Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Window Detector

The absorption of x-rays influences their detection in two important and obvious ways Absorption on the way to the detecting medium reduces the measured intensity and is undesirable this includes absorption by the detector window. Conversely, absorption within the detecting medium produces the effect to be measured and is desirable. [Pg.44]

The absorption problems for other detectors may be considered under three headings (1) attenuation along the beam path, (2) attenuation by the detector window, (3) absorption by the detecting medium. The results of absorption calculations (1.9) in Table 2 1 show the importance of these problems and suggest ways of dealing with them. [Pg.44]

Element Analyt- ical Line Wavelength, A Crystal Slit Width, in. Path Type of Detector Detector Path Length, cm Detector Window"... [Pg.260]

The ambitious goal of eBuilding is to maximize the comfort of the house owner at a minimum cost level. This will be achieved by using sensors in all sensitive areas, for instance, constant control of the air-conditioning or to cut off cost intensive peaks. Also sensors in smoke-detectors, windows, doors and other appliances will take care of the house security. [Pg.237]

An alternative to derivatizing carbohydrates is the use of indirect photometric detection. In this method, a detectable co-ion in the electrolyte is added to the buffer system generating a steady state absorbance signal in the detector. As the analyte ions migrate in front of the detector window, they displace the detectable co-ion and cause a decrease or negative response in the detector signal. This method provides universal detection of all anions or cations. Since most carbohydrates are not ionized... [Pg.51]

In practice, a radiation source impinges on an entrance slit S (Fig. 14.5) located before the dispersive system. The exit slit is located after the dispersive system, close to the detector window, which selects a narrow bandwidth of the spectrum (AA from 0.2 to 1 nm) and must not to be confused with the width of the exit slit or that of the image at the entrance slit. [Pg.258]

Of the three commonly used X-ray detectors—(1) Geiger counter, (2) scintillation counter, and (3) proportional counter—the latter is used most frequently for electron-probe microanalysis. In the wavelengths from 1 to 10 A, sealed proportional counters may be used. For longer-waveleiigtli analysis—in the range from 10 to 93 A—the thinnest possible detector window is required to limit spectral attenuation. Nitrocellulose windows have proved successful. Nondispersive detection systems using cooled Li-dnfted Si are also applicable. [Pg.1760]

In another expt with the same cartridge case, a level gauge was simulated by viewing the top of the case with a 1.6mm thick cadmium filter containing a horizontal slit, 3.5cm x 1.3cm. Here the top of the case and the cadmium slit were aligned with the center of the detector window. The case was first filled with proplnt grains to just below the bottom of the slit and successive 0.4-g increments of additional proplnt were added (Table 9)... [Pg.126]

FIGURE 10.6 Schematic diagram of flow and diffusion within the T-sensor at a 1 1 1 flow ratio. A reference solution enters the device from the left, a detection solution from the middle, and a particle-laden sample stream enters from the right. The inset shows (1) original flow boundaries, (2) reference stream, (3) particle-laden sample stream, (4) diffusion of detector substance into reference stream, (5) diffusion of reference substance into detector stream, (6) detection stream, (7) diffusion of sample analyte into detection stream, (8) diffusion of detector substance into sample stream, and (9) detector window [443]. Reprinted with permission from the American Association for the Advancement of Science. [Pg.343]

A recent innovation in thin-film detectors has appeared on the commercial market in which a standard Si(Li) detector is protected by an apparently strong, X-ray transparent and radiation insensitive thin-film. The exact nature and composition of this thin-film is yet to be disclosed in the open literature. However, according to manufacturer s reports [13] this detector window allows transmission of boron X-rays (0.185kv for Ka) and can be operated at atmospheric pressure. In a typical spectrum from this type of detector provided by the manufacturer, a strong oxygen Ka line as well as well-defined Si, Al and Na peaks can be observed for albite (NaAlSi Og). Since these detectors have only recently been installed in the field, details of performance have yet to be reported by users. [Pg.40]

Fig. 1.1. Schematic illustration of a CEC column consisting of a packed and an open segment. The latter is divided into pre-detection and post-detection open segments by the detector window. Fig. 1.1. Schematic illustration of a CEC column consisting of a packed and an open segment. The latter is divided into pre-detection and post-detection open segments by the detector window.
FIGURE 5.4 An x-ray spectrum from a normal skin sample obtained by PIXE-analysis. Detectable peaks are seen above the continuous background (Bremsstrahlung) radiation. Note that the Na peak is not detectable (due to absorption in the sample and the detector window). [Pg.49]

