) describes and compares diese three types of experimental apparatus for determining the MESG. [Pg.100]

S mg), dimer (peak I) and monomer (peak 2), ovalbumin (S mg) (peak 3), and cytochrome c (3 mg) (peak 4) was loaded onto a Fractogel EMD BioSEC column (600 X 16 mm) with a bed height of 600 mm. PBS (pH 7.2) was used as the eluent at a flow rate of I ml/min the sample volume was O.S ml. (B) The same protein sample as in A was injected onto a column of identical dimensions packed with unmodified Fractogel HW 6S. Without the tentacle modification the base matrix displays only a poor resolution of the test mixture. [Pg.223]

The preparations of luciferin (Ln, an electron acceptor) and soluble enzyme used were crude or only partially purified. The luciferase was an insoluble particulate material, possibly composed of many substances having various functions. Moreover, the luciferin-luciferase reaction was negative when both luciferin and luciferase were prepared from certain species of luminous fungus. It appears that the light production reported was the result of a complex mechanism involving unknown substances in the test mixture, and probably the crucial step of the light-emitting reaction is not represented by the above schemes. [Pg.270]

Their reaction was tested on the individual components of the test mixture indole, ergotamine tartrate, ergotaminine and ergobasine [ergometrine) [3]. The results obtained were as follows ... [Pg.228]

Figure 2.7 Activity test of an uncoated fused silica capillary after deactivation with poly(phenyliaethylhydrosiloxane), (A), and before deactivation, (B). Precolunn 15 x 0.20 m I.D. coated with SE-54. Test columns 10 a x 0.20 I.D. The column tandem was programmed from 40 to I80 c at a C/min after a 1 min isothermal hold with a hydrogen carrier gas velocity of 50 cm/s. The test mixture contained 10 n-decane, Cg-NH = l-aminooctane, PY 3,5-dimethylpyrimidine, C 2 n-dodecane, - 1-amlnodecane, DMA ... Figure 2.7 Activity test of an uncoated fused silica capillary after deactivation with poly(phenyliaethylhydrosiloxane), (A), and before deactivation, (B). Precolunn 15 x 0.20 m I.D. coated with SE-54. Test columns 10 a x 0.20 I.D. The column tandem was programmed from 40 to I80 c at a C/min after a 1 min isothermal hold with a hydrogen carrier gas velocity of 50 cm/s. The test mixture contained 10 n-decane, Cg-NH = l-aminooctane, PY 3,5-dimethylpyrimidine, C 2 n-dodecane, - 1-amlnodecane, DMA ...
Inject the test mixture under conditions that allow ca. 2 ng of a single test substance to enter the column (e.g., 1 microliter with a split ratio of 1 20 to 1 50, depending on injector design). [Pg.87]

A value for the column dead volume is required in most calculations. It is convenient to have one cosponent of the test mixture as an unretained solute. [Pg.184]

At least two components of the test mixture should have k values between 2 and 10. [Pg.184]

Having chosen the test mixture and mobile diase composition, the chromatogram is run, usually at a fairly fast chart speed to reduce errors associated with the measurement of peak widths, etc.. Figure 4.10. The parameters calculated from the chromatogram are the retention volume and capacity factor of each component, the plate count for the unretained peak and at least one of the retained peaks, the peak asymmetry factor for each component, and the separation factor for at least one pair of solutes. The pressure drop for the column at the optimum test flow rate should also be noted. This data is then used to determine two types of performance criteria. These are kinetic parameters, which indicate how well the column is physically packed, and thermodynamic parameters, which indicate whether the column packing material meets the manufacturer s specifications. Examples of such thermodynamic parameters are whether the percentage oi bonded... [Pg.184]

Figure 4.10 Typical routine column test chromatogram for a 30 cm X 4.6 mm I. D. column pacXed with an octadecylsiloxane bonded silica packing of lO micrometers particle diameter. The test mixture consisted of resorcinol (0.55 mg/ml), acetophenone (0.025 mg/ml), naphthalene (0.20 mg/ml) and anthracene (0.01 mg/ml) in acetonitrile, 10 microliters injected. The separation was performed isocratically at 23 C with acetonitrile-water (55 45) as the mobile phase at a flow rate of 1.5 ml/min. Detection was by UV at 254 nm (0.1 AUFS). Figure 4.10 Typical routine column test chromatogram for a 30 cm X 4.6 mm I. D. column pacXed with an octadecylsiloxane bonded silica packing of lO micrometers particle diameter. The test mixture consisted of resorcinol (0.55 mg/ml), acetophenone (0.025 mg/ml), naphthalene (0.20 mg/ml) and anthracene (0.01 mg/ml) in acetonitrile, 10 microliters injected. The separation was performed isocratically at 23 C with acetonitrile-water (55 45) as the mobile phase at a flow rate of 1.5 ml/min. Detection was by UV at 254 nm (0.1 AUFS).
Because of the invalidity of the classical procedure, several workers have attempted to devise a method that is free from interference by chloride. Chloride interference can be eliminated by preventing the concurrent oxidation of organic material and chloride. This can be effected in two ways - either by leaving the chloride in the test mixture but preventing its oxidation, or by removing the chloride prior to the chemical oxygen demand test. [Pg.507]

