Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Minimum acceptable criterion

Algorithm 1 requires the a priori selection of a threshold, s, on the empirical risk, /en,p( X which will indicate whether the model needs adaptation to retain its accuracy, with respect to the data, at a minimum acceptable level. At the same time, this threshold will serve as a termination criterion for the adaptation of the approximating function. When (and if) a model is reached so that the generalization error is smaller than e, learning will have concluded. For that reason, and since, as shown earlier, some error is unavoidable, the selection of the threshold should reflect our preference on how close and in what sense we would like the model to be with respect to the real function. [Pg.178]

No precise correlation is available between the immersion-compression test and results to be expected from pavements in the field. The test provides comparative guidance on water damage susceptibility. For the diabase aggregate, the Sulphlex-233 mixtures measured dry, and after immersign in the case of wet specimens, exceed the 300 psi (2.07 x 10 pascal) minimum compressive strength usually considered necessary for acceptable mixture performance in the field. Retained strengths generally fall below the 70 percent criterion considered acceptable for asphaltic mixtures. [Pg.218]

On the basis of the approach defined here, minimum values of Q consistent with an acceptable steady-state approximation may be estimated by comparing i(t)/Q with a(i, Xj) the former must be sufficiently small. An alternative approach to the estimation of minimum acceptable values of Q is to introduce simplified functions/(t), a(i), Le i), and WJW that enable equation (88) to be solved analytically [89]. Comparison of the solutions for finite values of Q with those for H = oo then provides an indication of how small Q can be if the steady state is to be reasonable. In this manner, Q > 100 has been obtained as an approximate criterion [89]. [Pg.177]

Our approach [265, 440, 445] enhances the simple selection mechanism of the CORBA trader to consider QoS. Here, an importer (which is the client using a trader to search for a service) describes a service by service tjrpe and service properties as before, but it is possible to specify roles for several values of the same property, namely a target value for the property, a lower bound (minimum acceptable quality) and an upper bound (maximum useful quality). This enhancement allows a client to formulate wishes and limits on service properties, see Fig. 4.16. Additionally, preferences for specifying the importance of properties can be expressed. These values, describing different aspects for the same quality criterion, are called Service Request Property. The importer can specify a whole vector of such properties to express its complete requirements on a service through different QoS aspects. In a similar way, a server offering a service can describe its limitations and capabilities by expressing upper and lower bounds for the quality aspects it can deliver. [Pg.409]

Some specifications require a minimum Charpy v notch impact requirement of 15 ft lb energy absorbed at the minimum expected service temperature. However, this does not mean that a test specimen exhibiting 60 ft lb is four times tougher than the minimum. The main value of notched bar impact testing is as a criterion for acceptance of materials where reliable correlation with service behavior has been obtained. [Pg.135]

We must take note of the behavior of the remaining materials in terms of chemical resistance. Consider the example of a salty aqueous fluid. The database includes a quahtative appreciation criterion, which means that we cannot cany out a classification or elimination on the basis of the limit values. We propose to visualize the hierarchical classification of the remaining materials in relation to this criterion in order to determine the maximum or minimum acceptable performance on the basis of that of a reference material (Figure 8.4). [Pg.137]

COD results can be used directly to evaluate and compare materials or to calculate defect sizes for particular structures. Two criteria are typically selected for COD evaluation. The first is the type of load-displacement record developed in the test. There are usually four types. These are shown in Fig. 2 in order of increasing resistance to crack extension. The second criterion is the magnitude of the critical COD, which occurs at maximum load or unstable crack extension. The value of 0.10 mm minimum COD is often cited as a criterion for acceptability, similar to 20-J CVN impact energy. [Pg.534]

The relevant International Standard for resin-modified glass-ionomer cements uses flexural strength as the criterion of acceptable mechanical performance [26], and further specifies a minimum flexural strength of 20 MPa for these materials. No distinction is made between values required for liner/base materials and full restoratives, or indeed for materials intended for any other application, such as luting, but most brands of resin-modified glass-ionomer cement comfortable exceed this minimum value [27]. [Pg.144]

Compliance with the EN standards is the minimum acceptable criterion for conformity, so I ll start by identifying the appropriate standards (also see Essential Requirements). The science of selecting the appropriate standards can be simple or complicated depending on your experience and product type. Exercise care when identifying the proper standards that may apply to your product design. The selection of a product standard may be a relatively simple task, as for many Low-Voltage Directive (LVD) products, or cumbersome, as for some machines. You must consider the product s intended usage and make sure the product falls within the standard s scope. [Pg.46]

