Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Scanning calibration

System Calibration A system calibration scan is obtained by placing the standard phantom containing a positron emitter in a phantom holder at the center of the FOV for uniform attenuation and exposure. The reconstructed images are checked for any nonuniformity. A bad detector indicates a decreased activity in the image, and warrants the adjustment of PM tube voltage and the discriminator settings of PHA. [Pg.105]

The procedure for calibration, scanning, and data acquisition is the same as that outlined in Subheading 3.1.4. [Pg.51]

A low-resolution mass measurement is a simple procedure and can be performed with most of the mass spectrometric systems discussed in Chapter 3. The instrument is set up at a resolving power (RP) above 1000. At this low resolving power, the molecular ions of most organic compounds that differ by a unit mass are well separated. The mass spectrometer is first mass-calibrated with an external calibration procedure. During the calibration scan, the computer stores the peak centroid (the center of gravity) time and the area of each peak. The mass of the ion is related exponentially to the peak centroid time ... [Pg.198]

For GC-HRMS (high mass resolution) systems, an internal mass calibration (scan-to-scan) for accurate mass determinations by control of the data system is employed. At a given resolution (e.g., 10 000), a known reference is used which is continuously leaked into the ion source during analysis. The analyser is positioned on the exact mass of the substance ion to be analysed relative to the measured centroid of the known reference. At the beginning of the next scan, the exact position of the centroid of the reference mass is determined again and is used as a new basis for the next scan (see Chapter 2.3.4.3 Lock-Plus-Cali Mass Technique). [Pg.325]

The calibration curve of each rosetta strain gauge was so obtained and ftg.5 shows the sum of the principal stresses at the measuring points versus pressure inside the vessel. Further tests were carried out to obtain the calibration factor and to check that it remained constant on the whole scan area of the test surface. This was achieved through additional measurements using the SPATE system on fixed points on the surface located very close to the applied rosetta strain gauges. This procedure gave the following results ... [Pg.411]

Documentation of area scanned, top view, side view and all calibration data of ultrasonic instrument and system (comes in a standard 3 page report form)... [Pg.776]

The Calibration of the positioning system is carried out using a bar with a given distance which is placed between the referenspoint on the microphone collar and the probe. The distance is then entered into the acquisition software together with informations of the air temperature close to the tested object, pipe dimension, type of UT-probe (probe height) and scanning direction. [Pg.863]

Sheiko S S, Moller M, Reuvekamp E M C M and Zandbergen H W 1993 Calibration and evaluation of scanning-force-microscopy probes Phys. Rev. B 48 5675... [Pg.1724]

The purity of a sample of K3Fe(CN)6 was determined using linear-potential scan hydrodynamic voltammetry at a glassy carbon electrode using the method of external standards. The following data were obtained for a set of calibration standards. [Pg.538]

Usually, in AFM the position of the tip is fixed and the sample is raster-scanned. After manual course approach with fine-thread screws, motion of the sample is performed with a piezo translator made of piezo ceramics like e. g. lead zirconate tita-nate (PZT), which can be either a piezo tripod or a single tube scanner. Single tube scanners are more difficult to calibrate, but they can be built more rigid and are thus less sensitive towards vibrational perturbations. [Pg.280]

Whilman, L.J. and Colton, R.J., Design and calibration of a scanning force microscope for friction, adhesion, and contact potential studies. Rev. Sci. fnsirum., 66, I (1995). Ba.selt, D.R. and Baldeschwieler, J.D., Imaging spectro.scopy with the atomic-force microscope. J. AppL Pliys., 76(1), 33-38 (1994). [Pg.217]

This, on the one hand, reduces the detection limit so that less sample has to be applied and, thus, the amounts of interfering substanees are reduced. On the other hand, the linearity of the calibration curves can also be increased and, hence, fewer standards need to be applied and scanned in routine quantitative investigations so that more tracks are made available for sample separations. However, the introduction of a large molecular group can lead to the equalization of the chromatographic properties. [Pg.57]

Fig. 38 Companson of manual dipping (A) with mechanized dipping (B) on the basis of scans and calibration curves [114] — 1 = cM-diethylstilbestrol, 2 = traw-diethylstilbestrol, 3 = ethinylestradiol Scanning curve 2 ng of each substance per chromatogram zone = 313 nm, /n > 390 nm Dipping solution water — sulfuric acid — methanol (85 + 15 + 1)... Fig. 38 Companson of manual dipping (A) with mechanized dipping (B) on the basis of scans and calibration curves [114] — 1 = cM-diethylstilbestrol, 2 = traw-diethylstilbestrol, 3 = ethinylestradiol Scanning curve 2 ng of each substance per chromatogram zone = 313 nm, /n > 390 nm Dipping solution water — sulfuric acid — methanol (85 + 15 + 1)...
Fig. 53 Fluorescence scan of femtogram quantities of 2,1,3-naphthoselenodiazole (A) and associated calibration curve (B). Fig. 53 Fluorescence scan of femtogram quantities of 2,1,3-naphthoselenodiazole (A) and associated calibration curve (B).
Double-beam spectrophotometers. Most modern general-purpose ultraviolet/ visible spectrophotometers are double-beam instruments which cover the range between about 200 and 800 nm by a continuous automatic scanning process producing the spectrum as a pen trace on calibrated chart paper. [Pg.667]

