Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Standard solutions,stability

Sample and standard solution stability knowledge are essential during the early stages of method development and optimization.The solutions must be stable during the trip to the lab, or an alternative procedure should be used to increase sample stability. For example, the sample may need to be immediately cooled or quenched at the time of the sample pull. If quenching the sample is a... [Pg.404]

Reaction solution and standard solutions stability must meet the process... [Pg.406]

Application of IP and NCS in conjunction with specification tolerance limits enables to substantiate acceptance criteria for linear regression metrological characteristics (residual standard deviation, correlation coefficient, y-intercept), accuracy and repeatability. Acceptance criteria for impurity influence (in spectrophotometric assay), solution stability and intermediate precision are substantiated as well. [Pg.340]

The analyte stability during storage and processing of samples or in standard solutions and extracts is not part of method validation in Germany. Therefore, insufficient stability will not be routinely detected and even then more or less only by chance. Also, separate tests for analyte homogeneity and extraction efficiency were not performed. [Pg.127]

Quantitation is performed by the calibration technique. A new calibration curve is constructed with each dinifroaniline standard solution. The peak area is plotted against the injected amount of standard. Each dinifroaniline in the sample is measured by using the peak area for each standard. Before each set of measurements, the sensitivity and stability of the GC and HPLC system is ascertained by injecting more than one standard solution containing ca 0.05-2 mg L of each compound. [Pg.394]

Recommendation Dilute the standard solutions twice with blank solutions prepared from each of the blank samples. Impurities in the blank samples reduce the thermal decomposition of the target analytes in the injection port and stabilize the profiles of ionization and fragmentation of the target analytes. [Pg.546]

Quantitation is performed by the calibration technique. Construct a new calibration curve with thenylchlor standard solutions for each set of analyses. The thenylchlor peak usually appears at a retention time around 4.5 min. Plot the peak area against the injected amount of thenylchlor. The injection volume (2 pL) should be kept constant as the peak area varies with the injection volume with NPD. Before injecting the sample solutions, check the stability of sensitivity of the GC system by injecting more than one standard solution containing ca 0.05-2 ng of thenylchlor. Recommendation inject standard solutions and sample solutions alternately rather than constructing the calibration curve in advance. [Pg.588]

In contrast to a mixture of redox couples that rapidly reach thermodynamic equilibrium because of fast reaction kinetics, e.g., a mixture of Fe2+/Fe3+ and Ce3+/ Ce4+, due to the slow kinetics of the electroless reaction, the two (sometimes more) couples in a standard electroless solution are not in equilibrium. Nonequilibrium systems of the latter kind were known in the past as polyelectrode systems [18, 19]. Electroless solutions are by their nature thermodyamically prone to reaction between the metal ions and reductant, which is facilitated by a heterogeneous catalyst. In properly formulated electroless solutions, metal ions are complexed, a buffer maintains solution pH, and solution stabilizers, which are normally catalytic poisons, are often employed. The latter adsorb on extraneous catalytically active sites, whether particles in solution, or sites on mechanical components of the deposition system/ container, to inhibit deposition reactions. With proper maintenance, electroless solutions may operate for periods of months at elevated temperatures, and exhibit minimal extraneous metal deposition. [Pg.228]

Hodgson DW, J.F. Thompson, Watts, RR. 1982. Accuracy of pesticide reference standard solutions. Part II Chemical stability under four storage conditions. J Assoc Off Anal Chem 65 94-102. [Pg.188]

Once the solution stability has been determined, a storage period should be determined based on the use of the method under real-world conditions. Data may show a standard solution to be very stable for 6 months at room temperature, but a quality control (QC) lab may not want to use this solution for such a long time. [Pg.211]

In developing a commercially viable method, the stability of samples, standards, and reagents used for the HPLC method must be considered. For the stability of standard solutions and reagents, long-term stability of up to weeks is desirable. For the stability of sample solutions, a minimum of 3 days is ideal. Generally, the reagents for standard and sample preparation should be the same or very similar to the mobile phase composition. [Pg.352]

The stability of sample and standard solutions over the course of the experiments must be demonstrated to ensure that the results are not erroneous owing to degradation. Bracketing of standard and sample injections can also help in this respect. [Pg.118]

Stabilize urine by adding 2 mL of an aqueous 30% citic acid solution to the bottle before collection. Place 50 mL of stablized urine in a 125 mL separatory funnel and spike the urine with 1 mL of MBOCA standard solution in methanol. Swirl to mix and let stand 5 minutes. [Pg.424]

Solution stability of standards The sample solutions are stable for... [Pg.48]

Stock Solution Stability. The stability of stock solutions of drug and the internal standard should be evaluated at room temperature for at least 6 h. If the stock solutions are refrigerated or frozen for the relevant period, the stability should be documented. After completion of the desired storage time, the stability should be tested by comparing the instrument response with that of freshly prepared solutions. [Pg.114]

One of the difficulties encountered with this method is in maintaining the stability and the sensitivity of the NMR instrument while the standard solutions are being analysed. [Pg.154]

It is neither necessary nor desirable to prepare large volumes of the working standard solution since there is always a question of the long-term stability of such solutions. Likewise, in most cases, the consumption of standard solution is in microliter units therefore, if it is desired to retain a stock solution, it should be the more concentrated original solution in small volume carefully kept in a freezer. From this, convenient small portions of working standard solutions can be prepared as needed. [Pg.397]

