Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Experimental data instrumentation

There are two fimdamental types of spectroscopic studies absorption and emission. In absorption spectroscopy an atom or molecule in a low-lying electronic state, usually the ground state, absorbs a photon to go to a higher state. In emission spectroscopy the atom or molecule is produced in a higher electronic state by some excitation process, and emits a photon in going to a lower state. In this section we will consider the traditional instrumentation for studying the resulting spectra. They define the quantities measured and set the standard for experimental data to be considered. [Pg.1120]

Specinfo, from Chemical Concepts, is a factual database information system for spectroscopic data with more than 660000 digital spectra of 150000 associated structures [24], The database covers nuclear magnetic resonance spectra ( H-, C-, N-, O-, F-, P-NMR), infrared spectra (IR), and mass spectra (MS). In addition, experimental conditions (instrument, solvent, temperature), coupling constants, relaxation time, and bibliographic data are included. The data is cross-linked to CAS Registry, Beilstein, and NUMERIGUIDE. [Pg.258]

Many modern instruments used in the analytical laboratory are interfaced with a computer and a printer provides a permanent record of the experimental data and the final result may even be given. This printout should be permanently attached to the observations page of the laboratory record book, and it should be regarded as normal practice to perform a rough calculation to confirm that the printed result is of the right order. [Pg.72]

Let us suppose that we have a particle counting instrument which sorts and counts the number of particles at a given particle size. The experimental data that we collect are ... [Pg.229]

For probing the nature of the acid sites by X-ray photoelectron spectroscopy (XFS), the samples were evacuated before gaseous pyridine was adsorbed. Excess pyridine was desorbed at 1S0°C, and then samples were pressed onto a sample stub imder Nj and loaded into the SCIENTA ESCA-300 instrument without exposure to air. Sample charging was minimized by using a Qood gun while acquiring the experimental data. [Pg.602]

Although the condensation of phenol with formaldehyde has been known for more than 100 years, it is only recently that the reaction could be studied in detail. Recent developments in analytical instrumentation like GC, GPC, HPLC, IR spectroscopy and NMR spectroscopy have made it possible for the intermediates involved in such reactions to be characterized and determined (1.-6). In addition, high speed computers can now be used to simulate the complicated multi-component, multi-path kinetic schemes involved in phenol-formaldehyde reactions (6-27) and optimization routines can be used in conjunction with computer-based models for phenol-formaldehyde reactions to estimate, from experimental data, reaction rates for the various processes involved. The combined use of precise analytical data and of computer-based techniques to analyze such data has been very fruitful. [Pg.288]

Although Equation (4) is conceptually correct, the application to experimental data should be undertaken cautiously, especially when an arbitrary baseline is drawn to extract the area under the DSC melting peak. The problems and inaccuracy of the calculated crystallinities associated with arbitrary baselines have been pointed out by Gray [36] and more recently by Mathot et al. [37,64—67]. The most accurate value requires one to obtain experimentally the variation of the heat capacity during melting (Cp(T)) [37]. However, heat flow (d(/) values can yield accurate crystallinities if the primary heat flow data are devoid of instrumental curvature. In addition, the temperature dependence of the heat of fusion of the pure crystalline phase (AHc) and pure amorphous phase (AHa) are required. For many polymers these data can be found via their heat capacity functions (ATHAS data bank [68]). The melt is then linearly extrapolated and its temperature dependence identified with that of AHa. The general expression of the variation of Cp with temperature is... [Pg.261]

Sections on matrix algebra, analytic geometry, experimental design, instrument and system calibration, noise, derivatives and their use in data analysis, linearity and nonlinearity are described. Collaborative laboratory studies, using ANOVA, testing for systematic error, ranking tests for collaborative studies, and efficient comparison of two analytical methods are included. Discussion on topics such as the limitations in analytical accuracy and brief introductions to the statistics of spectral searches and the chemometrics of imaging spectroscopy are included. [Pg.556]

Of course Phillips is correct to point out, as he does here, that even decorated scientists can unscientifically foreclose areas of investigation simply because of their own prejudices. Yet what explains the reluctance of scientists to affirm clairvoyant chemistry What Phillips fails to take into account is the rhetorical persuasiveness of the mass spectroscope. Science persuades on the basis of instruments and the visual displays they help create and on the reproducibility of experimental data from one similarly equipped laboratory to another. Individualized direct perception simply cannot match the rhetorical power of the modern laboratory. [Pg.94]

In principle, the three isotope method may be widely applied to new isotope systems such as Mg, Ca, Cr, Fe, Zn, Se, and Mo. Unlike isotopic analysis of purified oxygen, however, isotopic analysis of metals that have been separated from complex matrices commonly involves measurement of several isotopic ratios to monitor potential isobars, evaluate the internal consistency of the data through comparison with mass-dependent fractionation relations (e.g., Eqn. 8 above), or use in double-spike corrections for instrumental mass bias (Chapter 4 Albarede and Beard 2004). For experimental data that reflect partial isotopic exchange, their isotopic compositions will not lie along a mass-dependent fractionation line, but will instead lie along a line at high angle to a mass-dependent relation (Fig. 10), which will limit the use of multiple isotopic ratios for isobar corrections, data quality checks, and double-spike corrections. [Pg.17]

