Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Polymer instrumentation/measurement

Some viscoelasticity results have been reported for bimodal PDMS [120], using a Rheovibron (an instrument for measuring the dynamic tensile moduli of polymers). Also, measurements have been made on permanent set for PDMS networks in compressive cyclic deformations [121]. There appeared to be less permanent set or "creep" in the case of the bimodal elastomers. This is consistent in a general way with some early results for polyurethane elastomers [122], Specifically, cyclic elongation measurements on unimodal and bimodal networks indicated that the bimodal ones survived many more cycles before the occurrence of fatigue failure. The number of cycles to failure was found to be approximately an order of magnitude higher for the bimodal networks, at the same modulus at 10% deformation [5] ... [Pg.363]

An instrument designed to follow hysteresis losses in polymers by measuring the resistance to the rolling of small balls over the surface of the test piece it can investigate transitions in polymers to as low a temperature as -120 °C. Superseded by modem dynamic mechanical thermal analysis equipment. [Pg.54]

Evaluating the Polymer. The intrinsic viscosity (IV) of the polymer was determined at 25° C. with solutions in benzene or toluene containing 0.1 gram/liter polymer and calculated with the equation IV = (In rjTei)/C. In many cases some gel formation made I V determination for the whole polymer impracticable. Hoekstra viscosities of a number of polymers were measured at 100° C. with the Hoekstra (Wallace) plastometer (42). This instrument provides bulk viscosity data for rubbers on a scale 0 (low viscosity) to 100 (high viscosity). [Pg.137]

Typically, these instruments measure dynamic mechanical responses to sinusoidal input. To characterize the viscoelastic properties of a material, these tests must be repeated over a range of temperatures and frequencies. This is sometimes done at a fixed frequency while the polymer specimen is heated or cooled and... [Pg.92]

A high-pressure DTA assembled by the user is illustrated schematically in Figure 2.15 [27, 28). The pressure in the sample holder unit is increased by an electrical or mechanical pump, using either dimethylsilicone (maximum pressure 600 MPa) or kerosene (maximum pressure 1000 MPa) as the pressure medium. The temperature range of this instrument is 230-670 K at a heating rate of 1-5 K min" . The phase transiton behavior of various polymers is measured using the above type of high-pressure DTA [29-31]. [Pg.23]

Assuming a correct baseline is drawn, the question arises as to how to compare the MMD estimated from a tof mass spectrometer to that obtained from size exclusion chromatography (sec) using a UV or differential refractive index (DRI) detector. A sec is the instrument most commonly used to obtain the MMD for synthetic polymers. The most obvious difference between the two techniques is that the tof detector counts numbers of ions of n-mers while the DRI or UV detectors on sec instruments measures mass concentration of the polymer. The mass concentration of the polymer is proportional to the product of the molecular mass of the -mer and the number of molecules of that n-mer per unit volume. Thus, in the ms we obtain the number MMD while in the sec we normally obtain the mass MMD. Furthermore, the raw signal of each must be corrected for the transformation from time to mass, which is different for each instrument. This transformation has been discussed in detail (58). [Pg.4385]

Specimen preparation is simple, involving compressing a disc of the polymer sample for insertion in the instrument, measurement time is usually less than for other methods, and X-rays interact with elements as such, i.e., the intensity measurement of a constituent element is independent of its state of chemical combination. However, the technique does have some drawbacks, and these are evident in the measurement of cadmium and selenium. For example, absorption effects of other elements present, e.g., the carbon and... [Pg.47]

The mathematical model developed widi the assumption of a simple kinetic scheme and estimated kinetic parameters is instrumental to understand and to predict effects of different operating conditions on the polymer properties. Though the model results differ from the experimental results in terms of the polymer MWDs measured by GPC, the model predicted average polymer properties are in fairly good agreement with the experimental values. To improve the model predictions a better understanding of possible side reactions is most likely needed. Finally, in contrast to earlier literature results [29], no dependence of the pol)mier properties on the reactor pressure and/or pressure reduction rate is found through both the model simulations and experiments. [Pg.23]

The glass transition temperature of the grafted polymer was measured by the Perkin Elmer DSC-IB Instrument. [Pg.121]

Different polymer samples were used to test the performance of the instrument. Measurements with polyethylene glycol showed that on cooling from the melt, the crystallization can be totally suppressed with cooling rates of 1000 K s and above. [Pg.231]

The radiation and temperature dependent mechanical properties of viscoelastic materials (modulus and loss) are of great interest throughout the plastics, polymer, and rubber from initial design to routine production. There are a number of laboratory research instruments are available to determine these properties. All these hardness tests conducted on polymeric materials involve the penetration of the sample under consideration by loaded spheres or other geometric shapes [1]. Most of these tests are to some extent arbitrary because the penetration of an indenter into viscoelastic material increases with time. For example, standard durometer test (the "Shore A") is widely used to measure the static "hardness" or resistance to indentation. However, it does not measure basic material properties, and its results depend on the specimen geometry (it is difficult to make available the identity of the initial position of the devices on cylinder or spherical surfaces while measuring) and test conditions, and some arbitrary time must be selected to compare different materials. [Pg.239]