Other New Methods. Because the values obtained are dependent on the conditions of measurement, standard test procedures are under review by ISO for determination of cold-water solubility of water-soluble dyes determination of the solubility and solution stability of waler-soluhle dyes and determination of the electrolyte stability of reactive dyes. [Pg.527]

An advantage of the technique is the use of an electrical standard to replace chemical standards and the problems associated with their preparation and stability. The coulometric titration also permits the generation of reagents such as copper(I) or bromine, which are difficult to employ as standard solution, or others such as silver(II) or chlorine, which are virtually impossible to use in any other way. A disadvantage of the coulometric titration is its lack of specificity. [Pg.777]

The concentrations of these standard solutions should be checked regularly depending on the expected chemical stability. [Pg.1017]

Shin and Godber (228) used a silica column and a mobile phase composed mainly of isooctane and small amounts of ethyl acetate, acetic acid, and 2,2-dimethoxypropane. The acetic acid component reduced retention times of the late-eluting E vitamers, presumably by competing with water and polar material for binding to silanol groups on the silica surface. 2,2-Dimethoxy-propane reacts with water to form acetone and methanol, and its inclusion stabilized retention times and reduced the need for column regeneration. Chromatograms of vitamin E vitamers in a standard solution and in a saponified rice bran sample are shown in Fig. 13. [Pg.380]

The 3-(Diethoxymethylsilyl) propyl methacrylate was obtained from Union Carbide Corporation (Tarrytown, N.Y.). Hexamethylene diisocyanate was purchased from Eastman Kodak (Rochester, NY). The Tinuvin stabilizers were obtained from Ciba-Geigy Corporation (Ardsley, N.Y.). Standard solutions containing approximately IX by weight were prepared in toluene or methylene chloride for direct injection. [Pg.293]

E2, the theoretical Eh of the reference electrode, may be calculated from the stability constants. For each 1°C increase in temperature, the potential of K4Fe(CN)6-K3Fe(CN)6-KCl solution shows a decrease of about 2 mV The Eh of this redox standard solution for platinum electrode vs. Ag AgCl reference electrode may be calculated as follows ... [Pg.221]

Dealing with the chemical issues is usually much more demanding. In a typical chemical measurement there are two main components to the chemical chain of comparisons, as illustrated in Fig. 1. One component is established by measuring the response from one or more chemical standard(s), often in the form of standard solutions, prepared from pure substance RMs. The identity, purity, and stability of the standards are important issues. The calibration factor Fc and its uncertainty, Uc, describe this part of the chain. Such methods are often called ratio methods and include most of the... [Pg.86]

This example is of a traceability [1] protocol [2] for the chemical measurement of an element by a primary method of measurement [3]. It can be used for the certification of a single-element reference material by a national reference laboratory. This protocol relates to a very pure strontium nitrate solution, stabilized by 10% (by volume) nitric acid1. This solution is to be certified for the amount of strontium substance n(Sr) per unit mass of aqueous solution m (sol). The general measurement method described is based in part on the experience of certifying a currently available certified reference material (CRM) [4], Standard Reference Material (SRM) 3153a [5],... [Pg.192]

Number of theoretical plates Capacity factor Relative retention Solution stability Sample Standard... [Pg.160]

Experiment 1. Short-term analyte and IS bench top (i.e., room temperature or wet ice bath) solution stability. A minimum of one concentration such as ULOQ spiking solution must be tested. Some laboratories require solution stability at standard 2 (which has concentration typically twice of the LLOQ) or LLOQ spiking solution level. Test for minimum 6 h. [Pg.57]

Analytical measurements should be made with properly tested and documented procedures. These procedures should utilise controls and calibration steps to minimise random and systematic errors. There are basically two types of controls (a) those used to determine whether or not an analytical procedure is in statistical control, and (b) those used to determine whether or not an analyte of interest is present in a studied population but not in a similar control population. The purpose of calibration is to minimise bias in the measurement process. Calibration or standardisation critically depends upon the quality of the chemicals in the standard solutions and the care exercised in their preparation. Another important factor is the stability of these standards once they are prepared. Calibration check standards should be freshly prepared frequently, depending on their stability (Keith, 1991). No data should be reported beyond the range of calibration of the methodology. Appropriate quality control samples and experiments must be included to verify that interferences are not present with the analytes of interest, or, if they are, that they be removed or accommodated. [Pg.260]

Standard stability Stability of standard solutions should be established + +... [Pg.355]

While this approach has limitations related to the peptidic nature of the linker such as stability to harsh reaction conditions and requirement of specific functional groups to be coupled with the linker arms, its application to particular libraries and chemistries may be useful. Its biological utility was assessed in a bead-based antimicrobial assay on bacterial cells [54], which produced good correlations between the MIC (minimal inhibitory concentration) values for the compounds released in situ in the culture medium from the beads or tested as standard solutions. [Pg.213]


See other pages where Standard solutions,stability is mentioned: [Pg.237]    [Pg.237]    [Pg.150]    [Pg.101]    [Pg.245]    [Pg.39]    [Pg.202]    [Pg.475]    [Pg.211]    [Pg.372]    [Pg.407]    [Pg.340]    [Pg.351]    [Pg.21]    [Pg.367]    [Pg.522]    [Pg.527]    [Pg.278]    [Pg.872]    [Pg.406]   
See also in sourсe #XX -- [ Pg.397 ]




SEARCH



Solutions standard solution

Solutions standardization

Stability standards

Stabilizing solutes

Standard solution

Standardized Solutions

© 2024 chempedia.info