Point ion detectors

The major advantage of array detectors over point ion detectors lies in their ability to measure a range of m/z values and the corresponding ion abundances all at one time, rather than sequentially. For example, suppose it takes 10 msec to measure one m/z value and the associated number of ions (abundance). To measure 100 such ions sequentially with a point ion detector would necessitate 1000 msec (1 sec) for the array detector, the time is still 10 msec because all ions arrive at the same time. Therefore, when it is important to be able to measure a range of ion m/z values in a short space of time, the array detector is advantageous. [Pg.209]

There are two common occasions when rapid measurement is preferable. The first is with ionization sources using laser desorption or radionuclides. A pulse of ions is produced in a very short interval of time, often of the order of a few nanoseconds. If the mass spectrometer takes 1 sec to attempt to scan the range of ions produced, then clearly there will be no ions left by the time the scan has completed more than a few nanoseconds (ion traps excluded). If a point ion detector were to be used for this type of pulsed ionization, then after the beginning of the scan no more ions would reach the collector because there would not be any left The array collector overcomes this difficulty by detecting the ions produced all at the same instant. [Pg.209]

There is potential confusion in the use of the word array in mass spectrometry. Historically, array has been used to describe an assemblage of small single-point ion detectors (elements), each of which acts as a separate ion current generator. Thus, arrival of ions in one of the array elements generates an ion current specifically from that element. An ion of any given m/z value is collected by one of the elements of the array. An ion of different m/z value is collected by another element. Ions of different m/z value are dispersed in space over the face of the array, and the ions are detected by m/z value at different elements (Figure 30.4). [Pg.213]

Ions of a given m/z value are collected at one of the small point ion detectors ions of larger or smaller m/z values are collected at other point collectors placed on either side. [Pg.408]

Ion detectors can be separated into two classes those that detect the arrival of all ions sequentially at one point (point ion collector) and those that detect the arrival of all ions simultaneously along a plane (array collector). This chapter discusses point collectors (detectors), while Chapter 29 focuses on array collectors (detectors). [Pg.201]

An array ion collector (detector) consists of a large number of miniature electron multiplier elements arranged side by side along a plane. Point ion collectors gather and detect ions sequentially (all ions are focused at one point one after another), but array collectors gather and detect all ions simultaneously (all ions are focused onto the array elements at the same time). Array detectors are particularly useful for situations in which ionization occurs within a very short space of time, as with some ionization sources, or in which only trace quantities of a substance are available. For these very short time scales, only the array collector can measure a whole spectrum or part of a spectrum satisfactorily in the time available. [Pg.210]

In modem mass spectrometry, ion collectors (detectors) are generally based on the electron multiplier and can be separated into two classes those that detect the arrival of all ions sequentially at a point (a single-point ion collector) and those that detect the arrival of all ions simultaneously (an array or multipoint collector). This chapter compares the uses of single- and multipoint ion collectors. For more detailed discussions of their construction and operation, see Chapter 28, Point Ion Collectors (Detectors), and Chapter 29, Array Collectors (Detectors). In some forms of mass spectrometry, other methods of ion detection can be used, as with ion cyclotron instmments, but these are not considered here. [Pg.211]

Each element of an array detector is essentially a small electron multiplier, as with the point ion collector, but much smaller and often shaped either as a narrow linear tube or as somewhat like a snail shell. [Pg.409]

MID— multiple ion detector min—minute mol— mole m.p.—melting point MS— mass spectrometry m/z—mass to charge ratio... [Pg.1683]

Alternatively, ions of any one selected m/z value can be chosen by holding the magnetic field steady at the correct strength required to pass only the desired ions any other ions are lost to the walls of the instrument. The selected ions pass through the gas cell and are detected in the single-point ion collector. If there is a pressure of a neutral gas such as argon or helium in the gas cell, then ion-molecule collisions occur, with decomposition of some of the selected incident ions. This is the MS/MS mode. However, without the orthogonal TOF section, since there is no further separation by m/z value, the new ions produced in the gas cell would not be separated into individual m/z values before they reached the detector. Before the MS/MS mode can be used, the instrument must be operated in its hybrid state, as discussed below. [Pg.159]


See other pages where Point ion detectors is mentioned: [Pg.216]    [Pg.137]    [Pg.216]    [Pg.269]    [Pg.216]    [Pg.137]    [Pg.216]    [Pg.269]    [Pg.1334]    [Pg.154]    [Pg.158]    [Pg.201]    [Pg.203]    [Pg.205]    [Pg.206]    [Pg.209]    [Pg.213]    [Pg.279]    [Pg.322]    [Pg.408]    [Pg.180]    [Pg.14]    [Pg.273]    [Pg.191]    [Pg.323]    [Pg.154]    [Pg.158]    [Pg.201]    [Pg.203]    [Pg.205]    [Pg.209]   