A recent design of the maximum bubble pressure instrument for measurement of dynamic surface tension allows resolution in the millisecond time frame [119, 120]. This was accomplished by increasing the system volume relative to that of the bubble and by using electric and acoustic sensors to track the bubble formation frequency. Miller and co-workers also assessed the hydrodynamic effects arising at short bubble formation times with experiments on very viscous liquids [121]. They proposed a correction procedure to improve reliability at short times. This technique is applicable to the study of surfactant and polymer adsorption from solution [101, 120]. [Pg.35]

The presence of surface conductance behind the slip plane alters the relationships between the various electrokinetic phenomena [83, 84] further complications arise in solvent mixtures [85]. Surface conductance can have a profound effect on the streaming current and electrophoretic mobility of polymer latices [86, 87]. In order to obtain an accurate interpretation of the electrostatic properties of a suspension, one must perform more than one type of electrokinetic experiment. One novel approach is to measure electrophoretic mobility and dielectric spectroscopy in a single instrument [88]. [Pg.189]

Photomultipliers are used to measure the intensity of the scattered light. The output is compared to that of a second photocell located in the light trap which measures the intensity of the incident beam. In this way the ratio [J q is measured directly with built-in compensation for any variations in the source. When filters are used for measuring depolarization, their effect on the sensitivity of the photomultiplier and its output must also be considered. Instrument calibration can be accomplished using well-characterized polymer solutions, dispersions of colloidal silica, or opalescent glass as standards. [Pg.692]

The realization of sensitive bioanalytical methods for measuring dmg and metaboUte concentrations in plasma and other biological fluids (see Automatic INSTRUMENTATION BlosENSORs) and the development of biocompatible polymers that can be tailor made with a wide range of predictable physical properties (see Prosthetic and biomedical devices) have revolutionized the development of pharmaceuticals (qv). Such bioanalytical techniques permit the characterization of pharmacokinetics, ie, the fate of a dmg in the plasma and body as a function of time. The pharmacokinetics of a dmg encompass absorption from the physiological site, distribution to the various compartments of the body, metaboHsm (if any), and excretion from the body (ADME). Clearance is the rate of removal of a dmg from the body and is the sum of all rates of clearance including metaboHsm, elimination, and excretion. [Pg.224]

The study of acid-base interaction is an important branch of interfacial science. These interactions are widely exploited in several practical applications such as adhesion and adsorption processes. Most of the current studies in this area are based on calorimetric studies or wetting measurements or peel test measurements. While these studies have been instrumental in the understanding of these interfacial interactions, to a certain extent the interpretation of the results of these studies has been largely empirical. The recent advances in the theory and experiments of contact mechanics could be potentially employed to better understand and measure the molecular level acid-base interactions. One of the following two experimental procedures could be utilized (1) Polymers with different levels of acidic and basic chemical constitution can be coated on to elastomeric caps, as described in Section 4.2.1, and the adhesion between these layers can be measured using the JKR technique and Eqs. 11 or 30 as appropriate. For example, poly(p-amino styrene) and poly(p-hydroxy carbonyl styrene) can be coated on to PDMS-ox, and be used as acidic and basic surfaces, respectively, to study the acid-base interactions. (2) Another approach is to graft acidic or basic macromers onto a weakly crosslinked polyisoprene or polybutadiene elastomeric networks, and use these elastomeric networks in the JKR studies as described in Section 4.2.1. [Pg.134]


See other pages where Polymer instrumentation/measurement is mentioned: [Pg.73]    [Pg.14]    [Pg.306]    [Pg.59]    [Pg.17]    [Pg.66]    [Pg.109]    [Pg.28]    [Pg.359]    [Pg.563]    [Pg.163]    [Pg.287]    [Pg.359]    [Pg.617]    [Pg.46]    [Pg.479]    [Pg.236]    [Pg.733]    [Pg.187]    [Pg.388]    [Pg.435]    [Pg.184]    [Pg.112]    [Pg.39]    [Pg.128]    [Pg.33]    [Pg.198]    [Pg.63]    [Pg.552]    [Pg.330]   
See also in sourсe #XX -- [ Pg.315 , Pg.326 , Pg.327 , Pg.328 , Pg.329 , Pg.330 , Pg.331 , Pg.332 , Pg.333 , Pg.334 , Pg.335 , Pg.336 , Pg.337 , Pg.338 , Pg.339 , Pg.340 ]




SEARCH



Instrumentation measurement

Instruments measurements

Instruments, measuring measures

Polymer measurement

Polymers instrumentation

© 2024 chempedia.info