Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Ion collector

Stop must be used to prevent photons from the plasma flame reaching the ion collector, which would produce a spurious high background signal. [Pg.89]

After the skimmer, the ions must be prepared for mass analysis, and electronic lenses in front of the analyzer are used to adjust ion velocities and flight paths. The skimmer can be considered to be the end of the interface region stretching from the end of the plasma flame. Some sort of light stop must be used to prevent emitted light from the plasma reaching the ion collector in the mass analyzer (Figure 14.2). [Pg.95]

An AutoSpec-TOF mass spectrometer has a magnetic sector and an electron multiplier ion detector for carrying out one type of mass spectrometry plus a TOF analyzer with a microchannel plate multipoint ion collector for another type of mass spectrometry. Either analyzer can be used separately, or the two can be run in tandem (Figure 20.4). [Pg.154]

Alternatively, ions of any one selected m/z value can be chosen by holding the magnetic field steady at the correct strength required to pass only the desired ions any other ions are lost to the walls of the instrument. The selected ions pass through the gas cell and are detected in the singlepoint ion collector. If there is a pressure of a neutral gas such as argon or helium in the gas cell, then ion-molecule collisions occur, with decomposition of some of the selected incident ions. This is the MS/MS mode. However, without the orthogonal TOF section, since there is no further separation by m/z value, the new ions produced in the gas cell would not be separated into individual m/z values before they reached the detector. Before the MS/MS mode can be used, the instrument must be operated in its hybrid state, as discussed below. [Pg.159]

A fuller description of the microchannel plate is presented in Chapter 30. Briefly, ions traveling down the flight tube of a TOF instrument are separated in time. As each m/z collection of ions arrives at the collector, it may be spread over a small area of space (Figure 27.3). Therefore, so as not to lose ions, rather than have a single-point ion collector, the collector is composed of an array of miniature electron multipliers (microchannels), which are all connected to one electrified plate, so, no matter where an ion of any one m/z value hits the front of the array, its arrival is recorded. The microchannel plate collector could be crudely compared to a satellite TV dish receiver in that radio waves of the same frequency but spread over an area are all collected and recorded at the same time of course, the multichannel plate records the arrival of ions not radio waves. [Pg.197]

Ion detectors can be separated into two classes those that detect the arrival of all ions sequentially at one point (point ion collector) and those that detect the arrival of all ions simultaneously along a plane (array collector). This chapter discusses point collectors (detectors), while Chapter 29 focuses on array collectors (detectors). [Pg.201]

All mass spectrometers analyze ions for their mass-to-charge ratios (m/z values) by separating the individual m/z values and then recording the numbers (abundance) of ions at each m/z value to give a mass spectrum. Quadrupoles allow ions of different m/z values to pass sequentially e.g., ions at m/z 100, 101, 102 will pass one after the other through the quadrupole assembly so that first m/z 100 is passed, then 101, then 102 (or vice versa), and so on. Therefore, the ion collector (or detector) at the end of the quadrupole assembly needs only to cover one point or focus for a whole spectrum to be scanned over a period of time (Figure 28.1a). This type of point detector records ion arrivals in a time domain, not a spatial one. [Pg.201]

Three main types point ion collectors are in use for quadrupole, magnetic-sector, and TOF instruments, and they are discussed here. The multichannel plate collector (or time-to-digital converter)... [Pg.201]

A scintillator, sometimes known as the Daly detector, is an ion collector that is especially useful for studies on metastable ions. The principle of operation is illustrated in Figure 28.4. As with the first dynode of an electron multiplier, the arrival of a fast ion causes electrons to be emitted, and they are accelerated toward a second dynode. In this case, the dynode consists of a substance (a scintillator) that emits photons (light). The emitted light is detected by a commercial photon... [Pg.203]

An ion beam causes secondary electrons to be ejected from a metal surface. These secondaries can be measured as an electric current directly through a Faraday cup or indirectly after amplification, as with an electron multiplier or a scintillation device. These ion collectors are located at a fixed point in a mass spectrometer, and all ions are focused on that point — hence the name, point ion collector. In all cases, the resultant flow of an electric current is used to drive some form of recorder or is passed to an information storage device (data system). [Pg.204]

