Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Modes of detection

To determine moderate amounts of Cr(III) and Cr(VI) in samples that have both oxidation states present, Cr(VI) is analyzed by direct titration in one sample, and the total chromium is found in a second sample after oxidation of the Cr(III). The Cr(III) concentration is determined as the difference. Trace quantities of Cr(VI) in Cr(III) compounds can be detected and analyzed by (3)-diphenylcarbazide. Trace quantities of Cr(III) in Cr(VI) may be detected and analyzed either photometrically (102) or by ion chromatography using various modes of detection (103). [Pg.141]

Figure 8.19 Schematic diagram of the combination of multilayers (decreasing polarity) foi OPLC with different types of development (circular and anticircular ) and modes of detection (off-line and on-line). Figure 8.19 Schematic diagram of the combination of multilayers (decreasing polarity) foi OPLC with different types of development (circular and anticircular ) and modes of detection (off-line and on-line).
The TID design proposed Patterson consists of an alkali metal doped cerwlc cylinder, containing an embedded heater surrounded by a cylindrical collector electrode [100]. The ceramic thermionic emitter is biased at a negative potential with respect to the collector electrode, and it is heated to a surface temperature of 400-800 C, depending on the mode of detection. The response of the detector to different elements depends on the electronic work function of the thermionic surface (i.e., the... [Pg.652]

Figure 22 Pulse sequence of the HMBC-RELAY experiment. Filled and open bars represent 90° and 180° pulses, respectively. All other phases are set as x, excepted otherwise stated. A two-phase cycle x, —x is used for the pulse phases (j>, and Figure 22 Pulse sequence of the HMBC-RELAY experiment. Filled and open bars represent 90° and 180° pulses, respectively. All other phases are set as x, excepted otherwise stated. A two-phase cycle x, —x is used for the pulse phases (j>, and <p2 and the receiver phase. In order to separate the 2JCH and the nJCn spectra, two FIDs have to be acquired for each tn increment with the phase </)n set as x, — x and — x, x, respectively (interleaved mode of detection) and have to be stored separately. By using a composite 90°x — 180°y — 90°x pulse instead of a single 180° x H pulse, artefacts arising from misadjusted H pulse lengths are suppressed. The delays are calculated according to t/2 = [0.25/Vch]. 8 = [0.25/3Jhh] and A = [O.S/nJCH], The, 3C chemical shift evolution delay t, must be equal for both evolution periods.
In order to determine whether compounds identified in the primary HTS screen are specific, a counterscreen is required to identify and eliminate false positives that will arise in the primary screen. For protein—protein interaction screens, it is preferable to test an unrelated protein pair that uses the same mode of detection. For our purposes, we adapted a previously described TR-FRET assay that monitors the interaction between bacterial Staphylococcus aureus Dnal and phage protein 77ORF104 (Liu et al., 2004). [Pg.313]

For the cationic surfactants, the available HPLC detection methods involve direct UV (for cationics with chromophores, such as benzylalkyl-dimethyl ammonium salts) or for compounds that lack UV absorbance, indirect photometry in conjunction with a post-column addition of bromophenol blue or other anionic dye [49], refractive index [50,51], conductivity detection [47,52] and fluorescence combined with postcolumn addition of the ion-pair [53] were used. These modes of detection, limited to isocratic elution, are not totally satisfactory for the separation of quaternary compounds with a wide range of molecular weights. Thus, to overcome the limitation of other detection systems, the ELS detector has been introduced as a universal detector compatible with gradient elution [45]. [Pg.126]

In frequency-domain FLIM, the optics and detection system (MCP image intensifier and slow scan CCD camera) are similar to that of time-domain FLIM, except for the light source, which consists of a CW laser and an acousto-optical modulator instead of a pulsed laser. The principle of lifetime measurement is the same as that described in Chapter 6 (Section 6.2.3.1). The phase shift and modulation depth are measured relative to a known fluorescence standard or to scattering of the excitation light. There are two possible modes of detection heterodyne and homodyne detection. [Pg.361]

In the homodyne mode of detection, the modulation frequency of the excitation light is the same as that of the image intensifier. An example of data is shown in Box 11.1. [Pg.361]

