Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

The HPLC instrument

An HPLC instrument can be a set of individual modules or elements, but it can be designed as a single apparatus as well. The module concept is more flexible in the case of the failure of a single component moreover, the individual parts need not be from the same manufacturer. If you do not like to do minor repairs by yourself you will prefer a compact instrument. This, however, does not need less bench space than a modular set. [Pg.9]

Although the column is the most important part, it is usually the smallest one. For [Pg.9]

An HPLC instrument has at least the elements which are shown in Fig. 1.3 solvent reservoir, transfer line with frit, high-pressure pump, sample injection device, column, detector, and data recorder, usually together with data evaluation. Although the column is the most important part, it is usually the smallest one. For temperature-controlled separations it is enclosed in a thermostat. It is quite common to work with more than one solvent, thus a mixer and controller are needed. If the data acquisition is done by a computer it can also be used for the control of the whole system. [Pg.9]


The HPLC instrument is more expensive and thus costs more to run because it is used 12 to 16 hours per day, versus only 2 to 5 hours/day for the photometer, the hourly rates are not proportional to the initial investment and servicing needs. The cost structure given in Table 4.6 can now be derived. [Pg.188]

Most of the hplc instrumentation now in use is unsuitable for small bore columns. At the moment, the technique is used mainly in the applications laboratories of some instrument manufacturers (they are interested in selling it ). The method is potentially attractive in areas where sample sizes are very limited, for example in biochemical or life sciences applications, but whether or not it becomes widely accepted remains to be seen. [Pg.45]

The mobile phase in HPLC is called the eluent and is a liquid or a mixture of liquids. Common eluents are water, aqueous solutions, acetonitrile, and methanol. Almost any other common solvent compatible with the column packing and the detector may be used. In some cases, the HPLC instrument will be capable of making a mixture of eluents or changing the mixture of eluents during chromatography. If this is done, care must be taken to make sure that the eluent mixture is compatible with the detector. [Pg.281]

Examine the HPLC instrument to which you are assigned. Find the inlet line to the pump and place the free end of this line in the reservoir containing the mobile phase with the 90/10 composition. Trace the path of the mobile phase from the reservoir, through the pump, injection valve, column, and detector, to the waste container so that you identify and recognize all components of the flow path. Turn on the pump and detector and begin pumping the mobile phase at a rate... [Pg.386]

The promise of monolith is the achievement of a higher performance at a lower backpressure than a packed bed. While this is true in principle, current implementations are limited by the fact that the external wall to the structure is made from PEEK. At the time of this writing, the commercially available monoliths can only be used up to a pressure of 20MPa (200 atm, 3000 psi), while packed bed steel columns can be used up to double this pressure and higher. Also, the preparation of the monolith appears to be cumbersome. At the current time, the silica-based monoliths are available only with an internal diameter of 4.6mm. The speed is thus also limited by the flow rate achievable by the HPLC instrument. At the same time, the detector of choice today is the mass spectrometer, which can tolerate only much... [Pg.96]

Apparatus. The HPLC instrument used was a Water s Associates model 6000A pump for the solvent supply, a U6K septumless injector and a radial compression module with standard Radial Pak columns. Immediately after the column a low dead volume tee was inserted and another 6000A pump was used to deliver a solution of OPT for the post-column derivatization of histamine. Twenty feet of 9 thousandths (id) coiled stainless steel tubing was used as a mixing chamber and held at 60 C in a water bath. The reaction mixture then passed through a Water s 420 fluorescence detector which was connected to a recorder. The detector was equipped with a 340-nm excitation filter and a 440-nm emmission filter. [Pg.303]

The first ESI design at the end of the 1980s proved to work properly as the HPLC interface with mobile phase flow rates between 1 and lOpL/min. Meanwhile, the development of the HPLC instrumentation and columns was oriented in the mL/min flow rate mode. In addition, the nebulization process based only on the application of an electrical field does not produce a stable spray from aqueous mobile phases. A modified ESI source, called ionspray, was then introduced [39], in which the nebulization of a liquid solution is pneumatically assisted by a coaxial flow of nitrogen (sheath gas) that allows the formation of a stable aerosol at mobile-phase flow rates between 10 and 500 pL/ min and the use of aqueous mobile phases. When working at higher flow rates (500-1000 pL/min), an additional nittogen flow rate can be used (auxiliary gas) to assist the desolvation of the droplets. This modified source is called turboionspray. [Pg.239]

Inject the standard into the sample loop of the HPLC instrument and observe the output of the atomic fluorescence instrument on the chromatographic integrator. On elution of the first arsenic species, switch the mobile phase from 0.0001 M K2SO4 to 0.1 M K2SO4 (step gradient elution). [Pg.175]

It is imperative that the HPLC instrument, including the detector, is working correctly. The easiest way to check this is by first running a blank. If there is no response, one can move onto injecting the standards. If there is a response to the blank, the column may have been overloaded prior to this run. Refer to a troubleshooting guide for the specific HPLC system. The internet is also an invaluable source for troubleshooting (e.g., see Internet Resources). Keep in mind that the source of the problem may not be the system but may in fact be the column. [Pg.1127]

HPLC), commencing with batch 7. Further investigation revealed that the UV procedure was used for batch 14 as well, in this instance because the HPLC instrument was out of service. With the two populations properly grouped, consistency of the HPLC method to detect ingredient D2 becomes apparent. (See Table 10.)... [Pg.105]

Thus, in conventional installations the HPLC instrument is located at distances between 1.0 and 2.0 m from the cryomagnet, whereas with new available shielded cryomagnets the HPLC instrument can be directly hooked to the cryomagnet. [Pg.10]