An ion beam containing just two types of ion of m/z values 100 and 101 dispersed in space on passing through a magnetic field. After dispersal, ions of individual m/z value 100 or lOI are focused at points close to the entries of two elements of an array collector. Each element of the array is a point ion collector. [Pg.208]

An array ion collector (detector) consists of a large number of miniature electron multiplier elements arranged side by side along a plane. Point ion collectors gather and detect ions sequentially (all ions are focused at one point one after another), but array collectors gather and detect all ions simultaneously (all ions are focused onto the array elements at the same time). Array detectors are particularly useful for situations in which ionization occurs within a very short space of time, as with some ionization sources, or in which only trace quantities of a substance are available. For these very short time scales, only the array collector can measure a whole spectrum or part of a spectrum satisfactorily in the time available. [Pg.210]

In modem mass spectrometry, ion collectors (detectors) are generally based on the electron multiplier and can be separated into two classes those that detect the arrival of all ions sequentially at a point (a single-point ion collector) and those that detect the arrival of all ions simultaneously (an array or multipoint collector). This chapter compares the uses of single- and multipoint ion collectors. For more detailed discussions of their construction and operation, see Chapter 28, Point Ion Collectors (Detectors), and Chapter 29, Array Collectors (Detectors). In some forms of mass spectrometry, other methods of ion detection can be used, as with ion cyclotron instmments, but these are not considered here. [Pg.211]

Ions of different m/z values pass sequentially in time through the quadrupole mass filter to reach an in-line, single-point ion collector. [Pg.212]

By adjusting the magnetic field, the dispersed ion beam in Figure 30.2 can be moved up or down so that ions of specific m/z values can be focused at a point ion collector. [Pg.212]

Other types of mass spectrometer can use point, array, or both types of ion detection. Ion trap mass spectrometers can detect ions sequentially or simultaneously and in some cases, as with ion cyclotron resonance (ICR), may not use a formal electron multiplier type of ion collector at all the ions can be detected by their different electric field frequencies in flight. [Pg.212]

A second use of arrays arises in the detection of trace components of material introduced into a mass spectrometer. For such very small quantities, it may well be that, by the time a scan has been carried out by a mass spectrometer with a point ion collector, the tiny amount of substance may have disappeared before the scan has been completed. An array collector overcomes this problem. Often, the problem of detecting trace amounts of a substance using a point ion collector is overcome by measuring not the whole mass spectrum but only one characteristic m/z value (single ion monitoring or single ion detection). However, unlike array detection, this single-ion detection method does not provide the whole spectrum, and an identification based on only one m/z value may well be open to misinterpretation and error. [Pg.216]

Alternatively, the electric focusing potential E can be changed, but this method needs another ion collector sited at the electric-sector focus, and it must be a collector that can be raised out of the ion beam when the mass of the ion being examined is required. This arrangement is not convenient. A better solution is obtained by linked scanning of the E/V voltages (see later discussion). [Pg.239]


See other pages where Ion collector is mentioned: [Pg.52]    [Pg.68]    [Pg.89]    [Pg.154]    [Pg.154]    [Pg.154]    [Pg.158]    [Pg.159]    [Pg.160]    [Pg.160]    [Pg.195]    [Pg.197]    [Pg.197]    [Pg.201]    [Pg.201]    [Pg.202]    [Pg.203]    [Pg.205]    [Pg.205]    [Pg.206]    [Pg.206]    [Pg.209]    [Pg.210]    [Pg.212]    [Pg.219]    [Pg.237]    [Pg.237]    [Pg.238]    [Pg.239]   
See also in sourсe #XX -- [ Pg.2 ]

See also in sourсe #XX -- [ Pg.302 ]




SEARCH



Collector

Multiple ion collector ICP

Multipoint Collectors (Detectors) of Ions Arrays and MicroChannel Plates

Point Ion Collectors (Detectors)

Point ion collectors

The Basic Elements in Lithium-ion Batteries Electrodes, Electrolytes and Collectors

The ion collectors

Timing of Electrical Pulses Resulting from Ion Arrivals at the MicroChannel Plate Collector

Types of Point Ion Collector

© 2024 chempedia.info