The phosphate monitors in use can experience problems, as a result of chemical reagent quality and monitor reliability. In the Wales case study, this prompted a comprehensive review leading to site-specific improvements where necessary and the adoption of a minimum acceptable criterion for monitoring that the standard deviation of results should not exceed 0.2 mgA (P). The range in dosing performances that were experienced is illustrated by Figure 8.4, but were soon improved where necessary it was found that the application of ortho-phosphate at a constant or near constant dose is necessary to maximise its effect. [Pg.71]

Sensitivity by itself is not sufficient to completely evaluate an LCEC system for analytical purposes. The minimum detectable quantity (detection limit) is of more practical importance. The detection limit takes into consideration the amount of baseline noise as well as the response to the analyte. The detection limit is then defined as the quantity of analyte which gives a signal-to-noise ratio of three (a S/N of 3 is the generally accepted criterion although other values have been used). For a complete description of an LCEC application, both the sensitivity and detection limit, along with the S/N criteria used, should be provided. [Pg.24]

If r is outside any e-sphere, then count again the current configuration r (i.e., reject an implicitly attempted jump along Dkl because it would land outside the e-sphere for minimum /). The remainder of the simulation steps, on the average a fraction 1 - P of them, are local MC steps drawn from a uniform distribution and accepted or rejected according to the Boltzmann criterion. [Pg.292]

An approach for analyzing data of a quantitative attribute that is expected to change with time is to determine the time at which the 95% one-sided confidence limit for the mean curve intersects the acceptance criterion. If analysis shows that the batch-to-batch variability is small, it is advantageous to combine the data into one overall estimate by applying appropriate statistical tests (e.g., p-values for level of significance of rejection of more than 0.25) to the slopes of the regression lines and zero-time intercepts for individual batches. If it is inappropriate to combine data from several batches, the overall shelf life should be based on the minimum time a batch can be expected to remain within the acceptance criteria. [Pg.345]

Four blades are ordinarily considered a desirable minimum for mediumsized fans, with eight blades a minimum for large diameters. Twelve blades are a practical maximum. Fan diameters are basically determined by the volume of air to be handled. Air velocities through the fan range from 1200 to 2500 fpm. An economic design value is about 1800 fpm. Acceptable fan blade tip speeds, with low noise as the criterion, vary with the type tower, the type of fan cylinder and the blade tip clearance. [Pg.166]

This criterion may be used during a sequential optimization process (see chapter 5), leading to an acceptable result and to completion of the optimization process once the threshold value has been reached. Alternatively, it may be used to establish ranges of conditions in the parameter space for which the result will be acceptable. This latter approach has been followed by Glajch et al. [415], by Haddad et al. [424] and by Weyland et al. [425] and was referred to as resolution mapping by the former. Within the permitted area(s) secondary criteria are then required to select the optimum conditions. For example, the conditions at which the k value of the last peak (k is minimal while the minimum value for Rsexceeds 1 may be chosen as the optimum. Such a composite criterion can be described as... [Pg.141]

Allowable Spread in Residence Time. Other ways of stating the requirement of equal residence time of all parts of the reactant is that the flow through the reactor should approach plug flow or that the residence time distribution (RTD) should be equivalent to that in a large number of mixers in series. An often used rule of thumb is that this requirement is met when the equivalent number of mixers (N ) exceeds a certain value, say 5. However, this criterion is at best a semi-quantitative one, since the minimum value of is dependent upon the accepted deviation from the ideal reactor, and on the degree of conversion and the reaction order. [Pg.10]

Figure 1 shows the minimum Pe number and equivalent number of mixers for a first and second order reaction as a function of conversion depth according to Equation 1. It clearly follows that a fixed number of mixers (e.g, 5 or 10) as a criterion for an acceptable deviation from an ideal plug flow reactor is a gross oversimplification which can be misleading. [Pg.10]


See other pages where Minimum acceptable criterion is mentioned: [Pg.242]    [Pg.904]    [Pg.169]    [Pg.26]    [Pg.1207]    [Pg.981]    [Pg.149]    [Pg.210]    [Pg.214]    [Pg.451]    [Pg.398]    [Pg.254]    [Pg.11]    [Pg.15]    [Pg.37]    [Pg.340]    [Pg.12]    [Pg.561]    [Pg.841]    [Pg.267]    [Pg.142]    [Pg.108]    [Pg.110]    [Pg.121]    [Pg.34]    [Pg.4027]    [Pg.35]    [Pg.1001]    [Pg.470]   
See also in sourсe #XX -- [ Pg.46 , Pg.49 ]




SEARCH



Minimum criteria

© 2024 chempedia.info