Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrumentation factor correction

The ARC is a test instrument that is able to provide information on the runaway behavior of substances and reactions very quickly. Several publications are available regarding the applicability of the results of ARC tests [77, 126,127-132]. Most of the disadvantages of the ARC discussed are due to the high phi-factor of the equipment relative to plant operating conditions. For example, the phi-factor correction assumes that no additional or different reactions occur at higher temperatures that might be reached under realistic... [Pg.75]

Recalibration of the instrument response function reduces or eliminates most of the instrumental factors that lead to relative intensity variations over time. For example, a luminescent standard could be used at the beginning of each session as described in Section 10.3.3. Use of the same standard and correction procedure during qualification could establish the true value of one or more peak ratios for future reference. Table 10.9 shows results for this approach applied to the example of calcium ascorbate. The ratio of the 767- and 1587 cm" peak intensities was monitored after calibration of the response function with a luminescent standard. The standard deviations listed in Table 10.9 for the 767/1582 peak height ratio provide indications of the reproducibility of the response correction and sample spectra. [Pg.284]

It cannot always be assumed that Beer s law will apply, that is, that a linear plot of absorbance versus concentration will occur. Deviations from Beer s law occur as the result of chemical and instrumental factors. Most deviations from Beer s law are really only apparent deviations because if the factors causing nonlinearity are accounted for, the true or corrected absorbance-versus-concentration curve will be linear. True deviations from Beer s law will occur when the concentration is so high that the index of refraction of the solution is changed from that of the blank. A similar situation would apply for mixtures of organic solvents with water, and so the blank solvent composition should closely match that of the sample. The solvent may also have an effect on the absorptivity of the analyte. [Pg.503]

It is clear from the above that extreme care must be exercised in the characterization and rheological eva-luation of concentrated emulsions. Few, if any, com-mercial viscometers are designed to give reliable results for nonNew-tonian fluids. Not only are modifications of the hardware often called for, but also the software of automated instra-ments is generally incapable of dealing with yield-stress fluids, end effects, and wall slip. For example, to correct for end effects, it will not do to use a calibration or instrument factor for any but Newtonian fluids. Unfortunately, there are no shortcuts in this field ... [Pg.271]

Requirements for the establishment of the LODs from this study were that the instrument could produce a spectrum with peaks that were at least 3 times greater than the surrounding noise, that could be visually analyzed and attributed to the stimulant being sampled, and that the instrument could correctly match to the onboard library with a statistically significant difference from the next closest match. In this case, the library matching was the limiting factor, while the instrument produced spectra that could be visually matched to the simulants at even lower levels. [Pg.82]

Modulated beam experiments were carried out using a single sinusoidal frequency of 2.5 Hz, and the scattered Br product was detected for surface temperatures in the range 1050-1330 K. The temperature dependence of the phase shift of the Br product vector is shown in fig. 5 these phase shifts have not been corrected for the constant phase shift due to all instrumental factors, and this latter quantity was determined in the following manner ... [Pg.659]

Fluorescence is an extremely sensitive technique but it is not suitable as a general method to estimate natural DOC content due to the reason that it is impossible to find a reference material that would be common for all different natural waters. Characteristic for different fluorescence studies of NOM/DOM is that they may occasionally be somewhat surprising, contradictory, or laboriously explicable. The main reason for this incoherence is that fluorescence measurements are affected by many environmental factors, e.g., type of solution, pH, ionic strength, temperature, redox potential of the medium, and interactions with metal ions and organics. Several corrections are required to obtain a reliable and comparable spectrum, e.g., instrumental factors, Raman water peak, scattering effects (primary and secondary inner filter effects [31,32]), arbitrary fluorescence units should be standardized (dihydrate of quinine sulfate), etc. [Pg.441]

In fact, the incorrect choice of this component could result in a failure to observe signals that are otherwise easily visible. The strong dependence of the sensitivity of photomultipliers on the wavelength, to which is added, in lesser measure, the non-linear response of the emission monochromator, is the main cause of the fact that, as we will see in the next section, experimentally-obtained emission spectra must be corrected , i.e., they must be multiplied by an instrumental factor in order to be compared with spectra obtained under other experimental conditions. [Pg.104]

If the spectrofluorimeter is working in ratio mode (and, thus, corrects for the variations of emission intensity of the lamp), the only two instrumental factors that must be considered for the correction of the spectra are the different response of... [Pg.106]

As already mentioned G is purely an instrumental factor which must be known in order to perform correct FA measurements. Its determination is quite simple thanks to the fact that when excitation is performed with horizontally polarized light (instead than vertical) the vertical and horizontal component of the emitted light are expected to be equal, independently on the properties of the sample. A difference between the measured intensities is henee, in this exeitation condition, due exclusively to instrumental effect and it can be used to obtain the G factor. [Pg.158]

The SAXS method allows measure the diameter of nano-size structural defects and pores. The small-angle scattering of the cellulose sample was recorded, and intensity was measured as a function of the diffraction angle. To correct the intensity, /, the background scattering was subtracted and the experimental intensity was corrected on instrumental factor, absorption of X-ray and other factors. Porod s law / = Kq, was used for visualization of the small-angle X-ray scattering (Porod, 1951, 1952]. Linearization of the experimental curves was performed as + kq or Y = Yg + kX (Fig. 7.13], where Y = and... [Pg.224]

Most infrared monitoring systems or instruments provide special filters that can be used to avoid the negative effects of atmospheric attenuation of infrared data. However the plant user must recognize the specific factors that will affect the accuracy of the infrared data and apply the correct filters or other signal conditioning required negating that specific attenuating factor or factors. [Pg.799]

Requirements for standards used In macro- and microspectrofluorometry differ, depending on whether they are used for Instrument calibration, standardization, or assessment of method accuracy. Specific examples are given of standards for quantum yield, number of quanta, and decay time, and for calibration of Instrument parameters. Including wavelength, spectral responslvlty (determining correction factors for luminescence spectra), stability, and linearity. Differences In requirements for macro- and micro-standards are considered, and specific materials used for each are compared. Pure compounds and matrix-matched standards are listed for standardization and assessment of method accuracy, and existing Standard Reference Materials are discussed. [Pg.98]

Procedures for determining the spectral responslvlty or correction factors In equation 2 are based on radiance or Irradlance standards, calibrated source-monochromator combinations, and an accepted standard. The easiest measurement procedure for determining corrected emission spectra Is to use a well-characterized standard and obtain an Instrumental response function, as described by equation 3 (17). In this case, quinine sulfate dlhydrate has been extensively studied and Issued as a National Bureau of Standards (NBS) Standard Reference Material (SRM). [Pg.102]


See other pages where Instrumentation factor correction is mentioned: [Pg.739]    [Pg.135]    [Pg.150]    [Pg.337]    [Pg.286]    [Pg.100]    [Pg.703]    [Pg.141]    [Pg.172]    [Pg.168]    [Pg.321]    [Pg.221]    [Pg.406]    [Pg.297]    [Pg.568]    [Pg.912]    [Pg.364]    [Pg.1989]    [Pg.94]    [Pg.414]    [Pg.1002]    [Pg.97]    [Pg.98]    [Pg.106]    [Pg.33]    [Pg.20]    [Pg.453]    [Pg.605]    [Pg.205]    [Pg.1125]    [Pg.58]    [Pg.955]    [Pg.229]    [Pg.667]    [Pg.220]   


SEARCH



Correction factors

Instrument factor

© 2024 chempedia.info