Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Retention analyte

The retention times obtained at a concentration of 1 g are identical to the analytical retention times. Therefore, system behavior is linear at concentrations below 1 g L k When the concentration increases, the retention decreases, which are consistent with a Langmuir-type behavior. [Pg.269]

The simplest method for estimating the MDL and MQL would be to measure the peak-to-peak noise (Ap p) around the analyte retention time and then estimate the concentration (of the analyte in the matrix) that would yield a signal equal to three times the Ap p (estimated MDL). [Pg.69]

Ka and Kn derived from the knowledge analytical retention times1... [Pg.274]

Analyte retention times of 8.0, 0.7, 0.4, 12.0, and 17.6 min using 50 50 methanol watcr, 50 50 acetonitrile water, 70 30 methanol water, 50 50 methanol 0.1% formic acid, and 50 50 methanol 10 mM ammonium formate mobile phases, respectively. [Pg.47]

An accelerated retention window (ARW) starting from a certain proportion of organic (0A% acetonitrile) as the start of the gradient mobile phase composition was used to calculate preparative retention time on the basis of analytical retention time. The same procedure was used to adjust the retention times of all compounds to facilitate collection at the same predetermined retention time. This allowed set-up of a device for collecting any HPLC peaks that surpassed a certain threshold defined by UV or ELSD. [Pg.62]

Y.V. Kazakevich, R. LoBrutto, F. Chan and T. Patel, Interpretation of the excess adsorption isotherms of organic components on the surface of reversed-phase adsorbents. Effect on the analyte retention. J. Chromatogr.A 913 (2001) 75-87. [Pg.60]

The carbon content of a stationary phase is measured by an elemental analyser, as a weight balance before and after heating at 800 °C. Particle size, pore size, and surface area are measured by specific instruments, such as a particle size analyser, nitrogen adsorption porosimeter, and mercury depression analyser, respectively. The precision of the measurement of carbon content is high however, that of the other measurements is relatively poor. Therefore, it is difficult to relate the surface area of different silica gels to analyte retention factors. [Pg.43]

The discussion of IL-based stationary phases up to this point has centered around ILs that are either coated as a thin film on a capillary wall or on a solid support. Although ILs exhibit a variety of properties that allow them to be unique stationary phases, their most significant drawback lies with their drop in viscosity with increasing temperature. This results in an increased propensity for flowing of the IL within the capillary, which often produces pooling of the stationary phase and nonuniform film thickness throughout the column. These factors often contribute to diminished analyte retention time reproducibility as well as detrimental effects on separation efficiency. [Pg.158]

Further research on mixed IL stationary phases will allow for the chroma-tographer to tune the stationary phase composition to provide enhanced control over the separation selectivity and analyte elution order, particularly for complicated analyte mixtures. The development of models that correlate analyte retention with the IL composition will prove useful for multidimensional GC. Micellar GC utilizing IL solvents presents an exciting class of highly selective stationary phases. The development of CSPs will likely mature as more chiral ILs are synthesized and evaluated from the chiral pool. [Pg.162]

The packing material first described for direct injection of biological samples was prepared by simply saturating the accessible adsorption sites of a Cis reversed-phase silica with human plasma proteins (105). After saturation, the human plasma proteins were denatured at the external surface, and their native conformation was destroyed. With this treatment, the proteins formed a hydrophilic layer with weak ion-exchange properties, which provided protection from contact with the sample proteins, whereas the alkyl ligands inside the pores remained unchanged and thus served for analyte retention. The retention behavior of the saturated phase did not alter with this treatment, but the efficiency was reduced dramatically. Such protein-coated columns have shown a lifetime of several months (106). [Pg.606]

Every chromatographic method identifies analytes based on their retention time by comparing the target analyte retention time in the chromatogram of a sample to that of a standard. [Pg.214]

Fig. 6.6 Detection of matrix-suppression components at the analyte retention time and at late elution. The chromatogram of the drug analyte is superimposed on the post-column infusion trace of the matrix extract. The chromatography was reversed-phase with a short run time of 3 min. The retention time of the... Fig. 6.6 Detection of matrix-suppression components at the analyte retention time and at late elution. The chromatogram of the drug analyte is superimposed on the post-column infusion trace of the matrix extract. The chromatography was reversed-phase with a short run time of 3 min. The retention time of the...
The counterion associated with the sorbent when it is manufactured is replaced by another ion of like charge existing on the analyte to achieve retention. However, analyte retention is affected by the ionic strength of the sample matrix because other ions present will compete with the analyte of interest for retention by ion-exchange mechanisms [75],... [Pg.92]

As in analytical liquid chromatography (LC), analyte retention depends on sample concentration, solvent strength, and sorbent characteristics. An empirical approach to methods development initially involves screening the available sorbents. The first step is to determine which sorbents best retain the analyte. The second consideration is to evaluate the solvents needed to elute the compound and the compatibility of those sorbents to the chromatographic testing procedure. The third step is to test the blank sample matrix to evaluate the presence of possible interferents. Finally, recoveries of known quantities of analyte added to the sample matrix must be determined. [Pg.281]


See other pages where Retention analyte is mentioned: [Pg.550]    [Pg.125]    [Pg.270]    [Pg.263]    [Pg.57]    [Pg.67]    [Pg.308]    [Pg.403]    [Pg.417]    [Pg.758]    [Pg.1185]    [Pg.1186]    [Pg.292]    [Pg.189]    [Pg.284]    [Pg.343]    [Pg.345]    [Pg.26]    [Pg.33]    [Pg.39]    [Pg.53]    [Pg.59]    [Pg.62]    [Pg.63]    [Pg.64]    [Pg.68]    [Pg.83]    [Pg.387]    [Pg.576]    [Pg.217]    [Pg.871]    [Pg.22]    [Pg.125]    [Pg.270]    [Pg.164]    [Pg.153]   
See also in sourсe #XX -- [ Pg.79 ]

See also in sourсe #XX -- [ Pg.25 , Pg.56 , Pg.504 ]




SEARCH



Acidic analyte retention dependence

Acidic modifiers/buffers analyte retention

Analyte HPLC retention

Analyte capacity/retention factor

Analyte retention acetonitrile/water eluent

Analyte retention determination

Analyte retention factors affecting

Analyte retention modeling

Binary eluent analyte retention

Counteranion basic analyte retention

Ionizable analyte retention

Ionizable analyte retention/selectivity

Physicochemical parameters, retention analyte

Retention as a Function of Analyte Nature

Retention factor analyte

Retention of basic analytes

Solid phase extraction analyte retention

Terms Referring to the Retention of Analytes

Thermodynamic properties, analyte retention

© 2024 chempedia.info