Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Storage of Instruments

Wipe floors, countertops, furniture, and machinery housing with an intermediate- or low-level disinfectant to remove soil. Then rewipe with fresh disinfectant to complete the killing of microbes. Never rinse enviromnental surfaces unless the label directs or a surface might come into contact with eyes. [Pg.261]

International Association of Healthcare Central Service Materiel Management [Pg.261]


Storage of instruments, clean instrument disinfection (30min) Emergency instrument disinfection (2min)... [Pg.210]

Chlorhexidine occasionally causes skin sensitivity. Hemolysis has been reported following accidental intravenous administration.1123 It may damage safe tissues such as the eye. Chlorhexidine is an irritant and should not be used for sensitive tissues. Contact with the eyes should be avoided, as it will cause corneal damage. Syringes sterilized with chlorhexidine must be cleaned perfectly before using for other purposes. Aqueous solutions of chlorhexidine used for the storage of instruments should contain sodium nitrite (0.1%) to inhibit metal corrosion and should be required to be changed every week. [Pg.310]

QACs, e.g. cetyltrimethyl 0.1% Solution forwounds 0.1% Storage of instruments 0.01% Eye-drops... [Pg.293]

Its detergent action destroys the membranes that coat and protect the miCTOorganisms. Zephiran chloride is recoimnended as a disinfectant solution for skin and hands prior to surgery and for the sterile storage of instruments. The trade names of srme other anti-infectives that contain quaternary ammonium salts are Phemerol , Bactine , and Ceepryn . [Pg.204]

Over the years, many instruments have been developed for and used in the scientific laboratory. Today, the computer is used as a major tool in the scientific laboratory for the capture, manipulation, transfer, and storage of data. Consequently, the concern for data quality has shifted from the instruments that are used in the generation of the data to these electronic systems, often neglecting the fact that the data are only as accurate as the instrument measurements. For instance, many electronic systems can be used in chromatography analysis, from the electronic log book where the test substance inventory is kept, throughout data capture in the instrument, to the digitized electronic signal that is the raw data on the computer network. For crop residue samples, the... [Pg.1039]

Although the DL 40 was capable of performing Karl Fischer water titrations and Mettler developed a separate microprocessor-controlled push-button operated DL 18 KF titrator, they also introduced as an all-purpose apparatus the improved DL 40 RC (see Fig. 5.11) with a dual titration head and with a modified software program to handle the new two-component titrants for Karl Fischer titration (see pp. 204-205). The instrument can also be expanded into an automatic series titrator by connecting the RT 40 sample transport for 16 samples and storage of 50 sample weights from a connected balance this series routine can be interrupted at any time after completion of the titration in progress. [Pg.342]

A typical instrument (Figure 13.6) will have a certain amount of memory for the storage of methods and data, and will offer a choice of several modes of titration, e.g. [Pg.679]

This Second Edition continues the basic approach of the first with the addition of four chapters. Chapter 1 is an outline of the development of soil chemistry with specific reference to the development of instruments that have been essential to the present understanding of soil chemistry. Chapter 7 is a new chapter dealing with soil sampling, both in the field and in the laboratory, soil water sampling, sample transport, and storage. Chapter 8 discusses direct, modified, and indirect methods of soil analysis. Chapter 15 covers the recent development of hyphenated instrumental methods and their application to soil analysis. [Pg.13]

Beyond simple data storage and instrument control, modern data systems provide extensive data analysis capabilities, including fitted baselines, peak start and stop tic marks, named components, retention times, timed events and baseline subtraction. Further, they provide advanced capabilities, such as multiple calibration techniques, user-customizable information and reports and collation of multiple reports. If a Laboratory Information Management System (LIMS) is available, the chromatographic data system should be able to directly transfer data files and reports to the LIMS without user intervention. The chapter by McDowall provides a terse but thorough description of the... [Pg.476]

A. STORAGE OF PARAMETERS AND DATAON SAMPLES FOR SELECTED INSTRUMENTAL TECHNIOUES... [Pg.379]

Instrument cabinets should not have pockets, shelves or other devices that encourage the storage of drawings or other combustible materials inside the cabinets. Metal file cabinets or shelves preferably in the control room area should be provided for the storage of drawings, manuals, etc. Closed-door metal cabinets should be provided for the storage of any spare parts that must be kept in the area. [Pg.305]

The terms defined above are perhaps illustrated by using the example of the simple assay which we have mentioned before. The assay in Box 1.3 is laid out in the style of a SOP. This particular section of the operating procedure describes the assay itself but there would also be other sections in the procedure dealing with safety issues, the preparation and storage of the solutions used for extraction and dilution, the glassware required and a specification of the instrumentation to be used. [Pg.9]

A typical set-up is shown in Fig. 6. The analytical instruments are coupled to a mainframe computer over a decreasing hierarchy of microcomputers and instrument processors. Figure 7 shows schematically the information flows and storage of a... [Pg.15]

Dynamic techniques are used to determine storage and loss moduli, G and G respectively, and the loss tangent, tan 6. Some instruments are sensitive enough for the study of liquids and can be used to measure the dynamic viscosity rj. Measurements are made as a function of temperature, time, or frequency, and results can be used to determine transitions and chemical reactions as well as the properties noted above. Dynamic mechanical techniques for solids can be grouped into three main areas free vibration, resonance-forced vibrations, and nonresonance-forced vibrations. Dynamic techniques have been described in detail (242,251,255,266,269—279). A number of instruments are listed in Table 8. Related ASTM standards are listed in Table 9. [Pg.196]

The chemical, instrumental and sensory data presented above indicated that storage of cooked beef affects the lipid composition and concomitantly, the flavor of beef. The data also indicated that primary tastes like bitter and sour are affected by storage. [Pg.85]

However, nowadays some other different mass spectrometers are used for ICP-MS time-of-flight (TOP) systems for multielemental analysis of transient signals, ion trap analysers for ion storage, multicollector instruments for precise isotope ratio measurements and double-focusing sector field mass spectrometers for high mass resolution, but still the majority of instruments are equipped with quadrupole filters, which are simpler and cheaper. [Pg.24]


See other pages where Storage of Instruments is mentioned: [Pg.519]    [Pg.293]    [Pg.261]    [Pg.1075]    [Pg.519]    [Pg.293]    [Pg.261]    [Pg.1075]    [Pg.58]    [Pg.241]    [Pg.804]    [Pg.315]    [Pg.54]    [Pg.54]    [Pg.619]    [Pg.182]    [Pg.169]    [Pg.156]    [Pg.207]    [Pg.102]    [Pg.8]    [Pg.356]    [Pg.524]    [Pg.536]    [Pg.543]    [Pg.358]    [Pg.66]    [Pg.165]    [Pg.335]    [Pg.145]    [Pg.9]    [Pg.81]    [Pg.175]    [Pg.349]    [Pg.437]    [Pg.466]   


SEARCH



Instruments storage

© 2024 chempedia.info