Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sample separation

The second fraction is eluted with two bed volumes of a 20 % solution of dichloro-methane in n-hexane followed by the third fraction eluting with two bed volumes of dichlor-omethane. Two bed volumes of methanol finally elute the most polar compounds from the column. All fractions are collected in 50-100 mL conical flasks. [Pg.456]

The first fraction usually contains aUphatic hydrocarbons, polychlorinated biphenyls (PCB), and some monocychc aromatic hydrocarbons the bulk of aromatic hydrocarbons, l,l,l-trichloro-2,2-bis(p-chlorophenyl)ethane (DDT) and l,l-dichloro-2,2-bis(p-phenyl)-ethene (DDE) are found in the second fraction. Carbonyl compounds such as phenylal- [Pg.456]

After addition of suitable internal standards the fractions are concentrated by rotary evaporation to 1 mL. The rotary evaporator should be fitted with a solvent trap as shown in Fig. 18-2 to prevent refluxing solvent to flow back into the attached flask. The rotary evaporator air inlet should either be connected to a tank of compressed nitrogen or be fitted with a 5 A molecular sieve trap. The concentrated samples are transferred with a 100-250 mL syringe into glass ampoules as shown in Fig. 18-3 which are flushed with nitrogen and flame-sealed while being cooled to well below room temperature with a mixture of crushed ice cubes and table salt. [Pg.457]


Figure Bl.19.21. A plot of cantilever displacement as a function of tip sample separation during approach and retraction with an AFM. Note the adliesive forces upon retraction from the surface. Figure Bl.19.21. A plot of cantilever displacement as a function of tip sample separation during approach and retraction with an AFM. Note the adliesive forces upon retraction from the surface.
Many groups are now trying to fit frequency shift curves in order to understand the imaging mechanism, calculate the minimum tip-sample separation and obtain some chemical sensitivity (quantitative infonuation on the tip-sample interaction). The most conunon methods appear to be perturbation theory for considering the lever dynamics [103], and quantum mechanical simulations to characterize the tip-surface interactions [104]. Results indicate that the... [Pg.1697]

Analytical chemistry is more than a collection of techniques it is the application of chemistry to the analysis of samples. As you will see in later chapters, almost all analytical methods use chemical reactivity to accomplish one or more of the following—dissolve the sample, separate analytes and interferents, transform the analyte to a more useful form, or provide a signal. Equilibrium chemistry and thermodynamics provide us with a means for predicting which reactions are likely to be favorable. [Pg.175]

This, on the one hand, reduces the detection limit so that less sample has to be applied and, thus, the amounts of interfering substanees are reduced. On the other hand, the linearity of the calibration curves can also be increased and, hence, fewer standards need to be applied and scanned in routine quantitative investigations so that more tracks are made available for sample separations. However, the introduction of a large molecular group can lead to the equalization of the chromatographic properties. [Pg.57]

This volume is the seeond of a series of practiee-orientated TLC/HPTLC books published in excellent quality by VCH Publishers. As in the first volume, a series of reagents and detection methods have been reviewed with the intention of helping the practieal analyst increase the detection specificity of routine samples separated by thin-layer ehromatography. [Pg.471]

Select a set of compounds resolved on a given CSP, calculate the similarity indices between all possible molecule pairs, and then use these indices to build a similarity matrix containing relevant information about the structural diversity within the set of samples separated on this CSP. [Pg.113]

Note that the type of exchange experiment just discussed is now carried out infrequently. Quite clearly, the experiments can be tedious, requiring many samples, separations, and analyses. In many systems, but not all, comparable data can be obtained by the NMR methods described in Chapter 11. [Pg.58]

When an EPDM elastomer was examined, its tack was so low that no permanent cured mbber-to-rubber bond formed and the peel sample separated cleanly during test, at very low separation forces. No GD stage was necessary. [Pg.653]

The chemical compositions of the samples, obtained from chemical analyses are reported in Table 1. In order to check the chemical analyses, the mother and washing liquors were collected, analysed and their acidity was titrated. In all cases, the alkaline cations were detected only as traces. The acidimetric titration allowed us to determine the HPA amount remaining in the solution. On the other hand, the samples separated after precipitation and washings were weighted in order to calculate the precipitate yields. The results are reported in table 1 where the samples are designated as MxY (M being the alkaline or ammonium cation, Y the heteroatom, x the stoichiometry deduced from chemical analyses. [Pg.592]