The HPLC instrumentation available to the QC laboratory could only run linear gradients and could only control the flow rate to 0.1 ml/min (not 0.05 ml/min as the method would require). Additionally, the contract QC laboratory did not have adequate control of the room temperature so in the summer the temperature could rise as high as 32 C. Simply providing the development laboratory with the instrument specifications prior to method development would have allowed the developed method to be directly transferable to this QC laboratory. [Pg.149]

Perform prevalidation linearity study By analyzing sufficient data points to determine the linear range, sample preparation concentrations and related compound detection and quantitation limits can be estimated. Consideration must be given to the capability of the HPLC instrumentation that will eventually run the method. [Pg.150]

To study the abiotic formation of Me-Hg from Hg2+ as well as demethylation of Me-Hg to Hg2+ in biological tissues during treatment with TMAH, enriched isotope standards from Me198Hg and 201Hg2+ were added to the samples to monitor species transformation and to apply species-specific isotope dilution (SSID) calibration. The HPLC instrumental set-up with a C18 column and an aqueous phase eluent was directly coupled to ICP-MS [31]. [Pg.717]

The microdialysis probe is the heart of the method, as a chromatographic column is the heart of the HPLC instrument. Rigid CMA probes, Models 10, 11, and 12, are used for stereotaxic implantations into the brain, where the probe can be fixed (cemented) to the skull. A flexible probe design (CMA 20) allows the placement of such a catheter into the moving tissues (muscle) or peripheral organs for studies in freely moving animals. The technical difficulties of microdialysis experiments impose requirements for precise liquid delivery, minimized dead volumes, and the capability of handling small sample volumes. [Pg.119]

For absolute quantitative studies the gc instrument can be calibrated in exactly the same way as described for the hplc instrument. Some people use this technique to measure theoretical yields for reactions, although it is always preferable that isolated yields are quoted. [Pg.158]

Injection Repeatability. Precision is measured by multiple injections n = 10)22 of the reference standard at the 100% level and indicates the performance of the HPLC instrument using the chromatographic conditions on one particular day and in one lab. The relative standard deviation, RSD(%), as specified here, will determine the lowest variation limit of the analytical results. Injection repeatability indicates the performance of the HPLC instrument using the chromatographic conditions on one particular day and in one lab. [Pg.434]

General aspects of extra-column volumes in the HPLC instrument have been discussed in Section 2.6. Various formulas have been proposed which allow to calculate the maximum allowed capillary length with regard to band broadening an adaption after M. Martin et al. is as follows ... [Pg.70]

V = (g p X u)ID. For a constant v it is therefore necessary to work with a constant product 6 p X u. Thereby the pressure increases markedly. In Figure 7.3 the particle diameter was decreased from 10 jam to 3 jam. For an identical number of theoretical plates it was possible to reduce the column length to 30% of the original value (note that all the extra-column volumes of the HPLC instrument need to be adapted ). Despite the shorter column the pressure increases by a factor of ten when the condition... [Pg.262]

Use an electronic filter, install the HPLC instrument at another place. [Pg.391]

The general safety concerns in the HPLC Lab are in-line with most analytical laboratories dealing with small sample sizes.10 The high-pressure operation of the HPLC instrument usually does not pose a significant safety risk since small volumes of liquids are used and the units are designed for these conditions. Typical safety risks involve external conditions such as high electric voltages in close proximity with the mobile phase, which can be a flammable or combustible liquid. The operation and set-up of HPLC modules should be in compliance with local, state, and national fire codes such as NFPA 30, NFPA 45, and/or NFPA 70 (National Electrical Codes). [Pg.112]

We greatly appreciate support for this study by Waters Associates and advice from Kenneth Conroe. The National Science Foundation, via the Cornell Materials Science Center, provided the HPLC instrument. [Pg.26]

Which column do I have to install in the HPLC instrument ... [Pg.7]

Poor reproducibility in adsorption chromatography (normal-phase chromatography) is, aside from the limited solubility of polar compounds in organic solvents, the main problem of these systems. Almost always, the problem is with the (uncontrolled) water content in the total system, i.e. the HPLC instrument, the sample, the mobile phase and the column. [Pg.85]

When the HPLC instrument is converted from a reversed-phase mobile phase to a normal-phase mobile phase, care should be taken to flush all... [Pg.97]

Whenever buffers or other mobile-phase additives are used, you should carefully check the solubility in the mobile phase. This is especially true for gradient applications or when the mobile-phase composition is generated automatically by the HPLC instrument. Buffer predpitation is one the most common column problems. A predpitated buffer can be redissolved by slowly purging the column with a mobile phase, in which the buffer is soluble. [Pg.114]

Stagnant solvent in parts of the HPLC instrument can cause extra peaks. The pressure pulse from the injection may cause a smaU quantity of the stagnant solvent to enter the mobile-phase stream and ultimately create additional peaks. This is one of the more diflScuIt conditions to trace and to eliminate. [Pg.195]

The actual pressure measured on the HPLC instrument is usually somewhat higher, due to the flow resistance of the connection tubing. So we diould add an allowance for this effect. Also, we should take into consideration that the particle-diameter designation given by the manufactum may be be a nominal designation. Only if the column backpressure is significantly different from the calculation can we assume that the b kpressure is excessive and that we need to troubleshoot the problem. [Pg.196]


See other pages where The HPLC instrument is mentioned: [Pg.432]    [Pg.50]    [Pg.363]    [Pg.78]    [Pg.199]    [Pg.199]    [Pg.109]    [Pg.10]    [Pg.10]    [Pg.12]    [Pg.279]    [Pg.45]    [Pg.526]    [Pg.1694]    [Pg.57]    [Pg.701]    [Pg.9]    [Pg.460]    [Pg.104]    [Pg.32]   


SEARCH



HPLC instrumentation

The instrumentation

© 2024 chempedia.info