First we analyzed the states of a tip scanning along an ascent and descent surface on nanometer scale, and then we calibrated the lateral force obtained by the FFM we modified. It may be helpful to understand how to measure the true lateral force by an FFM. [Pg.208]

The emission of the indicator is reduced in places where there are substance zones that absorb at 2 = 254 nm present in the chromatogram. This produces dark zones (Fig 4A), whose intensity (or rather lack of it) is dependent on the amount of substance appUed. If the plate background is set to 100% emission the phosphorescence is reduced appropriately in the region of the substance zones. When the chromatogram is scanned peaks are produced, whose position with respect to the start can be used to calculate Rf values and whose area or height can be used to construct calibration curves as a function of the amoimt applied (Fig. 25). [Pg.521]

Relatively long time-constants of 1 to 60 s are used in order to minimize the noise inherent in c.d. spectra. Thus, the scanning rate is low, and as c.d. instruments tend to drift with time, they should be calibrated daily, and baselines should be measured for each sample. [Pg.78]

Atomic force microscope (AFM) is a powerful nanotechnology tool for molecular imaging and manipulations. One major factor limiting resolution in AFM to observe individual biomolecules such as DNA is the low sharpness of the AFM tip that scans the sample. Nanoscale 1,3,5,7-tetrasubstituted adamantane is found to serve as the molecular tip for AFM and may also find application in chemically well-defined objects for calibration of commercial AFM tips [113]. [Pg.233]

For the purpose of calibration and particle size calculations, it was decided to confirm the reported particle sizes of the various latices supplied by Dow and Polysciences, using scanning electron microscopy. Unfortunately, after subtracting the thickness of the gold layer with which the particles were coated from the size shown on the micrographs, some inconsistencies were noted with respect to the measured sizes of the particles and their elution behaviour. It was therefore decided to assume the reported sizes as true values with the exception of the 5T nm particle. [Pg.52]

Tables 6.27 and 6.31 show the main characteristics of ToF-MS. ToF-MS shows an optimum combination of resolution and sensitivity. ToF-MS instruments provide up to 40000 spectra s-1, a mass range exceeding 100000 (in principle unlimited), a resolution of 5000, and peak widths as short as 200 ms. This is better than quadruples and most ion traps can handle. Unlike the quadrupole-type instrument, the detector is detecting every introduced ion (high duty factor). This leads to a 20- to 100-times increase in sensitivity, compared to QMS used in scan mode. The mass range increases quadratically with the time range that is recorded. Only the ion source and detector impose the limits on the mass range. Mass accuracy in ToF-MS is sufficient to gain access to the elemental composition of a molecule. A single point is sufficient for the mass calibration of the instrument. ToF mass spectra are commonly calibrated using two known species, aluminium (27 Da) and coronene (300 Da). ToF is well established in combination with quite different ion sources like in SIMS, MALDI and ESI. Tables 6.27 and 6.31 show the main characteristics of ToF-MS. ToF-MS shows an optimum combination of resolution and sensitivity. ToF-MS instruments provide up to 40000 spectra s-1, a mass range exceeding 100000 (in principle unlimited), a resolution of 5000, and peak widths as short as 200 ms. This is better than quadruples and most ion traps can handle. Unlike the quadrupole-type instrument, the detector is detecting every introduced ion (high duty factor). This leads to a 20- to 100-times increase in sensitivity, compared to QMS used in scan mode. The mass range increases quadratically with the time range that is recorded. Only the ion source and detector impose the limits on the mass range. Mass accuracy in ToF-MS is sufficient to gain access to the elemental composition of a molecule. A single point is sufficient for the mass calibration of the instrument. ToF mass spectra are commonly calibrated using two known species, aluminium (27 Da) and coronene (300 Da). ToF is well established in combination with quite different ion sources like in SIMS, MALDI and ESI.

See other pages where Scanning calibration is mentioned: [Pg.64]    [Pg.64]    [Pg.64]    [Pg.64]    [Pg.166]    [Pg.739]    [Pg.778]    [Pg.1248]    [Pg.1916]    [Pg.242]    [Pg.426]    [Pg.154]    [Pg.403]    [Pg.81]    [Pg.105]    [Pg.96]    [Pg.393]    [Pg.51]    [Pg.682]    [Pg.6]    [Pg.359]    [Pg.833]    [Pg.132]    [Pg.360]    [Pg.362]    [Pg.496]    [Pg.113]    [Pg.46]    [Pg.203]   
See also in sourсe #XX -- [ Pg.209 ]

See also in sourсe #XX -- [ Pg.209 ]




SEARCH



Calibration of differential scanning calorimeters

Differential scanning calorimeters calibration

Differential scanning calorimeters enthalpy calibration

Differential scanning calorimeters heat capacity calibration

Differential scanning calorimeters temperature calibration

Differential scanning calorimetry calibration

Differential scanning calorimetry calorimetric calibration

Differential scanning calorimetry enthalpy calibration

Differential scanning calorimetry temperature calibration

Scan inherent calibration

Temperature calibration of differential scanning calorimeters

© 2024 chempedia.info