The nonlinear dependence of the reaction rate on the partial pressure of CO suggests that there are stimulated either a single reaction step with the established nonlinearity or at least two steps, one of which has a linear and the other one a nonlinear (nperiodic operation of the reactor at the 180°C level and at a middle oxidation state (pretreatment at p /p = 5,0).has been accomplished (see Figure 9) with a H fl/N testing mixture followed by a C0/N recuperation mixture. After an intermediate activity of the catalyst has been attained by this periodic operation, CO has been added in the testing mixture. [Pg.291]

Fig. 3.130. HPLC chromatograms of the test mixture detected by DAD (270 nm, upper lane) by APCI-MS-TIC (middle) and by ESI-MS-TIC (lower lane). Peak identification l=benzene sulphonic acid sodium salt 2=2-naphtalene sulphonic acid sodium salt 3=2-anthraquinone sulphonic acid sodium salt 4 = sulphorhodamine D sodium salt 5=crocein orange G 6=eriochrome black T 7=2,6-anthraquinone disulphonic acid disodium salt 8 = 1,5-naphtalene disulphonic acid disodium salt 9 = azophloxine 10 = 1,2-benzene disulphonic acid dipotassium salt. Reprinted with permission from G. Socher et al. [178]. Fig. 3.130. HPLC chromatograms of the test mixture detected by DAD (270 nm, upper lane) by APCI-MS-TIC (middle) and by ESI-MS-TIC (lower lane). Peak identification l=benzene sulphonic acid sodium salt 2=2-naphtalene sulphonic acid sodium salt 3=2-anthraquinone sulphonic acid sodium salt 4 = sulphorhodamine D sodium salt 5=crocein orange G 6=eriochrome black T 7=2,6-anthraquinone disulphonic acid disodium salt 8 = 1,5-naphtalene disulphonic acid disodium salt 9 = azophloxine 10 = 1,2-benzene disulphonic acid dipotassium salt. Reprinted with permission from G. Socher et al. [178].
FIGURE 8 Analysis of the test mixture with (A) the SDS MEKC system containing 10 mM phosphate buffer (pH 7.5), 60 mM SDS, and 10% acetonitrile and (B) the CTAB MEKC system containing 25 mM phosphate buffer (pH 7.5), 10 mM CTAB, and 10% acetonitrile. Peaks I formamide 2 pyridine 3 aniline 4 meta-cresol 5 phenyl acetate 6 nitrobenzene 7 benzoic acid 8 thiamine 9 ethyl benzoate. (Reprinted from reference 270, with permission.)... [Pg.291]

Most authors use a mixture of the same composition as the test mixture for the incubation control, except for omission of UDP-sugar. As outlined above, complete suppression of endogenous synthesis is preferred by the present authors for determination of enzyme activities. [Pg.258]

TTAB) as surfactant and diethylether, n-heptane, cyclohexane, chloroform, or octanol as oil phase. Table 5 summarizes the values of N for the test mixtures in microemulsions containing different organic solvents with cationic surfactants. [Pg.151]

Table 5 Number of Theoretical Plates (N) for the Test Mixture Component in Microemulsions Containing Different Organic Solvents with TTAB... Table 5 Number of Theoretical Plates (N) for the Test Mixture Component in Microemulsions Containing Different Organic Solvents with TTAB...
The etching rate of PECVD silicon nitride is comparable to PECVD TEOS oxide. Ti, TiN, and W present an acceptable etching rate whatever the pH. In the presence of copper, only HF-based chemistries can be used. For Al/Cu none of the tested mixtures are suitable. [Pg.190]

The latest column test procedure has been introduced by NIST [61]. The test mixture contains toluene and ethyl benzene to describe hydrophobicity, amitriptyline as the basic analyte and... [Pg.71]

LC columns are fairly durable unless they are used in some manner that is intrinsically destructive, as, for example, with highly acidic or basic eluents, or with continual injections of inadequately purified biological samples. It is wise to inject some test mixture into a column when new, and to retain the chromatogram. If questionable results are obtained later, the test mixture can be injected again under the specified conditions. The two chromatograms may be compared to establish whether the column is still useful. [Pg.665]

The use of the laser-light-scattering detector permits the record of excellent chromatograms in gradient elution. An analysis of the test mixture of standards is shown in Fig. 37B for comparison with the same analysis obtained in isocratic conditions. The analysis lasts about 30 min before the elution of BBB instead of 18 min in isocratic conditions, but the first part of the chromatogram is considerably improved, and a number of impurities in the standards can be resolved. [Pg.228]

