Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Retention times matrix

Continuing calibration for a Series Method is performed using calibration check compounds. Surrogate compounds are added to the matrix before sample preparation to evaluate recovery levels. To check GC retention times, internal standards are added to a sample after its preparation for analysis. [Pg.418]

Fig. 10. Wet high, intensity magnetic separator using cryogenically cooled coils and a stationary matrix where A is the feed control for top-fed or retention time control for underfed operation and B is the feed control for underfed or retention time control for top-fed operation. Fig. 10. Wet high, intensity magnetic separator using cryogenically cooled coils and a stationary matrix where A is the feed control for top-fed or retention time control for underfed operation and B is the feed control for underfed or retention time control for top-fed operation.
At this point, the solution containing the component to be measured (Ax) also contains any other compounds from the original matrix that are soluble in the solvent used in the analysis. For the analysis to be accurate, other components in the matrix cannot interfere by eluting at the same retention time as the components to be measured. For accurate MS analyses, the matrix component must not interfere with production of the ions being measured for either the internal standard or the component to be measured. In some cases, to eliminate interferences, it may be necessary to increase the resolution of the mass spectrometer by narrowing the mass window being monitored. Alternatively, MS/MS can be used to avoid chemical interference (see Chapter 1). [Pg.26]

Any data matrix can be considered in two spaces the column or variable space (here, wavelength space) in which a row (here, spectrum) is a vector in the multidimensional space defined by the column variables (here, wavelengths), and the row space (here, retention time space) in which a column (here, chromatogram) is a vector in the multidimensional space defined by the row variables (here, elution times). This duality of the multivariate spaces has been discussed in more detail in Chapter 29. Depending on the chosen space, the PCs of the data matrix... [Pg.246]

As explained before, the scores of the spectra can be plotted in the space defined by the two principal components of the data matrix. The appearance of the scores plot depends on the way the rows (spectra) and the columns have been normalized. If the spectra are not normalized, all spectra are situated in a plane (see Fig. 34.5). From the origin two straight lines depart, which are connected by a curved line. We have already explained that the straight line segments correspond with the pure spectra which are located in the wings of the elution bands (selective retention time... [Pg.260]

Contrary to EFA which calculates a PCA of a sub data matrix to which rows are added, in fixed-size window EFA a small window of rows is selected which is moved over the data set (see Fig. 34.30). Typically, a window of seven consecutive spectra is used. At each new position of the window a PCA is calculated and the eigenvalues associated with each PC are recalculated and are plotted as a function of the position of the window. This yields a number of eigenvalue-lines. Figure 34.31 shows the eigenvalue-lines obtained for a simulated pure LC-DAD peak. In the baseline zones (null spectra) all eigenvalue-lines are noisy horizontal lines. In the selective retention time regions (one component present) the eigenvalue-line associated with the first PC follows the appearance and disappearance of the... [Pg.279]

A dissimilarity plot is then obtained by plotting the dissimilarity values, dj, as a function of the retention time i. Initially, each p 2 matrix Y, consists of two columns the reference spectrum, which is the mean (average) spectram (normalised to unit length) of matrix X, and the spectrum at the /th retention time. The spectrum with the highest dissimilarity value is the least correlated with the mean spectrum, and it is the first spectrum selected, x, . Then, the mean spectrum is replaced by x, as reference in matrices Y, (Y, = [x j x,]), and a second dissimilarity plot is obtained by applying eq. (34.14). The spectrum most dissimilar with x, is selected (x 2) and added to matrix Y,-. Therefore, for the determination of the third dissimilarity plot Y, contains three columns [x, x 2 /]> wo reference spectra and the spectmm at the /th retention time. [Pg.295]

In summary, the selection procedure consists of three steps (1) compare each spectrum in X with all spectra already selected by applying eq. (34.14). Initially, when no spectrum has been selected, the spectra are compared with the average spectrum of matrix X (2) plot of the dissimilarity values as a function of the retention time (dissimilarity plot) and (3) select the spectrum with the highest dissimilarity value by including it as a reference in matrix Y,-. The selection of the spectra is finished when the dissimilarity plot shows a random pattern. It is considered that there are as many compounds as there are spectra. Once the purest spectra are available, the data matrix X can be resolved into its spectra and elution profiles by Alternating Regression explained in Section 34.3.1. [Pg.296]

It is important to note that the matrix effects, interferences, and variability in method efficiency are to be factored in when determining the MDL. If this was not done then only the background noise (see Figure 2, peak 13) would be considered in the definition of the MDL. In real-life samples there is a good possibility that matrix component peaks would either co-elute or elute at retention times close to... [Pg.67]

