Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Concentrate test fluids

Concentration test fluid (Igepal CO-630 or comparable) Temperature (°C)... [Pg.262]

The limit of determination [or limit of quantitation (LOQ)] is defined in Directive 96/46/EC as the lowest concentration tested at which an acceptable mean recovery (normally 70-110%) and acceptable relative standard deviation (normally <20%) are obtained. The specific requirements for LOQ in crops, food, feed, soil, drinking and surface water, air, body fluids, and tissues are described in Section 4. Because the abbreviation LOD usually means limit of detection rather than limit of determination, the authors prefer not to use this abbreviation here in order to avoid confusion, and LOQ is used throughout. According to Directive 96/46/EC no data with regard to the limit of detection must be given. [Pg.24]

As the name implies, the cup-and-bob viscometer consists of two concentric cylinders, the outer cup and the inner bob, with the test fluid in the annular gap (see Fig. 3-2). One cylinder (preferably the cup) is rotated at a fixed angular velocity ( 2). The force is transmitted to the sample, causing it to deform, and is then transferred by the fluid to the other cylinder (i.e., the bob). This force results in a torque (I) that can be measured by a torsion spring, for example. Thus, the known quantities are the radii of the inner bob (R ) and the outer cup (Ra), the length of surface in contact with the sample (L), and the measured angular velocity ( 2) and torque (I). From these quantities, we must determine the corresponding shear stress and shear rate to find the fluid viscosity. The shear stress is determined by a balance of moments on a cylindrical surface within the sample (at a distance r from the center), and the torsion spring ... [Pg.60]

According to USP 28 [1], the range of an analytical method can be defined as the interval between upper and lower levels (in the Pharmaceutical Industry usually a range from 80 to 120% of the target concentrations tested) of the analyte that have been demonstrated to be determined with a acceptable level of precision, accuracy, and linearity. Routine analyses should be conducted in this permitted range. For pharmacokinetic measurements, a wide range should be tested, where the maximum value exceeds the highest expected body fluid concentration, and the minimum value is the QL. [Pg.256]

The additional optimization of other ELISA factors may be necessary after determining the best concentration of antigen/enzyme conjugate to use (see Note 8). Three parameters may be varied to make improvement. Each one should be adjusted independently and an optimum dilution factor or value determined for routine use. Molecules of interest may be masked by components of the test fluid leading to unexpectedly low signals. This may be remedied by mixing the sample with PBS/Tween supplemented with 10% fetal bovine serum (see Note 9). [Pg.117]

Dimethyl sulfoxide (DMSO) test6 — This test measures the diffusional resistance of the horny layer. Equal quantities of three different concentrations of DMSO are applied to three plastic wells for 5 min. DMSO provokes whealing in human skin. The wheals are scored 10 min after removal of the test fluid using a scale. Individuals with high reactivity are those susceptible to whealing with the lowest concentration of DMSO. [Pg.493]

Sodium chloride is a component of all body fluids, including tears. A solution of 0.9% is approximately isotonic with tears. Of the various concentrations tested, 2% to 5% formulations have proven effective, with an irritation level acceptable to most patients. Studies comparing various hyperosmotic agents in human subjects have confirmed the usefulness of hypertonic sodium chloride in the treatment of corneal edema. Use of 5% sodium chloride in ointment form can be effective in reducing corneal thickness and in improving vision.The maximum reduction in corneal thickness occurs 3 to 4 hours after instillation of the ointment (Figure 15-1). [Pg.279]

The apex of a cone is brought into close proximity, but not in to contact, of a horizontal plate (Figure 3-7). Often, the apex is truncated slightly to eliminate a sharp point. The minimum gap between the cone and plate is usually of the order of 50 xm so that this geometry may not be suitable for dispersions containing larger diameter solids. The test fluid fills the gap between the cone and the plate, and because the gap is small, only a small volume (typically, 1-5 mL) of fluid is needed. The cone is rotated and the torque is measured at various speeds of rotation. A cone and plate viscometer can be used to obtain shear stress-shear rate curves and shear-stress versus time at constant shear rate curves as described above for concentric cylinder geometry. The... [Pg.70]

Concentric Cylinder Rotary Viscometer. These instruments are designed to have the test fluid in the annulus between two concentric... [Pg.137]

When pulmonary surfactant is present in amniotic fluid in sufficient concentrations, the fluid is able to form a highly stable surface film that can support bubbles. Other substances in the fluid, including proteins, bile salts, and salts of free fatty acids, are also capable of forming stable bubbles, but they can be removed from the film by ethanol, which competes with the other substrates for a position in the surface fihn. The test makes use of the principle that more surfactant activity is necessary to support a stable foam as the fraction of ethanol in the mixture is increased. Therefore a fixed volume of undiluted amniotic fluid is mixed with increasing volumes of ethanol, and the largest fraction of ethanol in which the amniotic fluid is stfll capable... [Pg.2192]