Single linear developments are mostly employed in the vertical mode. The apph-cabihty of the horizontal mode is discussed in Chapter 6. For circular and anticircular developments, the movement of the mobile phase is two-dimensional however, from the standpoint of sample separation it is a one-dimensional technique. Circular developments result in higher hRp values compared to linear ones imder the same conditions, and compoimds are better resolved in the lower-AR range. The same effect is noticed on plates with a layer thickness gradient (see Section 5.2.1). On the other hand, using antieircular developments, compounds are bettCT resolved in the upper-M range. [Pg.120]

Keep blank field samples separate from treated samples, both when in transit from the field and when in refrigerated storage awaiting processing. [Pg.76]

Table 1 summarizes several of the experimental methods discussed in this chapter. A need exists for new or revised methods for transport experimentation, particularly for therapeutic proteins or peptides in polymeric systems. An important criterion for the new or revised methods includes in situ sampling using micro techniques which simultaneously sample, separate, and analyze the sample. For example, capillary zone electrophoresis provides a micro technique with high separation resolution and the potential to measure the mobilities and diffusion coefficients of the diffusant in the presence of a polymer. Combining the separation and analytical components adds considerable power and versatility to the method. In addition, up-to-date separation instrumentation is computer-driven, so that methods development is optimized, data are acquired according to a predetermined program, and data analysis is facilitated. [Pg.122]

If the material sampled is made up of particles of different sizes, the large and small particles should be sampled separately and in the respective proportions in which they are present in the bulk. Moreover, each sample should contain as many particles as possible. [Pg.56]

Urine ( -aminolevuli nic acid) Acidification of sample separate -aminolevulinic acid on HPLC reaction with formaldehyde and acetylacetone HPLC/FL 10 pg/L No data Tabuchi et al. 1989... [Pg.446]

FIGURE 17.6 Chromatogram of a technical octylphenoxy PEO sample separated by LCCC, stationary phase RP-18, mobile phase methanol-water 86 14% by volume (reprinted from Adrian et al., 1998, with permission of Advanstar Communications, UK). [Pg.396]

As we hinted above, the most common use of the term validation involves simply retaining some samples separately from the main set of calibration samples and using those as a more-or-less independent test of the accuracy of the calibration model obtained. However, this definition is not universally agreed to. When the subject came up in the on-line discussion group, the following comment was made by Richard Kramer of the discussion group [1] ... [Pg.136]


See other pages where Sample separation is mentioned: [Pg.1677]    [Pg.1678]    [Pg.1696]    [Pg.1709]    [Pg.1716]    [Pg.186]    [Pg.25]    [Pg.220]    [Pg.445]    [Pg.246]    [Pg.96]    [Pg.46]    [Pg.210]    [Pg.32]    [Pg.48]    [Pg.48]    [Pg.9]    [Pg.684]    [Pg.323]    [Pg.502]    [Pg.765]    [Pg.32]    [Pg.34]    [Pg.36]    [Pg.262]    [Pg.294]    [Pg.74]    [Pg.74]    [Pg.74]   
See also in sourсe #XX -- [ Pg.162 ]

See also in sourсe #XX -- [ Pg.11 ]




SEARCH



Analytical separation, aqueous environmental samples

Automatic sample preparation and separation techniques

Capillary electrophoresis, separation samples

Compound group separation of seawater samples

Effects of Phase Separation, Sample Preparation, Grain Size

Effects of separating sample from the living system

Generic Sample Preparation and Uranium Separation Techniques

High-resolution separation, protein sample

Identification of Multicomponent Samples without Previous Separation

Parallel-plate separators isokinetic sample

Parallel-plate separators sample train

Residue samples, separation

Sample Preparations and Separations

Sample SCF Separations

Sample components, separation

Sample preparation separation methods

Sample preparation trace matrix separation

Sample separable rate

Samples impurities separation

Sampling phase separation

Separating the sample

Separation airborne sampling

Separation complex sample mixtures

Separation in a Macroscopic Sample Equilibrium State Diagram

Separation of Complex Sample Mixtures

Separation techniques sample preparation

Separator sample procedure

Solid/liquid separation samples

Terms Referring to Sample Component Separation

Weakly adsorbing samples, separation

© 2024 chempedia.info