Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Proportional counter pulse

Fig. 2-4. Mean pulse height versus quantum energy, to illustrate pulse-height selection of characteristic lines. Side window tube, 3-in. diameter 0.005-in. wire operating at 1275 v filled with argon plus 10% ethylene total pressure, 15 cm of mercury (proportional counter) gain, 3.5 X 104. (After Friedman, Birks, and Brooks, A STM Spec. Tech. Publ., No. 157, page 3. Copyright 1954. American Society for Testing Materials.)... Fig. 2-4. Mean pulse height versus quantum energy, to illustrate pulse-height selection of characteristic lines. Side window tube, 3-in. diameter 0.005-in. wire operating at 1275 v filled with argon plus 10% ethylene total pressure, 15 cm of mercury (proportional counter) gain, 3.5 X 104. (After Friedman, Birks, and Brooks, A STM Spec. Tech. Publ., No. 157, page 3. Copyright 1954. American Society for Testing Materials.)...
When a proportional counter is. used in conjunction with a pulse-height selector, the occurrence of an escape peak may vitiate the results. Assume that the counter filling contains argon, whose K edge is at 3.87 A, and suppose that the pulse-height selector is set to select an x-ray line 3f shorter wavelength the intensity of which is to be measured. This line will excite the K lines of argon. To the extent that these lines are... [Pg.54]

In the phosphor-photoelectric detector used as just described, the x-ray quanta strike the phosphor at a rate so great that the quanta of visible light are never resolved they are integrated into a beam of visible light the intensity of which is measured by the multiplier phototube. In the scintillation counters usual in analytical chemistry, on the other hand, individual x-ray quanta can be absorbed by a single crystal highly transparent to light (for example, an alkali halide crystal with thallium as activator), and the resultant visible scintillations can produce an output pulse of electrons from the multiplier phototube. The pulses can be counted as were the pulses-from the proportional counter. [Pg.59]

The tube of Figure 2-2 can be operated as an ionization chamber, as a proportional counter, or as a Geiger counter. The tube output differs radically from one case to another. Because of these differences, the electronic circuitry associated with the tube must also be different for each case if the pulses from the tube are to be reliably selected and counted. In particular, the circuitry will have to differ in characteristics such as stability, amount of amplification, and time of response. In all cases, linear amplification (amplifier output always proportional to tube output) is desirable. [Pg.59]

The apparatus as modified for x-ray emission spectrograph is also shown in Figure 11-1. The proportional counter may be used alone (pulse-height analysis Section 2.13) or a curved-crystal spectrometer can be employed to achieve better resolution. Analytical results were comparable to those quoted above, but localization of the area analyzed was considerably less sharp than the micron-diameter spot achieved in differential absorptiometry. [Pg.294]

Proportional counters are extremely sensitive, and the voltages are large enough so that all of the electrons are collected within a few tenths of a microsecond. Each pulse corresponds to one gamma ray or neutron interaction. The amount of charge in each pulse is proportional to the number of original electrons produced. The proportionality factor in this case is the gas amplification factor. The number of electrons produced is proportional to the energy of the incident particle. [Pg.45]

Proportional counters measure the charge produced by each particle of radiation. To make full use of the counter s capabilities, it is necessary to measure the number of pulses and the charge in each pulse. Figure 9 shows a typical circuit used to make such measurements. [Pg.47]

The BF3 proportional counter is used to monitor low power levels in a nuclear reactor. It is used in the "startup" or "source range" channels. Proportional counters cannot be used at high power levels because they are pulse-type detectors. Typically, it takes 10 to 20 microseconds for each pulse to go from 10% of its peak, to its peak, and back to 10%. If another neutron interacts in the chamber during this time, the two pulses are superimposed. The voltage output would never drop to zero between the two pulses, and the chamber would draw a steady current as electrons are being produced. [Pg.51]

Flat plates or concentric cylinders may be utilized in the construction of an ionization chamber. The flat plate design is preferred because it has a well-defined active volume and ensures that ions will not collect on the insulators and cause a distortion of the electric field. The concentric cylinder design does not have a well-defined active volume because of the variation in the electric field as the insulator is approached. Ionization chamber construction differs from the proportional counter (flat plates or concentric cylinders vice a cylinder and central electrode) to allow for the integration of pulses produced by the incident radiation. The proportional counter would require such exact control of the electric field between the electrodes that it would not be practical. [Pg.52]

B10 lined or BF3 gas-filled proportional counters are normally used as source range detectors. Proportional counter output is in the form of one pulse for every ionizing event therefore, there is a series of random pulses varying in magnitude representing neutron and gamma ionizing events. [Pg.88]

Electronic noise, identified via rise-time characteristics, shown rising to 30 times background rate for 15-mL gas proportional counter. A tiny fraction of such noise, not identified by means such as pulse shape analysis, could invalidate results which presume that the background is stable to < 10%. [Pg.171]

A proportional counter consists of a tube filled with a gas such as xenon, with positive and negative electrodes. The negative electrode is a thin wire maintained at a potential around -2 kV. Incoming photons ionise gas molecules. These drift towards the negative electrode, until the field enhancement around the thin wire is sufficient to multiply them by the cascade effect, and cause a charge pulse on the wire. The pulse is quenched by the addition of a quench gas, normally a halogen or hydrocarbon which reacts with the ions and stops the cascade. [Pg.44]

Radioactivity of uranium can be measured by alpha counters. The metal is digested in nitric acid. Alpha activity is measured by a counting instrument, such as an alpha scintillation counter or gas-flow proportional counter. Uranium may be separated from the other radioactive substances by radiochemical methods. The metal or its compound(s) is first dissolved. Uranium is coprecipitated with ferric hydroxide. Precipitate is dissolved in an acid and the solution passed through an anion exchange column. Uranium is eluted with dilute hydrochloric acid. The solution is evaporated to near dryness. Uranium is converted to its nitrate and alpha activity is counted. Alternatively, uranium is separated and electrodeposited onto a stainless steel disk and alpha particles counted by alpha pulse height analysis using a silicon surface barrier detector, a semiconductor particle-type detector. [Pg.958]

The electronic instrumentation necessary for the operation of the proportional counter is shown in Figure 18.6. Pulses from the detector pass through a preamplifier and amplifier, where they are shaped and amplified. Emerging from the amplifier, the pulses go to a discriminator. The discriminator is set so as not to trip on noise pulses but rather to trip on radiation pulses of any larger size. The number of discriminator pulses produced is recorded by the scaler. [Pg.546]


See other pages where Proportional counter pulse is mentioned: [Pg.150]    [Pg.109]    [Pg.150]    [Pg.109]    [Pg.341]    [Pg.51]    [Pg.53]    [Pg.55]    [Pg.60]    [Pg.60]    [Pg.61]    [Pg.61]    [Pg.62]    [Pg.62]    [Pg.67]    [Pg.303]    [Pg.350]    [Pg.38]    [Pg.66]    [Pg.67]    [Pg.67]    [Pg.149]    [Pg.155]    [Pg.346]    [Pg.459]    [Pg.105]    [Pg.44]    [Pg.45]    [Pg.46]    [Pg.67]    [Pg.123]    [Pg.456]    [Pg.74]    [Pg.1760]    [Pg.544]    [Pg.545]    [Pg.547]    [Pg.346]    [Pg.459]   
See also in sourсe #XX -- [ Pg.194 ]




SEARCH



Pulse counter

© 2024 chempedia.info