There is a need for improving the understanding of complex fluid and suspension flow behavior in mixing equipment. The concentration profiles obtained by MR imaging were used to monitor mixing with respect to axial position in a Kenics mixer, to which the two viscosity-matched test fluids were delivered at approximately the same volumetric flow rate. ... [Pg.440]

A major objective of fundamental studies on hollow-fiber hemofliters is to correlate ultrafiltration rates and solute clearances with the operating variables of the hemofilter such as pressure, blood flow rate, and solute concentration in the blood. The mathematical model for the process should be kept relatively simple to facilitate day-to-day computations and allow conceptual insights. The model developed for Cuprophan hollow fibers in this study has two parts (1) intrinsic transport properties of the fibers and (2) a fluid dynamic and thermodynamic description of the test fluid (blood) within the fibers. [Pg.75]

This viscometer consists of two concentric cylinders, the outer cup and the inner bob, with the test fluid in the annular gap. The radius of the bob is R, that of the cup is R, and the length of surface in contact with the sample is L. One of the cylinders (preferably the cup) is rotated at various controlled angular velocities (Q), and the resulting torque (T) is measured. The shear stress is... [Pg.398]

Rheometer Any instrument designed for the measurement of non-Newtonian as well as Newtonian viscosities. The principal class of rheometer consists of the rotational instruments in which shear stresses are measured, and a test fluid is sheared between rotating cylinders, plates, or cones. Various types of rotational rheometers are concentric cylinder, cone-cone, cone—plate, double cone—plate, plate—plate, and disc (16). [Pg.515]

Particle shape, size, and density polymer gelling-agent concentration solids concentration test temperature and fracture shear rate affect viscosity increase that results from the addition of a solid in the fracturing fluid. [Pg.566]

In experiments where the extent of hemolysis was needed, samples of test fluid were withdrawn from the circuit at various times, and the hemoglobin concentration in the supernatant was measured (18). [Pg.278]

The distribution of lithium between the intracellular and extracellular spaces occurs unequally, and reaches equilibrium at least 12 h after a single oral dose. Thus, the standardized 12-h serum lithium concentration test (12 h-stSLi) is used to adjust dosage and to monitor treatment (Amdisen and Nielsen-Kudsk 1986). Repeat administration of lithium normally leads to a steady state of serum lithium concentration within 4 to 7 days (Lehmann 1997b). The lithium concentration in extracellular fluids and blood is higher than that in the cell, which has a depot function (Birch 1997). [Pg.488]

The region of this curve between DES concentrations of 15 to 0.5 g/well can be plotted as a linear function. In this example, the correlation coefficient is r = 0.98 which demonstrates that the concentration of DES on the well is related to absorbance. Unloiown quantities of DES in the fluid can then be determined by reading the absorbance of the test fluid and using the standard curve to identify levels of DES. [Pg.285]

Any rotational viscometer, such as the Brookfield model DV, or the Anton-Paar Rheolab QC, can be used. The suggested test fluids of the pseudoplastic type can be paper pulp or orange juice concentrate, while the dilatant type can be concentrated starch or limestone suspensions. [Pg.360]

The permeability of the test fluid that has to be filtered is very difficult to predict, since it depends on other factors. On one hand, there are fouling and polarization problems, which can reduce dramatically the flow rate. These depend on the type of sample, the concentration, etc. On the other hand but related, the design of the membrane module also plays an important role. A correct design of the module can prevent stagnant flows inside the module (which reduces the effective membrane area) and can significantly increase the shear stress on the membrane, which can prevent the mentioned phenomena. [Pg.45]

Figure 3.5 shows the curve for an experiment carried out with a membrane obtained with 20% polysulfone in dimethyl formamide (DMF) and precipitated in a coagulation bath composed of 50% water and 50% DMF. The test fluid was obtained by mixing six dextrans of different molecular weights (17, 40, 70, 100, 200, 500 kDa) from Leuconostoc (Fluka). The concentration of each one in solution was 1.0 g/L. From the curve obtained, a MWCO of 28kDa could be determined. Transmembrane pressure (driving force) was set to 9 bar. [Pg.46]


See other pages where Concentrate test fluids is mentioned: [Pg.407]    [Pg.190]    [Pg.354]    [Pg.380]    [Pg.59]    [Pg.61]    [Pg.70]    [Pg.86]    [Pg.30]    [Pg.239]    [Pg.242]    [Pg.116]    [Pg.93]    [Pg.549]    [Pg.591]    [Pg.612]    [Pg.68]    [Pg.879]    [Pg.406]    [Pg.23]    [Pg.155]    [Pg.201]    [Pg.622]    [Pg.90]    [Pg.213]    [Pg.226]    [Pg.233]   
See also in sourсe #XX -- [ Pg.360 ]




SEARCH



© 2024 chempedia.info