The counting efficiency for the shown system approaches 52% for radionuclides with high maximum beta-particle energies (see Fig. 2A.2). This value exceeds the 39.6% based on the geometry of a 2.2-cm-dia. sample on a filter relative to the detector window. The counting efficiency exceeds the... [Pg.16]

Detection of separated components is done directly in the capillaries. A section of the polyimide coating is removed at one end of the capillary and this window is positioned in the light path of an optical detector. Bands of separated compounds are recorded as they migrate past the detector window. [Pg.291]

Figure 1 Schematic representation of isoelectric focusing in coated capillaries. Both patterns obtained in the focusing and mobilization steps are characteristic of the sample. Mobilization of the focused zones may occur at the anodic or the cathodic end by introducing an appropriate ion in the electrolyte vessel. The order of the appearance of the components in the detector window is opposite in the two steps. Figure 1 Schematic representation of isoelectric focusing in coated capillaries. Both patterns obtained in the focusing and mobilization steps are characteristic of the sample. Mobilization of the focused zones may occur at the anodic or the cathodic end by introducing an appropriate ion in the electrolyte vessel. The order of the appearance of the components in the detector window is opposite in the two steps.
EDAX analysis of these materials, as illustrated in Figures 2b and 3b, shew little difference between the samples with the exception of the silicon peak found in the carbon-silicon alloy. It should be noted that EDAX is inherently insensitive to the lower atomic number elements due to the low fluorescent yields of the lighter elements, internal absorption, and low transmission factors for these elements through the beryllium detector window of the instrument. Thus, carbon and oxygen are notably absent from the conventional EDAX spectra. [Pg.388]

First, the uranium must be present on the surface of the material being surveyed. Since uranium decays by emission of a particles, which travel only short distances in materials, any uranium that is imbedded in the surface being surveyed will be partially or completely masked. Secondly, when performing surveys, it must be possible to place the detector very close to the surface being surveyed (i.e., approximately one-quarter of an inch) (DOE 1988, 1994), and uneven surfaces that are unintentionally touched can tear the detector window, disabling the instrument. Additional information is available in MARSSIM (1997) on the use and usefulness of field survey instruments. [Pg.321]

Fig. 48. Change in the small angle scattering pattern of polyethyleneterephthalate yam at 21 C upon stretching in a fraction of a second Detector window was perpendicular to fibre axis covered one of the layer lines. The stretch occured between frames 3 and 5. Measuring time per frame 2s. Fig. 48. Change in the small angle scattering pattern of polyethyleneterephthalate yam at 21 C upon stretching in a fraction of a second Detector window was perpendicular to fibre axis covered one of the layer lines. The stretch occured between frames 3 and 5. Measuring time per frame 2s.
A w edge shaped detector window (with radial sides, circular inner and outer sides), centered on r and spanning r, to rj where h " I... [Pg.422]

If the detector window is fixed, both r, and a 2 will remain constant during a run, so that / and 6 will be constants independent of time. Note that while the ratio of dfd2 remains constant with time the difference between d and d2 decreases since the value of d, decreases with time ... [Pg.423]

The active area of the detector is usually between 10 and 30 mm. The collection angle is maximized by placing the detector window as close to the specimen as possible in the limited space available in the gap of the objective lens pole piece. Even so, the collection angle is a small fraction of the solid angle of 4ir steradian over which the x-rays are generated. [Pg.190]

Fig. 3.1 shows a schematic of a typical CE system set-up. Separations are achieved by filling a capillary with an electrolyte solution. A volume of sample is then injected into the end of the capillary furthest from the detector, usually performed by applying a pressure to the sample vial whilst the capillaiy is inserted into the sample vial. The capillary is then immersed in buffer reservoirs, which are placed at either end of the capillary. An electrical field is then applied (between 1 and 30 kV) which causes the compounds in the sample mixture to migrate along the capillary towards the on-capillary detection system. The smaller, higher-charged compounds will reach the detector window first. [Pg.102]


See other pages where Window Detector is mentioned: [Pg.188]    [Pg.200]    [Pg.202]    [Pg.216]    [Pg.258]    [Pg.260]    [Pg.102]    [Pg.629]    [Pg.455]    [Pg.415]    [Pg.122]    [Pg.195]    [Pg.611]    [Pg.168]    [Pg.126]    [Pg.455]    [Pg.55]    [Pg.71]    [Pg.145]    [Pg.50]    [Pg.72]    [Pg.62]    [Pg.126]    [Pg.16]    [Pg.171]    [Pg.171]    [Pg.182]    [Pg.146]    [Pg.426]   
See also in sourсe #XX -- [ Pg.208 ]




SEARCH



Ultrathin window detectors

Window detector circuit

© 2024 chempedia.info