Since all electrophoretic mobility values are proportional to the reciprocal viscosity of the buffer, as derived in Chapter 1, the experimental mobility values n must be normalized to the same buffer viscosity to eliminate all other influences on the experimental data besides the association equilibrium. Some commercial capillary zone electrophoresis (CZE) instruments allow the application of a constant pressure to the capillary. With such an instrument the viscosity of the buffer can be determined by injecting a neutral marker into the buffer and then calculating the viscosity from the time that the marker needs to travel through the capillary at a set pressure. During this experiment the high voltage is switched off. [Pg.43]

The variance of the instrumental spreading function, i.e. the spreading factor of monodispersed polymer in a SEC column was determined experimentally with narrow MWD polystyrene standard samples by the method of simultaneous calibration. The dependence of the spreading factor on the retention volume deduced from a simple theoretical approach may be expressed by a formula with four physically meaningful and experimentally determinable parameters. The formula fits the experimental data quite well and the conditions for the appearance of a maximum spreading factor are explicable. [Pg.281]

Today, ultrafast pulsed-laser techniques, high-speed computers, and other sophisticated instrumentation make it possible to measure the time evolutions of reactants, intermediates, transition structures, and products following an abrupt photoactivation of a starting material. Detailed theoretical calculations, experienced judgments based on the literature, and newly accessible femtosecond-domain experimental data providing observed intensities of chemical species versus time can provide insights on the atomic-scale events responsible for overall reaction outcomes. [Pg.903]

With appropriate calibration the complex characteristic impedance at each resonance frequency can be calculated and related to the complex shear modulus, G, of the solution. Extrapolations to zero concentration yield the intrinsic storage and loss moduli [Gr] and [G"], respectively, which are molecular properties. In the viscosity range of 0.5-50 mPa-s, the instrument provides valuable experimental data on dilute solutions of random coil (291), branched (292), and rod-like (293) polymers. The upper limit for shearing frequency for the MLR is 800 Hz. High frequency (20 to 500 K Hz) viscoelastic properties can be measured with another instrument, the high frequency torsional rod apparatus (HFTRA) (294). [Pg.201]

The more the precision of the instrument, and the more the points for the time unit in the acquired profile, the better the result of the fitting of experimental data. For this reason instruments with a low measure error and connectable to a computer for the automatic and continous aquisition of data are very much prefered. The UV-Vis spectrophotometer is by far the most used instrument in chemical kinetics. It has a good sensitivity and a good control of the temperature. It is connected or easily connectable to a computer and is available nearly everywhere. The absorbance has a very low dependence on the temperature so that, in the used temperature range, its variation can be neglected during the VTK experiments. [Pg.711]

Analytical practitioners place great faith in the readings and outputs from their instruments. When unexpected or out of specification results occur, the initial suspicion often falls on the sample, the preparation technique or the analytical standard employed. Rarely is the equipment questioned. Indeed, the whole underpinning of method validation assumes that the analytical equipment used to acquire the experimental data is operating correctly and reliably. [Pg.20]

Fig. 11.5 Measurement of lifetime of anthracene in solution by single photon time correlation technique. Fluorescence decay curve of 8 X10-4 M anthracene in cyclohexane in the absence (A) and presence (B) of 0.41 M CC14. Points experimental data Line best fitting single exponential decay convoluted with instrumental response function (C) Time scale 0.322 nsec per channel. (Ref. 13). Fig. 11.5 Measurement of lifetime of anthracene in solution by single photon time correlation technique. Fluorescence decay curve of 8 X10-4 M anthracene in cyclohexane in the absence (A) and presence (B) of 0.41 M CC14. Points experimental data Line best fitting single exponential decay convoluted with instrumental response function (C) Time scale 0.322 nsec per channel. (Ref. 13).

See other pages where Experimental data instrumentation is mentioned: [Pg.703]    [Pg.703]    [Pg.83]    [Pg.13]    [Pg.51]    [Pg.226]    [Pg.161]    [Pg.51]    [Pg.132]    [Pg.282]    [Pg.41]    [Pg.156]    [Pg.41]    [Pg.295]    [Pg.331]    [Pg.612]    [Pg.447]    [Pg.120]    [Pg.272]    [Pg.182]    [Pg.145]    [Pg.83]    [Pg.10]    [Pg.182]    [Pg.58]    [Pg.110]    [Pg.129]    [Pg.309]    [Pg.5]    [Pg.613]    [Pg.712]    [Pg.721]    [Pg.26]   
See also in sourсe #XX -- [ Pg.139 , Pg.140 ]




SEARCH



Experimental data instrumentation, conditions

Instrumental data

© 2024 chempedia.info