CE is frequently compared to two of the mainstream techniques in the protein laboratory HPLC and gel electrophoresis. The comparison to HPLC focuses on the instrumentation format and modes of detection, whereas gel electrophoresis shares its separation principles with CE. CE exploits the same molecular differences of the sample components to achieve separation as slab gel... [Pg.163]

Fluorescence detection offers the possibility of high sensitivity and, in the case of complex samples, improved selectivity. However, this mode of detection requires that the analyte exhibit native fluorescence or contain a group to which a fluorophore can be attached by chemical derivatization. Because only tryptophan and tyrosine exhibit significant native fluorescence, fluorescence detection of proteins usually requires derivatization. [Pg.173]

A majority of traditional NIR measurements are made on solid materials and these involve reflectance measurements, notably via diffuse reflectance. Likewise, in the mid-IR not all spectral measurements involve the transmission of radiation. Such measurements include internal reflectance (also known as attenuated total reflectance, ATR), external reflectance (front surface, mirror -style or specular reflectance), bulk diffuse reflectance (less common in the mid-IR compared to NIR), and photoacoustic determinations. Photoacoustic detection has been applied to trace-level gas measurements and commercial instruments are available based on this mode of detection. It is important to note that the photoacoustic spectrum is a direct measurement of infrared absorption. While most infrared spectra are either directly or indirectly correlated... [Pg.162]

The most important advantage of ACE/MS compared to ACE with UV detection is clearly the more specific mode of detection, thus facilitating peak assignment. Additional structural information is available using MS/ MS and MS", respectively. [Pg.351]

Suppressed conductivity detection is the most common mode of detection and differs from the previous approach for the use of an additional device, called suppressor, whose function is to reduce the background conductivity of the eluent prior to the conductivity cell and to increase the signal of the analyte. [Pg.407]

Indirect detection exploits the absorbance of the eluent ion (e.g., NOj , phthalate, benzoate, meth-ylbenzylamine) that should absorb more than the analyte. The eluting analyte exhibits a negative peak. Nevertheless, this mode of detection has limited utilization since the detection limits are often inadequate. [Pg.408]

In addition to an intense source and a well matched monochromator/mirror system, detectors require optimisation. For protein XAS, it is now well established that fluorescence detection is the preferred mode of detection Multi-detector... [Pg.79]

A promising recent development in the study of nitrenium ions has been the introduction of time-resolved vibrational spectroscopy for their characterization. These methods are based on pulsed laser photolysis. However, they employ either time resolved IR (TRIR) or time-resolved resonance Raman (TRRR) spectroscopy as the mode of detection. While these detection techniques are inherently less sensitive than UV-vis absorption, they provide more detailed and readily interpretable spectral information. In fact, it is possible to directly calculate these spectra using relatively fast and inexpensive DFT and MP2 methods. Thus, spectra derived from experiment can be used to validate (or falsify) various computational treatments of nitrenium ion stmctures and reactivity. In contrast, UV-vis spectra do not lend themselves to detailed structural analysis and, moreover, calculating these spectra from first principles is still expensive and highly approximate. [Pg.636]

Both types of detector systems can be used in two modes of applications, i.e., on-line and off-line. In the on-line mode the fractionation and detection systems are directly coupM. It is possible to use more than one detector either in series or in parallel. Examples of this mode of detection will be given in the section dealing with combined techniques. The off-line mode involves the collection of the fractions with subsequent determination of the constituents. The advantage of this approach is that further sample pretreatment procedures could be applied where necessary before the constituents are detected. In addition, quantitative estimates of the recoveries can be made. [Pg.163]

Zinc-5,10,15,20-tetraphenylporphyrin (ZnTPP) has been used as a coating material in ammonia sensors by immobilizing it on the surface of silicone rubber. Absorbance and fluorescence emission were the modes of detection. A spectral change is caused by the coordination of NH3 molecules to the Zn11 ion in the immobilized metalloporphyrins. Sensing films made from the ZnTPP immobilized in silicone rubber were found to be the most sensitive for NH3 sensing (20). [Pg.263]