Figure 6.1 UV chromatograms of the test mixture of four / -hydroxybenzoic acid esters (1, methyl 2, ethyl 3, propyl 4, butyl) after the column and after the NMR flow cell at flow rates of (a) 1.0 and (b) 0.1 ml/min conditions column, LiChrospher RP select B, 125 x 4 mm id, 5 Jim, spectrometer, Bruker DRX 600 probe head, 4 mm z-gradient LC probe, active volume 120 a1 eluents, acetonitrile (A) and D2O (B) gradient, t = Omin A/B (40/60), t = 8 min A/B (70/30) at a flow rate of 1.0 ml/min and t = 80 min A/B (70/30) at a flow rate of 0.1 ml/min... Figure 6.1 UV chromatograms of the test mixture of four / -hydroxybenzoic acid esters (1, methyl 2, ethyl 3, propyl 4, butyl) after the column and after the NMR flow cell at flow rates of (a) 1.0 and (b) 0.1 ml/min conditions column, LiChrospher RP select B, 125 x 4 mm id, 5 Jim, spectrometer, Bruker DRX 600 probe head, 4 mm z-gradient LC probe, active volume 120 a1 eluents, acetonitrile (A) and D2O (B) gradient, t = Omin A/B (40/60), t = 8 min A/B (70/30) at a flow rate of 1.0 ml/min and t = 80 min A/B (70/30) at a flow rate of 0.1 ml/min...
Fulminates (silver fulminate, AgCNO, can form in the reaction mixture from the Tolens test for aldehydes if it is allowed to stand for some time. This can be prevented by adding dilute nitric (V) acid to the test mixture as soon as the test has been completed.)... [Pg.420]

During processing of the test run data (injection of OPCW test mixture) AMDIS is searching a small, dedicated Chemical Standards Library (ASCII file onsite.csl). This library contains only data and spectra of the 16 compounds contained in the OPCW test mixture (see Annex 2, Table 1). A compound in the test mixture is defined as detected if the net match factor of the compound spectrum compared with library spectrum is >85. This threshold of identification is fixed in the on-site version of AMDIS. [Pg.54]

If the toxic effect of a chemical combination is tested and compared with the effect of the individual chemicals, it may happen that the effect of the tested mixture deviates from the effect predicted by CA or IA. This mixture can be considered as 1 combination of the endless number of other possible combinations in which these chemicals can be mixed. If more combinations of this specific set of chemicals are tested, it can happen that effects of a number of different combinations at low concentrations differ from CA or IA, but that the effects of high-concentration combinations are well predicted. Such a systematic deviation pattern may be relevant for risk assessment, or may provide insight into the modes of action. Three types of systematic deviations from CA or IA can be defined as biologically relevant, based on studies published in the literature ... [Pg.134]

Alternatively, response additivity (RA) for independently acting chemicals as a mathematical null model for testing observed responses (associated with the pharmacological concept of independent joint action) and with an assumed correlation of sensitivities of 0 also often fits the data well. Again, misfits occur (e.g., when the test mixture consists of compounds with the same MOA at concentrations below the individual compound s no-effect concentrations), and when they occur, they are often in the tails of the response curves. [Pg.178]

If the burning time of the test mixture is shorter than of the standard mixture, the sample is deemed hazardous. [Pg.136]

Once in the column, compounds in the test mixture are separated by virtue of differences in their capacity factors, which in turn depend on their vapor pressure and degree of interaction with the stationary phase. The capacity factor, which governs resolution and retention times of components of the test mixture, is also temperature dependent. The use of temperature-programmable column ovens takes advantage of this dependence to achieve efficient separation of compounds differing widely in vapor pressure. [Pg.837]

Figure 14 Separation of the test mixture using a step voltage gradient and a short packed column. Capillary 75 mm i.d., 8 cm packed with Dionex AS9-HC (8.5 cm to detector, 34.5 cm total). Mobile phase 2.5 mM hydrochloric acid (titrated to pH 8.05 with Tris). Flow is a combination of 10-bar pressure and EOF with -30 kV added at 1.3 min. All other conditions as given in Figure 1. (Reprinted from Ref. 75, with permission.)... Figure 14 Separation of the test mixture using a step voltage gradient and a short packed column. Capillary 75 mm i.d., 8 cm packed with Dionex AS9-HC (8.5 cm to detector, 34.5 cm total). Mobile phase 2.5 mM hydrochloric acid (titrated to pH 8.05 with Tris). Flow is a combination of 10-bar pressure and EOF with -30 kV added at 1.3 min. All other conditions as given in Figure 1. (Reprinted from Ref. 75, with permission.)...

See other pages where The test mixture is mentioned: [Pg.68]    [Pg.370]    [Pg.23]    [Pg.100]    [Pg.172]    [Pg.551]    [Pg.204]    [Pg.830]    [Pg.190]    [Pg.253]    [Pg.253]    [Pg.296]    [Pg.52]    [Pg.38]    [Pg.166]    [Pg.93]    [Pg.264]    [Pg.193]    [Pg.836]    [Pg.510]    [Pg.206]   


SEARCH



Special tests for mixtures of anions on the semimicro scale

Test mixture

© 2024 chempedia.info