The simplest method for estimating the MDL and MQL would be to measure the peak-to-peak noise (Ap p) around the analyte retention time and then estimate the concentration (of the analyte in the matrix) that would yield a signal equal to three times the Ap p (estimated MDL). [Pg.69]

Table 4.45 shows the main features of SEC. This technique has become an indispensable tool for polymer characterisation. SEC has some advantages over other LC methods, such as the predictability of the end of a chromatographic run and of the retention times in a calibrated chromatographic system. SEC is an attractive technique for prefractionation or sample clean-up prior to a more sensitive RPLC technique. This intermediate step is especially interesting for experimental purposes whenever polymer matrix interference cannot be separated from the peak of interest [647]. Disadvantages are that the whole separation must be eluted within the... [Pg.261]

When using microbial products for mammalian metabolite identification, it is suggested to compare all the analytical data available. For example, slight differences in MS2 or MS3 spectra may indicate that the microbial products are not the same as the mammalian metabolite. Owing to matrix effects, HPLC retention time often varies from run to run, so it is good practice to spike a comparable amount of purified microbial product into the in vitro, in vivo or purified samples that contain the mammalian metabolite of interest. If the microbial metabolite and the mammalian metabolite are the same compound, then they should co-elute under different HPLC conditions, including different solvent pH, and the MS and/or UV peak area would increase accordingly. [Pg.208]

UV detection, diode-array detector (DAD) and fluorescence have been the detection techniques used, coupled to HPLC for the analysis of OTC. UV detection is set at 355 nm [49-51], 350 nm [40], or at 353 nm [52], Using the diode array detector [49] offers advantages that the target peak can be identified by its retention time and absorption spectrum. Compared to UV detection, fluorescence detection is generally more specific and is less interfered by other compounds in the sample matrix [51]. A HPLC method with electrochemical detection has also been suggested recently. Zhao et al. [53] described HPLC with a coulometric electrode array system for the analysis of OTC, TC, CTC, DC, and methacycline (MC) in ovine milk. An amper-ometric detection coupled with HPLC was developed by Kazemifard and Moore [54] for the determination of tetracyclines in pharmaceutical formulations. [Pg.111]

In a bioanalytical method, analyses of blank samples (plasma, urine, or other matrix) should be obtained from at least six sources. Each blank sample should be tested for the possible interference of endogenous substances, metabolites, or degradation products. The response of the peaks interfering at the retention time of the analyte should be less than 20% of the response of a lower quantitation limit standard, and should be less than 5% of the response of the internal standard that was used [18, 19]. For dissolution studies, the dissolution media or excipients should not give a peak or spot that has an identical Rt or Rf value with the analyte [20]. [Pg.248]

The main advantages of CE are its ease of use, speed and efficiency, while disadvantages are matrix effects causing shifts in retention time and variations in peak area response, non-gaussian peak shapes for certain molecules which are difficult to integrate, and limited detector sensitivity. [Pg.124]

Chirica and Remcho first created the outlet frit, packed the column with ODS beads, and then fabricated the inlet frit. The column was filled with aqueous solution of a silicate (Kasil) and the entrapment achieved by heating the column to 160 °C [105,106]. The monolithic column afforded considerably reduced retention times compared to the packed-only counterpart most likely due to a partial blocking of the pores with the silicate solution. This approach was recently extended to the immobilization of silica beads in a porous organic polymer matrix [107]. [Pg.28]


See other pages where Retention times matrix is mentioned: [Pg.392]    [Pg.392]    [Pg.70]    [Pg.447]    [Pg.223]    [Pg.235]    [Pg.240]    [Pg.62]    [Pg.92]    [Pg.17]    [Pg.289]    [Pg.271]    [Pg.113]    [Pg.119]    [Pg.247]    [Pg.249]    [Pg.275]    [Pg.280]    [Pg.295]    [Pg.463]    [Pg.772]    [Pg.459]    [Pg.465]    [Pg.85]    [Pg.540]    [Pg.114]    [Pg.7]    [Pg.3]    [Pg.394]    [Pg.448]    [Pg.506]    [Pg.516]    [Pg.538]    [Pg.571]    [Pg.214]    [Pg.1076]   
See also in sourсe #XX -- [ Pg.335 ]




SEARCH



Matrix timing

Retention time

Retention times internal standards, sample matrix effect

© 2024 chempedia.info