The determination of diazinon in foods is important because this chemical is used as a pesticide on plant crops and, at least in some cases, in pesticide dips for the control of parasitic infestations in animals (Brown et al. 1987 Miyahara et al. 1992). Because animals are exposed to this compound, both via pesticide dips and by ingestion of crops to which diazinon has been applied, some methods have been reported for animal products. The majority of methods, however, deal with the determination of residues in plant products. Most of the analytical methods found that describe the extraction from, and determination of, diazinon residues in various crops (plant materials) were developed as part of multiresidue methods. They are based on homogenization of the sample with an organic solvent (polar or non-polar) the isolation of the residues from this initial extract and, usually, some additional cleanup prior to the analysis of the extract by GC. The most common non-MS modes of detection exploit the... [Pg.176]

Ultraviolet spectrophotometric detection can be carried out at wavelengths in the range 220-254 nm (517, 518, 520, 524, 525, 526). Fluorometric detection, which is particularly suitable for azaperol and carazolol, confers the advantages of selectivity and sensitivity. This mode of detection has been employed for the determination of carazolol residues in serum and plasma, using excitation and emission wavelengths at 330 and 360 nm, respectively (522). Fluorometric detection has also been applied to monitor carazolol and azaperol residues in swine kidney with excitation and emission wavelengths of 246 and 351 nm, respectively (524). [Pg.1102]

A variety of sequences exist, which differ with respect to the detected interaction ( J, or Jx ) and the mode of detection ( C or H detected, magnitude or phased mode, phase cycling or gradients for coherence selection). In view of the reduced sensitivity of heteronudear experiments with respect to homonuclear COSY experiments and the steadily decreasing sample amounts submitted for NMR experiments, there is no doubt that the inverse ( H) detected, gradient enhanced experiments are currently the best methods to apply. However on older type spectrometers, not equipped for inverse detection the old-fashioned direct C detected experiments are still in use. [Pg.67]

The signal-to-noise ratio of rows and columns obtained with 2D experiments vary. They are often different for rows and columns and depend on the mode of detection ( H/ C-detected), on the kind of experiment and its experimental parameters and on the sample amount. [Pg.171]

These detectors represent techniques that are widely used for trace analysis. Both modes of detection mentioned above can be used with capillary columns. [Pg.38]

Besides the spectrophotometric detectors seen in HPLC based on absorbance or fluorescence of UV/Vis radiation, another type of detector based on electrolyte conductivity can be used. This mode of detection measures conductance of the mobile phase, which is rich in ionic species (Fig. 4.6). The difficulty is to recognise in the total signal the part due to ions or ionic substances present in the sample at very low concentrations. In a mobile phase loaded with buffers with a high conductance, the contribution of ions due to the analyte is small. In order to do a direct measurement, the ionic loading of the mobile phase has to be as low as possible and the cell requires strict temperature control (0.01 °C) because of the high dependence of conductance on temperature. Furthermore, the eluting ions should have a small ionic conductivity and a large affinity for the stationary phase. [Pg.70]

HPLC analysis of polycyclic aromatic hydrocarbons (PAH) in drinking water is one of the current and classical applications of fluorescence. In this case, the detector contains a fluorescence flow cell placed after the chromatographic column. This mode of detection is specifically adapted to obtain threshold measurements imposed by legislation. The same process allows the measurement of aflatoxins (Fig. 12.11) and many other organic compounds (such as adrenaline, quinine, steroids and vitamins). [Pg.230]

The most popular electrochemical detectors to date have been based on the amperometric conversion of analyte in a cross-flow thin-layer cell. The basic functioning of this mode of detection is depicted schematically in Figure 27.2. [Pg.819]

The second approach (end-column detection) is best suited for capillaries of 25 pm or less [49]. This mode of detection is illustrated in Figure 27.20. In this case, the electrode is placed at the end of the capillary (but not inside) and no coupler is employed. A 70-cm, 5-pm-i.d. capillary filled with a zwitterionic... [Pg.848]


See other pages where Modes of detection is mentioned: [Pg.396]    [Pg.200]    [Pg.274]    [Pg.349]    [Pg.353]    [Pg.130]    [Pg.328]    [Pg.139]    [Pg.165]    [Pg.245]    [Pg.735]    [Pg.57]    [Pg.298]    [Pg.515]    [Pg.172]    [Pg.199]    [Pg.241]    [Pg.634]    [Pg.177]    [Pg.91]    [Pg.44]    [Pg.44]    [Pg.69]   


SEARCH



Application Modes of Conductivity Detection

Detection modes

© 2024 chempedia.info